HIV 2 Verification on Bio Rad Geenius Accuracy: We were going to test 3 Wadsworth samples and 3 spiked samples (total 6). Sample type- Serum, PC and NC passed Spike process: Let all samples thaw. Take 3ul non-reactive serum sample (Architect) and add 4ul reactive serum sample (from the Wadsworth). Vortex for 30 secs. Use it for testing.
Results Sample Collection date Geenius test date Wadsworth result KRL result Spiked result- KRL VL103 2/15/2018 4/2/2024 HIV pos untypable HIV2 pos with HIV1 cross reactivity HIV2 pos with HIV1 cross reactivity VL053 7/1/2021 4/2/2024 HIV2 pos with HIV1 cross reactivity HIV2 pos with HIV1 cross reactivity HIV2 pos with HIV1 cross reactivity VL042 12/14/2023 4/2/2024 HIV2 pos HIV2 pos HIV2 pos