10 and 20 electrode placement

19,984 views 15 slides Apr 22, 2018
Slide 1
Slide 1 of 15
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15

About This Presentation

EEG 10 - 20 electrode placement which help to technologist how to measure the placement.


Slide Content

Mohibullah Fazli Kakar
PT, NP Technologist
3
rd
year medical student
FMIC; April 2018
Electroencephalography
10 – 20 system electrode placement

What is 10 – 20 system
An internationally recognized method that allows EEG
electrode placement to be standardized.
OR
The 10-20 System of Electrode Placement is a method
used to describe the location of scalp electrodes

The 10-20 system is based on the relationship between
the location of an electrode and the underlying area of
cerebral cortex.
Each site has a letter and a number or another letter to
identify the hemisphere location.
The letters F, T, C, P, and O stand for Frontal,
Temporal, Central, Parietal and Occipital.
Even numbers (2,4,6,8) refer to the right hemisphere
Odd numbers (1,3,5,7) refer to the left hemisphere.
The z refers to an electrode placed on the midline.
What is 10 – 20 system

Four Skull Landmarks:
Nasion
Inion
Left Pre auricular point

Right Pre auricular point

What is 10 – 20 system

Pre – Auricular point

Measuring FPz or Reference Electrode:
Measure the distance from nasion to inion
 measure 10% of all distance from nasion and mark for
Reference electrode in the front.
Measuring Oz:
Measure the distance from nasion to inion
 measure 10% of all distance from inion and mark for Oz
in the back.
What is 10 – 20 system
Note:
•Every 10% = 20%
•Every 10% / 2 = 10%

Measurement of CZ:
Measure the distance from FPz to Oz and divide/ 2 = CZ
Measurement of FZ:
Measure the distance form CZ – Ref
divide by 2 = Fz
Measurement of PZ :
Measure the distance from Cz – Oz divide / 2 = PZ
What is 10 – 20 system
FPZ

Measurement of FP1 and FP2:
Measure the distance head circumference and take 10% of all
measurement divide / 2 mark it:
On left side for FP1
On right side for FP2
Measurement of F7 and F8:
Measure the distance head circumference and take 10% of all
measurement
Mark 10% by measuring for every led
What is 10 – 20 system

Measurement of T3 and T4:
Measure the distance between pre – auricular points and cross
Cz
Mark 10% of the distance from left and right pre – auricular
points of ears.
Left for T3
Right for T4
What is 10 – 20 system

Measurement of T5 and T6:
Measure the distance head circumference and take 10% of all
measurement
Mark 10% by measuring for every led
FPz 10%  F7 10%  T3 10%  T5 10%
FPz 10%  F8 10%  T4 10%  T6 10%
What is 10 – 20 system

Measurement of F3 and F4:
Measure the distance from F7 – Fz / 2
Mark on the left for F3
Mark on the Right for F4
What is 10 – 20 system

Measurement of C3 and C4:
Measure the distance form CZ – T3 / 2 = C3
Measure the distance form CZ – T4 / 2 = C4
Measurement of P3 and P4:
Measure the distance from T5 – PZ / 2
Mark on the left for P3
Mark on the Right for P4
What is 10 – 20 system

Measurement of O1 and O2:
 Measure the distance head circumference and take 10% of all
measurement mark it:
On Left side for O1 = T5 10% = O1
On Right side for O2 = T6 10% = O2
What is 10 – 20 system