Quality assurance in ultrasound machines

AhmadSanusiNalado 66 views 39 slides May 15, 2025
Slide 1
Slide 1 of 39
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39

About This Presentation

Quality assurance in ultrasound


Slide Content

Quality Assurance of Quality Assurance of
UltrasoundUltrasound

IntroductionIntroduction

Ultrasound is principally based Ultrasound is principally based
on reflection of very high sound on reflection of very high sound
frequencies (ultra sound). frequencies (ultra sound).

Acoustic energy is sent into the Acoustic energy is sent into the
body and the reflected energy is body and the reflected energy is
received and evaluated at the received and evaluated at the
point of transmission. point of transmission.

Ultrasound (U/S) Ultrasound (U/S)
Physical conceptPhysical concept

Ultrasound is a longitudinal plane Ultrasound is a longitudinal plane
wave.wave.

The wave can be displayed as a The wave can be displayed as a
sinusoid described by sinusoid described by
•AmplitudeAmplitude
•wavelength wavelength
•velocity of propagationvelocity of propagation

What is quality assurance in
ultrasound?

The purpose of a quality assurance
(QA) program in medical ultrasound
is to ensure that the equipment
and clinical practices of a facility
are safe, and that the
information obtained from a
clinical ultrasound procedure is
as accurate as can be determined

FDAFDA

The FDAThe FDA regulates all ultrasound regulates all ultrasound
equipment energies (for obstetrical equipment energies (for obstetrical
scanning this is below 100 milliwatts scanning this is below 100 milliwatts
per centimeter squared). Therefore, per centimeter squared). Therefore,
the manufactures of the equipment the manufactures of the equipment
are required and have designed their are required and have designed their
equipment in accordance to these equipment in accordance to these
FDA mandated levels.FDA mandated levels.

What are the 3 major
components of ultrasound
machine?

Any ultrasound system has three
basic components: a transducer, or
probe; the processing unit,
including the controls; and the
display

Phantoms used in ultrasound Phantoms used in ultrasound
equipmentequipment

Tissue-mimicking phantomTissue-mimicking phantom

Phantom designPhantom design

Phantom desiccationPhantom desiccation

MaterialsMaterials

The commercial phantom was made The commercial phantom was made
of a material (Zerdine,solid elastic of a material (Zerdine,solid elastic
waterbased polymer) which mimics

waterbased polymer) which mimics

the average acousticproperties of the average acousticproperties of
soft tissue: average speed of sound soft tissue: average speed of sound
(1540 m/s) andattenuation either (1540 m/s) andattenuation either
0.5 or 0.7 dB cm−1MHz−1)0.5 or 0.7 dB cm−1MHz−1)

Tissue-Mimicking PhantomTissue-Mimicking Phantom

Is a device that mimics soft tissues Is a device that mimics soft tissues
in its ultrasound transmission in its ultrasound transmission
characteristics, e.g. speed, characteristics, e.g. speed,
attenuation, echogenicity etc.. attenuation, echogenicity etc..

Phantoms have targets in known Phantoms have targets in known
positions for image comparisons positions for image comparisons
e.g. simulated cysts, echogenic e.g. simulated cysts, echogenic
structures and thin targetsstructures and thin targets

Tissue mimicking phantom

Phantom designPhantom design

The tissue mimicking material consist The tissue mimicking material consist
of a water-based gelatin mixed with of a water-based gelatin mixed with
microscopic graphite particlesmicroscopic graphite particles

Small scatterers are distributed Small scatterers are distributed
throughout the tissue mimicking throughout the tissue mimicking
materialmaterial

Many phantoms have simulated cysts Many phantoms have simulated cysts
and discrete reflectors to measure and discrete reflectors to measure
distancedistance

Phantom desiccationPhantom desiccation

Water-based gels loose water and shrinkWater-based gels loose water and shrink

Air might enter thus rendering it Air might enter thus rendering it
ineffective ineffective

Therefore, keep in a humid air-tight Therefore, keep in a humid air-tight
containercontainer

