405060940-s2000d-service-manual-pdf.pdf

417 views 184 slides Jul 10, 2022
Slide 1
Slide 1 of 439
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79
Slide 80
80
Slide 81
81
Slide 82
82
Slide 83
83
Slide 84
84
Slide 85
85
Slide 86
86
Slide 87
87
Slide 88
88
Slide 89
89
Slide 90
90
Slide 91
91
Slide 92
92
Slide 93
93
Slide 94
94
Slide 95
95
Slide 96
96
Slide 97
97
Slide 98
98
Slide 99
99
Slide 100
100
Slide 101
101
Slide 102
102
Slide 103
103
Slide 104
104
Slide 105
105
Slide 106
106
Slide 107
107
Slide 108
108
Slide 109
109
Slide 110
110
Slide 111
111
Slide 112
112
Slide 113
113
Slide 114
114
Slide 115
115
Slide 116
116
Slide 117
117
Slide 118
118
Slide 119
119
Slide 120
120
Slide 121
121
Slide 122
122
Slide 123
123
Slide 124
124
Slide 125
125
Slide 126
126
Slide 127
127
Slide 128
128
Slide 129
129
Slide 130
130
Slide 131
131
Slide 132
132
Slide 133
133
Slide 134
134
Slide 135
135
Slide 136
136
Slide 137
137
Slide 138
138
Slide 139
139
Slide 140
140
Slide 141
141
Slide 142
142
Slide 143
143
Slide 144
144
Slide 145
145
Slide 146
146
Slide 147
147
Slide 148
148
Slide 149
149
Slide 150
150
Slide 151
151
Slide 152
152
Slide 153
153
Slide 154
154
Slide 155
155
Slide 156
156
Slide 157
157
Slide 158
158
Slide 159
159
Slide 160
160
Slide 161
161
Slide 162
162
Slide 163
163
Slide 164
164
Slide 165
165
Slide 166
166
Slide 167
167
Slide 168
168
Slide 169
169
Slide 170
170
Slide 171
171
Slide 172
172
Slide 173
173
Slide 174
174
Slide 175
175
Slide 176
176
Slide 177
177
Slide 178
178
Slide 179
179
Slide 180
180
Slide 181
181
Slide 182
182
Slide 183
183
Slide 184
184
Slide 185
185
Slide 186
186
Slide 187
187
Slide 188
188
Slide 189
189
Slide 190
190
Slide 191
191
Slide 192
192
Slide 193
193
Slide 194
194
Slide 195
195
Slide 196
196
Slide 197
197
Slide 198
198
Slide 199
199
Slide 200
200
Slide 201
201
Slide 202
202
Slide 203
203
Slide 204
204
Slide 205
205
Slide 206
206
Slide 207
207
Slide 208
208
Slide 209
209
Slide 210
210
Slide 211
211
Slide 212
212
Slide 213
213
Slide 214
214
Slide 215
215
Slide 216
216
Slide 217
217
Slide 218
218
Slide 219
219
Slide 220
220
Slide 221
221
Slide 222
222
Slide 223
223
Slide 224
224
Slide 225
225
Slide 226
226
Slide 227
227
Slide 228
228
Slide 229
229
Slide 230
230
Slide 231
231
Slide 232
232
Slide 233
233
Slide 234
234
Slide 235
235
Slide 236
236
Slide 237
237
Slide 238
238
Slide 239
239
Slide 240
240
Slide 241
241
Slide 242
242
Slide 243
243
Slide 244
244
Slide 245
245
Slide 246
246
Slide 247
247
Slide 248
248
Slide 249
249
Slide 250
250
Slide 251
251
Slide 252
252
Slide 253
253
Slide 254
254
Slide 255
255
Slide 256
256
Slide 257
257
Slide 258
258
Slide 259
259
Slide 260
260
Slide 261
261
Slide 262
262
Slide 263
263
Slide 264
264
Slide 265
265
Slide 266
266
Slide 267
267
Slide 268
268
Slide 269
269
Slide 270
270
Slide 271
271
Slide 272
272
Slide 273
273
Slide 274
274
Slide 275
275
Slide 276
276
Slide 277
277
Slide 278
278
Slide 279
279
Slide 280
280
Slide 281
281
Slide 282
282
Slide 283
283
Slide 284
284
Slide 285
285
Slide 286
286
Slide 287
287
Slide 288
288
Slide 289
289
Slide 290
290
Slide 291
291
Slide 292
292
Slide 293
293
Slide 294
294
Slide 295
295
Slide 296
296
Slide 297
297
Slide 298
298
Slide 299
299
Slide 300
300
Slide 301
301
Slide 302
302
Slide 303
303
Slide 304
304
Slide 305
305
Slide 306
306
Slide 307
307
Slide 308
308
Slide 309
309
Slide 310
310
Slide 311
311
Slide 312
312
Slide 313
313
Slide 314
314
Slide 315
315
Slide 316
316
Slide 317
317
Slide 318
318
Slide 319
319
Slide 320
320
Slide 321
321
Slide 322
322
Slide 323
323
Slide 324
324
Slide 325
325
Slide 326
326
Slide 327
327
Slide 328
328
Slide 329
329
Slide 330
330
Slide 331
331
Slide 332
332
Slide 333
333
Slide 334
334
Slide 335
335
Slide 336
336
Slide 337
337
Slide 338
338
Slide 339
339
Slide 340
340
Slide 341
341
Slide 342
342
Slide 343
343
Slide 344
344
Slide 345
345
Slide 346
346
Slide 347
347
Slide 348
348
Slide 349
349
Slide 350
350
Slide 351
351
Slide 352
352
Slide 353
353
Slide 354
354
Slide 355
355
Slide 356
356
Slide 357
357
Slide 358
358
Slide 359
359
Slide 360
360
Slide 361
361
Slide 362
362
Slide 363
363
Slide 364
364
Slide 365
365
Slide 366
366
Slide 367
367
Slide 368
368
Slide 369
369
Slide 370
370
Slide 371
371
Slide 372
372
Slide 373
373
Slide 374
374
Slide 375
375
Slide 376
376
Slide 377
377
Slide 378
378
Slide 379
379
Slide 380
380
Slide 381
381
Slide 382
382
Slide 383
383
Slide 384
384
Slide 385
385
Slide 386
386
Slide 387
387
Slide 388
388
Slide 389
389
Slide 390
390
Slide 391
391
Slide 392
392
Slide 393
393
Slide 394
394
Slide 395
395
Slide 396
396
Slide 397
397
Slide 398
398
Slide 399
399
Slide 400
400
Slide 401
401
Slide 402
402
Slide 403
403
Slide 404
404
Slide 405
405
Slide 406
406
Slide 407
407
Slide 408
408
Slide 409
409
Slide 410
410
Slide 411
411
Slide 412
412
Slide 413
413
Slide 414
414
Slide 415
415
Slide 416
416
Slide 417
417
Slide 418
418
Slide 419
419
Slide 420
420
Slide 421
421
Slide 422
422
Slide 423
423
Slide 424
424
Slide 425
425
Slide 426
426
Slide 427
427
Slide 428
428
Slide 429
429
Slide 430
430
Slide 431
431
Slide 432
432
Slide 433
433
Slide 434
434
Slide 435
435
Slide 436
436
Slide 437
437
Slide 438
438
Slide 439
439

About This Presentation

manual


Slide Content

Technical
Publications
2302228-100
Revision 0
Seno
graphe 2000 D Acquisition System
Service Manual
sm
do not duplicate
Copyright
©
1999, 2000, 2001 by General Electric Co.

Page no. 2 COVER.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
X-RAY WARNING
WARNING.FM Page no. 3
X-RAY WARNING
ATTENTION
Les appareils à rayons X sont dangereux à la fois pour le patient et pour le manipulateur si les
mesures de protection ne sont pas strictement appliquees
Bien que cet appareil soit construit selon les normes de sécurité les plus sévères, la source de rayonnement X représente un danger lorsque le
manipulateur est non qualifié ou non averti. Une exposition excessive au rayonnement X entraîne des dommages à l’organisme.
Par conséquent, toutes les précautions doivent être prises pour éviter que les personnes non autorisées ou non qualifiées utilisent cet appareil
créant ainsi un danger pour les autres et pour elles-mêmes.
Avant chaque manipulation, les personnes qualifiées et autorisées à se servir de cet appareil doivent se renseigner sur les mesures de protec-
tion établies par la Commission Internationale de la Protection Radiologique, Annales 60 : Recommandations de la Commission Internationale
sur la Protection Radiologique et les normes nationales en vigueur.
WARNING
X-ray equipment is dangerous to both patient and operator unless measures of protection are
strictly observed
Though this equipment is built to the highest standards of electrical and mechanical safety, the useful x-ray beam becomes a source of danger
in the hands of the unauthorized or unqualified operator. Excessive exposure to x-radiation causes damage to human tissue.
Therefore, adequate precautions must be taken to prevent unauthorized or unqualified persons from operating this equipment or exposing them-
selves or others to its radiation.
Before operation, persons qualified and authorized to operate this equipment should be familiar with the Recommendations of the International
Commission on Radiological Protection, contained in Annals Number 60 of the ICRP, and with applicable national standards.
ATENCION
Los aparatos de rayos X son peligrosos para el paciente y el manipulador cuando las normas de
proteccion no estan observadas
Aunque este aparato está construido según las normas de seguridad más estrictas, la radiación X constituye un peligro al ser manipulado por
personas no autorizadas o incompetentes. Una exposición excesiva a la radiación X puede causar daños al organismo.
Por consiguiente, se deberán tomar todas las precauciones necesarias para evitar que las personas incompetentes o no autorizadas utilicen
este aparato, lo que sería un peligro para los demás y para sí mismas.
Antes de efectuar las manipulaciones, las personas habilitadas y competentes en el uso de este aparato, deberán informarse sobre las normas
de protección fijadas por la Comisión Internacional de la Protección Radiológica, Anales No 60: Recomendaciónes de la Comisión Internacio-
nal sobre la Protección Radiológica y normas nacionales.
ACHTUNG
Röntgenapparate sind eine gefahr fÜr patienten sowie bedienungspersonal, wenn die geltenden
sicherheitsvorkehrungen nicht genau beachtet werden
Dieser Apparat entspricht in seiner Bauweise strengsten elektrischen und mechanischen Sichereitsnormen, doch in den Händen unbefugter
oder unqualifizierter Personen wird er zu einer Gefahrenquelle. Übermäßige Röntgenbestrahlung ist für den menschlichen Organismus
schädlich.
Deswegen sind hinreichende Vorsichtsmaßnahmen erforderlich, um zu verhindern, daßunbefugte oder unqualifizierte Personen solche Geräte
bedienen oder sich selbst und andere Personen deren Bestrahlung aussetzen können.
Vor Inbetriebnahme dieses Apparats sollte sich das qualifizierte und befugte Bedienungspersonal mit den geltenden Kriterien für den gefahr-
losen Strahleneinsatz durch sorgfältiges Studium des Hefts Nr. 60 der Internationalen Kommission für Strahlenschutz (ICRP) vertraut machen:
Empfehlungen der Internationalen Kommission für Strahlenschutz und anderer nationaler Normenbehörden.

Page no. 4 WARNING.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
X-RAY WARNING This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
language warning.fm Page no. 5
WARNING
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN
CONSULTED AND IS UNDERSTOOD
If a customer's service provider requires a language other than English, it is the customer's responsibility
to provide translation services.
This Service Manual is available in English only.
Failure to heed this warning may result in injury to the service provider, operator or patient from electric
shock, mechanical or other hazards.
ATTENTION
NE PAS TENTER D'INTERVENIR SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N'A PAS ÉTÉ
CONSULTÉ ET COMPRIS
Ce Manuel de service n'est disponible qu'en anglais.
Si le technicien du client a besoin de ce manuel dans une autre langue que l'anglais, c'est au client qu'il
incombe de le faire traduire.
Le non-respect de cet avertissement peut entraîner chez le technicien, l'opérateur ou le patient des
blessures dues à des dangers électriques, mécaniques ou autres.
ATENCION
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE
MANUAL DE SERVICIO.
Este Manual de Servicio sólo existe en inglés.
Si algún proveedor de servicios ajeno a GEMS solicita un idioma que no sea el inglés, es responsabil-
idad del cliente ofrecer un servicio de traducción.
La no observancia del presente aviso puede dar lugar a que el proveedor de servicios, el operador o el
paciente sufran lesiones provocadas por causas eléctricas, mecánicas o de otra naturaleza.
WARNUNG
ERSUCHEN SIE NICHT DIESE ANLAGE ZU WARTEN, OHNE DIESE SERVICEANLEITUNG GELESEN UND
VERSTANDEN ZU HABEN.
Diese Serviceanleitung existiert nur in englischer Sprache.
Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es Aufgabe des Kunden für eine
entsprechende Übersetzung zu sorgen.
Wird diese Warnung nicht beachtet, so kann es zu Verletzungen des Kundendiensttechnikers, des Bedi-
eners oder des Patienten durch Stromschläge, mechanische oder sonstige Gefahren kommen.

Page no. 6 language warning.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
ATENÇAO
NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE
ASSISTÊNCIA TÉCNICA
Este Manual de Assistência Técnica só se encontra disponível em Inglês.
Se qualquer outro serviço de assistência técnica, quE não a GEMS, solicitar estes manuais noutro idi-
oma, é da responsabilidade do cliente fornecer os serviços de tradução.
O não cumprimento deste aviso pode por em perigo a segurança do técnico, operador ou paciente dev-
ido a` choques elétricos, mecânicos ou outros.
AVVERTENZA
SI PROCEDA ALLA MANUTENZIONE DELL'APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL
PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO
Il presente manuale di manutenzione è disponibile soltanto in inglese.
Se un addetto alla manutenzione esterno alla GEMS richiede il manuale in una lingua diversa, il cliente
è tenuto a provvedere direttamente alla traduzione.
Non tenere conto della presente avvertenza potrebbe far compiere operazioni da cui derivino lesioni
all'addetto alla manutenzione, all'utilizzatore ed al paziente per folgorazione elettrica, per urti meccanici
od altri rischi.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Transport Damage.
DAMAGE.FM Page no. 7
Transport Damage.
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “dam-
age in shipment” written on all copies of the freight or express bill before delivery is accepted or “signed
for” by a General Electric representative or a hospital receiving agent. Whether noted or concealed, dam-
age MUST be reported to the carrier immediately upon discovery, or in any event within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation company will
not pay a claim for damage if an inspection is not requested within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 785-5052/8*323-5052 immediately after damage is
found. At this time be ready to supply name of carrier, delivery date, consignee name, freight or express
bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bul-
letins.

