Installation Email: [email protected] Tel: +86 755 8654 0309 #6K, Building No.1, Baisha Emerging Industry Zone,Shahexi Road, Nanshan District, Shenzhen 518052 China
Pre-installation check Check the cartons and packages Correct quantity of the goods Without inversion or deformation Without mark of wet water Without trace of hit Without signs of being opened 2
Installation requirement Environment Temperature :10℃-30℃ Power supply : AC 220V , 50Hz , 250VA Relative humidity : 70% Atmospheric pressure : 76 kPa-106 kPa Altitude : 0 ~ 3000m Less dust, good ventilation environment. No corrosive, flammable and explosive gas or steam in the room Avoid direct sunlight or thermal radiation Good desktop level (inclination below 1/200) No body-perceivable shock. 3
Installation requirement No machinery transmitting high frequency nearby. Cautions on electromagnetic wave: Please don’t carry cellophane, walkie-talkie and other electric appliances transmitting electromagnetic wave into the room With protective earthling (below 10Ω earth impedance) 4
Not less than 20 cm Not less then 30cm Not less then 50cm Installation requirement 5
PC requirement CPU : mainstream CPU Storage : 4.0 GB Hard drive space : > 40G Operation system : Windows 7/10 Supporting software : Microsoft Word 97 and above At least 2 partitions required, install culture system in any disk except C partition No standby mode setting for computer No sleep mode setup on computer 6