Check List check area point(GAS & CALORIES).pdf
negriteluk
5 views
18 slides
Oct 28, 2025
Slide 1 of 18
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
About This Presentation
calories analizer
Size: 772.64 KB
Language: en
Added: Oct 28, 2025
Slides: 18 pages
Slide Content
Day Week
Month
3 Mon 6 Mon
Year
Filter / Contamination ○ △
Visual Check, Cleaning,
Replace Filter
Leakage ○ Visual Check
Heating Temp ○ Visual Check
Leakage ○ Visual Check
Heating Temp ○ Visual Check
Contamination ○ Visual Check, Cleaning
Back purge ○ Visual Check
Status of Movement ○ Visual Check, Cleaning
Leakage △
Visual Check,
Replace O-ring
Contamination ○ Visual Check, Cleaning
Leakage △
Visual Check,
Replace O-ring
Status of Movement ○ Visual Check
Leakage ○ △
Visual Check,
Replace Tube
Contamination ○ △
Visual Check,
Replace Tube
Cooling Temp ○ Visual Check
Contamination ○ Visual Check, Cleaning
Status of Movement ○ Visual Check, Cleaning
Cooling Fan ○ Visual Check, Cleaning
GAS ANALYZER CHECK LIST
Check Duration
Specification Check Point How to Remark NO.
1
Electric Cooler
Solenoid Valve
Drain Pump
Sample Line
Sample Probe
Gas Conditioner
2
3
4
5
6
Noise ○ Visual Check
Status of Movement ○ Visual Check
Flow Rate ○ Value Check, Record
Leakage ○ △ Replace Diaphragm
Value ○ Value Check, Record
Value ○ Value Check, Record
Value ○ Value Check, Record
Contamination ○ △
Visual Check,
Replace Filter
Check Drian Pump
Energy Value ○ Value Check, Record DIANOSTIC MENU
Absorbance Valve ○ Value Check, Record ABSORBANCE MENU
Cal Factor ○ Value Check, Record CRITICAL MENU
IR Energy ○ Value Check, Record CRITICAL MENU
Gas Outlet / Contamination ○ Visual Check, Cleaning OUTLET TUBE
In-Line Filter ○ △
Visual Check, Cleaning,
Replace Filter
Zero Gas ○ △
Residual Quantity Check,
Replace Cylinder
Span Gas ○ △
Residual Quantity Check,
Replace Cylinder
Moisture Sensor
Cal Gas Pressure
Sample Pressure
Vacuum Pressure
Sample Pump
Gas Analyzer
10
11
12
7
8
9
58
Module 2 – Analyzer Details
F
LOW
E
NERGY
T
EMPERATURE
C
ALIBRATION
It Boils down to the “FET C” Rule
35
Analyzer operation
Operationof software
36
Embedded hard/software
• Power supply filtering and overvoltage protection.
• LCD display.
• Capacitive keyboard.
• Industrial PC.
• Dedicated Control board.
• Barriers and Isolators for Intrinsically Safety
• Linux based software.
37
Alarms and
measuring
values
Digital in/outputsProcess
scheme
Large value
Graphs
Main display
38
Operation of the WIM Compas F software
• Menu driven control
• Select F1 to enter the menu
structure
• Select the options by selecting a
digit
• Select escape to return to
previous menu
• See the manual for a full
explanation of each menu
39
maintenance
Check’s and tasks to be
performed on regular intervals
40
Maintenance scheme
Once a week:
• Check the sample and bypass flows in the SCS.
• Check the sample and air pressure .
• When needed check and clean the restrictions
• After adjustment a new calibration is necessary.
Once a month:
• Perform a calibration (application dependant).
• Compare new slope and offset with previous.
– Check for 1-2 % O2 at maximum range
• Replace filters if needed.
• Check calibration gas bottle pressures
.
41
Once a year
• Clean the mixing chamber.
• Clean the flow meters with iso propanol.
• Cleaning of the gas- and air orifice with iso propanol. Replace Teflon ferrules.
• Remove the zirconium cell from the holder. Check the cell and replace if
needed.
• Remove the cell holder from the oven and add one NiCr grid.
• Clean the flame arrestors.
• Replace filters in air stations and sample filter.
• Check outlet of the oven for any deposit or blocking.
• Leak check the sample system.
• Check the analog and digital in- and outputs.
• Check air purge unit. Adjust purge pressure between 7 en 10 mbarg ( zone 2)
• Check the calibration gasses and check the correct values.
• Allow to stabilize and calibrate the analyzer.
• Check residual oxygen >1% at maximum wobbe-index value.
• Note critical settings (offset, slope).
42
Once a Year (options)
– Pump: replace membrane and valves.
– Fastloop filter: replace filters, support pads and o-
rings
– Sg cell: compare frequencies, if necessary clean
Once every Three years
– Perform yearly maintenance but now:
• Remove used NiCr grids from sample cell holder
• Replace Zirconium Cell
• Replace thermocouple of oven
43
Exploded view Cell plus housing
Cell housing
3x NiCr catalyst screen
Gas filled metal O-ring
Graphite seal tape
Ceramic insulation ring
Gas/air inlet
Vent
Zirconium cell
44
Insertion NiCr catalyst screens
Folded around the tube
12mm tubing
• Let the oven cool down!
• Unscrew nut at the back of the oven.
• Remove the signal clips from the cell.
• Unscrew the coupling at front of the oven.
• Remove bolts holding oven bracket.
• Pull forward the oven and release it.
• Unscrew the elbow coupling at the inlet.
• Pull the cell holder out of the oven.
• Unscrew the O
2
-cell from holder by using two
wrenches.
• Clean the cell holder.
• Insert new NiCr screens using a piece of 12mm
tubing. Push in the screens and push them backwards
with 12 mm tubing.
• Place a new seal ring over the O
2
-cell and screw
them back in cell holder. Tighten.
• Place holder with cell back in the oven.
• Place one turn graphite tape on the end thread of the
cell holder.
• Clean the thread of the elbow and re-install the elbow.
• Re-install the complete oven.
45
Zirconium cell
• Use the wrenches (size 22 mm) only on the cell holder and the small
nut.
• Check Zirconium cell for any damage.
• Use a new seal ring each time when reinstalling the cell.
Put the
wrench
here
Do not use a wrench here
Gas filled O- ring
Ceramic
isolation ring