basic of angiography physics and equipement.pdf

2,674 views 79 slides Nov 28, 2022
Slide 1
Slide 1 of 79
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79

About This Presentation

basic of angiography physics and equipment


Slide Content

Basic of angiographyphysics
Equipment and image modes
Dr. Naima SENHOU
King Saud bin Abdulaziz University for
Health Sciences
Ministry of National Guard Health Affairs
KAMC-Riyadh

A general term to describe the radiologic examination of
vascular structures within the body after the introduction of
an iodinated contrast medium.
Angiography
Angiography produces an angiogram, which is an image of the
blood vessels in the body. An angiogram looks a little like a
road map, showing the path of blood vessels and their
junctions. Any part of the body can be studied using
angiography so it is used for a wide variety of conditions.

The first angiogram was performed only
months after Roentgen's discovery of X rays.
Which was when?1895.
Two physicians injected mercury salts into an
amputated hand and created an image of the
arteries
HISTORY
Post mortem
injection of mercury
saltsin Jan,1896

Differentiate between angiography and radiographic examinations
List the basic components of the angiography system and identify
the function of each component
Describe a typical basic angiography image
Explain the design and operation of both image intensifier and flat
panel detector systems.
Describe the selectable factors which control both the mode of
operation radiation, dose and image quality.
Differentiate between amgiographymodes.
By the end of this Lecture the student will be able to:
Learning Objectives

Where the x-ray tube and
image intensifier are fixed
to c-arms.
Mostly used in surgical
theatres.
Single or bi-planar angiography system
The system consists of two
X-ray sources and
corresponding detector
panels positioned in a quasi-
orthogonal arrangement

Vascular studies usually require a room or
suite of rooms
specifically designed to accommodate the
sophisticated and accessory equipment needed
to perform angiography and interventional
procedures

Used for dynamic imaging
• X-ray tube operated at lower currents
–Factor of 50 to 100+ less
• X-ray quanta/cm
2
–Radiography: 10
7
/cm2
–Fluoroscopy: < 10
4
/cm2
• Statistical difference in the quality of the image
• Higher noise: lower signal
Fluoroscopy vsRadiography

Principles of X-ray Image Formation
•X-ray generation is inefficient
<1%of the electrical energy is
converted to X-rays. >99% heat.
•Cathode current (m A)= number
of X-ray photons
–Increasing mAincreases absorption
and increases patient dose.
•Tube voltage (k Vp) =energyof
X-ray photons
–Increasing kVpdecreases
absorption, and reduces patient
exposure.

•High-speedrotatinganodetubes.Theobjectofan
angiogramistoproducethehighestquality
radiographsintheshortesttimepossible
•a0.3mmsmallfocalspotwillproducethebest
detail.
•tuberatingcanbeexceededbecauseoftherapid
successionofexposuresneeded
•usualtohavea0.6-mmfocalspottube.
X-raytubeforangiographymachine

•Thismustbeathree-phaseorhigh-
frequency12-pulsemachineandatleast
1000mAtoaccommodatetherapid,short,
andhighexposurevaluesrequiredin
angiography
Generatorofangiographymachines

Cost
Single and bi-plane angiography equipment
Image Acquisition Speed
Specialization
Space Requirements
Site Preparation
Because bi-plane systems capture image data from detectors on two axes, they are able to
acquire 3D images faster.
bi-plane system has a larger footprint than a single-plane system,
There needs to be a ceiling support installed for the second c-arm, additional
rigging is necessary, and the system has an overall larger footprint.
Bi-plane cathlabs simply come with more "stuff",therefore they cost more than single-plane
systems.
Currently, service coverage for a bi-plane cathlab averages 25-30% higher than service for a
single-plane system.
If your facility is planning for a dedicatedNeuro-Angiography & Stroke Interventional Labs,
for example, you'd be best served by a bi-plane system.
If your work leans in a more general/blended direction (balloons, stents, angioruns, etc.) a
single-plane lab can be flexible

Current angiography systems fall into two distinct categories: image
intensifier and flat-panel detector (FPD).

