Check List check area point(GAS & CALORIES).pdf

negriteluk 5 views 18 slides Oct 28, 2025
Slide 1
Slide 1 of 18
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18

About This Presentation

calories analizer


Slide Content

Day Week
Month
3 Mon 6 Mon
Year
Filter / Contamination ○ △
Visual Check, Cleaning,
Replace Filter
Leakage ○ Visual Check
Heating Temp ○ Visual Check
Leakage ○ Visual Check
Heating Temp ○ Visual Check
Contamination ○ Visual Check, Cleaning
Back purge ○ Visual Check
Status of Movement ○ Visual Check, Cleaning
Leakage △
Visual Check,
Replace O-ring
Contamination ○ Visual Check, Cleaning
Leakage △
Visual Check,
Replace O-ring
Status of Movement ○ Visual Check
Leakage ○ △
Visual Check,
Replace Tube
Contamination ○ △
Visual Check,
Replace Tube
Cooling Temp ○ Visual Check
Contamination ○ Visual Check, Cleaning
Status of Movement ○ Visual Check, Cleaning
Cooling Fan ○ Visual Check, Cleaning
GAS ANALYZER CHECK LIST
Check Duration
Specification Check Point How to Remark NO.
1
Electric Cooler
Solenoid Valve
Drain Pump
Sample Line
Sample Probe
Gas Conditioner
2
3
4
5
6

Noise ○ Visual Check
Status of Movement ○ Visual Check
Flow Rate ○ Value Check, Record
Leakage ○ △ Replace Diaphragm
Value ○ Value Check, Record
Value ○ Value Check, Record
Value ○ Value Check, Record
Contamination ○ △
Visual Check,
Replace Filter
Check Drian Pump
Energy Value ○ Value Check, Record DIANOSTIC MENU
Absorbance Valve ○ Value Check, Record ABSORBANCE MENU
Cal Factor ○ Value Check, Record CRITICAL MENU
IR Energy ○ Value Check, Record CRITICAL MENU
Gas Outlet / Contamination ○ Visual Check, Cleaning OUTLET TUBE
In-Line Filter ○ △
Visual Check, Cleaning,
Replace Filter
Zero Gas ○ △
Residual Quantity Check,
Replace Cylinder
Span Gas ○ △
Residual Quantity Check,
Replace Cylinder
Moisture Sensor
Cal Gas Pressure
Sample Pressure
Vacuum Pressure
Sample Pump
Gas Analyzer
10
11
12
7
8
9

54
Module 2 – Analyzer Details
Checks, Preventive Maintenance and
Trouble-Shooting

55
Module 2 – Analyzer Details
Daily Checks

Successful Calibration

System in Measure

Reasonable Values

Gas Bottle Pressures

Instrument Air

Cabinet Temperature

56
Module 2 – Analyzer Details
Monthly Checks

MC3 Optical Energies

Sample Line Inspection

Vent Line Inspection

57
Module 2 – Analyzer Details
Quarterly Inspection

MC3 Internal Filter

Pump Diaphragms

Sample Flow

System Response and
Calibration

58
Module 2 – Analyzer Details

F
LOW

E
NERGY

T
EMPERATURE

C
ALIBRATION
It Boils down to the “FET C” Rule

35
Analyzer operation
Operationof software

36
Embedded hard/software
• Power supply filtering and overvoltage protection.
• LCD display.
• Capacitive keyboard.
• Industrial PC.
• Dedicated Control board.
• Barriers and Isolators for Intrinsically Safety
• Linux based software.

37
Alarms and
measuring
values
Digital in/outputsProcess
scheme
Large value
Graphs
Main display

38
Operation of the WIM Compas F software
• Menu driven control
• Select F1 to enter the menu
structure
• Select the options by selecting a
digit
• Select escape to return to
previous menu
• See the manual for a full
explanation of each menu

39
maintenance
Check’s and tasks to be
performed on regular intervals

40
Maintenance scheme
Once a week:
• Check the sample and bypass flows in the SCS.
• Check the sample and air pressure .
• When needed check and clean the restrictions
• After adjustment a new calibration is necessary.
Once a month:
• Perform a calibration (application dependant).
• Compare new slope and offset with previous.
– Check for 1-2 % O2 at maximum range
• Replace filters if needed.
• Check calibration gas bottle pressures
.

41
Once a year
• Clean the mixing chamber.
• Clean the flow meters with iso propanol.
• Cleaning of the gas- and air orifice with iso propanol. Replace Teflon ferrules.
• Remove the zirconium cell from the holder. Check the cell and replace if
needed.
• Remove the cell holder from the oven and add one NiCr grid.
• Clean the flame arrestors.
• Replace filters in air stations and sample filter.
• Check outlet of the oven for any deposit or blocking.
• Leak check the sample system.
• Check the analog and digital in- and outputs.
• Check air purge unit. Adjust purge pressure between 7 en 10 mbarg ( zone 2)
• Check the calibration gasses and check the correct values.
• Allow to stabilize and calibrate the analyzer.
• Check residual oxygen >1% at maximum wobbe-index value.
• Note critical settings (offset, slope).

42
Once a Year (options)
– Pump: replace membrane and valves.
– Fastloop filter: replace filters, support pads and o-
rings
– Sg cell: compare frequencies, if necessary clean
Once every Three years
– Perform yearly maintenance but now:
• Remove used NiCr grids from sample cell holder
• Replace Zirconium Cell
• Replace thermocouple of oven

43
Exploded view Cell plus housing
Cell housing
3x NiCr catalyst screen
Gas filled metal O-ring
Graphite seal tape
Ceramic insulation ring
Gas/air inlet
Vent
Zirconium cell

44
Insertion NiCr catalyst screens
Folded around the tube
12mm tubing
• Let the oven cool down!
• Unscrew nut at the back of the oven.
• Remove the signal clips from the cell.
• Unscrew the coupling at front of the oven.
• Remove bolts holding oven bracket.
• Pull forward the oven and release it.
• Unscrew the elbow coupling at the inlet.
• Pull the cell holder out of the oven.
• Unscrew the O
2
-cell from holder by using two
wrenches.
• Clean the cell holder.
• Insert new NiCr screens using a piece of 12mm
tubing. Push in the screens and push them backwards
with 12 mm tubing.
• Place a new seal ring over the O
2
-cell and screw
them back in cell holder. Tighten.
• Place holder with cell back in the oven.
• Place one turn graphite tape on the end thread of the
cell holder.
• Clean the thread of the elbow and re-install the elbow.
• Re-install the complete oven.

45
Zirconium cell
• Use the wrenches (size 22 mm) only on the cell holder and the small
nut.
• Check Zirconium cell for any damage.
• Use a new seal ring each time when reinstalling the cell.
Put the
wrench
here
Do not use a wrench here
Gas filled O- ring
Ceramic
isolation ring
Tags