Electro analytical chemistry


Slide Content

Nov 16, 2004
Introduction to Electroanalytical
Chemistry
Lecture Date: April 27
h
, 2008

Reading Material
●Skoog, Holler and Crouch:Ch. 22 (An Introduction to
Electroanalytical Chemisty)
●See also Skoog et al. Chapters 23-25.
●Cazes:Chapters 16-19
●For those using electroanalytical chemistry in their work,
the following reference is recommended:
A. J. Bard and L. R. Faulkner, “Electrochemical Methods”, 2
nd
Ed., Wiley, 2001.

Advantages of Electroanalytical Methods
Matched against a wide range of spectroscopic
and chromatographic techniques, the techniques
of electroanalytical chemistry find an important
role for several reasons:
–Electroanalytical methods are often specific for a
particular oxidation state of an element
–Electrochemical instrumentation is relatively
inexpensive and can be miniaturized
–Electroanalytical methods provide information about
activities (rather than concentration)

History of Electroanalytical Methods
Michael Faraday: the law
of electrolysis
–“…the amount of a substance deposited
from an electrolyte by the action of a
current is proportional to the chemical
equivalent weight of the substance.”
Walter Nernst:the Nernst
equation (Nobel Prize
1920)
Jaroslav Heyrovsky:the
invention of polarography:
(Nobel Prize 1959)
Walter Nernst
(1864-1941)
Michael Faraday
(1791-1867)
Jaroslav Heyrovsky
(1890-1967)

Main Branches of Electroanalytical Chemistry
Key to measured quantity: I = current, E = potential, R = resistance, G =
conductance, Q = quantity of charge, t = time, vol = volume of a standard solution,
m = mass of an electrodispensed species
Interfacial
methods
Bulk methods
Static methods
(I = 0)
Dynamic
methods
(I > 0)
Potentiometry
(E)
Conductometry
(G = 1/R)
Controlled
potential
Constant
current
Voltammetry
(I = f(E))
Amperometric
titrations
(I = f(E))
Based on Figure 22-9 in Skoog, Holler and
Crouch, 6th ed.
Electro-
gravimetry
(m)
Coulometric
titrations
(Q = It)

Main Branches of Electroanalytical Chemistry
Potentiometry: measure the potential of electrochemical
cells without drawing substantial current
–Examples:pH measurements, ion-selective electrodes,
titrations (e.g. KF endpoint determination)
Coulometry: measures the electricity required to drive an
electrolytic oxidation/reduction to completion
–Examples:titrations (KF titrant generation),
“chloridometers” (AgCl)
Voltammetry: measures current as a function of applied
potential under conditions that keep a working electrode
polarized
–Examples:cyclic voltammetry, many biosensors

Electrochemical Cells
Voltmeter
Salt bridge
(KCl)
Zn electrode
Cu electrode
e-
e-
Zn Zn
2+
(aq) + 2e
-
a
Zn
2+
= 0.010
Anode
Cu
2+
(aq) + 2e-Cu(s)
a
Cu
2+
= 0.010
Cathode
0.010M ZnSO
4
solution
0.010M CuSO
4
solution
Zinc (Zn) wants to ionize more than copper (Cu).
We can use this behavior to construct a cell:

e-
e-
reference electrode
counter electrode
working electrode
indicator electrode
detector electrode
Potentiometry: Measures equilibrium E
Amperometry: Control E, measures I as function of time
Coulometry: Control E, measure total Q over a period of time
control
measurement
Electrochemical Cells and Analytical Methods

Electrochemical Cells
Galvanic cell: a cell that produces electrical energy
Electrolytic cell: a cell that consumes electrical
energy
Chemically-reversible cell:a cell in which reversing
the direction of the current reverses the reactions at
the two electrodes

Conduction in an Electrochemical Cell
Electrons serve as carriers (e.g. moving from Zn
through the conductor to the Cu)
In the solution, electricity involves the movement of
cations and anions
–In the salt bridge both chloride and potassium
ions move
At the electrode surface: an oxidation or a
reduction occurs
–Cathode:the electrode at which reduction
occurs
–Anode:the electrode at which oxidation occurs

Oxidation occurs when a chemical species loses an electron.
LEO = lose electron is oxidation
Reductionis when a species gains an electron.
GER = gain an electron is reduction
For example, the chemical reaction
can be decomposed into two half reactions:
“Leo the Lion Says Ger”

