Qualification Of Autoclave

11,642 views 42 slides Dec 24, 2020
Slide 1
Slide 1 of 42
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42

About This Presentation

RCPIPER, SHIRPUR


Slide Content

Qualification of Autoclave 1 Guided By Dr.P.P.Nerker sir Assistant professor (Pharmaceutical Quality Assurance) Presented By Vijay d. Lonkar 1 st year M-pharm (sem-II) (Pharmaceutical Quality Assurance) R.C.Patel Institute of Pharmaceutical Education and Reaserch,Shirpur .

22/10/2020 2

Content… Introduction Need and importance Types of Qualification Risk management of Autoclave Operating an Autoclave Unloading an Autoclave Tests of Qualification of Autoclave Conclusion References 22/10/2020 3

What is Qualification.. Action of proving and documenting that equipment or ancillary systems are properly installed, work correctly, and actually lead to the expected results. Qualification is part of validation 22/10/2020 4

Introduction Definition of Autoclave An autoclave is a pressure chamber used to carry out industrial processes requiring elevated temperature and pressure different to ambient air pressure. Autoclave are used in medical application to perform sterilization, and in the chemical industry to cure coatings, vulcanize rubber and for hydrothermal synthesis. 22/10/2020 5

22/10/2020 6 Definition of Sterilization Sterilization can be defined as any process that effectively kills or eliminates transmissible agents (such as fungi, bacteria, viruses) from surface, equipment, foods, medications or biological culture medium. A sterility assurance level (SAL) of 10-6 means that there is less than or equal to one chance in million that particular item is contaminated or unsterile following sterilization process.

22/10/2020 7 The standard temperature and pressure of an autoclave Processes conducted at high temperatures for short time periods are preferred over lower temperatures for longer times. Some standard temperatures/ pressures employed are 115 °C/10 psi, 121 °C/ 15 psi, and 132 °C/27psi. (psi=pounds per square inch). In university autoclave, autoclaving generally involves heating in saturated steam under a pressure of approximately 15 psi, to achieve a chamber temperature of a least 121°C (250°F) but in other applications in industry For example, other combinations of time and temperature are sometimes used.

Need and importance The proper sterilization of medical devices, surgical instruments, supplies and equipment utilized in direct patient care and surgery Sterilization is very important in the medical industry. Without sterilization, infections would fly around and thousands of lives would be lost. Sterilization helps to prevent the development and spread of infection. In the pharmaceutical industry 22/10/2020 8

22/10/2020 9

Basic Qualification Approach User requirement specification (URS) User or customer of equipment has certain expectation about the equipment which wants to use. These expectations are generally in the form of his requirements. It is called as user requirement specifications. Example: Autoclave 22/10/2020 10

Types of Qualification .. 22/10/2020 11 Operational Qualification

22/10/2020 12 Design qualification may verify that design of equipment, system/facility is according to requirement of user and current good manufacturing practices. Design qualification (DQ) defines the functional and operational specifications of the instrument and details for the conscious decisions in the selection of the supplier. The steps that should be considered for inclusion in a design qualification of the autoclave:- Description of the analysis problem Description of the intended use of the equipment Description of the intended environment Design Qualification:-

22/10/2020 13 Continue... 4)Instrument tests(if the technique is new) 5)Preliminary selection of the suppliers 6)Final selection of the equipment 7)Final selection of the supplier and equipment The first qualification stage of an autoclave starts as soon as you formulate and document your requirements for the device to be purchased. This means you have already created your User Requirement Specification (URS). User Requirement Technical Qualification Report

22/10/2020 14 “Installation qualification establishes that the instrument is received as designed and specified, that it is properly installed in the selected environment, and that this environment is suitable for the operation and use of the instrument.” The qualification involves the coordinated efforts of The vendor. The operating department. The project team (which provide input into the purchase, installation, operation and maintenance of the equipment). Installation Qualification (IQ):-

22/10/2020 15 Installation qualification for Autoclave:- The IQ procedure is used to demonstrate that the autoclave as installed meets all specifications, is installed properly and that the supporting programme needed for ongoing operations(Ex:-SOPs) are in place.

