UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application �in Pharmaceutical Industry Dr. A. Amsavel
5,020 views
42 slides
Feb 04, 2021
Slide 1 of 42
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
About This Presentation
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application �in Pharmaceutical Industry Dr. A. Amsavel.
UV &Visible Spectroscopy-Absorption Theory
Electronic Transitions
Beer- Lambert Law
Chromophores & Auxochrome
Factors Influence the Absorption
UV-Vis Spectrophotomete...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application �in Pharmaceutical Industry Dr. A. Amsavel.
UV &Visible Spectroscopy-Absorption Theory
Electronic Transitions
Beer- Lambert Law
Chromophores & Auxochrome
Factors Influence the Absorption
UV-Vis Spectrophotometer-Instrumentation
Operation of the Spectrophotometer
Qualification & Calibration
Application
Size: 439.29 KB
Language: en
Added: Feb 04, 2021
Slides: 42 pages
Slide Content
Principle, Theory, Instrumentation and
UV-VISIBLE
SPECTROPHOTOMETRY
Principle, Theory, Instrumentation and
Application
in Pharmaceutical Industry
Dr. A. Amsavel, M.Sc., B.Ed., Ph.D.
An Overview
Introduction
UV &Visible Spectroscopy-Absorption Theory
Electronic Transitions
Beer-Lambert LawBeer-Lambert Law
Chromophores& Auxochrome
Factors Influence the Absorption
UV-Vis Spectrophotometer-Instrumentation
Operation of the Spectrophotometer
Qualification & Calibration
Application
UV-Vis Spectroscopy
UV-Visible spectrometry is one of the most widely used analytical
technique in Industries as well Academic Research. It is very
powerful and popular method for identification and estimation of
elements & organic molecules.
Advantage:
Readily available
Simple & Easy to operate
Relatively less expensive
Does not require complex sample preparation
Require small amount of sample & non-destructive method
To test wide verity of organic & Inorganic chemicals
Qualitative and quantitative analysis
UV-Vis Spectroscopy
Spectroscopy is the branch of science dealing the study of interaction of
electromagnetic radiation with matter. Absorption of molecule due to
electromagnetic radiation is based on the energy levels.
Ultraviolet-visible (UV-Vis) spectroscopy , electronic transition happens
due to interaction electromagnetic radiation and electrons. One or more
of the outer or the bonding electrons promote from a ground state into a of the outer or the bonding electrons promote from a ground state into a
higher-energystate.
Energy(E) =hν= (hc/λ) ×10
9
Where h= Planck's constant (6.63 ×10
-34
J · s)
ν= frequency (Hz), related to the energy changeΔE, induced when
electromagnetic radiation is absorbed (ΔE=hνper photon)
c= velocity of light (2.998 ×10
8
ms
-1
)
λ= wavelength (nm)
ElectromagneticRadiation
Wavelength
UV & Visible Light Absorption Theory
Ultraviolet andvisible(UV-Vis) Spectrophotometryis based on the
ability of atoms, molecules and ions to absorb light at wavelengths
in the ultraviolet (180-400nm) andvisible(400-800nm) range.
This absorption is associated with changes in electronic energy in
the form of temporary transitions of electrons to an excited state at
a higher energy orbital. As each energy level of a molecule or
molecular ion also has associated vibrationaland rotational sub-
levels.
Absorption band is characteristic of the functional groups and
bonds in a molecule.
Electronic Transitions
Most of the Organic compounds observed transitions of electrons
in σorornnon-bonding electron & orbitalsof atoms such as
H, C, N, O etc.
UV-Vis range only n → ∗& → ∗transition
UV-Vis Spectrophotometry
In UV-Vis spectrophotometry, transmittance T is a measure of the
attenuation of a beam of monochromatic light based upon the
comparison between the intensities of the transmitted light (I) and the
incident light (Io)
Transmittance (T)= I / IoTransmittance (T)= I / Io
Absorption (A) = log 1/T = log (Io/I)or = -log T
Specific Absorption coefficient Es =A/ Clor I= Io X10
–EsCl
Molar Absorption coefficient ε=A/Cl
Quantitative analysis: Laws of molecular absorption is used in all spectroscopic
quantitative Analysis
Beer-Lambert Law
French mathematician Lambert and German physicist Beer proposed a
hypothesis about absorption of molecules by electromagnetic radiation.
Lamberts Law: Absorbance (A) is proportional to the Path length (l) of the
absorbing medium.