Faradaic and Non-Faradaic Currents
Figure 22-2
Mass Transfer occurs by:
Convection
Migration
Diffusion
Faradaic (governed by Faraday’s law): direct transfer of
electrons, i.e. oxidation at one and reduction at the other
electrode
Non-Faradaic: increasing charge of the double layer

Fundamentals
Electrical charge, q, is measured in coulombs (C). The
charge associated with chemical species is related to the
number of moles through the Faraday constant,
F=96,485.3 (~96,500) C/mole.
Electrical current, I, is measured in Amperes (A). Current is
the amount of charge that passes in a unit time interval
(seconds).
Ohm's law relates current to potential (E) through the
resistance (R) of a circuit by E=IR. The potential is
measured in Volts (V) and the resistance in Ohms ().

Power (P) is measured in Watts (W = J/s) and is related to
the current and potential by P= IE.
The work is measured in Joules (J) and is related to the
potential and the amount of charge by work=q E.
The relationship between the standard Gibb's free energy
change, G°(J/mole), and the standard electromotive force
(EMF), E°(V), is given by
G°=-n F E°
where nis the number of electrons transferred and
superscript on E
0
refers to ‘standard state.’
Fundamentals

Fundamentals: The Nernst Equation
●The Nernst equation gives the cell potential E (in volts):
●Q (the activity quotient) is the ratio of products over reactants
as in equilibrium calculations. For the generic reaction:
●Q is given by:
●The A’s are activities. For low-concentration solutions (low
ionic strengths):
F = faraday (constant)
n = # moles electrons in process
E
0
= standard potential for cell

Electrode Potentials
The reactions in an electrochemical cell can be
thought of as two half-cell reactions, each with its
own characteristic electrode potential
–These measure the driving force for the
reaction
–By convention, always written as reductions
Standard electrode potential (E
0
):the measure
of individual potential of an electrode at standard
ambient conditions (298K, solutes at a
concentration of 1 M, and gas pressure at 1 bar).

Some Standard Electrode Potentials
Reaction E
0
at 298K (Volts)
Cl
2(g) + 2e
-
2 Cl
2 +1.359
O
2(g) + 4H
+
+ 4e
-
2 H
2O +1.229
Ag
+
+ e
-
Ag(s) +0.799
Cu
2+
+ 2e
-
Cu(s) +0.337
Hg
2Cl
2+ 2e
-
2Hg(l) + 2 Cl
2 +0.268
2H
+
+ 2e
-
H
2(g) 0.000
AgI(s) + e
-
Ag(s) + I
2 -0.151
Cd
2+
+ 2e
-
Cd(s) -0.403
Zn
2+
+ 2e
-
Zn(s) -0.763
See appendix 3 in Skoog et al. for a more complete list

The Standard Hydrogen Electrode (SHE)
A universal reference, but is really a hypothetical
electrode (not used in practice)
–Uses a platinum electrode, which at its surface
oxidizes 2H
+
to H
2gas.
–Very sensitive to temperature, pressure, and
H
+
ion activity
Because the SHE is difficult to make, the
saturated calomel electrode (SCE) is used
instead.
–Calomel = mercury (I) chloride

Q: What is the electrode potential for the
electrode Ag/AgI(s)/I
-
(0.01 M) ?
The overall reaction for this electrode is
This reaction cannot be found in tables of reduction potentials.
But the reaction is comprised of two components
Electrode Potentials

We can initially ignore the fact that the electrode contains
AgI and find E for the silver ion reduction.
Electrode Potentials

The Glass pH Electrode
●One of the most common
potentiometric measurements is pH
(a so-called “p-Ion” measurement).
●The common glass pH electrode
makes use of junction potentials to
determine the hydronium ion
concentration in a sample solution.
●A typical glass pH electrode is
configured as shown here:

The glass pH electrode is used with a Ag/AgCl reference
electrode. For most modern pH electrodes the reference
electrode is incorporated with the pH indicator electrode.
A small frit or
hole connects
the reference
electrode and
the sample
solutions
The Glass pH Electrode

pH Measurements
●A combination pH electrode combines the indicator and
reference into a single unit.
●The potential of this cell is:
●where E
ijand E
ojare the junction potentials at the inner
and outer layers of the glass membrane.
●Junction potential: occurs at the interface of two
electrolytes, caused by unequal diffusion rates of cation
and anions across the boundary (e.g the frit in a salt
bridge)

More About pH Measurements
●The surface of the glass is hydrated, which allows
exchange of hydronium ions for the cation in the glass
(sodium or lithium).
●There are four interface regions, the external solution and
hydrated glass, hydrated glass and dry glass on the
outside, dry glass and hydrated glass on the inside, and
hydrated glass and the internal solution.
●If the glass is uniform, the two hydrated glass/dry glass
interfaces should be identical and should have the same
junction potential.
●Since the glass interface junction potentials then cancel
each other, the junction potential is then the difference
between the internal and external solutions.