Rubber-based phantom solve the Rubber-based phantom solve the
problem partiallyproblem partially

Can be stored in an open environmentCan be stored in an open environment

..
..
Rubber-based tissue mimicking phantom

Quality Assurance & Quality Control Quality Assurance & Quality Control
TestsTests
1.1.Transducer choiceTransducer choice
2.2.System sensitivitySystem sensitivity
3.3.Photography and gray-scale hard copyPhotography and gray-scale hard copy
4.4.Routine image checksRoutine image checks
5.5.Scan uniformityScan uniformity
6.6.Distance measurementDistance measurement
7.7.spatial resolutionspatial resolution
8.8.Doppler testingDoppler testing
9.9.Doppler flow phantomDoppler flow phantom
10.10.DocumentationDocumentation

1-Transducer choice1-Transducer choice

Standardized transducer is Standardized transducer is
selected selected

Record its identificationsRecord its identifications

Future tests must be done with Future tests must be done with
the same transducerthe same transducer

2-System sensitivity2-System sensitivity

It refers to echo signal level that can be It refers to echo signal level that can be
detecteddetected

Achieved by controls that vary output and Achieved by controls that vary output and
amplify the received signals (gain)amplify the received signals (gain)

The maximum sensitivity is limited by The maximum sensitivity is limited by
electrical noise (e.g. from computer electrical noise (e.g. from computer
terminals)terminals)

Variations may be due to transducer Variations may be due to transducer
damaged cables or electronic driftdamaged cables or electronic drift

Commonly used in measuring maximum Commonly used in measuring maximum
depth of visualization depth of visualization

Results should be within 1 cmResults should be within 1 cm

Maximum depth of visualization
A- max. depth is 15.2 cm (4MHz)
B- max. depth is 16.8 cm (3.5 MHz)
C- max. depth cannot be determined
(2.5 MHz) because it is good all the
way down to the bottom of the
phantom

3-Photography and gray scale hard 3-Photography and gray scale hard
copycopy

A major acceptance testA major acceptance test

The display monitor should be set up The display monitor should be set up

Multiformat camera should be Multiformat camera should be
adjustedadjusted

Image display settings should not be Image display settings should not be
shifted routinelyshifted routinely

To avoid casuals, remove the knobsTo avoid casuals, remove the knobs

stepssteps

Set the contrast settings initially to Set the contrast settings initially to
the minimum and adjust the the minimum and adjust the
brightness brightness

Adjust the monitor contrastAdjust the monitor contrast

Adjust the image recording device to Adjust the image recording device to
obtain the same gray scale that obtain the same gray scale that
appear on the monitorappear on the monitor

Compare with the referenceCompare with the reference

4-Routine image check4-Routine image check

Should be performed weeklyShould be performed weekly

To ascertain that the brightness To ascertain that the brightness
variations are recordedvariations are recorded

The gray bar pattern should appear The gray bar pattern should appear
on the edge on the edge

Verify with the image taken recently Verify with the image taken recently

5-Scan Image Uniformity5-Scan Image Uniformity

If the unit gain settings are If the unit gain settings are
adjusted properly the image adjusted properly the image
should be uniform should be uniform

Deviation may be due to:Deviation may be due to:
1.1.Bad probe elementsBad probe elements
2.2.Inadequate compensation Inadequate compensation
3.3.Inadequate blending of pixel Inadequate blending of pixel
datadata

What is uniformity in
ultrasound?