Page no. 8 DAMAGE.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Transport Damage. This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Electrical Contractors.
CONTRACT.FM Page no. 9
Electrical Contractors.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT
All electrical installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. In addition, electrical feeds into the
Power Distribution Unit shall be performed by licensed electrical contractors. Other connections between
pieces of electrical equipment, calibrations, and testing shall be performed by qualified GE Medical per-
sonnel. The products involved (and the accompanying electrical installations) are highly sophisticated,
and special engineering competence is required. In performing all electrical work on these products, GE
will use its own specially trained field engineers. All of GE’s electrical work on these products will comply
with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, person-
nel of third–party service companies with equivalent training, or licensed electricians) to perform electri-
cal servicing on the equipment.

Page no. 10 CONTRACT.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Electrical Contractors. This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Table of Contents
FFDMSMTOC.FM Page no. 11
Table of Contents
X-RAY WARNING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Transport Damage.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Electrical Contractors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Chapter 1: Introduction
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Chapter 2: Installation
Job Card IST 000 - AWS System Configuration Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Job Card IST 001 - Installation Steering Guide.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Job Card IST 002 - Unpacking the Gantry and Generator Cabinet . . . . . . . . . . . . . . . . . . . 37
Job Card IST 003 - Unpacking the Cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Job Card IST 004 - Senographe Physical Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Job Card IST 005 - Cart Physical Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch). . . . . . . . . . . . 47
Job Card IST 007 - Cart Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Job Card IST 008 - AC Connection.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Job Card IST 009 - First System Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Job Card IST 010 - Line Resistance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Job Card IST 011 - Set Elevator Upper Travel Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Job Card IST 012 - Safety Labelling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Job Card IST 013 - Room Lamps and Door Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Job Card IST 014 - Automatic X-ray Tube Warm-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Job Card IST 015 - Installation of the CD-R writer (option). . . . . . . . . . . . . . . . . . . . . . . . . . 85
Job Card IST 016 - Modem installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Job Card IST 017 - Image Receptor Assembly/Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Job Card IST 018 - Bar Code Scanner (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Chapter 3: Configuration
Job Card CFG 001 - The Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Job Card CFG 002 - Gantry and Generator Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Job Card CFG 003 - Senographe Jumpers and Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Job Card CFG 004 - Check Supply Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Job Card CFG 005 - AWS Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Job Card CFG 006 - Declare Network Hosts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Job Card CFG 008 - Networking Connection and Configuration . . . . . . . . . . . . . . . . . . . . . 123
Job Card CFG 009 - Declare Printers (Filming Devices) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Chapter 4: Calibration
Job Card CAL 000 - Calibration Interactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141

Page no. 12 FFDMSMTOC.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Table of Contents Job Card CAL 001 - Heater Current Scale Factor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Job Card CAL 002 - X-Ray Tube Focal Bias Voltage Scale Factor . . . . . . . . . . . . . . . . . . . . 149
Job Card CAL 003 - X-Ray Tube Heater Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Job Card CAL 004 - X-Ray Tube mA Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Job Card CAL 005 - Calibration of kV Scale Factor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Job Card CAL 006 - Calibration of Compression Force Detector . . . . . . . . . . . . . . . . . . . . 163
Job Card CAL 007 - Calibration of Breast Thickness Measurement . . . . . . . . . . . . . . . . . . 165
Job Card CAL 008 - Compression/Decompression Minimum Force . . . . . . . . . . . . . . . . . . 167
Job Card CAL 009 - Compression Arm Angle Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Job Card CAL 010 - Tube Tilt Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Job Card CAL 011 - X-ray Beam Alignment Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Job Card CAL 012 - Collimator Format Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Job Card CAL 013 - Bad Pixel Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
Job Card CAL 014 - Detector Gain Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Job Card CAL 015 - Conversion Factor Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Job Card CAL 016 - Half-Value Layer Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Job Card CAL 017 - AOP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Job Card CAL 018 - ISC Monitor Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
Job Card CAL 019 - Light Beam Centering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Job Card CAL 020 - mAs Non-linearity Calibration (Automatic) . . . . . . . . . . . . . . . . . . . . . 205
Chapter 5: Functional Checks
Job Card VF 001 - Boot/Init and Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
Job Card VF 002 - Cart Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217
Job Card VF 003 - UPS Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
Job Card VF 005 - Senographe Functional Checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Job Card VF 008 – IQ Tools Flat Field Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
Job Card VF 009 - Image Acquisition and ACR Score . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
Job Card VF 010 - Acquisition in AOP Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
Job Card VF 011 - Light Beam Centering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
Job Card VF 012 - Dicom Printer Adjustments and Checks. . . . . . . . . . . . . . . . . . . . . . . . . 237
Job Card VF 013 - Disk Full Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241
Chapter 6: Planned Maintenance
Job Card PM 001 - Senographe 2000 D Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245
Chapter 7: Error Message Information
Job Card ERR 001 - IDC Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 249
Job Card ERR 002 - AWS Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251
Job Card ERR 003 - Gantry and Generator Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 253
Chapter 8: DIagnostic Information
Job Card DIAG 001 - AWS Sun-VTS Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
Job Card DIAG 002 - AWS Ethernet Port Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 293

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Table of Contents
FFDMSMTOC.FM Page no. 13
Job Card DIAG 003 - AWS File System and SCSI Board Diagnostics . . . . . . . . . . . . . . . . . 297
Chapter 9: Renewal Parts
Renewal Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 301
Chapter 10: Disassembly/Reassembly
Job Card D/R 000 - Calibration after FRU Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 341
Job Card D/R 002 - Paddle board 800PL8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343
Job Card D/R 003 - Magnification sensor board 800PL14 . . . . . . . . . . . . . . . . . . . . . . . . . . 347
Job Card D/R 004 - Inter-arm tilt sensor board 800PL15. . . . . . . . . . . . . . . . . . . . . . . . . . . . 349
Job Card D/R 007 - Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 353
Job Card D/R 008 - Compression Motor and Belts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355
Job Card D/R 009 - Gantry CPU Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 361
Job Card D/R 010 - Optical Fiber (Gantry to Generator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363
Job Card D/R 011 - X-ray Tube Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 365
Job Card D/R 012 - Centering Device Lamp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 369
Job Card D/R 013 - Mirror. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 371
Job Card D/R 015 - Filter wheel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 373
Job card D/R 016 - Gas Spring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375
Job Card D/R 019 - Back-up/Restore procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 379
AWS Parameters manual backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 385
Job Card D/R 021 - Generator CPU Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395
Job Card D/R 022 - Capacitor box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 397
Job Card D/R 025 - HV and kV/mA tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 399
Job Card D/R 026 - Magnification device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 401
Job Card D/R 027 - Arm Angle Sensor (potentiometer) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 403
Job Card D/R 101 - Removal of Covers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 405
Job Card D/R 102 - Image Receptor Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 411
Job Card D/R 103 - Image Receptor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 413
Job Card D/R 104 - IDC unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417
Job Card D/R 105 - Conditioner Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 421
Job Card D/R 106 - Conditioner Fill and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 425
Job Card D/R 107 - AWS unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 429
Job Card D/R 108 - Software reload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 433
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 435

Page no. 14 FFDMSMTOC.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Table of Contents

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
INTRODUC.FM Page no. 15
CHAPTER 1 INTRODUCTION
This chapter contains introductory information on the Senographe 2000 D system.

Page no. 16 INTRODUC.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction
introduction.fm Page no. 17 Chapter 1
Introduction Chapter 1
1. MAJOR COMPONENTS
The Senographe 2000 D system includes the major components shown in the illustration below:
Also included, but not shown in the illustration, are the Network Connectivity kit and a number of acces-
sories (additional screen, compression paddles, etc.)
ILLUSTRATION 1 - SENOGRAPHE 2000D BASIC CONFIGURATION
2. OPTIONAL EQUIPMENT
Optional equipment available for use with the system includes:
1. Mass Archiving system
2. Laser Printer.
3. CD-R (CD Recordable) writing system.
4. Additional accessories
Protective screen Generator Cabinet Gantry
AWS Cart
Digital DetectorControl Console

Chapter 1 Page no. 18 introduction.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction
3. PRESENTATION
Senographe 2000 D is the new Digital Mammography System from GE Medical Systems. It has been
designed to perform Screening examinations as well as Diagnostic Views (including Spot compression
Magnified and/or Coned views). It is a
modular system that eliminates the need for film cassettes, and
takes advantage of digital technology, including on-screen image display, Networking, Filming, and
Archiving.
The Senographe 2000D is equipped with a dual track X-ray tube (molybdenum/rhodium) and a digital
detector. The digital detector is a flat panel of amorphous silicon on which cesium iodide is deposited to
maximize detection of X-rays. Positioning operations and X-ray exposure are controlled by the Control
Panel which also controls power to all parts of the Senographe 2000D system.
1. Senographe 2000 D's digital technology offers the capability to acquire Images in near-real time and
to process them, with the ability to vary brightness and contrast levels and manipulate images.
2. It also offers high examination productivity as compared with screen/film, and introduces new appli-
cations such as Networking and Archiving.
3. Senographe 2000 D is built on the DMR Platform, recognized for Image Quality superiority. The
Rhodium spectrum of the Senographe 2000 D's tube is well adapted to Digital Imaging.
The Senographe 2000D includes an acquisition work station ("AWS") monitor, keyboard and mouse,
computer, electronics, accessory storage, and uninterruptible power supply. The AWS is used for
image acquisition, processing, and display. The AWS can also be used for database management,
and can send images to archive, review, or filming.
4. The Acquisition Workstation displays acquired images in the room, allowing immediate evaluation of
breast positioning and possible motion blur, or adjustment of brightness and contrast.
Archiving, Networking, and Filming are all possible from the Acquisition Workstation, which can pro-
duce any number of equally high quality film copies as needed.
A hard-copy laser-film printer is used for image interpretation. Printer window width and window level
are set automatically, based on the image content. Images are displayed per film 1 on 1.
Several options are available for use with the Senographe 2000D system. These options include a
Senographe 2000D review workstation (not to be used for final interpretation of examinations), a
mass archiving system, a laser camera, networking capabilities, and CD-ROM interchange media.
5. The optional Review Workstation is a stand-alone workstation, with its own dedicated computer and
image database, connected to the Acquisition workstation by a high speed link. It supports image dis-
play and manipulation.
This powerful computer is equipped with two dedicated, very high resolution B & W monitors and a
dedicated keypad. Networking is possible from this workstation, as well as printing and receiving
images from an archive device.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction
introduction.fm Page no. 19 Chapter 1
4. SYSTEM COMPONENTS
4-1. Overview
1. The Senographe 2000 D Gantry is equipped with a digital detector. Positioning operations and X-ray
exposure are controlled by the Control Panel, usually mounted on a protective screen, which also
controls power to all parts of the Senographe 2000 D system.
2. The Senographe 2000 D includes an Acquisition Workstation (AWS) monitor, keyboard and mouse,
computer, electronics, accessory storage, and UPS (Uninterruptable Power Supply).
3. Accessories (standard and optional).
The following sections describe each of these components.
4-2. Senographe 2000 D X-ray System
The Senographe 2000 D is equipped with a dual track X-ray tube (molybdenum/rhodium) and a digital
detector. Mammographic examinations can be made with patients standing or sitting; both contact and
magnification views are available.
Images are acquired by direct digitization; they are displayed immediately on the AWS monitor and are
stored for later diagnostic review. They can be processed and/or filmed.
AOP (Automatic Optimization of Parameters) and manual setting modes are provided for control of X-ray
parameters; the system provides auto-collimation and other leading features.
4-3. Digital Detector and Image Receptor
The Digital Detector is built into the Image Receptor, shown below. It is a flat panel of amorphous silicon
on which cesium iodide is deposited to maximize detection of X-rays and transmission of light photons.
The high definition digital images produced are sent to the acquisition workstation for visualization and
processing.
The upper surface of the Image Receptor is a removable grid (Bucky). When the grid is not required, it is
easily removed and replaced by an optional Breast Holder without grid.
Image Receptor
Compression
Paddle
Bucky

Chapter 1 Page no. 20 introduction.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction
4-4. Acquisition Workstation (AWS) Cart
4-4-1. Overview
The Senographe 2000 D AWS (Acquisition WorkStation) is used for image acquisition and display, data-
base management, and to send images to archive, review or filming.
The workstation is installed in the AWS Cart shown above.
Inside the cart are housed the workstation electronics (computer, image detection system), an accessory
storage unit, the Uninterruptible Power Supply (UPS), and the optional CD-R unit for interchange pur-
poses.
Monitor screen
Three-section table
Mouse pad
Keyboard
Light box
Accessory storage unit
Uninterruptible
Power Supply
Access to workstation
computer and optional
CD-R Interchange Me
dia unit

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction
introduction.fm Page no. 21 Chapter 1
4-4-2. Three-Section Table
The three-section table carries the workstation keyboard, light box and writing surface/mouse pad. It can
be raised for easy access to the Accessories which can be stored below the table.
The left and right sides of the table (light box and writing table/mousepad) can be lowered out of the way
if necessary.
• To lower a side table, pull it slightly upwards, push the release button beneath it to release the lock,
and lower it carefully (do NOT drop).
• To return a side table to its working position, lift it to the horizontal position, then pull it slightly
upwards until the lock springs back into place.
CAUTION
The three-section table is not designed to hold items in excess of 20 kg weight.
4-4-3. Light Box
You can use the light box for viewing comparison mammograms.
WARNING
The light box must not be used for final interpretation of examinations.
The ambient light conditions in the examination room, and the resulting light level of the
light box, are incompatible with its use for final interpretation .
WARNING
The workstation monitor must not be used for final interpretation of examinations. It is set up for optimum visualization with an ambient light level of 50 lux. Leaving the light box illuminated without a film in place may degrade the review quality of images displayed on the monitor.
Release button
Light box
Keyboard
Mouse and
writing table/mouse pad
Light switch Release button
AWS
Monitor