0.05916pHloglogE
thencomposition, fixedahassolution internalthe if
log
a
If the two solutions are identical
logloglog
,',,',mem
,',
,',
,,,,H
,,,',
,',,,
,,
,',
,',
,,
+
+=+-=
-=
=
-=--=
++
+
+
+
++
++
+
+
+
+
ka
F
RT
a
F
RT
a
a
F
RT
E
a
aa
aa
F
RT
a
a
F
RT
a
a
F
RT
E
innsolHoutnsolH
innsolH
outnsolH
mem
outglassHinglass
outglassHinnsolH
outnsolHinglassH
outglassH
outnsolH
innsolH
inglassH
mem

pH Measurements

pH Measurements
For a real electrode, the two surfaces will not be identical
and the constant k needs to be determined experimentally.
The constant k is termed the asymmetry potential. The
constant is termed the electromotive efficiency.

pH Measurements

Q: Why does the pH change the interfacial
potential of the glass/aqueous interface?
A: The motion of the sodium ions leave behind a
negatively charged glass layer that is neutralized to a
lesser or greater extent according to the pH.
More explanation about how a pH meter really
works: The sodium ions must move through the dry
part of the membrane and this process is slow. For
this reason, the membrane is made very thin. Also, a
nonperturbing (low-current) voltmeter is used to read
the cell voltage so that only a few sodium ions must
move through the dry glass in a given time period.
pH Measurements

Errors in pH measurements with glass electrodes arise from
the following effects:
–Calibration problems (e.g. drift, or error in the calibration)
–Junction potential
–High [Na+] interacting with electrode
–High acid concentration
–Equilibration time
–Temperature control
Typical electrodes have the following performance:
–Accuracy = +/-0.02 pH units
–Precision = +/-0.002 pH units
pH Electrodes: Errors, Accuracy and Precision

Modern pH electrodes are usually of the "combination"
type, meaning that a single cylinder contains both the
reference electrode, and a glass membrane electrode.
Schematically, the total cell may be expressed as
SCE//test solution ([H
3O
+
]=a1)/glass
membrane/[H
3O
+
]=a2, Cl-/AgCl(s)/Ag
The Combination pH Electrode

A Modern Combination pH Electrode

Electrochemical pH Measurements Concluded
Consider a typical problem related to the
use of the combination pH electrode.
Recall that
E
cell = L -0.0592 V pH
QUESTION:If E
cell= -0.115 V at a pH of 4.00,
what is the pH of a solution for which E
cellis
-0.352 V?

ANSWER: First, find L from the measurement of
the standard:
-0.115 V = L -0.0592 x pH
-0.115 V = L -0.0592 x 4.00
Therefore, L = 0.122 V
Second, use this value of L to find pH:
-0.352 V = 0.122 V -0.0592 V x pH
pH = (0.122 V -(-0.352 V))/0.0592
pH = 7.84

QUESTION: What does the pH meter read if the
pH is 7.00 in a 1 M salt solution having 1 M Na
+
ions present?
ANSWER:
[H
+
]
obs= 1 x 10
-7
+ 1 x 10
-12
Conclusion --the pH meter reads the true pH
under these conditions.

The Ion Selective Electrode (ISE)
●An ISE generally consists
of the ion-selective
membrane, an internal
reference electrode, an
external reference
electrode, and a
voltmeter.
●Example: an ISE for
fluoride (F
-
)

Automatic pKa and log P Determination
pKa (ionization constant) and log P (octanol/water partition) are
important physical parameters that play critical roles in determining how
compounds behave in physiological environments and how they
interact with enzymes, receptors and cell membranes
The Sirius
GLpKa system:
combination pH
electrode
sample tray
reagents
liquid
dispensors

Conductometry
Conductometry: Detection of electrical
conductivity
–Key analytical applications: conductometric detection
in ion-exchange chromatography (IEC or IC) and
capillary electrophoresis (CE)
Used to detect titration endpoints