Keratometry

69,008 views 41 slides Sep 05, 2018
Slide 1
Slide 1 of 41
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41

About This Presentation

Definition, Principle, Types, Parts, Optical system, Procedure , Instrument caliberation, Patient adjustment, Mires, Doublingechanism, Astigmatism, K values, Interpretations Automated keratometers, Bausch and Lomb Keratometer, Uses, Limitations , Source of error of Keratometer IOL power calculatio...


Slide Content

KERATOMETRY DR.ANKITA MAHAPATRA 1 ST YR PG,OPHTHALMOLOGY VIMSAR,BURLA

Keratometry ( Ophthalmometry ) Kerato = Cornea Metry = Measurement

definition Keratometry is measurement of curvature of the anterior surface of cornea across a fixed chord length,usually 2-3 mm, which lies within the optical spherical zone of cornea .

principle Anterior surface of cornea – CONVEX MIRROR From Image Size formed by anterior surface of cornea (1st Purkinje image) – radius of curvature of cornea can be calculated Greater the curvature of cornea, lesser is the image size.

For a convex mirror, M= i/o = v/u i - image size o - object size v - image distance from the mirror u - object distance from the mirror When object at infinity, i/o = f/ u i/o = r/2u So, R = 2ui/o i o v u u is constant for any instrument

Relationship b/w Radius of curvature and Diopteric Power of cornea D = n-1/r r=1.3375-1/D mm D – Dioptric power of cornea n – Index of refraction of cornea (1.3375) r – Radius of cornea in metres

Doubling Principle

Biprisms introduced into the optical system so that 2 images are formed . Lower edge of one image coincides with upper edge of the other. If eye moves – both moves The prism is moved until the images touch each other. Depending on the position of prism – if distance↓, doubling ↑

Eg. Bausch and Lomb keratometer Eg. Javal - Schiotz keratometer Keratometer is based on 2 concepts:

keratometers Ziess Oberkochen Keratometer

Humphrey Keratometer Javal Schiotz Keratometer

Bausch & Lomb Keratometer

parts

Optical system of keratometer

This image becomes the object for the remainder of optical system.

Beyond the diaphragm are two doubling prisms , one with its base up & other with its base out . Prisms can be moved independently, parallel to the central axis of instrument

Light passing through left aperture of diaphragm is made to deviate above the central optical axis by a base-up prism

Light passing through right aperture is deviated by base –out prism , placing the second image to the right of the central axis

Light passing through upper & lower apertures does not pass through either prism & an image is produced on the axis

Total area of upper & = Area of each of lower apertures the other two apertures Therefore, brightness of the images is equal. Upper and lower apertures also act as Scheiner’s disc doubling the central image, whenever the instrument is not focused precisely on central mire image. Thus, image-doubling mechanism is unique in Bausch and Lomb keratometer , in that double images are produced side by side as well as at 90 from each other.

This allows the measurement of the power of cornea in two meridia , without rotating the instrument. Therefore, it is also known as ‘one-position keratometer ’

PROCEDURE INSTRUMENT CALIBERATION Instrument is calibrated before use White paper held in front of objective lens & a black line is focused sharply on it Keratometer is then calibrated with steel balls Steel ball of known radius of curvature is placed before keratometer & its value is set on the scale or dial

2 .PATIENT ADJUSTMENT: Seated in front of the instrument. Chin on chin rest & head against head rest. Eye not being examined is covered with occluder . Chin raised or lowered till patient’s pupil & projective knob are at the same level. Mires are focused by clockwise & anticlockwise movement of eyepiece through trial & error When mires are in focus, the calibration is complete.

Instruct the patient to : Keep eyes open wide and blink normally. Try not to move the head nor speak. Look at the reflection of own eye in the keratometer barrel

3. Focusing of mires : Mire is focused in the centre of cornea Patient’s view of mire The central image is doubled, indicating that instrument is not correctly focused on the corneal image of the mire.

Bausch & Lomb Keratometer Range – 36.00 to 52.00 D Normal values – 44.00 to 45.00 D To increase the range – Place +1.25 D lens in front of aperture to extend range to 61 D ADD 9 D Place -1.00 D lens in front of aperture to extend range to 30D SUBTRACT 6 D

Automated keratometers Focuses reflected corneal image on to an electronic photosensitive device, which instantly records the size & computes the radius of curvature. Target mires are illuminated with infrared light, & an infrared photodetector is used. ADVANTAGES: Compact device Very short time consuming Comparatively easy to operate

No doubling device is needed. Measures angle size in many meridians so it computes angle as well as power in many meridians. Absence of annoying glare of brightly illuminated mires. Do not calculate clarity of cornea

Clinical uses of keratometers 1.Helps in measurement of corneal astigmatic error 2.Helps to estimate radius of curvature of the anterior surface of cornea  Use in contact lens fitting. 3. Monitors shape of the cornea  Keratoconus Keratoglobus 4.Assess refractive error in cases of hazy media . 5. IOL power calculation . 6.To monitor pre- & post-surgical astigmatism . 7.Used for differential diagnosis of axial versus curvatural anisometropia .

Limitations of keratometry Measurements of keratometer based on false assumption that cornea is a symmetrical spherical or spherocylindrical structure,with 2 principal meridia separated from each other by 90 Measures refractive status of small central cornea (3-4 mm) Loses accuracy when measuring very flat or very steep cornea Small corneal irregularities preclude use of keratometer due to irregular astigmatism. One-position instruments assume regular astigmatism. Distance to focal point is approximated by distance to image.

Sources of error in keratometry