Angiography with image intensifiers
What is an Image Intensifier ?
A complex electronic imaging device
that receives the remnant beam and converts it
to light and increases the intensity of the light.
The image intensifier tube is contained
in a glass envelope in a vacuum and mounted in
a metallic container which provides protection
for the components

Image Intensifier
Schematics
Angiography with image intensifiers
1. II input window
•Convex metal shield that covers the input
face of the II
•Usually made of aluminium or titanium foil
(low Z metal) to allow x-ray beam to enter
with minimum attenuation
•Provides protection for sensitive input
components of the tube and maintains the
vacuum

Input
Phosphor
•Layer of sodium activated cesiumiodide (CsI:Na)
for good x-ray absorption efficiency (70-90%)
•Channelled into tiny needle-like crystals (5µm in
diameter) with fibreoptic-like characteristics
•CsI:Nausually 400-500µm thick
•Each x-ray photon produces ~3000 light photons
in the blue spectrum
1.Constructed of cesium iodide.
2.Responsible for converting the incident
photon’s energy to a burst of visible light
photon.
◦Similar to intensifying screens in cassettes.
3.Standard size varies from 10 -35 cm.
◦Normally used to identify the II tubes.

Photocathode
Thin metal layer bonded directly to the
input phosphor.
Usually made ofCesium and
Antimonycompounds that respond to
light stimulation.
Responsible forPhotoemission.
Electron emission after light
stimulation
The number of electrons emitted is
directly proportional to the intensity of
light intensity of the incident x-ray
photon.

Electrostatic Focusing Lenses
II electron optics
The input screen is maintained at a negative voltage
with respect to the anode (output screen) with a
potential difference of 25 kV. This means the electrons
produced are accelerated across the II tube and carefully
focused on the output screen. The output screen is 1/10
the diameter of the input screen and, therefore, a
minified and inverted image is produced.
A series of lenses inside the II tube to
maintain proper focus of the photoelectrons
emitted from the photocathode.
They contain a positive charge.
They are located along the length of the II
tube.
The focusing lenses assist in maintaining
the kinetic energy of the photoelectrons to
the output phosphor.

Output Phosphor
Output Screen
Thin layer of silver-activated zinc cadmium sulphide
(ZnCdS:Ag)crystals deposited on the inner surface of the
output window that convert the electrons into light
photons. The output image is intensified significantly by the
acceleration of the electrons and the minificationof the
image that occurs in the II tube.
The screen is normally 25-35 mm in diameter and a few
micrometres thick.
This surface of the output screen is coated in a very thin
layer of aluminiumthat:
•Forms part of the anode structure
•High speed electrons travel through the
aluminium layer
•The layer is opaque preventing the light emitted
by the phosphor from back-illuminating the
photocathode and degrading II performance. The
light is reflected back towards the output
increasing the gain of the II tube.

X-ray photons enter tube through aluminium or titanium window.
Hit input phosphor layer of sodium activated caesium iodide and
release light photons.
Light photons detected by photocathode, then release electrons
into the tube.
Electrons accelerated and focused onto the output screen (silver-
activated zinc cadmium sulphide crystals) as a minified and
inverted image.
Light photons released, then leave through the output window.
Summary

Patient tables must provide strength to support patients and are rated by the
manufacturer for a particular weight limit.
It is important that the table not absorb much radiation to avoid shadows, loss of signal
and loss of contrast in the image.
Carbon fiber technology offers a good combination of high strength and minimal
radiation absorption, making it an ideal table material.
Foam pads are often placed between the patient and the table for added comfort, yet
with minimal radiation absorption.
Patient Table and Pad

Beam Filtration
It is common for fluoroscopic imaging systems to be equipped with
beam hardening filters between the X-ray tube exit port and the
collimator.
Added aluminum and/or copper filtration can reduce skin dose at the
patient’s entrance surface, while a low kVpproduces a spectral shape
that is well-matched to the barium or iodine k-edge for high contrast in
the anatomy of interest
In addition to beam shaping filters, many fluoroscopy systems have
“wedge” filters that are partially transparent to the X-ray beam. These
moveable filters attenuate the beam in regions selected by the operator
to reduce entrance dose and excessive image brightness

Variable X-ray BeamFiltration

In angiography, the collimation may be circular or
rectangular in shape, matching the shape of the image
receptor.
Collimation
When the operator selects a field of view, the collimator blade
positions automatically move under motor control to be just a
bit larger than the visible field.
When the source-to-image distance (SID) changes, the
collimator blades adjust to maintain the field of view and
minimize “spillover” radiation outside of the visible area. This
automatic collimation exists in both circular and rectangular
field of view systems.

