Anaesthesia machine

229,230 views 84 slides May 07, 2011
Slide 1
Slide 1 of 84
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79
Slide 80
80
Slide 81
81
Slide 82
82
Slide 83
83
Slide 84
84

About This Presentation

No description available for this slideshow.


Slide Content

The Anesthesia Machine Dr. Smriti Gupta

Introduction The anesthesia gas machine is a device which delivers a precisely-known but variable gas mixture, including anesthetizing and life-sustaining gases. Original Boyle was made by the firm COXTERS. There are several differences between newer and older anesthesia machines. Advanced ventilators are the biggest difference between newer and older gas machines.

The differences between older gas machines -such as the Ohmeda modulus, Excel, or Aestiva and the Dräger Narkomed GS, Mobile, MRI, 2B, 2C, 3 or 4- are less than their similarities . While the differences between new models from GE Healthcare (ADU, Aisys , Aespire , Avance ) and Dräger ( Fabius GS, Narkomed 6000/6400, Apollo ) are more than their similarities .

Types of anesthesia machine Intermittent -Gas flows only during inspiration Egs : Entonox apparatus , Mackessons apparatus Continuous -Gas flows both during inspiration and expiration. Egs : Boyle Machine Forregar Dragger

History

Anesthesia work station Integrates most of the components necessary for administration of anesthesia into single unit. Standard guidelines have been given to manufacturers for minimum performance, design, characteristics and safety requirements of machine. The current standard for anesthesia workstation as( promulgated by American society for testing and materials) (ASTM ) is F1850 . European standard is EN740

Components of anesthesia machine

Pressure units to remember 1 atm pressure = 1 bar = 760 mm Hg = 14.7 psi = 100 kilopascals

Components of pressure systems

Limitations of check valve assembly The check valves are not designed to act as permanent seals for empty yokes. Small amounts of gases can escape if the yoke is empty or an empty cylinder (or cylinder with low pressure) and valve open is present in the yoke.

In order to minimize such losses – Yokes should not be left vacant for extended periods An empty cylinder should be replaced as soon as possible , if not then, An yoke plug can be used to prevent gas leak or An empty cylinder can be left behind after closing the valve

Yoke Block(Dummy Cylinder plug) Yoke block is a solid piece of metal that has a conical depression on one side to fit into the retaining screw and a hollow area on the other side to fit over the nipple. They are pin indexed. Uses Prevents gas leak from the machine when placed in an empty yoke. Connect cylinders larger than size E to the machine. Connect pipeline supply to the machine that does not have pipeline inlet connections.

Limitations of yoke block Can be responsible for crossover of gases in the machine. Pressure regulators in the machine are designed to accept gases at cylinder pressure may not function properly when supplied with gases at reduced pressure(pipeline supply through yoke block)

Limitations of pin index safety system A wrong cylinder can be placed in yoke if- 2 washers are placed on the port. Pins on the yoke are broken. Holes on the cylinder valve are too deep.

Pin index safety system

Bourdon pressure gauge

Safety features in cylinder pressure indicator Gauge is usually color coded. Name and symbol of gas are written over dial. If bourdon tube ruptures gas is vented from back side Gauges are angled and placed in such a way that it can be easily read by anesthetist. Instructions like “use no oil’’ “open the valve slowly’’ are written on the gauge.

Electronic cylinder pressure indicator Light emitting diodes(LED’S)in electronic pressure gauge indicate Cylinder valve is close –Dark color Cylinder valve is open – Pressure adequate –Green Pressure inadequate-Red

Types of pressure regulators Adam’s pressure regulator valve- Used in past Working principle is same differs in internal structure from newer valves(toggle levers instead of push rod) Fins on Adam’s regulator for N2O which are not present in newer regulators. Mackesson’s regulator Medishield valve Endurance valve Modern preset type of reducing valve-Direct and Indirect .

Safety features on pressure regulator Pressure regulators have safety relief valves If due to any reason there is build up of pressure in pressure regulator then the safety valve blow off at a set pressure of 525 k pa(70psi)

Oxygen failure safety valve Pressure sensor shut off valve Based on threshold principle Cuts off N2O supply when O2 pressure falls below 25 psi. MOA-O2 pressurises and holds open shut off valve that interrupts the supply of N2O and other gases if O2 pressure<threshold setting. oxygen failure protection device Based on proportioning principle Gas loaded regulator When O2 pressure decreases there is a proportional decrease in N2O supply and complete cut off seen at<12psi. MOA-O2 pressure regulator(primary regulator) controls secondary(slave) regulator located in N2O line

Master switch Turning the master switch to the ‘on’ position activates both pneumatic and electrical functions of the machine as well as certain alarms and safety devices.

Electronic flow meter Prefer digital system Solenoid valves Control flow on or off valves Computer controlled .

Link-25 .

Oxygen ratio monitor controller Sensitive oxygen ratio controller Linear resistors [3:1 ratio for N2O & O2)between O2 andN2O flow control valves. Ensure 25% O2 by limiting N2O flow. ORMC shuts off N2O if ratio of O2 flow falls below 30% S-ORC-newest hypoxic guard . installed in Fabius -GS by Drager.ensures aFiO 2 of 23%. O2 flow <200ml/min . .

Limitations of proportioning system Anti hypoxic device may also deliver hypoxic mixture under following conditions- Wrong supply of gas. Defective pneumatics/mechanics. Inert gas administration(3 rd gas-He,N2,CO2). Leaks downstream.