Chapter 1 Page no. 22 introduction.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction 4-4-4. Writing Table/Mouse Pad
The space to the right of the keyboard, with the integrated mouse pad, can be used for manipulating the
mouse, and as a writing surface. There is a recess for holding pens and pencils.
Note:
For the comfort of left handed users, the mouse can also be operated on any other horizontal sur-
face of the three-section table. The mouse cable can be connected at either side of the keyboard
according to user requirements.
4-4-5. Accessory Storage Unit
The Accessory Storage Unit is located inside the drawer at the bottom of the Senographe 2000 D cart.
Open the drawer by depressing the small catch in the upper part of the opening. The unit is designed to
store items removed from the Senographe 2000 D, or other accessories. Access to the compartments of
the storage unit can be from the front of the cart, or the unit can be turned 90 degrees either way for
access from the side.
4-4-6. Uninterruptible Power Supply (UPS)
To assure system safety in the case of disturbances in the mains supply, the Senographe 2000 D system
incorporates an Uninterruptible Power Supply (UPS), housed in the lower part of the Senographe 2000 D
cart.
During power fluctuations or brief interruptions, the UPS assures an unchanging supply to the worksta-
tion and the acquisition system, thereby preventing mains disturbances being transmitted to the system.
When a power failure occurs, and also when the Senographe 2000 D system is switched off from the
Gantry Control Console, the UPS continues to supply power to the workstation during the time needed
for the workstation to shut down in an orderly manner.
UPS front panel

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction
introduction.fm Page no. 23 Chapter 1
5. ACCESSORIES AND OPTIONS
5-1. Senographe 2000 D Accessories
The Senographe 2000 D is delivered with a set of breast compression paddles, including an axillary pad-
dle, and a magnification stand. Accessories available as options include:
• Graduated biopsy compression paddle (for 2D localization) with optical localizer.
• Extra X-ray shield
• X-ray remote control hand switch
• Hydraulic chair
5-2. System Options
System options available include:
• Review workstation.
• Mass archiving system. When this option is installed and connected to the Senographe 2000 D,
acquired images can be sent to the mass archiving device for permanent storage, either automati-
cally or on request. A list of all patients ever imaged on the Senographe 2000 D system can be kept
on the mass archiving device, making future retrievals fast and easy.
• Laser Camera. To provide hard copies of images, the Senographe 2000 D system can be connected
to a laser camera for film output.
WARNING
Only images produced by GE-recommended laser cameras can be used for final interpre-
tation of examinations. for compatible printers, see the latest product data sheets for this
system, which uou can obtain from your local sales representative.
• Networking. The Senographe 2000 D is DICOM compliant, allowing it to be connected in a network
with other compliant devices for the exchange of images. Networking allowing you to transmit images
acquired with the Senographe 2000 D system to other DICOM-compatible review stations, using the
­Network Push" function of the AWS Browser. In some cases, additional studies will be needed for
the implementation of customized connections.
• CD-R Interchange Media. A CD-R (CD-Recordable) unit can be installed in the Senographe 2000 D
cart, allowing selected sets of images to be saved on CD-ROMs for communication purposes (e.g.,
recording images for referring physicians, training, personal image library, etc.). It is NOT recom-
mended for permanent archiving.

Chapter 1 Page no. 24 introduction.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Introduction This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
INSTALL.FM Page no. 25
CHAPTER 2 INSTALLATION
This chapter contains information for the installation of the Senographe 2000 D system.

Page no. 26 INSTALL.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 000 - AWS System Configuration Form
ist000.fm Page no. 27 Chapter 2
Job Card IST 000 - AWS System Configuration Form Chapter 2
1. INTRODUCTION
Complete a copy of this form at the end of installation.
Leave it attached to the site Service Manual; upgrade it for each configuration change (new option, new
software release, etc.).
MACHINE TYPE: ULTRASPARC 60
SERIAL NUMBER:
HOSTID:
INTERNET ADDRESS:
NETMASK:
SDC PASSWORD:
HOSPITAL NAME:
USER’S INTERFACE LANGUAGE :
APPLICATION SUPPORT TELEPHONE :
MISCELLANEOUS:
SOFTWARE INITIAL REVISION:
SOFTWARE UPGRADE REVISION:
SOFTWARE PROTECTION KEY:
CD-R PROTECTION KEY:

Chapter 2 Page no. 28 ist000.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 000 - AWS System Configuration Form
NETWORK HOST #1 :
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST#2
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST #3
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST #4
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST #5
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST #6
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST #7
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK HOST #8
Hostname: IP address:
Protocol: Port Number:

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 000 - AWS System Configuration Form
ist000.fm Page no. 29 Chapter 2
A.E. Title: Comments:
NETWORK or DICOM PRINTER #1 :
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK or DICOM PRINTER #2
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK or DICOM PRINTER #3
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
NETWORK or DICOM PRINTER #4
Hostname: IP address:
Protocol: Port Number:
A.E. Title: Comments:
ROUTING TABLE:
route add net:
route add net:
route add net:
route add net:
route add net:

Chapter 2 Page no. 30 ist000.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 000 - AWS System Configuration Form This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 001 - Installation Steering Guide.
IST001.FM Page no. 31 Chapter 2
Job Card IST 001 - Installation Steering Guide. Chapter 2
1. INTRODUCTION
This chapter provides information for installation of the Senographe 2000 D equipment.
2. STEERING GUIDE
Refer to Illustrations 1 through 4 for a Steering Guide outlining the recommended sequence of installa-
tion, followed by checks and calibration.
3. ROOM LAYOUT AND WIRING
Illustration 5, System Block Diagram, gives an overview of system connections

Chapter 2 Page no. 32 IST001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 001 - Installation Steering Guide.
ILLUSTRATION 1 - .INSTALLATION STEERING GUIDE - PREPARATION AND INSTALLATION
Check Room Preparation
(see p.i.m. inspection
Unpack the Senographe
Job Card IST 002
Unpack the Cart
Job Card IST 003
DMR Physical Install.
Job Card IST 004
Cart Physical Install.
Job Card IST 005
Preparation
report)
Apply non-European safety labels
Job Card IST 012
Schedule appointments
(e.g., with printer supplier
representative)
If the Senographe network
is to be connected to the
hospital network, request IP
addresses from the hospital.
Physical Installation
Install Protective
Screen
( see instructions supplied)
if required
Notes

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 001 - Installation Steering Guide.
IST001.FM Page no. 33 Chapter 2
ILLUSTRATION 2 - INSTALLATION STEERING GUIDE - CONNECTIONS AND POWER-UP
Connect Gantry to Generator,
Job Card IST 006
Connect Cart to Gantry
Job Card IST 007
Make a.c. power connections
Job Card IST 008
First System Power-on
Job Card IST 009
Connections and power-up
Carry out line resistance check
Job Card IST 010
Remove covers for access
Job Card D/R 101
(Gen. Cab., Gantry, Cart rear)
Console, and Footswitches
Check supply voltages
Job Card CFG 004
Coolant Fill
Job Card D/R 106
Connect and Configure Room
Job Card IST 013
Lamps and Door Switch
Notes
Unpack/install image
Job Card IST 017
receptor

Chapter 2 Page no. 34 IST001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 001 - Installation Steering Guide.
ILLUSTRATION 3 - INSTALLATION STEERING GUIDE - CONFIGURATION, ETC.
Configuration
Configure Gantry and Generator
Job Card CFG 002
Declare Network Hosts
Job Card CFG 006
AWS Configuration
Job Card CFG 005
Network Connection and Config.
Job Card CFG 008
Declare Printers
Job Card CFG 009
Install Modem
Install Modem
Job Card IST 016
Install CD-R Writer if present
Job Card IST 015
CD-R Writer option
Set Gantry upper travel limit
Job Card IST 011
Senographe Functional Checks
Job Card VF 005
Configuration
Notes

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 001 - Installation Steering Guide.
IST001.FM Page no. 35 Chapter 2
ILLUSTRATION 4 - INSTALLATION STEERING GUIDE - CALIBRATION AND CONFIGURATION
Refit covers
Job Card D/R 101
Acquisition in AOP mode
Job Card VF 010
Clean up room
Final checks
and complete paper-work
Completion
Flat Field check
Job Card VF 008
Refit covers
Verify X-ray beam Geometry
Job Card CAL 011
Image Acquisition & ACR Score
Job Card VF 009
NotesNetworked Options
(refer to supplied manuals):
Advantage Workstation
Mass Archiver
LaserCam
Calibrate AWS Monitor
Calibrate Monitor
Install required options
For ISC monitor, perform:
Job Card CAL 018
For Hi-Bright 21" monitor,
Brightness and Contrast Adjustment
refer to monitor s.m., perform:
Complete AWS System
Configuration Form
Job Card IST 000

Chapter 2 Page no. 36 IST001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 001 - Installation Steering Guide.
ILLUSTRATION 5 - SYSTEM BLOCK DIAGRAM
Conditioner
Water
Air
AC UPS Conditioner control
IDC-DMR link
DB9F
HE10-50F
Ethernet
crossover
RJ45M
RJ45M
DB9M
RJ45M
RJ45M
RJ45M
Connector for
network
Connector for
phone (Insite)
DB25F
AWS-DMR link
Fiber optic
Gantry DC power (+ve)
Gantry DC power (-ve)
Emergency stop
HV
Anode return (+HT)
Anode Starter
Gantry ground
AWS-DMR link
IDC-DMR link
AC
Detector PS Mains
distribution link
DB9F
Mains in:
Line 1
Line 2/Neutral
Ground
Keybd
Ethernet
Monitor
(for debug)
Monitor
Fiber
Ethernet
straight
through
RJ45M
AWS
Cart
Gantry
S1
Monitor
Generator
Generator Cabinet
XJ6
thru
XJ9
XJ11
XJ10
XJ4
XJ8
XJ7
XJ6
XJ10
XJ2
ECM
Filter
Console
XP4
XP2
Main Distribution
Rack
Mains
Circuit
Breaker J9
XP5
Keyboard
400PL2
200PL2
HV Tank
300PL2
Ground bar
IEC
320C1
4/10A
Plug
Detector
Bulkhead
Bulkhead
IEC 320C1
4/10A plug
Detector
PS
Com1
Com2
Can2
Aurora
Port 0
Port 1
AWS
IDC
IEC 320C1
4/10A Plug
UPS
Printer, AWS, CAD, etc.
Network
modem
Fiber optic
AC
Conditioner
Conditioner
Air/Water pipes
XJ10A
300PL2
XJ10B
IEC 320C1
4/10A plug
Foot switches

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 002 - Unpacking the Gantry and Generator Cabinet
IST002.FM Page no. 37 Chapter 2
Job Card IST 002 - Unpacking the Gantry and Generator CabinetChapter 2
1. SUPPLIES
None
2. TOOLS
1. Ratchet wrench with 22 mm socket.
2. 6 mm Allen wrench.
3. Claw hammer or other suitable tool for removing packing nails.
3. SAFETY PRECAUTIONS
CAUTION
- Follow standard safety practices for handling and moving large machines.
- When using a fork-lift to move the loaded pallet, always insert it on the Generator
Cabinet side; the other is blocked.
- Never lift or move the gantry by its handles.
- Do not remove the securing straps from the Generator Cabinet until it to be
removed from the pallet
4. PREREQUISITES
None
5. UNPACKING THE EQUIPMENT
The objective is to unpack the gantry, generator and accessories and move them to the mammography
room. Do not remove the packing on the Image Receptor.
WARNING
The Image Receptor is delicate, and can be easily damaged if subjected to any shock. DO NOT remove the packing on the Image Receptor at this stage. It will be removed after all
movement and installation work has been completed.
1. Inspect for damage and perform initial unpacking.
THE SENOGRAPHE WAS COMPLETELY INSPECTED FOR PROPER OPERATION AND APPEAR-
ANCE BEFORE SHIPMENT. HOWEVER, IT IS NECESSARY TO INSPECT THE PRODUCT AFTER
THE SHIPMENT IS RECEIVED. VISUALLY INSPECT THE PACKAGES FOR ANY APPARENT DAM-
AGE. IF THERE ARE SIGNS OF DAMAGE, REFER TO THE
Dama
ge in Transportation STATE-
MENT IN THE FRONT OF THIS MANUAL.
Open the packages and refer to the Product Delivery Instructions (PDI). Verify that items on the list
are present in the package. Carefully examine the contents for small parts.
2. Perform initial unpacking.
3. Remove the outer cardboard covering from the crate.
4. Remove ramp from crate and put it in place.
5. Remove the protective screen and the other accessories from the crate and move them into the

Chapter 2 Page no. 38 IST002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 002 - Unpacking the Gantry and Generator Cabinetmammography room.
5-1. Unpack gantry
1. Use the 6 mm Allen wrench to remove the four gantry baseplate shipping hold-down screws.
2. Thread the four special jacking bolts (provided in the accessory pouch) into their respective holes in
the gantry baseplate. Place the four flat round jack feet (provided in the accessory pouch) on the floor
of the crate with their nipples facing upwards. Continue turning the jacking bolts downwards until their
hollowed ends engage with the nipples of the jack feet.
3. Use the ratchet wrench with a 22 mm socket to progressively jack the gantry up via the four jacking
bolts.
4. When the gantry baseplate is high enough to be clear of the two wooden support blocks, remove
these blocks.
5. Lower the gantry progressively via the four jacking bolts until it is fully supported by the castors.
6. Leave the four jacking bolts threaded into the gantry baseplate and set the jack feet aside (they will
be used again when installing the gantry).
7. Carefully roll the gantry down the ramp and into the mammography room.
CAUTION
Never lift or move the gantry by its handles.
5-2. Unpack generator cabinet
1. Remove the securing straps from the generator cabinet. 2. For each of the two red clamping brackets securing the cabinet to
the pallet:
a. Remove the two bolts which hold the bracket to the pallet.
b. Remove the protective film from the strips of adhesive tape on
the bracket and on the bottom edge of the cover panel immedi-
ately above the panel.
c. Push the bracket firmly upwards as shown (C) so that it is held
up by the adhesive tape.
3. Remove the screws (B) which hold the two wooden blocks under-
neath the cabinet; remove the blocks.
4. Carefully roll the cabinet down the ramp and into the mammography
room.
CAUTION
Take care that the generator cabinet does not catch and jump while moving down the
ramp; note that its center of gravity is higher than that of a DMR cabinet.
5. Leave the two brackets in place (held to the covers by the adhesive tape) until you are ready to
remove the covers and connect the system. Then remove them and place them with the other pack-
ing mterials for return.
5-3. Accessories
The box of accessories is packed on the pallet with the AWS Cart.
A B
C

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 003 - Unpacking the Cart.
IST003.FM Page no. 39 Chapter 2
Job Card IST 003 - Unpacking the Cart. Chapter 2
1. SUPPLIES
None
2. TOOLS
Standard toolkit with metric wrenches.
3. SAFETY PRECAUTIONS
Follow standard safety practices for handling and moving large machines.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Reception
The AWS Cart is delivered with all items secured to a single pallet. Systems for overseas shipment are
protected by a wood and cardboard cover with shock indicators. Systems for road shipment do not have
this cover.
1. Examine the system on reception. Check any shock indicators present. If possible (this requires
removal of the cover, if present), make a visual check of all items on the pallet (monitor in carton, box
of accessories, AWS Cart).
2. If any problems are found (items missing, signs of shipping damage, excessive shock, etc.), note
them on the delivery note before signing for reception.
3. Move the pallet into or close to the examination room, ready for unpacking.
Notice:!
The Cart must be unpacked inside or close to the examination room; it is not a mobile, and cannot
be moved over bumps such as cables and elevator entrances.
If necessary, it may be moved over small steps with the aid of the ramps provided.

