pH Glass Electrode

Uses of the Glass Electrode
•pH measurement
•analyze foods
•Cosmetics
•environmental regulations
•soil acidity
It is fast responding and functions well in physiological systems
(Biological Systems).

Construction of pH Glass Electrode
Ag/AgCl
reference
electrode
Glass pH
electrode
+ -

Two Electrode System

The Gel Layer
•The glass has a lithium
silicate skeleton that forms
a thin hydrated layer on
both sides of the
membrane.
•Ions can penetrate this thin
layer and alter the
electrochemical potential.
•Without the hydrated layer
no pH measurements would
be possible
•The structure of the glass
has been optimised so that
virtually only H
+
ions can
enter the gel layer

Principle
•H
+
ion selective electrode
•Nernst Equation
•the electrode measures
the activity of hydrogen
ions and that pH is the
negative log of the
activity of hydrogen ions
•The Nernst equation
predicts that for every
decade change in H
+
ion
activity (1 unit change in
pH) the potential will
change by about 60 mV
( )
C25 @ pH 0591.0
log
303.2
o0
0
+=
-=
+
EE
a
nF
RT
EE
H

THE END