Beers law: Absorbance (A) is proportional to the Concentration (C) of the Beers law: Absorbance (A) is proportional to the Concentration (C) of the
sample.
Beer-Lambert Law -Absorbance is (A) proportional to Concentration
and Path length of the sample.
A Cl; A = εCl
C-Concentration (Moles /litre) ; l-Path length (cm) & ε -Molar absorption coefficient
(molar absorptivity)
Absorption for a mixture of two compounds A = l x (ε1C1+ ε2C2)= A1 +A2
Beer-Lambert Law
“When monochromatic light passes to the transparent,
homogeneous medium, absorption of light is
proportional to the Concentration and Path length of
the sample”.the sample”.
Beer-Lambert Law will obey:
the light used must be monochromatic
the concentrations must be low
the solution must not be fluorescent or heterogeneous
the solute must not undergo to photochemical transformations
the solute must not undertake variable associations with the
solvent.
Chromophores
What are the atoms or groups in the molecules shows
responses to UV-Vis spectrum?
Group /part of a molecule responsible for itsColouris called
Chromophores.
The functional groups containing multiple bonds capable of
absorbing radiaons above 200 nm due to n → π* & π → π*
transitions.
Chromophoregroup:
Eg-C=C-, -C=O, -C=N, -C≡N, -C=S, -NO2, -N=O, etc
Auxochrome
Auxochrome:
The functional groups attached to a Chromosphoreswhich
modifies the ability of the Chromosphoress to absorb light ,
altering the wavelength or intensity of absorption.
Auxochromeeg. –OH, -NH2, -OCH3, -halogens etc
Benzene absorption -λmax = 255 nm
-OH group in Benzene -Phenol λmax = 270 nm
-NH2group in Benzene Aniline λmax = 280 nm
Factors Influence the Absorption
BathochromicShift (Red Shift):
Absorption maxima (λmax) of a compound shifts to longer
wavelength.
It is due to presence of an auxochromeor by the change of solvent.
p-Nitrophenol-λ=255nm, but in alkaline medium λ= 265 nmp-Nitrophenol-λmax=255nm, but in alkaline medium λmax= 265 nm
HypsochromicShift (Blue Shift):
Absorption maxima (λmax) of a compound shifts to shorter
wavelength.
It is due to presence of an group causes removal of conjugation or by
the change of solvent.
Aniline (λmax280nm) & shows blue shift in acidic medium λmax
265nm. It loses conjugation.
Factors Influence the Absorption
HyperchromicEffect:
An increase in molar absorptivity(ε) of a compound.
If Auxochromeintroduces in the molecule , the intensity of
absorption increasesabsorption increases
Eg.Pyridineλ max = 257nm & 2-Methyl pyridine-λ max = 260 nm
HypochromicEffect:
An increase in molar absorptivity(ε) of a compound.
Eg. Naphthalene ε = 19000 & 2-Methylnaphthalene ε= 10250
Wavelength Shift
RED SHIFT
BLUE SHIFT
HYPERCHROMIC SHIFT
Absorbance (A)
BLUE SHIFT
HYPOCHROMIC SHIFT
Absorbance (A)
Wavelength ( λ)
λmax
UV & Visible Spectrum Range
Wavelength Range
Absorbed (WL-nm)*
Colour
Absorbed
Colour
Seen By Eye
380 -450 Violet Yellow -Green
450 -495 Blue Yellow450 -495 Blue Yellow
495 -570 Green Violet
570 -590 Yellow Blue
590 -620 Orange Green -Blue
620 -760 Red Blue -Green
UV-Vis Spectrophotometer
UV-Vis Spectrophotometerconsists:
Light source
Monochromatoror Polychromator
Sampling areaSampling area
Detector
Computer with processing
UV-Vis Spectrophotometers are available as single beam or double beam
UV-Vis Spectrophotometer
Light source : Deuterium lamp for the UV region and , a tungsten-
halogen lamp for thevisibleregion or a xenon lamp to cover the
entire UV-Vis range.
Filters or Monochromators: Wave Selectors
Note: Basic models will have Filters. Gelatin colouredwith organic dyes that
are sealed between Glass plates. Monochromatorsare commonly used.
Sample Holder Area: To hold the Cuvette(s ) for blank /sample on
the path of monochromatic light.
Cuvettes: Normally 1 cm Rectanlecell made in high-purity quartz
or sapphire transparent to UV-Vis radiation.