Chapter 2 Page no. 40 IST003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 003 - Unpacking the Cart.
5-2. Removal from pallet
Unpack the equipment:
1. Remove the cover, if present.
2. A box containing accessories (A) is secured above the monitor box in front of the Cart. Remove it.
3. Remove the box containing the monitor (B) from the pallet in front of the Cart.
4. Release the webbing strap (C), which secures the Cart and a wooden unloading ramp.
5. Remove the unloading ramp (D) and the packing material which protects the top of the Cart.
Position the ramp at the front of the pallet.
6. Remove the Cart securing box (E) after unscrewing the two bolts
which hold it. Remove the box containing the cassette support from
the securing box.
7. The front of the Cart is supported on two foam blocks. Lift the front of
the Cart slightly; remove the blocks by pulling the two tapes (F).
8. Remove all foam packing pieces from around the Cart.The Cart can
now be rolled down the ramp.
Notice:!
Avoid shocks to the Cart! In addition to the unloading ramp, a small triangular wooden ramp (30
mm x 50 mm) is provided for use if the Cart must be moved over steps. This ramp is taped at the
rear of the pallet.
5-3. Unpacking
1. Unpack the boxes containing the monitor and accessories .
2. Remove items packed in the accessory storage area at the front of the Cart;
3. Examine all items for damage. If there are any signs of damage, refer to the
Damage in Transporta-
tion
statement at the front of this manual. If there are no signs of damage, set them aside until
required for connection.
4. Collect all packing material for disposal.
D
C
A
B
E
F

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 004 - Senographe Physical Installation.
IST004.FM Page no. 41 Chapter 2
Job Card IST 004 - Senographe Physical Installation. Chapter 2
1. SUPPLIES
None
2. TOOLS
1. Ratchet wrench with 22 mm socket
2. 8 mm Allen wrench
3. Drill
4. Screwdrivers
3. SAFETY PRECAUTIONS
CAUTION
Follow standard safety practices for handling and moving large machines.
Never lift or move the gantry by its handles.
4. PREREQUISITES
• The procedures described in the Installation Job Card Unpackin
g the Gantry and Generator Cab-
inet must have been completed
.
5. PHYSICAL INSTALLATION
The objective is to place and anchor the gantry in its permanent position, and to assemble the protective
screen.
Note:
The generator is not permanently installed in this Job Card because it must remain mobile until the
wiring between it and the gantry is complete. The generator must therefore remain on its castors
for the moment.
The "CE" marking label, located on one of the large side panels of the generator, must remain visible
after installation of the generator.
6. INSTALL AND ANCHOR THE GANTRY IN ITS PERMANENT POSITION.
CAUTION
Do not remove the packing from the Image Receptor until the conditions specified in Sec-
tion 5-2 have been met.
• Before proceeding, it is recommended that you perform a final verification of the information given
in the Pre-Installation Manual, particularly concerning clearances around the gantry when
installed in its permanent position.
• Anchoring is recommended for Senographe 2000 D systems, but is not mandatory except in ses-
mic areas.
1. If anchoring is required, drill and prepare the securing holes:
a. Place the drilling template (provided in the installation kit) flat on the floor, align it with the desired

Chapter 2 Page no. 42 IST004.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 004 - Senographe Physical Installation. permanent position and mark the three gantry mounting hole centers on the floor.
b. Drill the three gantry mounting holes (depth and diameter depend on the type of insert to be used
- see below).
c. Place the appropriate type of insert (not provided) into each of the three mounting holes (depend-
ing on the type and thickness of floor). The inserts should be able to receive the 10 x 80 mm
screws (provided), or, for U.S. installations, the closest U.S. equivalent (not provided).
2. Wheel the gantry to its correct position (if it is to be anchored, position it with the three baseplate
mounting holes aligned with the three holes drilled in the floor).
3. Place the four flat round jack feet on the floor with their nipples facing upwards.
4. Turn the four jacking bolts downwards until their hollowed ends engage with the nipples of the four
jack feet.
5. Use the ratchet wrench with a 22 mm socket to progressively jack the gantry up via the four jacking
bolts until the castors are clear of the floor.
6. Use the 8 mm Allen wrench to remove the castors.
7. Lower the gantry progressively via the four jacking bolts until it is fully supported by its baseplate.
Remove the four jacking bolts completely.
Note:
The four jack feet remain below the gantry baseplate. If the floor is not perfectly even, level the
DMR base using the four screws p/n 45562657 (provided in installation kit). Insert them in place
of the jacking bolts to push on the jack feet.
8. To provide access to all three gantry baseplate mounting holes (one of the holes is located inside the
gantry column), remove the two column side panels (see D/R Job Card
Removal of Covers if neces-
sary).
9. If required, bolt the gantry to the prepared inserts in the floor.
10. Cover all unused gantry baseplate holes with the plastic plugs provided.
7. ASSEMBLE THE PROTECTIVE SCREEN.
See the mounting instructions delivered with the X-ray shield (one to three panels). Anchoring to the floor
is mandatory in seismic areas.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 005 - Cart Physical Installation
IST005.FM Page no. 43 Chapter 2
Job Card IST 005 - Cart Physical Installation Chapter 2
1. SUPPLIES
None
2. TOOLS
Standard toolkit with metric wrenches
3. SAFETY PRECAUTIONSWARNING
The monitor is heavy; two people must be available to lift it.
4. PREREQUISITES
The checks specified in the Installation Job Cards describing unpacking of the Gantry, Generator, and
Cart must be complete.
The room must be clean, and as far as possible free of unwanted material.
5. PROCEDURE
• Two procedures are given here, one for the ISC (Image Systems Corporation) monitor, and one for
the Hi-Bright 21" AWS Monitor.

Chapter 2 Page no. 44 IST005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 005 - Cart Physical Installation
6. INSTALLATION - ISC MONITOR
Notice:!
When installing the monitor, take care not to damage the support paint.
6-1. Monitor
1. Place the monitor in its position on the top of the Cart. Note that
the support is not square; it must be positioned with the small hole
(for cables) at the rear.
2. Remove the magnetic covers from the four monitor clamps (A).
3. Position the clamps and secure them to hold the monitor in posi-
tion (5 mm Allen key).
4. Replace the four magnetic covers on the clamps.
6-2. Monitor Connections
1. Remove the monitor rear cover.
2. Pass the monitor cables (power cable and five-way BNC signal cable, leaving the cart through the
cable cover behind the monitor) through the slot at the rear of the monitor.
3. Route the power cable so that the ferrite ring is to the right of the cable space (to allow the maximum
space for routing the signal cables), and connect it to the power input socket.
4. Identify and connect three of the BNC connectors to the monitor signal inputs (B) as follows:
a. Video input: Green cable (marked
G).
b. Synch input; Grey cable (marked
H).
c. Vertical input: Black cable (marked
V)
d. Use 75-ohm BNC terminators to terminate the remaining two BNC connectors (Blue and Red).
Leave them unconnected.
CAUTION
When routing the signal cables, position
them with the ferrite ring to the left of the
cable space, so that the cables are bent as
gently as possible (as shown here). Focus
problems can occur if the cables are bent too
tightly.
5. Use tie-wraps to secure the cables (including those
unconnected).
6. Set the monitor switch to ON and install the monitor
rear cover.
7. In the Cart, connect the monitor power cable to W18
(the upper connector of UPS outputs).
A

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 005 - Cart Physical Installation
IST005.FM Page no. 45 Chapter 2
7. INSTALLATION - HI-BRIGHT MONITOR
Notice:!
When installing the monitor, take care not to damage the support paint.
7-1. Monitor
• The Hi-Bright 21" AWS Monitor is delivered with its adaptor plate
already fitted, and the power and signal cables already connected.
1. Place the monitor on top of the AWS cart as shown (A and B).
2. Pass the a.c. mains connector through the hole in the top of the
cart (it must be turned to pass through sideways) and pass the
cable behind the AWS unit.
3. Push the mains cable to the side of the hole and pss the video
connector through the hole.
4. Move the monitor and adapter plate assembly back to position it as
shown (C). The four fixing holes in the adapter plate should corre-
spond to the four holes in the top of the AWS Cart.
5. Secure the monitor to the cart using the four screws provided.
6. Connect the monitor video cable to the 9-pin connector at the bot-
tom right of the AWS unit.
7. Plug the monitor power cable into a free socket on the UPS power
output strip at the bottom left of the cart.
8. Position and secure the cables as required.
A
B
C
D
9-pin video connector

Chapter 2 Page no. 46 IST005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 005 - Cart Physical Installation This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 47 Chapter 2
Job Card IST 006 - Connections (Gantry, Generator, Console, Foot-
switch) Chapter 2
1. SUPPLIES
• HV grease C333633
2. TOOLS
• Standard set of socket wrenches and nut drivers
• Standard set of open end wrenches
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
The system components must have been installed in their final positions in the room.
5. CONNECTIONS
Notice:!
The connection from the hospital a.c. mains supply must be switched by a mains isolator switch
external to the Senographe 20000 D system, accessible in case of emergency.
This module describes the connection of cables between the Generator, Gantry, and Console, and the
connection of footswitches. Connection of the X-ray lamp, room lamp and room door is described in
Installation Job Card
Room Lamps and Room Door Connections.
An overview of cable connections is given in the diagram below; cables and tubes to be connected are
listed in Tables 1 through 4.
A block diagram giving an overview of system connections is given in the Installation Steering Guide at
the beginning of this chapter.
Connection of the incoming a.c. supply cable to the Generator is described in the Installation Job Card

AC Connection.


DMR
Gantry
Generator
Cabinet
AC connection
Console
Harness 2 (black)
Harness 1 (white)
Console
cable
Phone
line
Refer to
Cart Connections
AWS Cart
Network
connection(Insite)
Footswitch
cables
Footswitches

Chapter 2 Page no. 48 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
TABLE 1 - GANTRY TO GENERATOR CABINET CABLES - HARNESS 2 (BLACK)
TABLE 2 - GANTRY TO GENERATOR CABINET CABLES - HARNESS 1 (WHITE)
TABLE 3 - CONSOLE TO GENERATOR CABINET CABLE
TABLE 4 - FOOTSWITCH TO GANTRY CABLES
Note:
Most of the illustrations in this Job Card show the Generator assembly only; they do not show the
MDR (Main Distribution Rack) which is mounted at the top of the Generator Cabinet.
Cable Reference Connects: See Section:
Detector ac supply 2225467 Gantry to Distribution Rack XP5 6-3
UPS ac supply 2225470 Gantry to Distribution Rack XP4 6-3
Water tubes 2266811, 2266812 To connectors on side of Conditioner 7
Air tubes n/a To connectors on side of Conditioner 9
Conditioner Control 2226611 To upper connector on side of Conditioner 6-4
AWS - Generator Link 2226623 Gantry to Generator Cabinet 400PL2 XJ4 6-1
IDC - Generator Link 2226613 Gantry to Generator Cabinet 400PL2
XJ10
6-1
Cable Color Connects to Generator Cabinet: See Section:
HV Grey HV Tank 8-1
Anode Return (+HT) Black HV Tank 8-2
Gantry DC Power Brown 200PL2: XJ6 (+ve); XJ7 (-ve) See Warning in
section 8-3
Gantry Ground Green/Yellow Ground Bar 8-3
Fiber-Optic cables Black 400PL2: XJ6 thru XJ9 See Caution in
section8-4
Anode Starter White 300PL2 XJ2 * 8-5
The Starter Motor cable screen terminates in a metal grounding clip, which must be securely fixed using the
screw pre-installed on the panel.
Emergency Stop Grey To connectors TB1 on Autotransformer 8-6
Cable Reference Connects: See Section:
Console Cable 2150063 Console to Generator Cabinet 400PL1 XJ8 6-2
Cable Reference Connects: See Section:
Footswitch Cable
(2-pedal)
2234874
(pedal assy.)
Footswitches to Gantry Power Board 800PL1
X10A/B
11-1
Footswitch Cable
(4-pedal option)
2234879
(pedal assy.)
Footswitches to Gantry Power Board 800PL1
X10 A/B, and XJ13 A/B
11-2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 49 Chapter 2
5-1. Route Cables from the Gantry to the Generator Cabinet
These cables are grouped in two harnesses. One is contained in a white flexible cover, the other is run
through a semi-flexible black tubular cable conduit. They are already connected in the Gantry Cabinet - it
is only necessary to route and connect them to the Generator Cabinet. Do not secure cables with cable
ties until all have been routed and connected.
1. Do not route and connect cables until the Gantry and Generator Cabinet are in their normal working
positions.
2. Remove the Generator Cabinet covers (refer to the D/R Job Card
Removal of Covers if necessary).
3. Remove the two red brackets which clamped the cabinet to the pallet for transport (now held to the
bottom of the front and rear cover panels by adhesive tape).
4. Run the cable harnesses from the Gantry to the Generator Cabinet through floor or wall ducts as
appropriate. Facing the front panel (with switches), cable entry is at the bottom of the left-hand side of
the cabinet.
5. When you are satisfied that all equipment is in its
correct position and unlikely to be moved, and that
the cables have been correctly routed, open up
any excess length of the black flexible conduit at
the Generator Cabinet end. Cut it to the correct
length for clamping at the cabinet entry point.
6. Clamp the white cable harness as shown. Secure
the black conduit in the U-shaped clip mounted
under the clamp.
7. Separate the cables and tubes contained in the black
harness. The electrical cables (A) will be routed first.
The air tubes (clear plastic; B) and water tubes (red
reinforced tubing, insulated at the conditioner end; C)
will be run separately.
A
B
C

