Intra Aortic Balloon Pump by Rubina Shehzadi RN

8,360 views 33 slides Sep 24, 2019
Slide 1
Slide 1 of 33
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33

About This Presentation

An intra-aortic balloon pump (IABP) is a type of therapeutic device which helps heart to pump more blood. You may need it if your heart is unable to pump enough blood for your body. The IABP consists of a thin, flexible tube called a catheter. Attached to the tip of the catheter is a long balloon.


Slide Content

IABP by Rubina Shehzadi 1

Intra aortic Balloon Pump(IABP)
By: RubinaShehzadi
(BSN,Mphil,PublicHealth)

Objectives
At the end of this presentation you will be able to learn:
Define Intra Aortic Balloon Pump(IABP)
Mechanics of IABP
Indications & Contra indications
IABP Settings
IABP wave form Interpretation
Variations in IABP
IABP frequency
Timing Errors and manifestations
Complication & prevention of IABP
Nursing care plan of Patient with IABP
IABP by Rubina Shehzadi 3

Intra Aortic Balloon Pump
(IABP)
An intra-aortic balloon pump(IABP) is a type of therapeutic
device which helps heart to pump more blood. You may need
it if your heart is unable to pump enough blood for your body.
The IABPconsists of a thin, flexible tube called a catheter.
Attached to the tip of the catheter is a long balloon.
“Theintra-aorticballoonpumpisamechanicaldevicethat
increasesmyocardialoxygenperfusionandindirectlyincreases
cardiacoutputthroughafterloadreduction.Itconsistsofa
cylindricalpolyurethaneballoonthatsitsintheaorta,
approximately2centimeters(0.79in)fromtheleftsubclavian
artery”
IABP by Rubina Shehzadi 4

IABP by Rubina Shehzadi 5

Intra Aortic Balloon Pump
(IABP)
It consists of two parts: a ballooninserted into the
aorta, one of the large arteries through which blood
passes from the heart to the rest of the body; and a
machine/console outside the body.
It gives temporary support for the left ventricle by
mechanically displacing blood within aorta.
It is the most common and widely used methods of
mechanical circulatory support.
It is used in surgical and non-surgical patients of
cardiogenic shock.
IABP by Rubina Shehzadi 6

Insertion of IABP
The device is introduced from the femoral artery and
guided into the aorta till its tip is just distal to the left
subclavian artery takeoff from the aortic arch by using an
X-ray camera to move it.
The device is hooked up to a machine that tells the balloon
when to get bigger and when to get smaller.
It uses helium to blow up the balloon because helium won’t
cause problems in body if it leaks.
Therefore, when properly positioned, the IABP occupies
the entire descending thoracic aorta and much of the
suprarenalabdominal aorta.
Although fluoroscopy/echocardiography can help guide
placement, they are not absolutely necessary in emergent
situations.
IABP by Rubina Shehzadi 7

IABP by Rubina Shehzadi 8

IABP by Rubina Shehzadi 9

How long can an intra
aortic balloon pump stay in?
Sometimes the Intra-aortic-balloon pump(IABP) stays in
for up to one week
in rare instances, staying in for up to two weeks,
therefore increasing the risk of infection.
IABP by Rubina Shehzadi 10

Indications of IABP
IABP therapy is a short-term treatment. It is mostly used
to treat cardiogenic shock. That’s when your heart can’t
pump enough blood to meet the needs of your body.
Some heart problems can cause cardiogenic shock.
These include:
Unstable angina
Heart attack
Certain abnormal heart rhythms
Heart failure
Heart defects
As a bridge to cardiac transplantation
IABP by Rubina Shehzadi 11

Contra Indications
Aortic Insufficiency
Aortic aneurysm
Aortic dissection
Limb ischemia
Thromboembolism
IABP by Rubina Shehzadi 12

Core Principle of IABP
Therapy
Synchronized counterpulsationis the core principle of
IABP therapy.
This describes inflation in diastole and deflation in
systole of a balloon situated in the descending aorta.
The overall aim is to improve myocardial function by
myocardial oxygen supply
myocardial oxygen demand
Input and removal of helium gas causes inflation and
deflation of the balloon. The timing of balloon inflation
and deflation is based upon the aortic pressure
waveform and the electrocardiogram.
IABP by Rubina Shehzadi 13

IABP by Rubina Shehzadi 14

Cont….
The balloon inflation occurs immediately after aortic
valve closure and balloon deflation just before opening of
the aortic value.
Inflation and deflation of the balloon have two major
effects:
Inflation during diastole causes blood displacement into
the proximal aorta, resulting in increased coronary
blood flow, while Deflation during systole reduces aortic
volume and afterload through a vacuum effect
IABP by Rubina Shehzadi
.
15

