INTRODUCTION TO EVAR & TEVAR Dr Amrit Kumar Senior Resident Dept of CTVS
ENDOVASCULAR AORTIC REPAIR (EVAR) First Endovascular Abdominal aortic aneurysm repair done by Parodi,199 Performed by inserting graft components folded and compressed within a delivery sheath through the lumen of an access vessel, usually the common femoral artery Upon deployment, the endograft expands, contacting the Aortic wall proximally and distally to exclude diseased part Mainly Used in Treatment of - Aortic An eurysm ( an abnormal dilatation of the aorta greater than 50% of the normal proximal segment) - Aortic Dissection - Penetrating Aortic Ulcer
A dvantages over Open-surgical repair Avoidance of long incisions in the thorax or abdomen No cross-clamping of the aorta Lower incidence of visceral, renal, and spinal cord ischemia Shorter hospital stay R educed recovery time L ower complications such as blood loss and wound infections. It is more suitable for patients who are considered too high-risk for open surgery due to factors such as age or underlying health conditions.
INDICATION OF ANEURYSM REPAIR Abdominal Aortic Aneurysm Symptomatic (tenderness or abdominal or back pain, evidence for embolization, rupture) Aneurysm size ≥5.5 cm AAA that has expanded by more than 0.5 cm within a six-month interval Thoracic aortic aneurysm Size >6 cm R apidly enlarging more than 0.5 cm within a six-month interval - S ymptoms such as chest pain, and diagnosis of aortic rupture or dissection
Infrarenal – originates below the renal arteries Juxtarenal – originates at the level of the renal arteries Suprarenal – originates above the renal arteries Majority of Endovascular aneurysm repairs are performed on the I nfrarenal aneurysm
CT angiography with 3-D reconstruction preferred 2D CT- aortic diameter measurements will be overestimated if the aorta is Angulated DSA cannot evaluate the True lumen diameter, extent of thrombus, plaque, or degree of calcification MR angiography fails to depict vessel wall calcification, which has implications for vascular access Pre-Operative Imaging
Aortic neck diameter Aortic diameter at the lowest renal artery The required endograft diameter 15 to 20% more than the measured aortic neck diameter This provides sufficient radial force to prevent device migration Over-sizing lead to kinking of the device, thrombus formation or endoleak • Under sizing- inadequate seal
- A conical neck is present when the diameter of the aorta 15 mm below the lowest renal artery is ≥10 percent larger than the diameter of the aorta at the lowest renal artery Solution- Supra-renal Fixation Balloon-expandable Stent Fenestrated or Branched Endograft Conical/reverse tapered aortic neck
The distance from the lowest renal artery to the origin of the aneurysm Should be at least 13-15 mm to provide an adequate proximal landing zone for endograft fixation Aortic neck length
The angle formed between points connecting the lowest renal artery, the origin of the aneurysm, and the aortic bifurcation Ideally, the aortic neck angle should be < 60 Greater lead to difficulties in Implantation, Kinking, Endoleak , and Distal device migration Angle >60 is generally considered to be a contraindication Aortic neck angulation
Iliac artery & access vessel morphology Should have No significant Stenosis, Mural thrombus Calcification and Tortuosity CIA is a preferred distal attachment site EIA diameter of 7mm is needed to allow safe passage of the endograft delivery sheath CIA between 8 and 22 mm L ength of normal diameter CIA into which the limbs of the endograft will be fixed should be at least 15-20 mm to achieve an adequate seal In Case of Diffuse narrowing or significant calcification –an iliac conduit can be created
Proximal neck length >10mm Neck Diameter <28mm Aortic neck angulation <60 Iliac artery diameter >7mm and <15mm Minimal to moderate tortuosity No mural thrombus at attachment sites Minimal calcification No associated mesenteric occlusive disease Ideal case for EVAR
Endograft Principle -Self-expanding metallic stent framework with a high outward radial force that allows attachment to the artery wall and graft fabric to creates a new conduit for blood flow. There are significant variation in Endovascular graft design Three types of components are common to all: A delivery system Main body device Extension (limb)
delivery system Typically delivered through the femoral artery, either percutaneously or by direct surgical cutdown If the femoral artery is too small to accommodate the delivery system, access can be obtaind by suturing a synthetic graft to the iliac artery ( ie , iliac conduit) through a retroperitoneal low abdominal incision The size of the delivery system varies depending upon the device diameter
Endografts main device The main body device is usually bifurcated. Endovascular grafts rely primarily upon outward tension in the proximal graft to maintain the position of the graft Fixation system may also include barbs or a suprarenal uncovered extension Endograft Extensions Bifurcated abdominal aortic grafts require adjunctive placement of iliac artery limbs to complete the graft Iliac Limbs on the main body device vary in length depending upon weather the graft is a 2 or 3 component graft
Stent-graft design Commonaly used Graft system are AneuRx (Medtronic, Inc., Minneapolis, MN) Talent (Medtronic, Inc., Minneapolis, MN) Zenith (Cook, Inc., Bloomington,IN ) Excluder (W.L Gore and Associates, Flagstaff, AZ) Endurant (Medtronic, Inc., Minneapolis, MN) Powerlink ( Endologix , Irvine,CA )
AneuRx®device - A B ifurcated stent graft composed of a nitinol exoskeleton and polyester(Dacron) lining. It is deployed just below the renal arteries and relies on radial force to fix the device into place. Distal and proximal extension cuffs are available. The Gore Excluder - B ifurcated device, with a Nitinol exoskeleton and a PTFE graft. It has proximal barbs to anchor into the proximal infrarenal aorta. The Powerlink®system - A unibody device made of PTFE and a Cobalt chromium alloy s keleton . It has a long main body and sits on the anatomic bifurcation. The Zenith®device - The bare proximal stents allow for suprarenal fixation. The device also has barbs to allow for more secure attachment into the suprarenal aorta. It has a stainless steel exoskeleton.