Chapter 2 Page no. 50 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
5-2. Cabinet internal routing
WARNING
It is important that the instructions given for routing the two water tubes are carefully fol-
lowed. Only the insulated parts of the tubes should pass through the cabinet; any excess
must be coiled below the cabinet as specified. If these instructions are not followed, the
heat of the cabinet can affect the temperature of the detector cooling water and make cor-
rect detector temperature control impossible.
This section gives requirements for routing cables and tubes in the Generator Cabinet. Before perma-
nently securing any item, be sure that you have studied and understood all the requirements of the con-
nections described in this job card. It may be necessary to try different arrangements for excess cable
before finally deciding on the best cable routes.
• Any excess length of water tubes is coiled on the floor in the space beneath the bottom of the cabinet.
• Any excess length of electrical cables is coiled in the space above the cabinet cable entry, over the
guide provided at the top of the space and down to the clamping bracket at the bottom of the cabinet.
• Any excess length of air tubes is coiled with the electrical cables.
Take great care not to allow the air tubes to pinched by the weight of cables.
• Make connections and route cables in the order given:
- Black harness and Console cables; see section
Black Harness and Console Cable Connection.
- Water tubes; see section
Water Tube Connection.
- White Harness cables; see section
White Harness Connections.
- Air Tubes; see section
Air Tube Connection.
- Room door wiring; see section
X-ray On and Room Door Wiring.
- Footswitch connections; see section
Footswitch Connection.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 51 Chapter 2
6. BLACK HARNESS AND CONSOLE ELECTRICAL CONNECTIONS
ILLUSTRATION 1 - AWS, IDC, AND CONSOLE CABLES
6-1. Connecting the AWS and IDC Link Cables
1. AWS - Generator Link. Route the cable up the side of the generator and connect it to Generator Inter-
face Board 400PL2 connector XJ4.
2. IDC - Generator Link. Route the cable up the side of the generator and connect it to Generator Inter-
face Board connector 400PL2 XJ10
6-2. Connecting the console cable.
1. Route the console cable (provided in the accessory pouch) from console to generator.
2. Route the cable up the side of the generator,
along the inside of the middle generator cross-member,
and connect it to generator command board 400PL1 connector XJ8.
3.
Route and connect the other end of the console cable to the console.
CAUTION
Tighten the connector retaining screws finger tight only at each end of the console cable.
400 PL1
400 PL3
400 PL2
200 PL2
701 PL1
Power
Line
Filter
XJ 10
XJ8 XJ10
XJ4

Chapter 2 Page no. 52 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ILLUSTRATION 2 - DETECTOR AND UPS AC SUPPLY CABLES
6-3. Connecting the Detector and UPS supply cables
From the cable entry point, run the cables up the rear of the cab-
inet, along the side cross-member and the front cross member
to the right hand side of the cabinet. Plug them into the supply
outlets XJ4 and XJ5 as shown (A).
6-4. Connecting the Conditioner Control cable
From the cable entry point, run the cable up the rear of the cabinet, along
the side cross-member and up to the side of the Conditioner in the MDR.
Plug and secure it in the upper connector as shown (B).
6-5. Excess cable
Coil any excess cable in the space above the cabinet cable entry, over the
guide provided at the top of the space and down to the clamping bracket at
the bottom of the cabinet.
7. WATER TUBE CONNECTION
Run the insulated sections (only) of the tubes up the rear of the cabinet
and plug them into the Conditioner connectors (C). Take care not to
obstruct the tubes by bending them too sharply, or by securing them too tightly. Note that the tubes are
not identified as inlet or outlet; they are interchangeable.
Coil any excess length of water tube on the floor in the space beneath the cabinet and secure it using
cable ties when installation is complete.
A
B
C

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 53 Chapter 2
8. WHITE HARNESS CONNECTIONS
ILLUSTRATION 3 - HV TANK CONNECTIONS
8-1. Connecting the HV cable
1. Remove the protective covering from the HV cable connector and apply HV grease to the connector.
2. Plug the HV cable (2) into the HV unit and carefully tighten the connecting ring.
3. To ensure proper grounding of the HV cable, tighten the 1.5 mm screw on the top of the cable locking
ring.
Note:
If the HV cable is not properly grounded, intermittent console lock-up and random error messages
may occur - if this occurs, check the grounding at both ends of the cable.
8-2. Connecting the anode return cable
4. Connect the X-ray tube anode return wire marked BLOC +HT (1) to the threaded stud marked with a
ground symbol on the top of the HV unit.CAUTION
Do not tighten the nut excessively - doing so may damage the thread or break off the stud. If this happens the sealed HV unit must be returned to the factory for repair.
2
1
200 PL4
TB1

Chapter 2 Page no. 54 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ILLUSTRATION 4 - GANTRY D.C. SUPPLY AND GROUND CONNECTIONS
8-3. Connecting the gantry power and grounding cables
1. Route and connect the gantry 27 V power cables to 200PL2 as shown: the +ve connector to XJ6, the
0 V connector to XJ7 (cables are routed under the capacitor box, through the middle of the genera-
tor).
2. Route the green/yellow ground cable in the same way; connect it to the grounding bar (1).
WARNING
These cables must be routed through the opening close to the capacitor box.
WARNING
Be very careful to connect the 27 V power cables to XJ6 & 7 on 200 pl2; (NOT to XJ6 & 7 on
200 pl1)
- the 200pl1 connectors carry line level ac voltages which will cause considerable
damage to components if connected to the gantry.
400 PL1
400 PL3
400 PL2
200 PL2
701 PL1
Power
Line
Filter
1
XJ6
XJ7
Capacitor
box
Power
Line
Filter

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 55 Chapter 2
ILLUSTRATION 5 - FIBER-OPTIC LINKS CONNECTION
8-4. Connecting the fiber-optic cables
CAUTION
Fiber-optic cables are more fragile than ordinary copper-stranded cable and you must
therefore treat them with greater care - bend or twist them only as much as is strictly nec-
essary to install and connect them.
Note:A small protective cap protects the cable end; remove it before connecting.
1. Route the black plastic tube protecting the fiber-optic cables as shown in the illustration, taking care
to keep a smooth radius on all bends (fibers are routed under the capacitor box, through the middle of
the generator).
2. Identify the four connectors XJ6 through XJ9 on the generator interface board 400PL2 and connect
the four fiber-optic cables to their respective connectors.
400 PL1
400 PL3
400 PL2
200 PL2
701 PL1
Power
Line
Filter
XJ6 through
XJ9
Fibre- optic
cables
Power
Line Filter

Chapter 2 Page no. 56 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ILLUSTRATION 6 - ANODE STARTER MOTOR CABLE CONNECTION
8-5. Connecting the anode starter cable
1. Route the anode starter cable (marked 300 PL2 XJ2) as shown; connect it to XJ2 on the anode
starter board 300PL2(on rear of the unit; not illustrated here).
2. Remove the metal grounding clip; pass the anode starter cable inside it and bolt the clip back to the
panel.
400 PL1
400 PL3
400 PL2
200 PL2
701 PL1
Power
Line
Filter
Anode starter
cable
Metal
grounding
clip here

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 57 Chapter 2
ILLUSTRATION 7 - EMERGENCY STOP CABLE CONNECTION
8-6. Connecting the Emergency Stop cable
Route the emergency stop two-wire cable (marked AUTOTRANSFORMER TP1 OR 200PL2 XJ10) over
the HV tank; connect the two wires to the clip connectors TB1 on the top of the Autotransformer. The
leads may be connected either way round. TB1 is connected through the autotransformer thermal cutout
to the emergency stop connector on board 200PL1.
8-7. Excess cable
Coil any excess cable in the space above the cabinet cable entry, over the guide provided at the top of
the space and down to the clamping bracket at the bottom of the cabinet.
200 PL4
HV tank
701
TB1
200 PL4
HV tank
701

Chapter 2 Page no. 58 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)9. AIR TUBE CONNECTION
Run the air tubes up the rear of the cabinet and plug them into the Condi-
tioner connectors (D). Take care not to obstruct the tubes by bending them
too sharply, or by securing them too tightly. Note that the tubes are not
identified as inlet or outlet; they are interchangeable.
Coil any excess length of water tube with the electrical cables, in the space
above the cabinet cable entry. It should be coiled last, to avoid any danger
of being pinched by the weight of the cables.
10. X-RAY ON AND ROOM DOOR WIRING
Refer to the Installation Job Card Room Lamps and Room Door Connection for information on connect-
ing the X-ray ON lamp and room-door wiring
D

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
ist006.fm Page no. 59 Chapter 2
11. FOOTSWITCH CONNECTION
11-1. 2-Pedal Footswitch basic hardware (Compression only):
1. Remove gantry column left side cover to access 800 PL1 Gantry Power Board.
2. Route the two footswitch cables along gantry main harness (use cable ties).
3. Attach each footswitch cable to the Gantry chassis using a clamping plate (4),
4. Connect the two footswitch cables to the fixed connectors XJ10A and XJ10B on the Gantry 800PL1
Power Board (2). The “flying” connectors XJ13A and XJ13B (3), if present, are not used.
11-2. 4-Pedal Footswitch optional hardware (Compression + Elevator):
1. Remove gantry column left side cover to access 800 PL1 Gantry Power Board
2. Route the two footswitch cables along gantry main harness (use cable ties).
3. Attach each footswitch cable to the Gantry chassis using a clamping plate (4).
4. Connect the two compression footswitch cables to fixed connectors XJ10A and XJ10B on the Gantry
800PL1 Power Board (2).
5. Connect the two elevator control footswitch cables to the “flying” connectors XJ13A and XJ13B (3).
ILLUSTRATION 8 - FOOTSWITCH CABLE CONNECTION
XJ10A XJ10B
800PL1 Gantry Power Board
(4) CLAMPING PLATES
(2) CONNECTORS FOR COMPRESSION CONTROL
BASIC (COMPRESSION ONLY):
RIGHT
LEFT
XJ13A
XJ13B
(3) CONNECTORS FOR ELEVATOR CONTROL
Route Footswitch
cables along the
main harness
XJ10A XJ10B
800PL1 Gantry Power Board
RIGHT
LEFT
OPTION (COMPRESSION + ELEVATOR):
(2) CONNECTORS FOR COMPRESSION CONTROL

Chapter 2 Page no. 60 ist006.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 006 - Connections (Gantry, Generator, Console, Footswitch)
12. COMPLETION
12-1. Final cable dressing and tying
When all cables have been correctly routed and connected, use cable ties to secure them as required.
Specific recommendations are as follows:
1. Secure the HV cable to the lower and middle cross-members with two cable-ties (one on each cross-
member).
2. Arrange the remaining cables, then attach them to the lower and middle cabinet cross-members with
two cable-ties (one on each cross-member).
Note:
Secure the anode current return wire BLOC +HT to the lower cross-member only.
3. Secure the X-ray lamp and room lamp wires (if used) to the lower and middle cross-members with
two cable-ties (one on each cross-member).
4. Leaving sufficient slack to avoid stress on the fibre-optic connectors, attach the black plastic tubing to
the middle cross-member with a cable-tie.
5. Attach the two gantry power leads and the protective grounding cable to the lower cabinet cross-
member with a cable-tie
6. Secure the excess cable loops as required
12-2. HV tank
• In order to release pressure during long exposures, loosen the vent screw of the HV tank by 3 mm.
VENT SCREW

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 007 - Cart Connections.
IST007.FM Page no. 61 Chapter 2
Job Card IST 007 - Cart Connections. Chapter 2
1. SUPPLIES
•None
2. TOOLS
• Standard set of socket wrenches and nut drivers
• Standard set of open end wrenches
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
The Cart and Gantry must be in their final positions in the room.
5. PROCEDURE
5-1. Cable routing
Cart to Gantry cables are already connected in the cart; they leave the cart through a black tubular con-
duit.
1. Run the cable conduit from the cart to the gantry (through floor or
wall ducting as appropriate).
2. If the conduit is too long, remove the securing ring (A), cut off the
excess at the gantry end, and replace the securing ring.
3. Starting with the biggest connectors (to ease the task of routing
the last cables), pass the cables through the opening (B) at the
bottom of the gantry.
4. Identify and connect the cables as described in the following sec-
tions.
5. After all connections have been made, secure the conduit by fit-
ting the securing ring (A) into the fork at the bottom of the gantry
as shown. Coil any excess cable inside the gantry and secure
with cable ties.
5-2. Gantry Connections
Identify and connect the cables in the gantry, using the following table and illustrations.
WARNING
Fiber-optic cables are more easily damaged than electrical cables; handle them with care. Note:A small protective cap protects the cable end; remove it before connecting. The connectors are keyed; if they are difficult to connect, check that the key is
correctly oriented.
When disconnecting, be sure to pull on the connector and not the cable.
A
B

Chapter 2 Page no. 62 IST007.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 007 - Cart Connections.
:
Cable Reference Connects to: Illus. Ref.
Power connections:
AC UPS Power 2231555 UPS IECC14/10A to IEC 320/10A
Gantry Bulkhead
C
Detector PS-IDC 2226621 IDC to Power Supply J2 D
Signal connections:
ADS-DMR Link 2226625 - XJ7 ADS Port 0 to Gantry Bulkhead XJ7
UPS-DMR Link 2226615 - XJ6 UPS to Gantry Bulkhead XJ6
IDC--DMR Link 2240227 - XJ5 IDC COM2 & RT line to Gantry
Bulkhead
XJ5
Conditioner Control 2226612 - XJ2 IDC COM1 to Gantry Bulkhead XJ2
Fiber-Optic Cable
(see Warning)
n/a IDC to Gantry Bulkhead E
C
D
E
XJ2
XJ6
XJ7
XJ5

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 007 - Cart Connections.
IST007.FM Page no. 63 Chapter 2
5-3. Modem and Network Connections
Connect the modem and network cables to the connectors at the rear of the AWS cart (see illustration):
• Connect the modem cable from the telephone wall socket to the modem connector.
• Route and connect the cable from the network hub (provided in the Senographe 2000 D networking
kit) to the Ethernet network connector.
• Note that the third socket is intended for connection of a local printer, but is not currently used.
Modem connection Ethernet connection
Printer connection
(not used)

Chapter 2 Page no. 64 IST007.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 007 - Cart Connections. This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 008 - AC Connection.
IST008.FM Page no. 65 Chapter 2
Job Card IST 008 - AC Connection. Chapter 2
1. SUPPLIES
None
2. TOOLS
•AC voltmeter
• Torque wrench (optional)
Note:
The recommended torque for all lug nuts is 4 Nm
3. SAFETY PRECAUTIONS
Follow standard safety techniques and procedures for handling equipment and circuits carrying AC sup-
ply voltages.
4. PREREQUISITES
A suitable mains isolator switch must have been installed for connection of the Senographe 2000D sys-
tem to the hospital supply.
Remove covers from the Generator Cabinet.
Note:
Jumpers and switches not dependent on site voltages etc. are set in the factory, and normally need
not be checked. If you suspect that they have been disturbed, refer to to the Configuration Job
Card
Senographe Jumpers and Switches.
5. PROCEDURES
The procedures described in this document and listed below must be followed carefully before applying
power to the system:
1. AC Supply Check.
2. Voltage Selection - Mains Distribution Rack.
3. Voltage Selection - Generator.
4. Configure Fuses F1, F2.
5. Connect the AC Supply Cable.
• When all of these procedures have been satisfactorily completed, and before switching on the
Generator Cabinet and Mains Distribution Rack circuit breakers, follow the instructions for
First Power On in the Installation Job Card System Power On.