The image uniformity is defined as
the equipment ability to give
ultrasound echos with the same
amplitude and deepness when
the brightness is fixed. This
quality test is also important to
verify if the transducer crystals have
a well functioning

....
Image uniformity tests
Good uniformity
Problematic transducer

6-Distance measurement6-Distance measurement

Done periodically Done periodically

Distance indicator include 1-cm Distance indicator include 1-cm
depth markerdepth marker

Two measurements are executed:Two measurements are executed:
vertical and horizontalvertical and horizontal

Vertical distanceVertical distance
Steps:Steps:

Ensure that the reflectors are clearly Ensure that the reflectors are clearly
imaged imaged

Use the caliper to measure the Use the caliper to measure the
distance between each two distance between each two
reflectorsreflectors

Accuracy must be within 1 mmAccuracy must be within 1 mm

..
..
Vertical distance measurement. The caliper reading 9.9 mm
compare to 10 mm

Horizontal DistanceHorizontal Distance

Similar to the vertical accuracy Similar to the vertical accuracy

Should agree with the Should agree with the
manufacturers i.e. within 3 mmmanufacturers i.e. within 3 mm

..
..
Horizontal distance measurements. The caliper reading is
30.5 mm compare to the actual 30 mm

7-Spatial Resolution Tests:7-Spatial Resolution Tests:

Usually an acceptance testUsually an acceptance test

It allows intercompensations of It allows intercompensations of
scannerscanner

It involve axial resolution, lateral It involve axial resolution, lateral
resolution and focal lesion resolution and focal lesion
resolutionresolution

Axial resolutionAxial resolution

It is a measure of how close two It is a measure of how close two
reflectors are and still be resolved reflectors are and still be resolved
as separate reflectorsas separate reflectors

The axial resolution in the following The axial resolution in the following
phantom, are 2 mm 1 mm 0.25 mmphantom, are 2 mm 1 mm 0.25 mm

Comparison must be made wit the Comparison must be made wit the
reference reference

..
..
Axial resolution measurement

Lateral resolution:Lateral resolution:

Is a measure of how close two Is a measure of how close two
reflectors can be to one another reflectors can be to one another
perpendicular to the beam axis perpendicular to the beam axis
and still be distinguished as and still be distinguished as
separate reflectors on an imageseparate reflectors on an image

Just like axial resolution test it Just like axial resolution test it
depends on the power and gaindepends on the power and gain

..
..
Lateral resolution measurements
The horizontal size of the pin target is 1.2 mm

8-Doppler Testing8-Doppler Testing
String testString test::

This is a thin string wound This is a thin string wound
around a pulley and driven by a around a pulley and driven by a
motormotor

The string is echogenicThe string is echogenic

It provides evaluation of velocity It provides evaluation of velocity
measurement of Doppler devicemeasurement of Doppler device

A target of a precise velocity can A target of a precise velocity can
be detected be detected

9-Doppler Flow Phantom9-Doppler Flow Phantom

Consist of one or more hollow Consist of one or more hollow
tube inserted in tissue-mimicking tube inserted in tissue-mimicking
materialmaterial

A blood-mimicking fluid (made of A blood-mimicking fluid (made of
water and glycerol with plastic water and glycerol with plastic
particles ) is pumped inside the particles ) is pumped inside the
tube to simulate blood flowtube to simulate blood flow

The test are:The test are:

..

The maximum depth at which flow The maximum depth at which flow
waveform can be detectedwaveform can be detected

Alignment of the pulse Doppler Alignment of the pulse Doppler
sample volumesample volume

Volume flow accuracyVolume flow accuracy

Velocity accuracyVelocity accuracy

Color flow indicationColor flow indication

Any changes overtime such as Any changes overtime such as
greater than 1 cm indicate a greater than 1 cm indicate a
change in sensitivity of the change in sensitivity of the
instrument instrument

..
. .
Color flow image of a Doppler flow phantom for presentation I color

10-Documentation10-Documentation

Results of tests should be kept in a log Results of tests should be kept in a log
bookbook

One transducer is selected to do the job One transducer is selected to do the job
each time (preferably the most used one)each time (preferably the most used one)

Any compensation done must be Any compensation done must be
registeredregistered


According to industry experts, on According to industry experts, on
average, ultrasound machines should average, ultrasound machines should
be replaced be replaced every 5-7 yearsevery 5-7 years
11
. .
Knowing that an ultrasound system Knowing that an ultrasound system
typically lasts 5-7 years can be typically lasts 5-7 years can be
instrumental when purchasing a new instrumental when purchasing a new
machine.machine.