22/10/2020 16 The Installation Qualification involves following checklist.. Mechanical equipment specifications (chamber,valves,traps,filters,regulators,vaccumpump,heat exchanger,condenser etc) Control and instrumentation specification (printer recorder, control valves,pressure and temp. Transmitter, résistance temperature device,switches,level sensors,interlock etc) Drawing verification Construction materials Change part Instrument calibration

22/10/2020 17 Installation qualification(IQ) is conducted to prove that equipment/system has been installed as per user and manufacturer recommendation and verifying that all required utilities have provided safe operation of equipment/system. · Utilities specification · Drawing specification (electrical, mechanical) · Construction material in product contact · Operating and maintenance manual

22/10/2020 18 The operational qualification process is intended to demonstrate that the components are operating properly and ready for performance or load testing. Operational qualification shall be done “without load” Operational qualification (OQ):- "Operational qualification (OQ) is the process of demonstrating that an instrument will function according to its operational specification in the selected environment”. The proper operation of equipment is verified by performing the test functions specified in the protocol. A conclusion is drawn regarding the operation of equipment after the test functions are checked and all data has been analyzed.

22/10/2020 19 Operational Qualification for Autoclave OQ testing typically includes the following tests: Empty chamber temperature mapping Simulated load chamber temperature mapping and, if required verification of sterilization efficacy using biological indicators (BI) Alarm conditions and expected results

22/10/2020 20 Operational Qualification of Autoclave involves points to be checked... Check of alarms • Temperature too high or too low • Pressure too high or too low (pressure variations) • Time limits of process steps • Utility supply • Cable break of sensors Filter sterilization Saturated steam check Leak/steam penetration test Empty chamber test

22/10/2020 21 Heat distribution check of the empty chamber (Identification of cold spots):- Method:- • Repeated measurement of the empty chamber (e.g. 3 times) • Use of in minimum 10 to 12 sensors / m3 of chamber volume • One sensor should be near to the control sensor respect. near to the condensate drain • Documentation of the exact localisation of the used sensors

22/10/2020 22 " Performance Qualification (PQ) is the process of demonstrating that an instrument consistently performs according to a specification appropriate for its routine use". PQ should always be performed under conditions that are similar to routine sample analysis. PQ should be performed on a daily basis or whenever the equipment is being used. In practice, PQ can mean system suitability testing, where critical key system performance characteristics are measured and compared with documented. Performance Qualification (PQ):-

22/10/2020 23 Performance qualification is documented evidence to prove that equipment/system is performing under specified condition. It involve in taking trial under “loaded condition”

22/10/2020 24 Performance Qualification of Autoclave It includes test: Loaded chamber steam penetration test:- It involves determination of load items difficult to sterilize and worst case location. While performing the test load configuration is taking in to consideration.

Performance of autoclave It involve in taking trial under “loaded condition”. The calibration of instruments, apparatus carried out at suitable intervals in accordance with an established written program containing specific directions, schedules, limits for accuracy and precision. Operating range, approval standard operating procedure is used for verification. Performance qualification is integrate, procedure, personal, system and material verify pharmaceutical grade utility, and environment, equipment, and system produce required output 22/10/2020 25

22/10/2020 26 Autoclave to protect against scalds and burns: Heat-insulating gloves that provide complete coverage of hands and forearms. Closed-toed footwear. Packaging and Loading use approved autoclave bags prepare and load material to ensure steam penetration ensure all containers including bags are vented do not overfill containers (prevent spill and boil over) ensure sufficient water in load to allow steam penetration use secondary containers label all material (name, contents) ensure material is permitted to be autoclave do not mix clean and contaminated material in the same load complete “Daily Autoclave Log” do not allow bags to touch or strap sides of autoclave

22/10/2020 27 Operating an Autoclave ensure the autoclave is operating properly before commencing determine the appropriate exposure time for the load, consider the many factors effecting exposure time ensure the autoclave attains the desired temperature (normally 121oC) and pressure (minimum 15 psi) for the desired time (minimum 30 min.) record information in “Daily Autoclave Use Log” undertake weekly testing using a biological indicator B.Stearothermophilus. Record results on “Biological Test Indicator Results” form.

22/10/2020 28 Unloading the Autoclave wait until the chamber pressure gauge reads zero before opening wait 10 minutes for the contents of the autoclave to cool remove the waste in manner reduce the risk of spillage, use a trolley verify temperature and duration of exposure has been met

22/10/2020 29 Risk management:- Autoclaves are inspected every 3 months and certified annually by spore strips. The inspection, service and repair records are maintained in the lab. The name of the person responsible for the autoclave shall be posted near the autoclave. An authorized training session must be successfully completed by users prior to use of autoclaves. It is the supervisor's responsibility to ensure employees are trained before operating any autoclave unit. Procedural and instructional documents must be followed. Personal protective clothing and equipment must be worn when loading and unloading the autoclave.