Chapter 2 Page no. 66 IST008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 008 - AC Connection.
5-1. AC supply Check
Determine the nominal voltage and frequency of the available power supply.
5-1-1. Check nominal line voltage
Consult the user to determine the nominal line voltage and frequency, then measure the open-circuit line
voltage to verify that it corresponds to the nominal line voltage, ±10%. Allowable nominal line voltages
are 200, 208, 220, 240, 380, 415 V single-phase.
5-1-2. Line frequency
The
Senographe operates with line frequencies of 50 or 60 Hz without adjustment.
5-2. Voltage Selection - Mains Distribution Rack
Check the position of the input connection wire
on the Mains Distribution Rack transformer ter-
minal block (reference A in the illustration).
If the tap connection does not correspond to the
nominal (
not the measured) line voltage you
determined in section 5.1, then remove and
reconnect the wire to the correct terminal.
A

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 008 - AC Connection.
IST008.FM Page no. 67 Chapter 2
5-3. Voltage Selection - Generator
5-3-1. Position the voltage selection board
Referring to the illustration and Table 1
(below), install the voltage selection board
IN2 on the AC Distribution board 200–PL1
according to the nominal AC supply voltage.
5-3-2. Voltage selection
Check the position of the SEL wire on the AC
Distribution board 200–PL1 (see illustration).
By default, the SEL wire is connected to the
240v transformer tap. If the tap connection
does not correspond to the nominal (
not the
measured)
line voltage you determined in
section 5.1, then remove and reconnect the
SEL wire to the correct terminal.
Notice:!
The SEL wire must NOT be connected
to the 0V terminal. This terminal is re-
served exclusively for use while per-
forming the measurement described in
Installation Job Card
Line Resistance
Measurement
. The Senographe can-
not function normally with the SEL wire
in this position.
TABLE 1 - POSITION OF VOLTAGE SELECTION BOARD IN2 AND SEL WIRE
Nominal single-
phase line voltage
Position of IN2 Position of SEL wire
415 V HT 415 V
380 V HT 380 V
240 V BT 240 V
220 V BT 220 V
208 V BT 208 V
200 V BT 200 V
440 V Use transformer reference 45563370
and 380 V configuration.
480 V
HT for >380v
AC input
200 PL1
HV tank 701
BT for <240v
AC input
IN2 board
position
SEL wire

Chapter 2 Page no. 68 IST008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 008 - AC Connection.
5-4. Configure fuses F1, F2.
To configure these fuses (see illustration 2 for location), follow the indications on the label located on the
bottom of the generator cabinet, just below the AC supply switch S1.
WARNING
FOR SAFE OPERATION OF THE SENOGRAPHE, PAY PARTICULAR ATTENTION TO THE INFORMATION ON THIS LABEL.
DETERMINE WHETHER THE AC SUPPLY CONSISTS OF TWO PHASES OF A THREE–
PHASE SYSTEM, OR A SINGLE PHASE AND NEUTRAL.
WHEN THE SUPPLY IS SINGLE–PHASE WITH NEUTRAL, A METAL CYLINDER (delivered
with the DMR) MUST BE INSTALLED INSTEAD OF F1.
ILLUSTRATION 1 - LOCATION OF FUSES

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 008 - AC Connection.
IST008.FM Page no. 69 Chapter 2
5-5. Connecting the AC supply cable.
1. Verify that generator AC supply Switch 200-S1 is in the 0 position and that AC supply is OFF.
2. Connect the AC line and neutral wires to the L & N terminals and the ground wire to the yellow/green
terminal as shown in illustration 2.
3. Cable-tie the cable to the lower cross-member of the cabinet, using the holes provided in the cross–
member.
4. Route and connect the other end of the supply cable to the AC supply source.
ILLUSTRATION 2 - CONNECTING THE AC SUPPLY CABLE
200 PL4
200 PL2
200 PL1
HV tank
701
AC supply cable

Chapter 2 Page no. 70 IST008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 008 - AC Connection. This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 009 - First System Power On
IST009.FM Page no. 71 Chapter 2
Job Card IST 009 - First System Power On Chapter 2
1. SUPPLIES
None
2. TOOLS
None
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
• The installed components (AWS, IDS, Conditioner, etc.) must be in a safe state, ready to receive
power, and all of their power switches must be set to ON.
• The instructions given in the Configuration Job Card
Check Supply Voltage must have been carried
out to ensure that the Generator and Mains Distribution Rack power supplies have been correctly
configured.
• The detector cooling circuit must have been filled with coolant and checked for leaks (refer to the D/R
Job Card
Conditioner Fill and Maintenance, and carry out the instructions given for Coolant Fill).
5. FIRST POWER ON
After installation of the system, it must be powered up before installing options or configuring the system.
CAUTION
When you are ready to apply power to the Senographe 2000 D system for the first time,
carry out the following steps:
Note:
TIf any problem is encountered during the first power up, refer to the checklists for power up/down
given in the Verification Job Card
Boot/Init and Shutdown.
1. Set the circuit breaker connecting the hospital supply to the Senographe 2000 D system to ON.
2. Set the two circuit breakers in the Generator Cabinet (one in the Generator Rack, the other in the
Main Distribution Rack) to ON
3. To power up the system, press the power switch I on the DMR Control Console. Power is applied to
the whole system (DMR, Cart, Digital Detector, etc.); green power indicators are lit on the Generator
Cabinet, the Cart, and the monitor.
4. A series of system initialization messages appears on the screen, then the
Login: prompt is dis-
played. Type sdc (all lower case) followed by <Enter>. When the
Password prompt appears, type
adw3.1<Enter>. The AWS Browser should be displayed after a short delay.
If the screen is still blank after about 30 seconds, check that the monitor is switched on (the green
LED should be lit).
When the system has been successfully powered up, it must be powered down ready for the next steps
in the installation process.
Follow the normal shutdown procedure described below.

Chapter 2 Page no. 72 IST009.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 009 - First System Power On
6. POWERING DOWN THE SYSTEM
• Normal Shutdown Procedure:
Switch power OFF at the AWS Monitor by clicking the
System Shutdown button on the Browser
window. Wait for complete system and UPS shut down (about two minutes); check that no LEDs
remain lit on the UPS before continuing.
• To power down the Gantry and Generator system
only, press the switch STOP on the DMR Control
Console.
• NEVER switch off at the UPS directly except in emergency (risk of data loss).

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 010 - Line Resistance Measurement
IST010.FM Page no. 73 Chapter 2
Job Card IST 010 - Line Resistance Measurement Chapter 2
1. SUPPLIES
None
2. TOOLS
• Voltmeter (a.c.)
3. SAFETY PROCEDURES
During the procedures in this Job Card, you will be in close proximity to circuits operating at line voltage -
take appropriate precautions.
4. PREREQUISITES
You must follow the chronological order given in the Installation Job Card Installation Steering.
Note:
Make sure the console is connected to the DMR generator.
5. PROCEDURE
If the resistance at the line input to the Senographe generator is too high, there will be a significant volt-
age drop when the generator draws maximum power. This procedure checks that the resistance is within
acceptable limits.
5-1. Preparation
1. Switch off the DMR; set the generator power switch 200-S1 to the 0 position.
2. Disconnect voltage selection wire SEL from its lug on the generator board 200-PL1; reconnect it to
the 0 V lug.
Notice:!
The 0 V position of the SEL wire is reserved exclusively for this line resistance measurement. The
Senographe cannot function normally when the SEL wire is in this position. Do not attempt to
switch on the Senographe at this point - follow the instructions below.
3. Connect the ac voltmeter to the input end of the power switch 200-S1; Note and write down the
observed value of the open-circuit line voltage V1:
V1 = ......................................
5-2. Procedure
1. Turn the AC supply on: set the generator switch 200-S1 to the 1 position.
Note:
Read the next step completely before performing it. You will have only 3 seconds to read the volt-
meter; if you miss the reading you will have to wait about 15 minutes to repeat it.
2. While looking at the voltmeter, press the power-on switch on the generator board 200-PL2. During a
period of three seconds, a resistive load (power resistors R1 and R4 on the generator board) is con-
nected across the line input, causing a drop in the voltage reading on the voltmeter. Note and write
down this closed circuit voltage:

Chapter 2 Page no. 74 IST010.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 010 - Line Resistance Measurement
V
2
= ....................................
Note:
After this three-second period the red LED DE3 on the generator board light, indicating that heating
of the resistive load prevents an immediate repeat reading. After about 15 minutes, the LED goes
out, the reading can then be repeated if necessary.
3. Calculate line resistance R
L
according to the formula:
R
L
= [(V
1
- V
2
) / V
2
] R
INT
where R
INT, the internal resistive load, depends on the nominal a.c. supply voltage V:
Voltage (V) R
INT
(W)
200 11
208 11
220 11
240 11
380 44
415 44
4. Turn off the DMR: set the generator power switch to the 0 position and switch off AC power.
5. Disconnect the line voltage selection wire SEL from the 0 V lug on the generator board; reconnect it
to the lug corresponding to the nominal (not measured) AC line voltage. Tighten the lug nuts to a
torque setting of 4 Nm.
Check that the calculated line resistance R
L
does not exceed the maximum value given in Table 1 for
the nominal a.c. supply voltage.
If the calculated value exceeds the maximum value given in Table 1, consult with the hospital electri-
cal department and your OLC. The problem must be resolved before the Senographe 2000 D sytem
is put into service.
TABLE 1 - MAXIMUM PERMISSIBLE LINE RESISTANCE
Nominal AC voltage Maximum Line Resistance
415 V 1.19
380 V 1.00
240 V 0.40
220 V 0.34
208V 0.30
200V 0.28

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 011 - Set Elevator Upper Travel Limit
IST011.FM Page no. 75 Chapter 2
Job Card IST 011 - Set Elevator Upper Travel Limit Chapter 2
1. SUPPLIES
None.
2. TOOLS
• Screwdriver 4mm.
• Allen wrench 2.5mm.
3. SAFETY PRECAUTION
• No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. SET ELEVATOR UPPER TRAVEL LIMIT
5-1. Preparation
1. Measure room overall height, floor to ceiling.
2. Remove left cover from column housing (see D/R Job Card
Removal of Covers).
5-2. Determine optical sensor position
Determine the position of the optical sensor as follows:
1. Ceiling height 2.5m (8.22ft.) or greater :
- Install the optical sensor in the upper mounting hole (1 in Illust.1 ).
Note:
The DMR is delivered with the optical sensor is installed in this position.
2. Ceiling height 2.4m (7.9ft.) or greater, but less than 2.5m (8.22ft.):
- Install the optical sensor in the middle mounting hole (2 in Illust.1).
3. Ceiling height 2.3m (7.5ft.) or greater, but less than 2.4m (7.9ft.):
- Install the optical sensor in the lower mounting hole (3 in Illust. 1).
4. Ceiling height less than 2.3m (7.5ft.):
- In this case, the DMR cannot be installed without special modification.
Note:
The adjustment of the metal plate which blanks off the upper and lower sensors only affects the
elevator lower travel limit.
5-3. Reposition the optical sensor (if required)
1. Remove the tyraps on the cable connected to the optical sensor.
2. Remove the two 2.5mm Allen screws securing the sensor.
3. Remove the optical sensor and reinstall it in its new position, as determined in 5.2 above.
4. Reinstall and tighten the two 2.5mm Allen screws, in the new position.
5. Reinstall tyraps on the sensor cable, as required.