A) Fenestrated Graft- Useful when Aneurysm involving the visceral vessels proximally or associated with hypogastric artery aneurysms Have openings in the fabrics of the endograft, which allow flow into the visceral arteries. Also Can be used when the proximal aortic neck is short ( ie , <10mm) Special Design Graft
B) Branched- Grafts have a separate small grafts sutured to the basic endovascular graft for deployment into a vessel to preserve flow into it. Branched grafts have been designed to accommodate the hypogastric and renal arteries.
C) Chimney grafts technique- Stent placed parallel to the aortic stent graft Used to preserve perfusion to branch vessels In the absence of available fenestrated or branched grafts, chimney grafts remain a feasible endovascular option for high risk patients.
Choice of graft The choice is based upon multiple factors, including patient anatomy, operator preference and cost Bifurcated grafts are more often, but are not appropriate for patients with unilateral severe iliac stenosis or occlusion Unilateral iliac stenosis-unibody grafts known as aorta- uni -iliac (AUI) devices are used. Iliac artery occluded followed by Distal Femoro-Femoro Bypass Grafting to maintain Distal limb perfusion.
preparation Antithrombotic therapy- Moderate to high risk for DVT. The incidence of DVT following EVAR is 5.3% spite of pharmacologic thromboprophylaxis Antibiotic prophylaxis- A first generation Cephalosporin or, in the case of penicillin allergy Vancomycin is recommended Anesthesia - can be performed under GA or local anaesthesia with conscious sedation Vascular access- Bilateral femoral access is needed – via surgical cutdown or percutaneously.
procedure Gaining vascular access Placement of arterial guidewires and sheaths Imaging to confirm aortoiliac anatomy Main body deployment Gate cannulation (bifurcated graft) Iliac limb deployment Graft ballooning Completion imaginig
Once vascular access is established and landmarks for positioning the device are obtain with angio , the main device is positioned with particular attention paid to the location of the opening for the contralateral iliac limb (“ contralateral gate”) The aortic neck is imaged, slight degree of craniocaudal and left anterior oblique angulation may improve imaging of the renal ostia With the proximal radiopaque markers of the graft positioned appropriately The body of the graft is deployed A guidewire is advanced through the contralateral access site into the contralateral gate
Once the contralateral guidewire is positioned within the main body of the endograft, the deployment of the endograft at the neck of the aneurysm is completed followed by deployment of the contralateral Once the Endograft component are in place, the attachment sites and endograft junctions are gently angioplastied with a compliant or semi compliant balloon Completion angiography is performed to evaluate the patency of the renal arteries and evaluate for endoleak Guidewire access is maintained throughout the procedure but is particularly important when removing the main graft body device sheath since disruption of the access vessels by an oversized sheath may not become apparent until after sheath has been removed.
Endoleak Type I -Persistent flow at proximal (a) or distal (b) attachment sites Type II -Retrograde flow from side branches Inferior mesenteric Lumbar arteries Type III- Graft defect Type:-IV- Graft porosity Type V- Continued aneurysm sac expansion without a demonstrable leak on any imaging modality. Endoleak is a term that describes the presence of persistent flow of blood into the aneurysm sac after device placement
Untreated type I and III endoleaks are at high risk of rupture Type II are the commonest endoleaks , affecting up to 43% of cases, associated with low (0.52%) risk of rupture and a significant rate of spontaneous closure. Treatment is required only for endoleaks that persist for more than a year in an aneurysm of increasing size. Type IV endoleak typically resolves in 24 hours. It has not been associated with any long-term adverse events and does not require any treatment.
