CT Artifacts including their appearance , causes and solutions.
Size: 337.97 KB
Language: en
Added: Dec 02, 2020
Slides: 15 pages
Slide Content
CT ARTIFACTS Presented by: Charlene.T.Mutambanengwe Course : BSc Radiology and Imaging Technology[BRIT] NOIDA INTERNATIONAL UNIVERSITY (NIU)
Definition – Artifacts are defined as anything that appears on the image that is not on the object being scanned CLASSIFICATION 1.Physics based artifacts – Result from the physical processes involved in the data acquisition. Examples are photon starvation and beam hardening. 2.Patient based artifacts- Caused by patient movement or presence of metallic objects in the patient body. Example of these type of artifacts are metallic artifacts and motion artifacts 3.Scan based artifacts –Results from imperfection in scanners functioning. Examples are ring artifacts
A).BEAM HARDENING ARTIFACTS Caused by a polychromatic beam . This beam hardening artifact also produces cupping artifact. The center of the object is usually the thickest and therefore, the beam will be harder in the center than at the periphery of the image and is assigned lower Hounsfield units. This can be corrected by beam correction algorithms. Filtering can also reduce beam hardening by creating a monochromatic x-ray beam. Prepatient filter – absorbs soft x-rays. Bow filter – prehardens the x-ray beam.
B).METALLIC ARTIFACTS.
C).PHOTON STARVATION Parts of individual projections can be very noisy due to insufficient photons passing through the widest part of the patient. When these views are reconsructed,the noise is magnified, resulting in streaks. Example is scan across the shoulders. Smaller signals reach the detector, therefore large proportion of noise. Solution :Adaptive filtering- the regions in which the attenuation exceeds a specified level are smoothed before undergoing back projection . Milliampere (mA) modulation –higher mAs is used for the more attenuation profile to reduce the effect of photon starvation.
D).PARTIAL VOLUME ARTIFACTS. Partial volume effect occurs when two objects of different densities are occupying the same plane. This may cause the object of interest to not fill the plane resulting in an underestimated CT number. Partial volume artifacts occur when an off center object is occupying part of the scan plane. The inconsistences between views causes streaks. Solution : Use thinner slices.
E). RING ARTIFACTS. Ring artifacts occur due to malfunctioning of one or more of the detectors in the array. Due to failure of a particular detector or miscalibration, incorrect data in every projection will appear as a ring in the image. Solution : recalibrate the detectors. If rings persist even after recalibration ,call service.
F).CONE BEAM ARTIFACTS. Only in multi detector CT systems. As the section being scanned increases per rotation, a wider collimation is used. Because of this the x-ray beam becomes cone shaped instead of fan shaped and area imaged by each detector as it rotates around the patient is now a volume instead of a flat plane . It appears as lines in windmill formation particularly at the edges of the image. Solution :Use pitch selection recommended by the manufacturer. Most modern scanner cone beam reconstruction algorithms correct this artifact.
G). TUBE ARCING
H).STEP ARTIFACTS Stair step artifacts occur around edges of structures in three dimensional (3D) and multi planar reconstruction (MPR) images when wide collimations and non-overlapping reconstruction intervals are used. They are found in straight structures which are oriented obliquely with respect to movement of the table. Solution : Reduce collimation. Use overlapping reconstruction in helical imaging. Thinner acquisition slices.
I).MOTION ARTIFACTS. Motion artifacts are caused by the voluntary or involuntary movement of the patient during data acquisition. It appears as shading, streaking and blurring. Solution : use shortest scan time as possible. Immobilization aids. Consider sedation. Beta blockers for cardiac protocols. Special software correction which reduces weighting of views obtained towards the beginning and end of the scan.
J).OUT-OF-FIELD ARTIFACTS. These occur when a portion of the patient is outside the field of view (FOV). It is seen on the image as streaks. Common problem areas include shoulders, abdomen and lower pelvis. Solution : adjust patient position (L/R). Expand the FOV if possible. K).SPIRAL INTERPOLATION ARTIFACTS. These are caused by images that are created from views that are not in the same plane. This results in inaccuracies in CT number. Appearance : streaks Solution : Lower pitch.
L).ALIASING ARTIFACTS Also known as under sampling. Refers to the error in the accuracy component of the ADC during image digitization. When sampling, the brightness of each pixel in the image is measured and creates an analog signal. The more the samples, the more accurate the representation of signal will be,hence,if a lack of sampling has occurred the computer will process an inaccurate image resulting in an aliasing artifact. Appearance : Fine lines. Solution :reduce pitch Acquiring the largest possible number of projections per rotation.(on some scanners this can be achieved by using slower rotation speed)
M).EDGE GRADIENT EFFECT The edge gradient effect results in shading arising from irregularly shaped objects that have different densities. Occurs when barium and air lie adjacent to each other in the stomach . Appearance : straight lines radiating from high-contrast areas. Solution :Use thinner slices, Use low or neutral HU value oral contrast in place Barium.