Chapter 2 Page no. 76 IST011.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 011 - Set Elevator Upper Travel Limit6. Switch on the DMR. Carry out a column raising and lowering test and check that the tube housing
cover does not collide with the ceiling (check for a minimum safety clearance of at least 80 mm to
90 mm).
ILLUSTRATION 1 - OPTICAL SENSOR POSITIONS
COLUMN FRONT VIEW
POSITIONS FOR
UPPER OPTICAL
SENSOR
1
2
3
1
2
3

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 012 - Safety Labelling
IST012.FM Page no. 77 Chapter 2
Job Card IST 012 - Safety Labelling Chapter 2
1. SUPPLIES
Envelope containing safety labels (supplied with the system).
2. TOOLS
None
3. SAFETY PRECAUTIONS
Notice:!
The attachment of appropriate safety labelling to the equipment is a LEGAL REQUIREMENT for
operation of an X-ray producing unit.
4. PREREQUISITES
None
5. SAFETY LABELLING
The objective is to select and attach appropriate safety labels to the Senographe Control Console,
according to local safety regulations concerning the use of X-ray producing equipment.
• Locate and open the safety label envelope taped underneath the console. It contains two radiation
warning labels and a surface hygiene warning label.
5-1. Radiation Warning Labels
• Identify the radiation warning labels illustrated below; select the one that corresponds to local regula-
tions.
ILLUSTRATION 1 - SAFETY LABEL
CEI (EUROPE) & ASIA
REDAC (CANADA)
CDRH (USA)

Chapter 2 Page no. 78 IST012.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 012 - Safety Labelling Note:
The CEI warning symbols are already screen-printed on the console.
• If one of the other warning labels is required, peel off the self-adhesive backing from the selected
label and affix it over the CEI warning symbols in the lower right-hand corner of the console as shown
below.
5-2. Surface Hygiene label (for US only)
1. If your installation is in the USA, locate the surface hygiene label similar to that shown below.
2. Peel off the self-adhesive backing from the label, and affix the label in the center of the upper part of
the console, as shown below.
kV mAs
Affix label here
CAUTION: Patient contact surfaces should be cleaned and disinfected between patients
kV mAs
Affix label here
CAUTION: Patient contact surfaces should be cleaned and disinfected between patients

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 013 - Room Lamps and Door Configuration
IST013.FM Page no. 79 Chapter 2
Job Card IST 013 - Room Lamps and Door Configuration Chapter 2
1. SUPPLIES
• The necessary hardware (wires, lamp, switches, etc.) is field and/or customer supplied.
2. TOOLS
• None.
3. SAFETY PRECAUTIONS
• None.
4. PREREQUISITES
• None.
5. PROCEDURE
1. Refer to Illustration 1; connect the room and lamp wiring to XJ1, XJ2, XJ3 on the 200PL4 board (see
Illustration 2 for locations).
2. Route the wires in the generator cabinet, along with the mains supply cable.
ILLUSTRATION 1 - ROOM LAMPS AND DOOR WIRING
Note:
The load current ratings of the relays K1 and K2 on 200-PL4 are 5A/30 VDC
Relay K3 must be supplied with 24 V ac.
AC or DC
30 V max
AC or DC
30 V max
K1
K2
K3
XJ1
XJ2
XJ3
Seno ON lamp
X-ray ON lamp
Room door switch 24v
(external power
CUSTOMER-SUPPLIED HARDWAREDMR+ GENERATOR
Board
200PL4
~

Chapter 2 Page no. 80 IST013.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 013 - Room Lamps and Door Configuration3. Locate board 400-PL1. Refer illustration 3; and set jumpers JP1 and JP2 as required according to the
table below. Place unused jumpers in the parking positions JP3 and JP4.
Note:
If there is no door switch, jumpers JP1 and JP2 should both be present (case 4 in the table).
TABLE 1 - JP1 AND JP2 JUMPER SETTINGS
0 = No jumper
1 = Jumper present
Note:
When the room door switch is present and configured for use, the signals valid-PR (1st trigger or
prep) and valid-GR (RADIOGRAPHY) are enabled through the switch.
JP1
(PREP)
JP2
(RADIOGRAPHY)
Door closed Door opened
1 0 0 PREP and RADIOGRAPHY
enabled
PREP and RADIOGRAPHY
disabled
2 0 1 PREP and RADIOGRAPHY
enabled
PREP and RADIOGRAPHY
disabled.
NOTE:
if the door opens during
an exposure, the exposure is
NOT aborted and completes
normally
)
3 1 0 PREP and RADIOGRAPHY
enabled
PREP enabled
RADIOGRAPHY disabled
4 1 1 PREP and RADIOGRAPHY
enabled
PREP and RADIOGRAPHY
enabled

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 013 - Room Lamps and Door Configuration
IST013.FM Page no. 81 Chapter 2
ILLUSTRATION 2 - LOCATION OF ROOM WIRING CONNECTORS
ILLUSTRATION 3 - LOCATION OF JP1 TO JP4
200 PL4
XJ2
XJ3
XJ1Room
lamps
and door
wiring

Chapter 2 Page no. 82 IST013.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 013 - Room Lamps and Door Configuration This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 014 - Automatic X-ray Tube Warm-Up
IST014.FM Page no. 83 Chapter 2
Job Card IST 014 - Automatic X-ray Tube Warm-Up Chapter 2
1. SUPPLIES
None
2. TOOLS
None
3. SAFETY PRECAUTIONS
CAUTION
These procedures produce X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
This procedure should only be used when the performance of Installation Job Card Calibration of X-ray
Tube Heater Current
has been impossible due to excessive arcing.
5. PROCEDURE
This procedure is to be used for re-starting an X-ray tube which was known to be good upon leaving the
factory, but now exhibits arcing after a long period of inactivity.
The principle is to take a series of 12 pre-determined exposures, starting at low kV values and progres-
sively increasing towards high kV values. If arcing occurs during an exposure, some or all exposures are
automatically repeated. If persistent arcing occurs, the warm-up procedure is automatically terminated.
Proceed as follows:
4. Start from application mode. Change the position of the installation menu enable switch (switch B1 on
generator CPU board 400-PL3).
Note:
This must only be done if the Senographe DMR has been switched off since the last access to
SETUP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL.
5. Select SETUP/GENE/INSTAL/GENE/GENE/TUBE/WarmUp on the console. The track/kV combina-
tion for the first exposure should appear on the console display.
6. Press the 2nd trigger button and HOLD IT DOWN while the exposures are taken.
Note:
There is a time delay of 10 seconds between exposures. You can release the 2nd trigger button
during this 10 second period, then press it again to continue, but the minimum time between ex-
posures will still be 10 seconds. If the tube overheats at some point, exposures will be inhibited
until it cools down again. During the cooling period, the 2nd trigger button can be released; as soon
as the display shows that the X-ray tube has sufficiently cooled, press the 2nd trigger button to con-
tinue.
7. When the warm-up procedure is finished, the message
Tube OK normally appears on the console
display. If persistent arcing occurred during the procedures, the message
Too many spits appears on
the console display.

Chapter 2 Page no. 84 IST014.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 014 - Automatic X-ray Tube Warm-Up This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 015 - Installation of the CD-R writer (option)
Ist015.fm Page no. 85 Chapter 2
Job Card IST 015 - Installation of the CD-R writer (option) Chapter 2
1. SUPPLIES
CD-R (CD-ROM Writer) unit.
2. TOOLS
• Allen keys (metric).
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
None
5. PROCEDURE
5-1. Configure the CD-R unit
Two switches at the rear of the CD-R unit must be configured before installation. The correct configura-
tion is:
• Terminator switch: "ON".
• CDR-ID: "2".
5-2. Install and connect the CD-ROM Writer
1. Power OFF the system.
2. Remove cart rear cover (see D/R Job Card
Covers Removal).
3. Locate the equipment tray second from top (i.e., under the AWS tray); remove the screw which
secures it to the Cart left upright.
4. Disconnect the ground wire from the terminal at the right of the tray.
5. Pull out the tray and place the CD-R unit on it, facing to your left.
6. Connect the SCSI cable (supplied with the CD-R unit) between the AWS SCSI connector (on slot
PCI 1) and the CD-R SCSI connector "SCSI 1".
7. Locate an unused AC power supply cable (connected to the UPS) and plug it into the CD-R AC power
input socket.
8. Push the tray into its normal position, with the CD-R unit placed so that it is accessible through the
lower sliding hatch on the right of the cart (viewed from the front).
9. Reinstall the tray securing screw and the ground wire.
10. Set the switch on the front face of the CD-R unit to ON.
11. Power up the system, and check that the CD-R unit LEDs are lit.

Chapter 2 Page no. 86 Ist015.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 015 - Installation of the CD-R writer (option)
5-3. Software installation
A software procedure must be run in order to declare the presence of the CD-R writer to the AWS. Pro-
ceed as follows:
1. Open a command window (right-click on screen background and select
Service Tools/Command
Window).
2. Type: reboot -- -r <Return> ( type: reboot, space, dash, dash, space, dash, r)
3. The workstation automatically restarts and goes to the
login: prompt. Log in as sdc with adw3.1 as
password; wait until the station is correctly initialized
4. Open a command window.
5. Type:
su - <Return> followed by root password operator
cd ~sdc/install <Return>
install.cdr [cdr_key] <Return> (where
[cdr_key] is the CD-R software key)
6. You will see a message
CDR OPTION SUCCESSFULLY INSTALLED
7. Type again: reboot -- -r <Return> ( type: reboot, space, dash, dash, space, dash, r)
The CD-R unit is now installed.
5-4. Verification of the installation
• A new icon CD-R writer should be visible in the upper left corner of the browser window.
• Insert a blank CD-ROM in the unit and write a test image on it.
5-5. Error actions
If a problem is encountered after the installation (either the message CDR OPTION SUCCESSFULLY
INSTALLED
does not appear, or the verification of the installation fails), proceed as follows:
1. Open a command window
2. Type:
cd ~sdc/install <Return>
uninstall.cdr <Return>
3. Check connections and configuration and repeat the software installation procedure (section 5-3).

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 016 - Modem installation
Ist016.fm Page no. 87 Chapter 2
Job Card IST 016 - Modem installation Chapter 2
1. SUPPLIES
Modem
2. TOOLS
None.
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
None
5. PROCEDURE
1. Power OFF the system.
2. Remove the cart rear cover (see D/R Job Card
Removal of Covers).
3. Refer to the illustrations. Route modem cables to the
modem in its position on the shelf below the AWS. The
modem should be installed with the LEDs on the front
face visible from the rear of the cart:
- Connect the lead marked
Modem from the telephone
socket at the rear of the cart to the modem RJ-11
LINE socket; use an adaptor (RJ45/RJ11) if
required. The
PHONE socket is not used.
- Connect the modem serial cable from the AWS serial
port A (25-pin female connector on the AWS) to the
25-pin
RS232 connector on the modem.
- Plug the modem AC adaptor input lead into the UPS
output power strip. Connect its output to the
POWER
input on the modem.
4. Secure the modem and AC adaptor in their positions
using the Velcro pads and straps supplied, and use tie-
wraps to secure the cables as necessary.
AWS
IDC
Modem
UPS power out
Serial A
AC adaptor
Telephone connectors
View on rear of cart
UPS

PHONE LINE
View on rear of Multitech modem
POWER RS232

Chapter 2 Page no. 88 Ist016.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 016 - Modem installation This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
ist017.fm Page no. 89 Chapter 2
Job Card IST 017 - Image Receptor Assembly/Unpacking Chapter 2
1. SUPPLIES
Required if the Image Receptor is not pre-installed:
• Image receptor 2224137.
• Compression arm central cover (plate with 2 rails).
• Screw kit for image receptor assembly:
2. TOOLS
• Standard service toolkit.
3. SAFETY PRECAUTIONS
Power to the Gantry should be OFF before beginning the procedure.
4. PRE-REQUISITES
• The Gantry has been moved to its final position, and secured to the floor if required.
• All procedures likely to require movement and disturbance in the area of the Tube and Image Recep-
tor have been completed.
• Physical installation, inter-connection and a.c. connection of the Gantry, Generator Cabinet, and
AWS Cart have been successfully completed.
Name Reference Quantity Used for:
Screw CHC 6x25 2185581-7 4 Attaching the image receptor to the gantry column
Nut 1
Connecting the image receptor
Washer AZ5 99132748 2
Screw CHC5x10/10 99137572 1
Screw TF 4x12 99107110 2
Screw CHC5x50 99030316 2
Attaching cover assemblies – lateral covers
Stove bolt 4x10 99116465 2
White caps 45559940 2

Chapter 2 Page no. 90 ist017.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
5. PROCEDURE - UNPACKING A PRE-INSTALLED IMAGE RECEPTOR
Notice:!
If the Image Receptor has been pre-installed, its packing must be removed before any movement
of Gantry components (tube arm rotation, etc.).
Do not do this before:
• The Gantry has been moved to its final position, and secured to the floor if required.
• All procedures likely to require movement and disturbance in the area of the Tube and Image
Receptor have been completed.
When these conditions have been met:
• Carefully remove the packing from the Image Receptor
.
6. PROCEDURE - IMAGE RECEPTOR INSTALLATION
Note:
If the system is delivered with the image receptor not installed, it is shipped in a separate box with
additional thermal protection. The Senographe 2000D system is assembled and calibrated as usu-
al on the manufacturing line; the image receptor is then disconnected from the system and pack-
aged separately.
If the Image Receptor is not pre-installed, install it after:
• The Gantry has been moved to its final position, and secured to the floor if required.)
• All procedures likely to require movement and disturbance in the area of the Tube and Image
Receptor have been completed.
6-1. Remove the front column lateral covers
• Remove both compression arm lateral covers (see D/R Job Card Removal of Covers).
6-2. Unpack the image receptor
• Open the image receptor case and remove all packaging.
• Take the image receptor out of its plastic covering and place it carefully on a stable stand.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
ist017.fm Page no. 91 Chapter 2
6-3. Attach the image receptor to the gantry column
Notice:!
The procedure requires the tube arm to be rotated for ac-
cess to locating screws behind the gantry
EITHER: Power the Gantry ON when the tube is to be
rotated, so that the arm can be turned in the normal way.
If this is done, it must be powered OFF again before
making electrical connections.
OR: Use a screwdriver to release the arm locking pin
(A) while moving the arm.
CAUTION
The image receptor is heavy. Two people are
required to install it.
1. Rotate the tube to the right (see note above), to allow
access behind the left of the gantry.
2. Identify the two alignment pins (B) in the gantry column.
3. Carefully insert the image receptor into its alignment pins.
WARNING
WARNING: Take care not to pinch the cables behind
the image receptor.
4. While one person continues to support the image receptor
(C) in position, the other inserts two securing screws (D) at
the rear of the image receptor. Do not tighten them com-
pletely.
5. Rotate the tube to the left to allow access behind the right of
the gantry.
6. Insert the two securing screws at the rear right of the image
receptor, but do not tighten them completely.
7. Rotate the tube to the right again; tighten the screws.
8. Rotate the tube to the left again; tighten the screws.
A
B
D
C

Chapter 2 Page no. 92 ist017.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
6-4. Make electrical connections
Notice:!
It is important that electrical connections are made before connecting the cooling pipes. This is to
minimize any risk of coolant spill onto electrical connectors, which can result in corrosion damage.
1. Locate the bucky connector plate (C) on the end of a
cable descending through the gantry. Remove its
protection, and insert it in the corresponding holes of
the image receptor interconnection plate (D).
2. Secure it to the image receptor with two screws (E).
3. Remove the protection from the set of image receptor cables and pipes. This includes two water
pipes, two air pipes, two fiber-optic cables, and the power supply connector.
4. Connect one end of the ground cable 2248904 to the stud
located at the rear of the image receptor connector panel,
using a washer and a nut (F).
5. Connect the other end of this ground cable, and the
ground cable 45559796 (included in the gantry cables),
to the receptor metal chassis (G), using a screw and a
washer.
6. Plug the power supply connector into its socket (H) at the
front of the image receptor connector panel and secure it
by tightening the two screws.
E
C
D
F
G
H

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
ist017.fm Page no. 93 Chapter 2
6-5. Make air, water, and optical connections
1. Connect the fiber-optic cables to their connectors (H) on the image receptor connector panel.
Note: one of these cables is marked with the letter L. Connect the
other, unmarked cable to the con-
nector colored red (see sketch).
2. Connect the air pipes to the smaller pair of connectors (J). The order is not important.
3. Connect the water pipes to the larger pair of connectors (K). The order is not important.
CAUTION
If any coolant is spilled, it must be cleaned up immediately.
Red marker
Fiber-optic cable marked L
J
K
H

Chapter 2 Page no. 94 ist017.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
5.5 Assembling the covers
Note:
You may choose to perform the cover installation at this point, or leave the arm uncovered and put
all the covers on at the end of the installation.
6-5-1. Central cover
Notice:!
This cover is delicate, take care when installing it.
1. Remove the paddle and turn the arm to the right until it reaches the horizontal position.
2. Move the paddle support as close as possible to the tube.
3. Slide the central cover,under the paddle support until it clears the top of the image receptor.
4. Slide it back into position behind the image receptor.:
5. Connect the ground cable to the cen-
tral cover using pliers:
6. Move the paddle support to the mid position to keep the central
cover in position while preparing to install the lateral covers
12

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking
ist017.fm Page no. 95 Chapter 2
6-5-2. Lateral covers
Install both compression arm lateral covers (see D/R Job Card
Removal of Covers).
6-5-3. Check the central cover assembly
Push with your finger on different points of the central cover.
It should remain in place and not move towards the back of the col-
umn. If this is not the case, the three covers are not well assembled;
they must be removed and re-installed, taking care to use the lateral
cover guides.