device migration Device movement of >10mm relative to anatomic landmark Multifactorial- Aortic neck length, Angulation, Thrombus in the aortic neck Short neck and Late neck dilation after EVAR is a major cause of concern because of the potential loss of proximal fixation and seal Superstiff -Guidewire Technique: Maneuvers like bending the guidewire before introduction, in order to align it with the axes of the aneurysm and the neck, could be helpful in prevention. The Anaconda device: undergoing clinical trials .This is the only graft system that enables repositioning of the graft after deployment. It is highly flexible and has good torque control
POSTOPERATIVE SURVEILLANCE 30 days : CTA and X-ray Abdomen PA+Lateral 6 months: CTA can be omitted if no prior endoleak and good component overlap 1 year: CTA and X-ray Abdomen PA+lateral If No Endoleak and stable/shrinking Aneurysm Annual Doppler US with plain radiographs CTA if increasing diameter or new endoleak
Thoracic endovascular aORTIC repair ( tevar ) TEVAR refers to the percutaneous placement of a stent graft in the descending thoracic or thoracoabdominal aorta in patient 1994, Dake first reported the initial Stanford experience with 13 patients undergoing endovascular therapy of descending thoracic aortic aneurysms, had 100 percent technical success and no mortality. TEVAR was first approved by the FDA for Thoracic aneurysm repair following the Gore TAG pivotal trial in 2005
Thoracic aortic aneurysm Penetrating aortic ulcer Blunt thoracic aortic injury T ype B aortic dissection ( Type A Unsuitable for TEVAR due to involvement of Aortic valve and Proximity to Coronaries) Aortobronchial fistul a Co- arctatio n Pseudoaneurysm INDICATION
The “gold standard” for preoperative evaluation is a computed tomography (CT) angiogram that includes the thorax, abdomen, pelvis, and femoral arteries Procedure is typically done under GA A lumbar drain is placed in the L3-L4 disc space for drainage of CSF in cases where extensive coverage of the thoracic aorta is anticipated where interruption of contributing blood supply is high Lumbar drainage of cerebrospinal fluid to decrease the pressure in the subarachnoid space and in increase the spinal cord perfusion pressure PROCEDURE
The thoracic aorta is of larger Caliber than that of the infrarenal aorta so needs larger diameter stent grafts & Creation of a conduit to the femoral or iliac artery may be necessary to achieve adequate access There is high force of blood flow in the thoracic aorta so requires a longer seal zone (20 mm both proximal and distal) In severe cases of tortuosity of Ileo-femoral vessel, B rachiofemoral access may be required to perform “Body F lossing” with an stiff wire. A long, stiff wire is guided from the brachial artery and retrieved through the femoral artery. Gentle traction on both the brachial and f emoral sites will straighten out the tortuosity.
LANDING ZONE Unless revascularization is performed, proximal landing in zones 0 and 1 is unacceptable because of the occlusion of the Left CCA in zone 1 & the innominate artery in zone 0. Proximal landing in zone 2 is commonly used with either partial or total occlusion of the left subclavian artery. Zone 3 landing is dependent on the exact anatomical neck at the arch. Proximal landing in zone 3 can lead to angulation of the graft, which provides inadequate sealing of the proximal graft. Zone 4 landing is usually straightforward because of the lack of angulation and distance from the arch vessels To view the landing areas in the arch, it is necessary to view the arch with a 60-to-90 degree left anterior oblique (LAO) projection
A pigtail catheter via the femoral or brachial/radial artery is used to perform an angio , After the angiogram is performed, the proximal neck is evaluated. The length and the diameter of the proximal and distal neck are measured. Based on these measurements, the stent graft is chosen. The patient is heparinized to an activated clotting time of 250 seconds. .
The stent graft is flushed with heparinized solution and advanced into the proximal neck. A repeat angiogram is commonly performed to reconfirm the positioning of the device within the aorta and the landing zone
Prior to device deployment, different manufacturers of TEVAR devices recommend the induction of transient cardiac arrest or a significant decrease in blood pressure to obtain precise deployment of the device, avoiding migration secondary to forward arterial blood flow. During deployment, another option is to place a transfemoral venous pacing wire, and rapid pacing can be done to prevent ejection. The ventilator is stopped shortly for device deployment in to eliminate motion artifact
After the deployment, the stent graft is ballooned to decrease the possibility of E ndoleaks A completion angiogram is performed to confirm lack of gross Endoleak
The distal pulses are checked to assure they correspond with the preoperative baseline. This maneuver is important as embolic events to the lower extremities need to be realized and acted on immediately
Stent grafts Gore-TAG device -made of e-PTFE and an exoskeleton made of nitinol. The proximal and distal ends of the graft have scalloped flares Medtronic Talent thoracic stent graft system- Four are available: proximal main, proximal extension, distal main, and distal extension. The proximal configurations and the distal extension are offered with a bare-spring design (Free-Flo design). The bare-spring design allows for placement of the device crossing the arch vessels proximally and the celiac artery distally for suprasubclavian and infraceliac fixation It consists of a woven polyester graft with a nitinol endoskeleton The Cook TX2 stent graft- is a two-piece modular endograft system made of proximal and distal tubular endografts. The proximal endograft is covered and has stainless steel barbs, allowing for active fixation to the aortic wall. The distal component has at its distal end a bare metal. The TX2 is made of Dacron fabric covered by stainless steel Z-stent
Gore-TAG device The Cook TX2 stent- Medtronic Talent
complication Risk of Perioperative stroke 4% to 8% The risk of spinal cord ischemia 3 to 11% Viseral ischemia can occur with coverage of the celiac axis
LOCAL Groin hematoma I nfection L ymphocele ACCESS ARTERY INJURY T hrombosis Dissection P seudoaneurysm