Seminar on “ THERMOGRAVIMETRIC ANALYSIS AND DIFFERENTIAL SCANNING CALORIMETRY ”
Thermal Analysis Differential Scanning Calorimetry (DSC) Measure heat absorbed or liberated during heating or cooling Thermal Gravimetric Analysis (TGA) Measure change in weight during heating or cooling
Differential Scanning Calorimetry (DSC ) Calorimeter-Heat flow in sample Differential calorimeter-heat flow in sample vs referance as function of time mJ /sec.
• Differential Scanning Calorimetry (DSC) measures the temperatures and heat flows associated with transitions in materials as a function of time and temperature in a controlled atmosphere. • These measurements provide quantitative and qualitative information about physical and chemical changes that involve endothermic or exothermic processes , or changes in heat capacity . DSC: The Technique
Conventional DSC Metal 1 Metal 2 Metal 1 Metal 2 Sample Empty Sample Temperature Reference Temperature Temperature Difference = Heat Flow A “linear” heating profile even for isothermal methods
Modes and principles of operation (1) (b) Power compensated DSC: Temperature differences between the sample and reference are ‘compensated’ for by varying the heat required to keep both pans at the same temperature. The energy difference is plotted as a function of sample temperature
(c) Heat flux DSC ultilizes a single furnace. Heat flow into both sample and reference material via an electrically heated constantan thermoelectric disk and is proportional to the difference in output of the two themocouple junctions
Modes and principles of operation (3) a) DTA: difference in temperature between the sample and reference is plotted against sample temperature
A DSC apparatus is built around - a differential detector - a signal amplifier - a furnace - a temperature controller - a gas control device - a data acquisition device Diagram of a DSC apparatus
DSC measures:- Glass transitions Melting and boiling points Crystallisation time and temperature Percent crystallinity Heats of fusion and reactions Specific heat capacity Oxidative/thermal stability Rate and degree of cure Reaction kinetics Purity
Main Sources of Errors Calibration Contamination Sample preparation – how sample is loaded into a pan Residual solvents and moisture. Thermal lag Heating/Cooling rates Sample mass Processing errors
Sample preparation Form of sample : bulk solid, powder (pressed), liquid. Amount of sample : 3-5mg. DSC Pan : Al, Pt, stainless steel, Ag, Cu, Al 2 O 3
Operation procedures Calibration of instrument Temperature, heat of reaction, heat capacity scale using high purity standards (In, Sn , Bi, Pb,Au ) Baseline correction for a given scan rate (1 - 40K/min). Weight samples before (and maybe after) experiment .
Type of DSC experiments Dynamic heating - thermodynamic properties Isothermal heating - kinetic parameters Dynamic heating :- Constant heat rate mode. (e.g. heat flow vs. temperature). What can we characterise ? Information: 1. Transformation temperature (e.g. onset, peak). 2. Transformation enthalpy (e.g. area under the peak). 3. Activation energy for transformation (e.g. Kissinger analysis) 2.)Isothermal heating mode :- Heating at a fixed temperature over a time interval.(e.g. heat flow vs. time)
THERMOGAVIMETRIC ANALYSIS (TGA) Principle: TGA measures the amount and the rate of weight change of a material with respect to temperature or time in controlled environments. A TGA consists of three major parts a furnace, 1. A microgram balance, 2. An auto sampler and 3. A thermocouple. .
Instrument: Instrument used for thermogravimetry is “ Thermobalance ”. Data recorded in form of curve known as ‘ Thermogram ’ The furnace can raise the temperature as high as 1000°C which is made of quartz. The auto sampler helps to load the samples on to the microbalance. The thermocouple sits right above the sample. Care should be taken at all times that the thermocouple is not in touch with the sample which is in a platinum pan.
A technique that permits the continuous weighing of a sample as a function of temperature and/or as a function of time at a desired temperature
TGA Curve of Calcium Oxalate
Sample Preparation Sample preparation has a significant effect in obtaining good data. It is suggested that maximizing the surface area of the sample in a TGA pan improves resolution and reproducibility of weight loss temperatures. The sample weight affects the accuracy of weight loss measurements. Typically 10-20mg of sample is preferred in most applications. Whereas, if the sample has volatiles 50-100mg of sample is considered adequate. It is to be noted that most TGA instruments have baseline drift of ±0.025mg which is ±0.25% of a 10mg sample. Experimental Conditions Heating Rate Purge gas
Experimental Conditions - Heating Rate Samples are heated at a rate of 10 or 20°C/min in most cases. Lowering the heating rates is known to improve the resolution of overlapping weight losses. Experimental Conditions -Purge gas Nitrogen is the most common gas used to purge samples in TGA due to its inert nature. Whereas, helium provides the best baseline. Air is known to improve resolution because of a difference in the oxidative stability of components in the sample. Vacuum may be used where the sample contains volatile components, which helps improve separation from the onset of decomposition since the volatiles come off at lower temperatures in vacuum. e.g. oil in a rubber tire product.
Calibration Blank test Calibration of mass changes Calibration of temperature
Applications of TGA Determination of the bound and unbound water in the suspension of Milk of Magnesia ( MoM ), used as a laxative. In an overview of thermal analysis testing it is always preferable to do a TGA experiment on unknown samples before doing a DSC experiment (especially for pharmaceuticals). Decomposition of pharmaceuticals renders products which are insoluble and generally sticky on the inside of a DSC cell. These products will lower the life use of a DSC cell. Therefore, know the decomposition temperatures of all drugs and heat in a DSC evaluation to 50°C below those temperatures.
Applications of TGA Evaporation of free (unbound) water begins at room temperature due to dry gas flowing over the sample. Dehydration/ Desolvation of bound water almost always begins at temperatures above room temperature and typically 125°C. Decomposition can have multiple stages (weight losses) but the presence of multiple weight loss steps can also indicate the presence of multiple components in the sample.