Mirena evo insertion

azaki5 3,240 views 27 slides Mar 25, 2014
Slide 1
Slide 1 of 27
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27

About This Presentation

CONTRACEPTION- MENORRHAGIA


Slide Content

Mirena IUCD, new inserter AHMED ZAKI, MD, MRCOG Head of OB/GYN Department Chairman of Scientific Committee Bakhsh Hospital

Contraceptive museum

Dr. Lippe in Yr.2000 at world congress

First generatin I.U.C.D.is Lipes loop Second generation I.U..C.D.is Copper T-200 Third generation I.U.C.D.is Gyne -T 380 With 4 years Intra uterine life

. 4 th generation IUCD

Mirena IUCD Menorrhgia Contraception HRT Endometriosis Reproductive benefits

Mirena ® old inserter Open the sterile package completely. Release the threads. Rotate the inserter so that the centimeter scale is upwards (Fig. 1a). Check that the arms of the system are in horizontal position (Fig. 1b). Holding the slider in the furthest position, pull on the threads to draw the IUS system into the insertion tube (Fig. 2a). If the knobs at the ends of the arms do not meet properly (Fig. 2b), release the arms by pulling the slider back to the mark. Re-load the product by aligning the open arms on a sterile surface (Figure 1a). Return the slider to its furthermost position and pull on both threads. Check for proper loading (Fig. 2b). Fix the threads in the cleft at the near end of the inserter shaft (Fig. 3). Figure 1a Figure 1b Figure 2a Figure 2b Knobs at the end of the arms Figure 2b

Why changes????

Packaging Size of secondary package smaller Evo Inserter ™ is positioned horizontally in the primary package Primary package contains a sticker

EVO Inserter Threads are released when the slider is pulled downwards and past the mark The IUS system is already in the horizontal position in the package. No need to check the horizontality. New shape of the slider Scale on both sides of the insertion tube Loading of the IUS system is achieved by only pushing the slider upwards. The removal threads are thinner and softer, and are inside the inserter handle: no handling of threads required for the loading of the IUS system.

Insertion procedure Less preparatory steps: simplified loading by pushing the slider forward to its furthest position Insertion technique remains the same as before!

Mirena ® Evo Inserter Open the sterile package completely (Figure 1). Push the slider forward in the direction of the arrow to the furthest position to load Mirena ® into the insertion tube (Figure 2). Figure 1 Figure 2

(Click the picture to see video) Preparatory steps Loading of Mirena ® : note the arrow in the slider which indicates the direction of movement

What has changed? Dimensions Mirena ® old inserter Mirena ® Evo Inserter™ Insertion tube cross-section (mm) 4.8 4.4 Removal thread lenght (cm) 40 27 Removal thread thickness (mm) 0.30 0.27

Modified knobs to form a smooth sphere when loaded New Old T-body of Mirena ®

What is different about this inserter?  threads are now inside the inserter handle, eliminating the handling of threads  p reloaded in the horizontal position  The arms of the LNG-IUS can be loaded into the inserter using a one-step technique (slider pushed forward)  The diameter of the insertion tube, reduced from 4.8mm to 4.4mm  There is now a centimetre scale on both sides of the insertion tube  The LNG-IUS cannot be reloaded into the inserter if it is released prematurely (important to avoid pulling the slider down before the inserter is correctly positioned in the uterus)

To make it easy to insert in the correct position

GOLDEN RULE IF YOU USE EXCESSIVE FORCE TO DO ANYTHING, YOU ARE DOING SOMETHING WRONG

CONCLUSION Mirena I.U.C.D., if properly inserted with aseptic precautions, by an experienced gynecologist, chances of its expulsion, failure and misplacement are rare and few. Some patients experience pain during insertion as the device is relatively bulky Doctors must follow the instructions of application

THANK YOU From: AHMED ZAKI To: ALL OF YOU