Image IntensifierMagModes

Anti-scatter grids are standard components in angiography systems, since a large
percentage of fluoroscopic examinations are performed in high-scatter conditions, such as
in the abdominal region. Typical grid ratios range from 6:1 to 10:1. Grids may be circular
(XRII systems) or rectangular (FPD systems) and are often removable by the operator.
Anti-Scatter Grid

Convert X-ray energy into visible light image
• Concentrates energy from the input phosphor onto
smaller output phosphor
–“minification” gain
• Adds kinetic energy
–15-50 X more brightness
• Increases brightnessper unit area by several
thousand
X-Ray ImageIntensifier: Gain

X-Ray ImageIntensifier: Gain
Minificationgain, GM
G
m=(D
input)2/ (D
output)2 ~ 16 –200 …..>100
• Electronic (flux) gain, GE , produced by electron
acceleration
GE ~ 50
• Total gain, GT
GT = GM x GE ~ 5000

Flat panel detectors utilise the same technology as digital
radiography in that there is a flat panel of detectors that provide a
direct electronic readout instead of requiring the conversion of
analogue to digital as is seen in the IITV.
Similar to digital radiography dynamic FP detectors can be direct or
indirect. However, they are more commonly indirect with a CsI:Tlx-
ray scintillator layer which is superimposed onto an a-Si high
resolution active matrix.
Flat panel detector

DIRECT AND INDIRECT FLAT PANEL DETECTORS
In indirect FPD,
The scintillator converts X-ray photons to
visible light usually an amorphous silicon (a-
Si) thin-film-transistor (TFT)
The intensity of the light is proportional to
the intensity of the X-ray photons. The
light emitted is detected by the TFT array,
which converts light energy toelectrical
signal.
Direct FPD
One of the most common photoconductors isamorphous
selenium
(a-Se)
When X-ray photons traverse the photoconductor, they are absorbed
and electron-hole pairs proportional to the X-ray intensity are generated in the
solid state material.
The electrons and holes are directed by the electric field towards the TFT elements

Thin-film transistor(TFT)

Caesiumiodide or cesium iodide (chemical formula CsI) is the ionic compound of
caesiumand iodine. It is often used as the input phosphorof an X-ray image

Flat PanelMagModes
FP pixelated to finest resolution but readout binning
occurs for large FOVs
–Maximum presentation is 10242 x10242 during
angiography

Resolution improves until pixel matrix equals display matrix
–16-20 cm FOV
• Further FOV
reductions, no inherent change in resolution
• No change in gain!
Flat PanelMagModes

Flat Panels: DoseRatevsFieldSize
Increases with 1/(field size)
–15 x15 cm2 has 4x dose over 30 x 30
cm2
• No change in gain
• Empirically determined
• No change from II based systems

Pixel-binning is used in flat-panel (FP) x-ray detectors for
fluoroscopy to increase signal-to-noise ratio (SNR) and lower
digital data rates.
Flat panel Pixel-binning

This is a portion of a Mars Observer
Camera image of gullies cut into a
crater wall. The portion shown is
about 1 km across. The left hand
image has not been binned. The
right hand image has been binned 4
x 4. Note that binning reduces the
spatial resolution and the finest
details can no longer been seen.
No binning 4x4 binning
Effect of pixel binning

• Dynamic range of flat panel is greater than an
image intensifier
• Issues such as burnout (blooming) and
blackout (saturation into black) regions in
image is not as significant an issue with FD
as it is with II.
Dynamic Range

Smaller equipment.
Video signal emerges in digital form, reducing
electronic noise
Square or rectangular field (unlike circular field in IITV)
= better coverage in the corners
Better temporal resolution with matrix size of 2048 x
2048 pixels
Greyscale of 12 or 14 bits per pixel
Produces better quality images than IITV
Fewer artefacts such as geometrical distortion,
vignettingor contrast loss
Detective quantum efficiency 10-20% better than IITV
so can afford to reduce patient dose
Zoom option available (but doesn't increase spatial
resolution as it does in IITV)
Benefitsof Flat panel detector

X-Ray Absorption Efficiency

TV MONITORS
•This practical and efficient viewing system was employed
because of the limitations of the mirror optic viewing
system.
TV monitors:
1.Enables viewing by multiple persons.
2.Monitors may be located in remote locations other than
the radiographic room.
3.Image brightness and contrast can be manipulated.
4.Images may be stored on different medium for reviewing
at a later time.

