Goldman applanation tonometry

47,602 views 44 slides Feb 20, 2014
Slide 1
Slide 1 of 44
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44

About This Presentation

No description available for this slideshow.


Slide Content

TONOMETRY
H.Kangari O.D.

TONOMETRY/H.KangariO.D. 2
Different technique in measurement of the
Intra Ocular Pressure(IOP):
Goldmann Applination Tonometry
Perkins Tonometer
Non-Contact Tonometry
Digital Evaluation

TONOMETRY/H.KangariO.D. 3
Goldmann Applination Tonometry is the
most accurate method of measuring
Intraocular Pressure.
Measuring the intraocular pressure is
important in diagnosis and management
of Glaucoma and Ocular Hypertension.

TONOMETRY/H.KangariO.D. 4
CONTRAINDICATIONS TO IOP
MEASUREMENT WITH GOLDMAN
TECHNIQUE
1)Active Infection
2)Patients with recurrent corneal erosions
3) Corneal abrasions that are not totally
healed

TONOMETRY/H.KangariO.D. 5
The Intraocular
Pressure is
determined by the
amount of force
required to flatten a
constant corneal
surface area

TONOMETRY/H.KangariO.D. 6

TONOMETRY/H.KangariO.D. 7
Major components of the tonometer
1)Contact Prism
2)Contact Probe
3)Housing
4)Calibrated
measuring drum
5)Tonometer Arm

TONOMETRY/H.KangariO.D. 8

TONOMETRY/H.KangariO.D. 9

TONOMETRY/H.KangariO.D. 10
CHECKING THE
INSTRUMENT
To calibrate the instrument a control weight bar
is used. Only a trained technician can calibrate
the tonometer.
To grossly check the calibration:
Turn the measuring drum below zero (the arm
should move backward)
Then turn the measuring drum past the zero
mark and the arm should move forward.
The arm should move in the same degree in
both direction, above and below the zero mark.

TONOMETRY/H.KangariO.D. 11
PROBE STERILAZATION
Always clean the tonometer probe before and
after each use.
Remove the probe from contact holder (hold the
arm before you remove the probe)
Wet a tissue with alcohol and apply to probe in a
circular fashion for 5 seconds.
Rinse the probe with Saline Solution and blot dry
with a tissue.

TONOMETRY/H.KangariO.D. 12
PROBE STERILAZATION
OR insert the probe into Hydrogen
Peroxide 3% for 5 to 10 minutes.
Rinse the probe with Saline Solution and
blot dry with a tissue.

TONOMETRY/H.KangariO.D. 13
CLEANING & STERILIZING THE
PROBE
some factors to be considered such as the following:
(1) 3% H2O2 will ruin the probes in a short period of time and if not properly
dried can cause marked pain and a secondary anterior uveitis that is very
difficult to resolve. Minimum of 5 minutes maximum of 10 minutes.
(2) 70% Isopropyl alcohol will also destroy the probes and if not properly
dried can also cause marked pain and a secondary anterior uveitis that is
very difficult to resolve.
When Using Either Of The Above Recommended Procedures Make Sure
You Always Rinse The Probe Thoroughly With Saline Solution Then Dry It
Completely Before Using It On The Patient's Eye. Never Leave The Probe
In The Hydrogen Peroxide For Longer Than 10 Minutes.

TONOMETRY/H.KangariO.D. 14
Hold the arm, while
inserting the probe.
ALIGN the probe- if the
patient has less than 3.00
D of astigmatism, position
the probe so that the
patient’s minus cylinder
axis is aligned with the
WHITE LINE on the prism
holder.

TONOMETRY/H.KangariO.D. 15
ALIGN the probe- if the
patient has greater than
3.00 D of astigmatism,
position the probe so that
the patient’s minus
cylinder axis is aligned
with the RED LINE on the
prism holder.
KEEP in mind, as you
rotate the probe
according to cylinder axis
the mires will tilt with the
direction of axis.

TONOMETRY/H.KangariO.D. 16
PATIENT PREPARATION
After you have completed a full refractive,
binocular and slit lamp evaluation proceed to
tonometry.
Anesthetize the cornea. Inform the patient that
you are going to put a drop in their eye which will
burn and their lids will feel heavy afterwards.
Place a drop of Tetracaine 0.5% in each eye
 Wet a strip of fluorescein with sterile saline and
place it is superior or inferior bulbar conjunctiva.

TONOMETRY/H.KangariO.D. 17
Position the patient comfortably at slit
lamp.
With the COBALT BLUE filter scan the
cornea
Place the measuring drum at 10 mmHg
Use Low Magnification

TONOMETRY/H.KangariO.D. 18
Give proper instructions to the patient:
“ I want you to keep your forehead and Chin in
the rest all the time. I want you to look at this
point ( based on the eye being tested give an
appropriate place for the patient to look at ).
Keep your eyes open as wide as you can and try
not to blink. I am going get close to your eye with
this blue light.