Note: Normal Glass absorbs will absorb uvradiation
UV-Vis Spectrophotometer
Slit –entry of polychromatic light from the
source.
Collimating device –lens or mirror which
helps in reflecting the polychromatic light
Monochromatorsor Polychromators:
Concave
lens
helps in reflecting the polychromatic light
to the dispersion device.
A wavelength resolving device –Grating.
Design single beam or double-beam
spectrophotometers requirement
A focussinglens or mirror
Exit slit
Grating Monochromator
Focal plane
λ2
λ1
UV-Vis Spectrophotometer
UV-VIS detectors:
A detector produces an electric signal when it is struck by photons.
Phototube emits electrons from a photosensitive, negatively charged
surface (the cathode) when struck by visible light or ultraviolet radiation.
The electrons flow through a vacuum to a positively charged collector The electrons flow through a vacuum to a positively charged collector
whose current is proportional to the radiation intensity.
Photoelectric detectors, are the most common. It generate an electric
current that is directly proportional to the intensity of the radiant energy
incident upon them.
Photosensitive semiconductor devices, either discrete detectors, linear or
two-dimensional arrays, or photomultipliers or photodiodes.
Data processing: connected to Suitable computeriseddata processing &
evaluation systems.
Operation of the Spectrophotometer
Set spectrum mode or Photometry mode .
Select the scan range or wavelength max
Zero correction before starting the analysis .
Baseline line flatness : ±0.001abs or Photometric noise < 0.001 Baseline line flatness : ±0.001abs or Photometric noise < 0.001
Select a suitable spectroscopic blank e.g. air, blank solvent, solid
material
Blank run value may differ by NMT ±2nm
Quantitative measurements: Absorption values less than 2.0
Wavelength must not exceed 0.4 and is preferably < 0.2nm.
To improve resolution or sensitivity, use derivative spectra
Operation of the Spectrophotometer
Software used shall compliance with 21 CFR Part 11
Ensure the calibration is performed and within validity date
Use 1cm Cell unless otherwise stated.
Cell used must be clean, no finger print, no deposit or contaminantCell used must be clean, no finger print, no deposit or contaminant
Perform system suitability is required.
Check the lamp energy
Use only spectroscopic reagents / solvents
Qualification
Ensure the Qualification is performed and documented
before use of the Instrument.
Design Qualification / URS
Installation QualificationInstallation Qualification
Operational Qualification
Performance Qualification
Reference guideline for qualification:
PA/PH/OMCL (07) 11 Def Corr: Qualification Of Equipment-
Annex 3: Qualification Of UV-Visible Spectrophotometers
Performance Check of Instrument
Method
Wavelength
Accuracy
Absorbance
Accuracy
Photometric
Linearity
Stray
Light
Quantitative or limit test*
Based on measurement of the
absorbance at one or more
identified wavelengths (e.g. assay
or impurities test)
X X X X
or impurities test)
Identification test
Based on wavelength of
absorption maxima & minima
X - - X
Based on absorption
measurement and wavelength
of absorption maxima
X X - X
Based on comparison of
spectrum with that of
reference substance
X X - -
*Resolution/spectral bandwidth: As required in the monograph
Control of Wavelength Accuracy
Material Wavelengths (nm)
*
Solutions :
Didymium in Perchloricacid511.8; 731.6; 794.2
Ensure the control the wavelength accuracy for appropriate number of
bands in the intended spectral range using below reference materials.
511.8; 731.6; 794.2
Holmium in Perchloricacid
(Generally used )
241.1; 287.2; 361.3; 451.4; 485.2; 536.6; 640.5
Solid filters :
Holmium Glass
(Generally used )
279.3; 360.9; 453.4; 637.5
Lamps : Deuterium 486.0; 656.1
Lamp : Mercury (low pressure)
184.9; 253.7; 312.5; 365.0; 404.7; 435.8; 546.1; 577.0;
579.1
Acceptance Limit: 200-400 ±1nm & above 400nm ±3nm .
Control of Absorbance
Absorbance Accuracy:
Nicotinic acid solution: Dissolve 57.0-63.0mg ofNicotinic acid
(RS/CRS) in 0.1M hydrochloric acid solution and dilute to 200mL.
Dilute 2.0mL of the solution to 50mL. (final conc-2mg/L).
Measured the absorbance at 213nm & 261nm.
Acceptance criteria: The difference between the measured
absorbance and the absorbance of the standard value (RS) should
be ±0.010 or ±1per cent, whichever is greater.
