Multi-slice CT
Developed in late
1990’s
Multiple rows of
detectors
Detector array
segmented in z-
direction
Simultaneous
acquisition of
multiple slices
Single Slice vs. Multislice Detector
Z-Axis
Collimated Beam Thickness
Single slice
detector
Multislice
detector
Multi-Slice Detectors
Many detectors going around patient
Many detector rows in z-direction
Slice thickness determined by
Collimation
electronic detector selection
“Z” Direction
Single Multi
Distribution of detectors in axial direction
varies with manufacturer
All detectors
same width
“Z”
Direction
Variable widths
Multi-slice CT
Uniform Detector Thickness
Multiple detectors in axial
direction
Size must accommodate
thinnest slice
Detector signals can be
used
Individually
In groups
1234
Four thin slices
1 2 3 4
Four thicker slices
“z” direction
Detectors vs. Channels
# Physical Detectors not necessarily equal to #
of possible Slices
Maximum # slices limited by Digital
Acquisition System (DAS) channels
Electronic counters
Imaging speed bottleneck
How fast data can be received from
detector arrays
Detectors vs. Channels Example
16 detectors
4 channels
Multi-Slice Detector Example
16 Detector Rows, 4 Channels
Detectors vs. Channels
4 X 1.25 mm
Beam collimated to 4
detector rows
1 detector row per
DAS channel
Effective
Detector
Detectors vs. Channels
4 X 2.5 mm
Beam collimated to 8
detector rows
2 detector rows per
DAS channel
Effective
Detector
Detectors vs. Channels
4 X 3.75 mm
Beam collimated to 12
detector rows
3 detector rows per
DAS channel
Effective
Detector
Detectors vs. Channels
4 X 5 mm
Beam collimated to 16
detector rows
4 detector rows per
DAS channel
Effective
Detector
Capture Efficiency
Fraction of detector area
that is active detector
Equal-width Detectors
Disadvantage
Many gaps
Gaps are dead space
Reduce capture
efficiency
Multi-slice CT
“Adaptive Array Detectors”
Some scanners
use detectors of
various widths
Post-collimators
used to partially
block wider
elements for
thinner slices
“z” direction
123
Three thicker slices
124
Four thinner slices
3
Post-
collimators
Variable Width Detectors
Center
detectors
thinner
Thicker
detectors can
function as
thinner ones
using
collimation
Thinner
detectors can
function a
thicker one by
combining
signals
Single Slice Pitch Definition
table motion during one rotation
Slice Pitch = ---------------------------------------
slice thickness
Beam Pitch
Defined only for Multi-slice scanners
table motion during one rotation
Beam Pitch = ---------------------------------------
Beam thickness
Beam thickness
Beam Pitch
Defined only for Multi-slice scanners
Beam Pitch = 1Beam Pitch > 1
Example
table motion during one rotation
Beam Pitch = ---------------------------------------
Beam thickness
5 mm slices
4 simultaneous
slices
Beam pitch = 1
1 revolution / sec.
Beam thickness?
Table speed?
Beam Thickness
table motion during one rotation
Beam Pitch = ---------------------------------------
Beam thickness
5 mm slices
4 simultaneous
slices
Beam pitch = 1
1 revolution / sec.
Beam Thickness = 5 X 4 = 20 mm
Table Speed
table motion during one rotation
Beam Pitch = ---------------------------------------
Beam thickness
5 mm slices
4 simultaneous
slices
Beam pitch = 1
1 revolution / sec.
20 mm beam
thickness
Table speed = 20 mm rotation (1 sec) = 20 mm / sec
Slice Thickness Defined at
Rotational Center
Tube
Rotational
Center
Detector Field must be Larger than Slice
Thickness at Rotational Center
Rotational
Center
Beam Divergence More of a Problem for
Multi-Slice
Rays diverge
No longer
essentially parallel
Leads to Cone Angle
Artifact
Significant for 16, 32,
64 … data channels
Requires use of
special reconstruction
algorithms to
compensate
Multislice CT Doses
Can be 10-30% higher than for single slice units
(ICRP #47)
Cause
Divergent beam
Other considerations
Tendency to cover more volume (anatomy)
Better availability of equipment
Other Reasons for High CT Doses
Repeat Exams
No adjustment of technique factors for different
size patients
No adjustment for different areas of body
Multislice CT Advantage?
Speed!
Single slice / Multislice Images about the
same!
Multi-slice CT Imaging
Clinical Advantages
1.Thinner slices for improved z-direction resolution
2.Same acquisition in shorter time
3.Scan larger volumes in same time
Multi-slice CT Imaging
Clinical Advantages
Thinner slices
Improvement in CTA of neck, aorta, renal
vessels
Better reconstructions
Sagittal, coronal, oblique
3-D
Fundamental Trade-off
“z” axis resolution vs. image noise
Multi-slice CT Imaging
Clinical Advantages
Improved x-ray tube utilization
Reduced x-ray tube loading
4 slices acquired with same tube loading previously
used for 1
Less need to pause of tube cooling
Reduced wear & tear
Other anticipated benefits
CT endoscopy
Diagnosis of pulmonary embolism
Multi-slice CT Imaging
Clinical Advantage: Angiography
Simplifies contrast bolus timing
Continuous observation of target vessel
Can reduce amount of contrast required
Coverage from aorta to lower extremities
Runoff
Continuous CT Imaging
Interventional Procedures
Biopsy & drainage
Neuro
Chest
Abdominal
Spine
Catheter and tube placement
Helps operator avoid critical structures near
path of biopsy needle
Better visualizing of moving structures
Respiration
Functional CT
Brain perfusion
Implications of 256+ Slices
Full organ coverage in single rotation
12-16 cm coverage
Improved temporal resolution
Reduced artifacts
Whole-organ function (perfusion) studies
Functional data perfectly registered to anatomic
data
The Future
More slices
Flat panel area detectors
???
Multi-slice challenges: More Slices
Computer issues
More archival capacity
Requires faster computer systems
Requires faster communications for remote
viewing
Radiologist responsible for all images
Acknowledgement
Many drawings obtained from
www.impactscan.org website