Augmentation on a balloon
pump
Ballooninflation and diastolic augmentation. As the
aortic valve closes in end-systole (this correlates with
the dicroticnotch) the IABP ballooninflation pushes
blood against the closed valve. ... Thus, as
“the ballooninflates in diastole, it creates a peak of
pressure, which is the diastolic augmentation”
IABP by Rubina Shehzadi 16

IABP counter-pulsation rapidly stabilizes patients in
cardiogenic shock.
The greatest improvement in cardiac index and
pulmonary capillary wedge pressure occurs in patients
with mechanical defects complicating acute myocardial
infarction (MI), ie, mitral regurgitation(MR) or
ventricular septal defect(VSD)
IABP by Rubina Shehzadi 17

IABP balloon waveform
The normal IABP balloon waveform
The balloon itself has a pressure transducer, and it
generates a waveform.
IABP by Rubina Shehzadi 18

ECG triggering of the IABP
The basic principles are:
Inflation of the balloon is triggered by the thebeginning of
diastole, which correlates with the middle of the T-
wave.The balloon is timed to deflate at the very end of
diastole. This correlates with the R-wave on the ECG, and
this is the most commonly used trigger for balloon
deflation.
In atrial fibrillation, the ECG trigger is timed to deflate on
the R wave as usual, but the R-R interval (which governs
the timing of theballoon remaining inflated) varies. The R
wave timing can also be of the"pattern" type, where
normal QRS morphology is expected, or "peak" type where -
if your QRS is monstrously misshapen -the IABP will choose
the maximum voltage peak and use that instead.
IABP by Rubina Shehzadi 19

IABP by Rubina Shehzadi 20

IABP by Rubina Shehzadi 21

IABP by Rubina Shehzadi 22

Normal Balloon Pressure
Waveform
IABP by Rubina Shehzadi 23

Variation in balloon pressure
waveforms(Heart Rate)
IABP by Rubina Shehzadi 24

Variation in balloon pressure
waveforms(Rhythum)
IABP by Rubina Shehzadi 25

Variation in balloon pressure
waveforms(Gas leak)
IABP by Rubina Shehzadi 26

Variation in balloon pressure
waveforms(CathetarKinking)
IABP by Rubina Shehzadi 27

Balloon too large syndrome
IABP by Rubina Shehzadi 28

Complications of IABP
Infection at site of insertion
Hematoma and bleeding at insertion site
Limb ischemia, absent pulses
Thrombocytopenia
Coagulation disturbances
Aortic dissection
Displacement of the balloon catheter obstructing left
subclavian artery or renal artery perfusion
Vascular complications
Balloon leak, rupture, gas loss from the balloon
Timing issues
IABP by Rubina Shehzadi 29

Limb Ischemia & Bleeding
Cause of Limb ischaemiaare obstruction of a small or
diseased femoral artery by the catheter, formation of
thrombus from direct arterial injury during IABPinsertion
and thromboembolism.
Bleeding at the insertion site is due to anticoagulation
therapy or associated thrombocytopenia.
Patients receiving IABP therapy are normally maintained on
an anticoagulation regimen to avoid thrombus formation .
Baseline full blood count and anticoagulation screen should
be reserved prior to and during therapy to observe for
changes such as a decrease in haemoglobinand to monitor
the effect of anticoagulation therapy prescribed.
IABP by Rubina Shehzadi 30

Weaning and Removal of IABP

Once the patient’s condition has stabilized the IABP is
considered for removal. This is preceded by a process of
weaning the pump from 1:1 to 1:2 and finally 1:3
Observing the patient for ischaemicchest pain or
development of heart failure symptoms such as
breathlessness, hypotension and tachycardia. Once the
patient is stable the IABP can be removed.
IABP by Rubina Shehzadi 31

Nursing care For IABP patients
Post removal the patient is kept on bed rest with the
leg kept straight as per local guidelines.
Nursing care involves observing for recurrence of
symptoms as well as checking the insertion site for
haematomaformation ooze or symptoms suggestive of
retroperitoneal bleed every 15 minutes for the first
hour, half hourly for the next two hours and thereafter
hourly or as clinical acuity dictates. In addition lower
limb perfusion is assessed at these times confirming
adequate perfusion and presence of pedal pulses.
IABP is a form of circulatory support for those
presenting with ischaemiaor heart failure. Nursing care
involves care of the pump as well as assessing patient
from a cardiovascular and hemodynamic perspective .
IABP by Rubina Shehzadi 32

IABP by Rubina Shehzadi 33