Photometric Linearity :
Measure the Absorbance at 5.0 -40.0 mg/L Nicotinic acid solution
of min 4 concentration and determine coefficient.
Acceptance criteria: The coefficient of determination (R²) is not
less than0.999.
Limit of Stray Light
Stray light is determined at an appropriate wavelength
using suitable solid or Solution
Use 1cm cell andwateras reference liquid.
Absorbance of Potassium chloride12g/L solution at Absorbance of Potassium chloride12g/L solution at
198nm is not less than 2.0
Sodium iodide10g/L solution at 220nm
Potassium iodide10g/L solution at 250nm or
Sodium nitrite-50g/L solution at 340nm and 370nm
Acceptance Limit: Absorbance must not Less than 3.0
Cuvettes/Cells
Cell used should be 1cm path length.
Measured value obtained may not differ by ±2nm, unless
otherwise prescribed in monograph
Quantitative measurements relying on absorption values above 2.0
should be avoided.
Acceptance criteriafor Cuvettesas per Eu. Pharm
The apparent absorbance is not greater than 0.093 for 1cm quartz
cuvettes(UV region) and 0.035 for 1cm glass cuvettes
(visibleregion);
The absorbance measured after rotation (180°) does not differ by
more than 0.005 from the value previously obtained.
Control of Resolution
Measure the resolution of the equipment as per
monograph using reference materials
Alternately record the spectrum of a 0.02% (v/v) solution
ofTolueneinHexaneorin Heptane.
Use Hexane/ Heptaneas the compensation liquid.
Acceptance criterion:
The minimum ratio of the absorbance at the maximum
(269nm) to that at the minimum (266nm) is stated in the
monograph.
Calibration
Spectrophotometer shall be Calibrated to ensure that Instrument
is performing well and measurement is accurate & reliable.
Internal calibration
Match pairing of cells (Cuvettequalification)
Control of wavelengthControl of wavelength
Control of absorbance
Limit of Stray Light
Resolution Power
Linearity study
Use reference standard for calibration is NIST traceable or certified
Refer for further detail Ph. Eur. 10 2.2.25 and USP 42-NF 37 GC <857>
Calibration: Match Pairing Cells
Internal Calibration of UV Spectrophotometer :
Perform the internal Calibration as per manufacturer’s instruction
using software
Match Pairing of Cells (CuvetteQualification):
Fill the cell with distilled water and measure the absorbance
against air at 240 nm for quartz cells and 650 nm for glass cells.
Absorbance should not be greater than 0.093 for 1 cm quartz cells
(UV region) and 0.035 for 1 cm glass cells (Visible region).
Rotate the cell in its holder (180°) again check the absorbance
Difference not greater than 0.005 from initial.
Calibration :Control of Wavelength:
Take the UV spectrum of 4%w/v Holmium oxide in 1.4 M Perchloric
acid solution from 200 nm to 600 nm against the 1.4 M Perchloric
acid as a blank.
Wavelength shall be check for the peak detection of Holmium Wavelength shall be check for the peak detection of Holmium
Oxide at 241.15 nm, 287.15 nm, 361.5 nm, 486.0 nm & 536.3 nm.
The permitted tolerance limit shall be ±1 nm for the range of 200
nm to 400 nm (UV range) and ±3 nm for the range of 400 nm to
800 nm.(Visible range)
Calibration: Control of Absorbance:
Take the spectrum of the Potassium dichromate(60ppm) solutionbetween
200 nm to 400 nm using 0.005 M Sulfuric acid as a blank.
Measure the absorbance of peak detection at 350 nm & 257 nm and Valley
detection at 313 nm & 235 nm.
Absorbance of the Potassium dichromate60ppm at 430 nm using 0.005
M Sulfuric acid as a blank in photometric modeM Sulfuric acid as a blank in photometric mode
Control of absorbance =(AbsorbanceX10000 ) / Wt. Taken in mg.
Control of absorbance (for λ 430 nm) =(Absorbance X 1000) / Wt (g)X100
Wavelength Maximum Tolerance
235 nm 122.9 nmto 126.2 nm
257 nm 142.8 nm to 146.2 nm
313 nm 47.0 nm to 50.3 nm
350 nm 105.6 nm to 109.0 nm
430 nm 15.7nmto 16.1nm
Calibration: Stray Light
Limit of Stray Light :
Prepare Potassium chloride of 12,000 ppmin distilled water or Use
certified standard solution
Measure the absorbance of the potassium chloride solution against
distilled water as a blank between 220 nm and 190 nm in scan mode.distilled water as a blank between 220 nm and 190 nm in scan mode.