TONOMETRY/H.KangariO.D. 19
APPLINATION
From the outside instrument get as close
as you can to the cornea and center the
probe on the cornea.
If you now look in through the microscope,
you will see two faint blue mires (semi
circles)
If you don’t see this, AGAIN center the
probe from the outside of the instrument.

TONOMETRY/H.KangariO.D. 20
When you see two faint mires, move in the
instrument with the joy stick to touch the
cornea.
When you touch the cornea, the mires will
turn fluorescent green.

TONOMETRY/H.KangariO.D. 21
If semicircles move freely with the rotation
of the measuring drum, the initial contact
is correct.

TONOMETRY/H.KangariO.D. 22
The width and height
of two rings must be
equal and centered in
the field of view.
Rotate the measuring
drum, until the inner
borders of two
semicircles just touch.

TONOMETRY/H.KangariO.D. 23
Back off the tonometer and read the
pressure from the measuring drum.
The scales of measuring drum is marked
from 0-8 grams of force. To convert to
mmHg, multiply the result by 10.
After completion of the procedure,
reevaluate the corneal integrity with cobalt
blue light.

TONOMETRY/H.KangariO.D. 24
Correct position of
mires when
measuring the IOP

TONOMETRY/H.KangariO.D. 25
If this is the left eye,
the contact of the
probe is too much to
temporal side.
Move the probe to
nasal side

TONOMETRY/H.KangariO.D. 26
Top mire is superiorly
positioned.
The probe contact is too
inferior.
Move the probe
superiorly.

TONOMETRY/H.KangariO.D. 27
The contact is too
low.
Move the instrument
upward.

TONOMETRY/H.KangariO.D. 28

 ?

TONOMETRY/H.KangariO.D. 29
 ?

TONOMETRY/H.KangariO.D. 30
Inappropriate contact
The instrument has been
moved too much toward
the patient.
The mires will not move
by changing the drum.
Move backward and
reapplinate.

TONOMETRY/H.KangariO.D. 31
Mires are too thin.
Add more flourescein.

TONOMETRY/H.KangariO.D. 32
After applination, this
type of staining is
indicate of applying
too much force.

TONOMETRY/H.KangariO.D. 33
DIURNAL VARIATION
Generally, the variation in intraocular pressure over a 24
hour period is considered to be 3 to 5 mmHg with the
highest readings being about (6:00 a.m.).
However, there is recent evidence that known glaucoma
patient's highest pressure findings are in the afternoon.
Therefore, it would be best to monitor any questionable
patients. Take diurnal pressure measurements during
the day looking for any pressure spikes with variations
greater than 5 mmHg. Example: 14 mmHg O.U. @ 8:30
a.m. and 21 mmHg O.U. @ 3:30 p.m. is diagnostic.
Differences in pressure readings between the two eyes
of 3 mmHg or more must be questioned, this is not
normal.

TONOMETRY/H.KangariO.D. 34
Latest information:
Intraocular pressures (IOP) following Laser In Situ Keratomileusis
(LASIK) and Photorefractive Keratectomy (PRK) are
underestimated.
It depends on the patient's Pre-Op refractive prescription and how
much tissue needs to be removed. There is about 10 microns of
tissue removed per diopter of refractive error. A patient with a
refractive error of 3 diopters of myopia will have approximately 30
microns of tissue removed. For this patient the IOP findings would
be underestimated by about 2mm Hg and more for higher refractive
errors. This underestimation has been reported for Goldmann
applanation tonometry. It is an important new finding and appears to
be related to changes in corneal thickness. This might help explain
normal tension glaucoma where these patients may have thinner
corneas.

TONOMETRY/H.KangariO.D. 35
PERKINS TONOMETER
For patients with physical constraints who
cannot be positioned in the slit lamp or
bedriden patients Perkins tonometer (a
hand held tonometer can be used)

TONOMETRY/H.KangariO.D. 36

TONOMETRY/H.KangariO.D. 37
DIGITAL INTRAOCULAR
PRESSURE ASSESSMENT
This technique is used in those individual,
where the slit-mounted instrument or the
hand-held tonometer cannot be used.
In very young children, extremely anxious
patient or developmentally delayed
individuals.
The firmness of the globe is subjectively
evaluated as soft, meduim, or hard. The
harder the globe the higher the IOP.

TONOMETRY/H.KangariO.D. 38
Digital IOP assessment is contraindicated
for eyes with a recent history of blunt
trauma, penetrating ocular injury, or
intraocular surgery.
Sedation might be recommended for more
accurate measurement of the pressure.

TONOMETRY/H.KangariO.D. 39
Ask the patient to look
down and gently rest
the tip of the fingers
on the center of the
upperlid

TONOMETRY/H.KangariO.D. 40
Indent the globe
slightly with one finger
tip.
The opposite fingertip
will rebound slightly

TONOMETRY/H.KangariO.D. 41
NON-CONTACT TONOMETER

TONOMETRY/H.KangariO.D. 42

TONOMETRY/H.KangariO.D. 43

TONOMETRY/H.KangariO.D. 44