Angiography modes

Manual Mode
◦Allow the use to select the exact MA and KVprequired
AEC Mode
◦Allow the unit to drive the KVpand MA to optimize
dose and image quality
Pulsed Digital mode
◦Modifies the fluoroscopic output by cutting by cutting
out exposure between pulses
◦With the pulsed mode, it can be set to produce less than
the conventional 25 or 30 images per second. This
reduces the exposure rate.
Fluoroscopy -Modes of operation

57
Set the default fluoroscopy mode
to LOW
Lowest input dose needed to
generate a USABLE image

Continuous Basic form of fluoroscopy; continuously ON x-ray beam
High dose rate
• Allows exposure rates of up to 20 R/min
• Used to reduce noise in images
Variable frame rate pulsed
30, 15, and 7.5 frames/sec operation allows lower temporal resolution for
parts of procedure
Angiography Modes of Operation

59
30 images in 1 second
X rays
In conventional continuous-beam angiography there is an
inherent blurred appearance of motion because the exposure
time of each image lasts the full 1/30
th
of a second at 30 frames
per second.
Continuous stream of X rays produces blurred
images in each frame
Images
Continuous mode

62
Lesson: Variable pulsed fluoroscopyis an important
tool to manage radiation dose to patients but the
actual effect on dose can be to enhance, decrease or
maintain dose levels. The actual effect must be
estimated by a qualified physicist so that variable
pulsed fluoroscopy can be properly employed.
Variable Pulsed

63
Understanding Variable Pulsed Fluoroscopy
Background: dynamic imaging captures many still
imagesevery second and displays these still-frame
images in real-time succession to produce the
perception of motion. How these images are captured
and displayed can be manipulated to manage both dose
rate to the patient and dynamic image quality.Standard
imaging captures and displays 25 -30 images per
second.
Pulsed mode

64
Each X ray pulse shown above has greater intensity
than continuous mode, but lasts for only 1/100
th
of a
second; no X rays are emitted between pulses; dose to
patient is same as that with continuous fluoroscopy
Pulsed fluoroscopy, no dose reduction
Images
Pulsed fluoroscopy produces sharp appearance of motion
because each of 30 images per second is captured in a pulse
or snapshot (e.g., 1/100
th
of a second).
X rays
30 images in 1 second

65
Each angiographic ‘run’ consists of multiple still images taken in quick succession.

67
Pulsed fluoroscopy, dose reduction at 15 pulses per second
Sharp appearance of motion captured at 15 images per second
in pulsed mode. Dose per pulse is same, but only half as many
pulses are used, thus dose is reduced by 50%.The tradeoff is a
slightly choppy appearance in motion since only half as many
images are shown per second
Images
X rays
15 images in 1 second

68
Pulsed fluoroscopy, dose enhancementat 15 pulses per second
Dose per pulse is enhanced because pulse intensity and
duration is increased. Overall dose is enhanced.
Images
X rays
15 images in 1 second
Reproduced with permission from Wagner LK, Houston, TX 2004.
Images
X rays
15 images in 1 second

69
Pulsed imaging controls:
Displaying 25–30 picture frames per second is usually
adequate for the transition from frame to frame to
appear smooth.
This is important for entertainment purposes, but not
necessarily required for medical procedures.
Manipulation of frame ratecan be used to produce
enormous savings in dose accumulation.
Pulsed mode

70
Pulsed fluoroscopy at 7.5 images per second with
only 25% the dose
Pulsed fluoroscopy, dose reduction at 7.5 pulses per second
Images
X rays
Average 7.5
images in 1
second

Patient DoseRate

Frame averaging
Fluoroscopy images generally are noisy
Sometimes beneficial to compromise resolution for lower noise images
Digitize fluoroscopic images and perform real-time averaging in
computer memory for display

73
Influence of operation modes: from lowmode to
cine, radiation / scatter dose rate could increase in
a factor of 10-15

Automatic brightness control
Purpose of ABC is to keep brightness of the image
constant at the monitor.
Accomplished by regulating the X-Ray exposure rate
incident on the image receptor
Video signal itself can be used to sense light output
ABC can adjust both tube (mA) current and generator
voltage (kVp)

When the fluoroscopist takes his or her
footoffof the fluoroscopy pedal, rather than
seeing a blank monitor, last-image-hold
enables the continues display of the last
captured fluoroscopic image.
The image will by displayed until the
fluoroscopy beam is turned on again.
Last-Image-Hold

Fluoroscopy vs. Cineangiography
•Fluoroscopy
–A real-time X-ray image when it is
not necessary to record it.
–Requires less image quality than
does acquisition (cine)
–Images seen in motion, neuro-
psychology of vision integrates
frames effectively reducing
perceived image noise.
–With more noise tolerated, input
doses can be lower.
•Cineangiography
–Images are obtained at higher X-
ray input doses for acquisition
–Most units are calibrated such
that patient dose is 10-15x
greater than fluoroscopy
–Thus, a single frame in cine is
equal to about one second of
fluoro
–A typical acquisition frame rate is
15 frames/sec

THANK YOU