Check absorbance at 198 nm by keeping cursor.
Absorbance steeply increases between 220 nm to 200 nm and shall be
more than 2.0 at 200nm
Resolution Power:
Measure Toluene solution in Hexane (0.02%v/v) refer performance check
Calibration :Linearity
Linearity:
Prepare potassium dichromate using 0.005 M Sulfuric Acid at
20ppm, 40ppm, 60ppm, 80ppm & 100ppm
Measure the absorbance of the solutions at 257 nm by using Measure the absorbance of the solutions at 257 nm by using
0.005 M sulfuric acid as a blank.
Plot a graph of absorbance verses concentration.
Co-relation co-efficient R
2
shall be > 0.999.
Tips to Handle the Cells
Carefully clean & store properly to avoids contamination.
Clean Cells with high purity water, if required clean with 1% (v/v) nitric
acid, do not use chromic acid for cleaning.
Do not use cracked or scratched cells.
Wipe cells carefully with a soft, clean, lint-free cloth while use.Wipe cells carefully with a soft, clean, lint-free cloth while use.
Contaminated cells are major source of error
Cells should never be handled by the optical polished faces.
Should rinse off residual or spilled solution.
Do’tuse Strong Acid or Alkaline solutions. Impact will be based on the pH
and contact time.
Try to used in the cell same beam ref/ test by marking
Application
Method : External standard & Internal standard (calibration), Standard
addition and etc
Enzymatic Analysis in biochemical and Clinical Lab. EgSugar, acids, or their
salts, alcohol etc
Analysis of nucleic acids, proteins and bacterial cell cultures. To detremineAnalysis of nucleic acids, proteins and bacterial cell cultures. To detremine
Concentration & Purity of nucleic acids –DNA and RNA
Enzyme base test kits are readlyavailable
Food analysis
Absorbance of co-enzyme NADH or NADPH (340nm)
Chloesterolin mayonnaise by oxidation method
Application
Qualitative & Quantitative Analysis:
Identification / characterizing aromatic compounds and conjugated
olefins.
Detection of impurities in organic compound and solvents.
Detection of isomers are possible.Detection of isomers are possible.
Determine of assay , molar concentration of the solute .
Determination of pKa
Determination of molecular weight using Beer’s law.
Determination of most of metal ions, by preparing colouredcomplex with
suitable ligand. Eg. Iron –1,10 Phenanthroline
Analysis of Pharmaceuticals
Analysis of Pharmaceutical ingredients
Pseudoephedrine hydrochloride
Triprolidinehydrochloride
Codeine , Morphine
Vitamins
Pharmaceutical dosage formPharmaceutical dosage form
Simultaneous Equation methodfor determination of binary / ternary mixture in
Rabeprazolesodium and Levosulpirideat 284 nm, 232 nm methanol
Tenofovir, Efavirenz, and Lamivudineat 260 nm, 347 nm, 272 nm (methanol)
Difference Spectrophotometry
Pioglitazoneand Metforminphosphate buffer (pH 9) and chloride buffer (pH 2) 228.1 nm
and 228.2 nm
Zero crossing technique to analysis of binary mixtures
Gatifloxacinand Prednisolone348 and 263 nm
Reference
Analytical Chemistry -7
th
Edition. Gary D. Christian, PurnenduK. (Sandy)
Dasgupta& Kevin A. Schug
Quantitative Chemical Analysis. 7
th
Edition Daniel C. Harris
Chemical Analysis: Modern Instrumentation Methods and Techniques
Francis and AnnickRouessacand Steve Brooks, 2007-John Wiley & Sons Ltd,.
Vogel’s –Quantitative Chemical Analysis-6
th
editionVogel’s –Quantitative Chemical Analysis-6
th
edition
PA/PH/OMCL (07) 11 Def Corr: Qualification Of Equipment-Annex 3:
Qualification Of UV-Visible Spectrophotometers
European Pharmacopeia General Chapter 2.2.25. Absorption
Spectrophotometer, Ultraviolet AndVisible
USP General Chapter <857> and <1857>Ultraviolet-Visible Spectroscopy &
Theory And Practice
Ultraviolet Spectroscopy and its Pharmaceutical Applications-A Brief review
DipaliM Atoleet al ; Asian J PharmClinRes, Vol11, Issue 2, 2018, 59-66