report presentation on students’ industrial work experience scheme ( siwes ) 1 BY Ogar Helen Onyangbe 2020/9100 U ndertaken at asi ukpo memorial hospital calabar 1 24/07/2023
OUTLINE
BRIEF HISTORY OF SIWES The Students Industrial Work Experience Scheme (S.I.W.E.S)was established by the federal Government of Nigeria’s Industrial Training Fund (ITF) in 1973 in order to bridge the gap existing between the theoretical based learning and the practical form of learning. This is necessary to solve the problem of lack of adequate practical skills.
HISTORY OF ASIUKPO MEMORIAL HOSPITAL ASI UKPO Memorial Hospital is located in the city of Calabar. It started operations in 2013,with the sole aim of making medical services affordable and easily accessible to all. It grew into a full-fledged hospital in 2018. Fig.1 Asiukpo Memorial Hospital
Description of asi ukpo hospital HOSPITAL DEPARTMENT Radiology department Oncology H istopathology department Machine/Equipment’s MRI (Magnetic Resonance Imaging) Multi-sliced Computed Tomography scan (CT SCAN) Digital x-rays Ultrasound machine Linear accelerator Echocardiography,
ASIUKPO MEMORIAL HOSPITAL ORGANOGRAM
WORK DONE BY THE ENGINEERING TEAM Engineering morning routine. Engineering closing routine. Carrying out preventive maintenance Trouble shooting and repairs of equipment
MAINTENANCE LINEAR ACCELERATOR MAINTENANCE CT SCAN MAINTENANCE Fig. 2 Linear Accelerator Fig. 3 CT Scan Machine
repairs CT SCAN REPAIRS Problem : H ardware shutdown error message, the machine wasn’t able to beam. Cause : Improper shutdown of machine Fault : faulty contactor Solution : R eplaced contactor Fig. 5 CT Scan Machine
REPAIRS Auto Immune Assay Analyzer Problem : W asn’t able to detect fluid Cause : N on functioning pump Fault : F aulty pump Solution : Re place pump Fig. 6 Auto Immune Assay Analyzer
repairs Repair of the chemistry analyzer Problem : displayed cooling unit error message Cause : leakage of the refrigerant, disconnected tube Solution : connecting the tube, adding more refrigerant to the tube and removing the air bubble Fig. 7 Selectra Chemistry Analyzer
REPAIRS CENTRIFUGE REPAIR problem : The machine was unable to spin FAULT : Broken rotor and worn out carbon brush Solution : The rotor was welded and the carbon brush was replaced SPIROMETER REPAIR PROBLEM : Display buttons weren’t working CAUSE : Disconnection of the wires SOLUTION : Used soldering iron and lead to solder the disconnected wires
Hemo -scale Repair: Problem : the machine being unable to turn on. Fault : the diode in the power-pack was faulty Solution : the diode was replaced with a new one. BLOOD BAG TUBE SEALANT MACHINE REPAIR PROBLEM : the machine was unable to seal the blood bag tube FAULT : faulty diode SOLUTION : replaced the faulty DIODE.VV Fig. 8 Blood Bag Sealant Machine Fig. 7 Hemo -Scale
Fabrication and Installation Fabrication of water bath C ircuit diagram include a heating element, a temperature sensor, a control unit (such as a thermostat ), a power supply, and indicator lights. Design consideration I nsulation T emperature control H eating element C ontrol and monitoring system R esistance to corrosion
RECOMMENDATION I recommend incorporating interactive workshops, fostering stronger industry connections, and providing mentorship opportunities. additionally, offering specialized training sessions, enhancing communication channels, and creating a supportive learning environment can greatly enhance the program. organizing industry-led seminars and talks, fostering a culture of innovation and creativity, and offering financial assistance.
RECOMMENDATION In order to solve the problem of students not being able to find place of attachment or a good place to learn , the school should get a directory containing list of it’s friendly companies and their location to enable students to get placement easily and assign students to different work places where they will be able to acquire more knowledge on their field of study.