The different tests are follows for qualification of autoclave are 22/10/2020 30 Vacuum leak test

Vacuum Leak Test Objective: To verify the leakage in sterilization chamber during vacuum hold time when the sterilizing chamber is empty. Principle: These tests are designed to show that the sterilizer chamber does not leak in empty chamber. Leakage of air into the chamber is not acceptable for two reasons: 22/10/2020 31

22/10/2020 32 Continue.. 1. The presence of air inhibits penetration of the load by the sterilant (Steam) and prevents sterilization. 2. Air leaking into the chamber during the sterilization and drying cycle are not passed through the bacteria retentive filter, and therefore there is a risk of contamination of the load. The test is performed by measuring the change in vacuum in the chamber when all the valves leading to it have been closed and the vacuum source is isolated Acceptance criteria: Vacuum leak rate should be NMT 0.013 bar / 10 minutes. Frequency: Run the test for 3 consecutive cycles at the time of Initial Qualification.

Bowie-Dick Test Objective: To ensure that the vacuum pulses applied before the sterilization hold period are sufficient to remove the entrapped air or non-condensable gases so as to facilitate the event and rapid steam penetration into all parts of load and maintaining this condition during sterilization holding time. 22/10/2020 33

22/10/2020 34 Principle: Sterilization is achieved by the rapid and even penetration of steam into all parts of the load and the maintenance of these conditions for the specified holding time. To ensure this, it is essential to remove air from the chamber and load, and to provide a steam supply which contains a minimal volume of Non-condensable gases. The Bowie Dick test shows whether or not steam penetration is taking place by testing the presence of Non condensable gases in the chamber, but it does not confirm that the sterilization condition in the load is achieved or not.

22/10/2020 35 Set Parameters: 17 minutes cycle at a temperature of 121°C Place of keeping Bowie -Dick Indicator: Place the Bowie dick indicator approximately 100 mm to 200 mm above the sterilization chamber base. Acceptance criteria: The Bowie dick indicator should show uniform color change (Yellow to Brown / black) after the cycle. No change or no uniform change or air entrapment (bubbles) spot on the test sheet indicates inadequate air removal from the sterilization base chamber. Frequency: Run the test for 3 consecutive cycles at the time of Initial Qualification.

Heat distribution Study Objective: To verify the temperature uniformity throughout the chamber and to locate the cold spot in Empty Chamber. The sterilizer is capable attaining a temperature of 121° C throughout the sterilizing hold period in Empty Chamber. 22/10/2020 36

22/10/2020 37 Acceptance criteria: Temperature distribution within the chamber must be between 121°C to 123°C at all location during the sterilization period (dwell time) There should not be any slowest heating point (cold spot) in the autoclave chamber and equilibrium time should not be more than 30 second.

Heat Penetration Study Objective In order to verify sterilizing temperature has been reached in each load subjected to moist heat sterilization, it is necessary to conduct heat penetration studies. This study is conducted to ensure that the coolest unit within a pre-defined loading pattern (including minimum and maximum loads) will consistently be exposed to sufficient heat lethality (minimum "F0"). 22/10/2020 38

22/10/2020 39 Acceptance criteria Temperature distribution within the chamber must be between 121°C to 123°C at all location during the sterilization period (dwell time) Sterilization temperature should be maintained for NLT 15 minute for minimum 10 thermo couple during hold time. Biological indicator ( Geobacillus Stearothermophillus ) should show complete sterilization (i.e. no growth after incubation)

22/10/2020 40 Conclusion Qualification is a fundamental concept of cGMP. Where autoclave is used for sterilization of the garments, cleaning aids filters, utensil, vial filling machine parts, rubber stopper etc This was followed by performing the qualification of the equipment which describes the entire test right from vacuum leak test, bowie-dick test, heat distribution study (empty chamber, loaded chamber) and heat penetration test. All the parameters and processes which are described were found within the acceptance criteria. Hence autoclave is considered to be qualified and can be routinely used.

22/10/2020 41 References:- http://www.validation-online.net/processqualification.html https://www.validation-online.net/autoclave-validation.html Ramesh Sawant and Sandip Hapse,Fundamentals of Quality Assurance Techniques,Text and Reference book,Career Publications,2 nd edition:May 2016,Page No. 194-198 B.T.Loftus & R.A.Nash,”Pharmaceutical Process Validation”, Drugs and Pharm sci.Series,Vol.129,3 rd edition,Page No. 445-452 Manohar A.Potdar,’’Pharmaceutical Quality Assurance’’,Nirali Prakashan,5 th edition:Aug 2016,Page NO. 8.13-8.19

22/10/2020 42
Tags