Chapter 2 Page no. 96 ist017.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 017 - Image Receptor Assembly/Unpacking This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 018 - Bar Code Scanner (Option)
ist018.fm Page no. 97 Chapter 2
Job Card IST 018 - Bar Code Scanner (Option) Chapter 2
1. SUPPLIES
Bar Code Scanner and connecting cable, p/n 225842 (FRU 2).
2. TOOLS
None
3. SAFETY PRECAUTIONS
None
4. PRE-REQUISITES
The system must have been correctly installed and normal application software must be operational.
5. PROCEDURE
5-1. Introduction
The FFDM scanner P/N 2258242 is a fast and accurate tool
for scanning bar code symbols in AWS applications. It allows
patient information to be entered rapidly and accurately
Available with an 80mm CCD scanning window and an inte-
grated keyboard decoder, the scanner captures an entire bar
code symbol instantly. The integrated decoder then transmits
the data to the AWS workstation just as if the data were
entered from the AWS keyboard.
Supported bar code types are:
- Code 3 of 9 (standard & full-ASCII).
- EAN 13.
- Code 128.
To pTOP
BOTTOM
SIDE
LED and CCD beam window
Handle light
Scanner trigger

Chapter 2 Page no. 98 ist018.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 018 - Bar Code Scanner (Option)
5-2. Installation
Refer to the connection diagram below when connecting the bar code reader:
1. Power down the AWS workstation.
2. Unplug the keyboard cable from the AWS workstation rear panel.
3. Withdraw the keyboard cable from the cart and lift the keyboard from its normal position.
4. Pass the DIN 8-pin male branch of the scanner Y-cable to the AWS workstation, following the path of
the removed keyboard cable.
5. Connect the DIN 8-pin male connector to the keyboard connector on the workstation rear panel.
6. Connect the RJ45 connector on the long branch of the Y-cable to the bar code scanner handle con-
nector.
7. Connect the keyboard cable to the DIN 8-pin female connector on the Y- cable.
8. After checking the security of cable connections, route the keyboard cable and connectors inside the
cart, with the scanner cable exiting in a convenient position for use. Replace the keyboard in its nor-
mal position.
5-3. Check Installation
• Before trying to use the scanner, make sure that all cable connections are secure.
• Power up the AWS workstation and wait until the login banner is displayed.
5-3-1. Read a bar code symbol
1. Hold the scanner handle horizontally directly over the bar code printed below.
2. Press the scanner trigger located below the scanner handle; the scanner red LED beam lights. The
Keyboard
AWS Workstation
bar code
scanner
RJ45
DIN8F
DIN8M
DIN8M

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 018 - Bar Code Scanner (Option)
ist018.fm Page no. 99 Chapter 2
scanner has successfully read the symbol if:
- You hear a beep.
- The LED beam turns off.
- The handle light turns red.
5-3-2. Set default configuration
• Use the scanner to read the symbols below in the order given to activate the scanner and decoder
default configuration. Refer to the scanner documentation and/or the AWS operator manual to acti-
vate other functions if necessary.
1. Reset the bar code scanner to factory settings. After scanning, wait until you hear 3 beeps:.
2. Start Settings. You should hear 2 beeps after scanning:
3. Set Full ASCII transmission to allow extended Code 39 characters transmission:
4. Set Extended Full ASCII to allow Function keys transmission on two characters:
5. Configure the keyboard emulation model setting:
6. Enter code 719 to support the SUN Type 5/6 QWERTY keyboard (the unsupported AZERTY key-
board code is 718; refer to the bar code and decoder user guides for more information on how to con-
figure alternate keyboards):

Chapter 2 Page no. 100 ist018.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 018 - Bar Code Scanner (Option) 7. Force keyboard to CAP LOCK OFF mode after transmission:
8. End Settings. You should hear 2 beeps after scanning.
5-3-3. Complete a Medical Procedure Card
Perform the following procedure to check for correct installation and configuration of the scanner.
1. Login as sdc on the AWS:
Login: sdc <Return>
Password: adw3.1 <Return> (factory loaded default sdc password)
2. Wait until AWS application has started launched.
3. Request a new exposure and open the Medical Procedure Card (MPC).
4. Click on the first text input field
Last Name with the mouse left button to set the keyboard focus.
5. Complete the MPC by scanning all the symbols given on the following page in order, scanning left-to-
right/top-down. As the MPC user interface may change from one AWS release to another, the layout
and entries given in these symbols are templates only: field values are not meaningful and can have
any value. To go to (change focus to) the next input field, scan the $I bar code symbol.
6. On completion, the Medical Procedure Card fields should appear as shown below:

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 018 - Bar Code Scanner (Option)
ist018.fm Page no. 101 Chapter 2

Chapter 2 Page no. 102 ist018.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card IST 018 - Bar Code Scanner (Option) 5-4. What If…
• If nothing happens when you follow the operating instructions:
- Check the system power.
- Check for loose cable connections.
• If the scanner does not read a bar code symbol:
- Be sure that the scanning system is programmed to read the type of bar code you are trying to
scan. Refer to section 5-3-2 for default setting.
- Check that the bar code symbol is not defaced.
- The scan head must touch and cover every bar and space on the symbol as shown below. You
can tilt the scanner forward or backward to achieve a successful decode. Simple practice quickly
shows what tolerances to work within.
• If the AWS workstation returns to displaying the monitor
ok prompt:
- Connecting/disconnecting or resetting the bar code scanner may cause the workstation go back
to the SUN monitor
ok prompt. Type the go command to resume to the normal working mode.
- If necessary, refresh the screen using the appropriate root window menu option.
Right Wron g

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
CONFIG.FM Page no. 103
CHAPTER 3 CONFIGURATION
This chapter contains information for configuration of the Senographe 2000 D, including a description of
basic functions available from the Gantry Control Console.

Page no. 104 CONFIG.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 001 - The Control Console
CFG001.FM Page no. 105 Chapter 3
Job Card CFG 001 - The Control Console Chapter 3
1. INTRODUCTION
• This Job Card provides an introduction to use of the Gantry Control Console, its menu structure, etc.
ILLUSTRATION 1 - SENOGRAPHE CONTROL CONSOLE
2. HOW TO USE THE CONTROL CONSOLE
2-1. Basic functions
Console functions for the normal application are described in the Operator Manual; other operating
modes are summarized below.
•
Access to functions. Press one of the console function keys to select a displayed function or menu
choice, which in turn may display a new set of menu selections. Continue selection until the required
function is displayed.
•
Exit. To go up (back) one level in the menu structure, press the SETUP key on the console once.
•
Selection functions. These functions require simple selection of a menu choice from a display of
options or values; confirm the choice by selecting VALID.
•
Modification functions. These functions allow a displayed value or parameter to be changed by rota-
tion of the console kV knob. When the required value or parameter is displayed, confirm the selection
by selecting VALID.
•
Calibration functions. Some calibration functions require the 2nd trigger button to be pressed in order
to perform the calibration.
WARNING
NEVER modify a Senographe 2000 D parameter value that has already been calibrated
unless you have a specific reason. If you do, it will usually be necessary to enter the
default parameter value and completely re-calibrate the generator or gantry for the
affected parameter(s).
2-2. Application and Installation modes
Throughout the service documentation, reference is made to application mode and installation mode.
•Application mode. The normal operating mode; the Gantry enters application mode when it is pow-
ered up, regardless of previous settings. In application mode, the six function keys below the display
window carry out the functions indicated by their symbols (focal spot, track, filter, etc.).
For information on using the Console in application mode, refer to the Operator Manual.
kV mAs

Chapter 3 Page no. 106 CFG001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 001 - The Control Console •Installation mode. Provides special setup functions. To enter or leave installation mode, press the
SETUP key on the console once.
In installation mode, the six function keys below the display window carry out functions or menu
selections which depend on the current procedure. The choice is indicated by a label displayed above
the key. The illustration below shows typical console displays before (left) and after (right) entering
installation mode. The normal key symbols (shown shaded) no longer apply, and in the example
shown two of the keys (shown shaded) have no function; the remaining four carry out the functions
displayed above them (e.g., the left hand key now gives access to the GENE menus).
• Some special
installation menus are displayed only after changing the position of the installation
enable switch (see below).
2-3. Installation Menus
Installation Menus are special menus which allow specific Gantry or Generator installation procedures to
be performed. To access installation menus from application mode, proceed as follows:
1. Enter installation mode (press the SETUP key on the console once).
2. Change the position of the installation menu enable
switch (switch B1-8 on Generator CPU Board 400-
PL3).
Note that the original position of the switch is
not importan
t. When the system is powered up it
automatically enters application mode with installa-
tion menus disabled; the switch position at that time
is memorized as the "installation-menu-disabled"
position.
3. Two function keys are now labelled (on the display) as GENE and ARM. Select Gantry or Generator
installation menus as follows:
a.
For Gantry installation menus: press the key labelled ARM, then press the VALID key. The two
right-hand keys are now labelled INSTAL and MAINT. Press the INSTAL key to access installation
menus (the MAINTenance menu is not available during installation).
b.
For Generator installation menus: press the key labelled GENE. The VALID key is not required.
The two right-hand keys are now labelled INSTAL and MAINT. Press the INSTAL key to access
installation menus (the MAINTenance menu is not available during installation).
Notice:!
When you have finished using installation mode, you must switch off the Gantry before handing
over the machine to the user, to lock out access to installation mode.
+0SMALL AUTO AUTO A
SELECT ARM/GENE INSTALL
GENE ARMMEDICAL LANGUAGE
?
GENE CPU BOARD 400-PL3
B1
P1 P2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 001 - The Control Console
CFG001.FM Page no. 107 Chapter 3
2-4. Console Self-Test
To carry out a console self-test you must press five keys simultaneously. Refer to the illustration below.
Press and hold the two keys A, and press the three keys B.
2-5. Saved Parameters Checksum
Parameters resulting from installation or non-proprietary maintenance procedures are saved in memory
zones whose contents are not lost when the Gantry is switched off.
To perform the saved parameter checksum function CKSUM, select either SETUP/GENE/INSTAL/
CKSUM/CKSUM (for Generator parameters) or SETUP/ARM/VALID/INSTAL/CKSUM/CKSUM (for
Gantry parameters). You can now return to application mode without receiving a checksum error mes-
sage.
It is not mandatory to do a checksum each time you leave a menu. There is no risk of losing data if you
switch off the DMR displaying a checksum error.
2-6. Re-loading Default (Initial) Parameters
WARNING
These functions are not normally used after the Senographe 2000 D has left the factory.
To re-load generator or gantry default parameters (prior to any installation or calibration procedures), use
the appropriate CLEAR menu. This executes a software power-up comparable to the first hardware
power-up of the generator or gantry.
When the Generator CLEAR menu is used, the DMR returns to application mode (to return to installation
mode, you must change the position of the installation menu enable switch again).
When the Gantry CLEAR menu is used, the DMR returns to the gantry installation entry level (SETUP/
ARM/VALID).
After use of the CLEAR menu, a CHECKSUM ERROR message is displayed. All generator or gantry
calibration procedures must be performed, followed by a CKSUM to enable use of the new param-
eters.
WARNING
The generator CLEAR menu causes the recorded thermal condition of critical components
(tube, tube housing, generator, etc.) to be reset to the cold condition. This means that an
excessive temperature rise at the time that this menu is used may not be immediately
detected by the software thermal protection.
kV mAs
AB

Chapter 3 Page no. 108 CFG001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 001 - The Control Console 2-7. Representation of Numeric Parameters
When displayed on the console, most numeric values appear in the form of real numbers in scientific
notation floating point format with four significant figures, as shown in the examples below:
-1.250E+3 (= -1250.00); +4.345E-1 (= 0.4345) +1.000E+0 (= 1.000)
2-8. Display of Parameter Values
To display parameter values on the Control Console, access the calibration menu for the procedure that
computes the parameter. Typically this is done at a menu choice between CALIB and PARAM. Select
PARAM to display the parameter value.
If the calibration procedure computes more than one parameter, turn the kV dial to display the
parameters in sequence.
2-9. Modification and entry of values
At menu points allowing entry of a measured value or modification of a parameter, there are two possible
selections: NEXT and VALID. Select NEXT to allow the first character to be entered (by changing the dis-
played character). Select NEXT again to enter the following character, and so on.
Use the kV dial to change the character being entered. Numeric or alpha-numeric characters are
displayed, according to the function. Parameters are entered or modified in scientific notation form (see
above). Select VALID when all characters are correctly entered. If you make a mistake, select NEXT
repeatedly until you arrive at the faulty value, then correct it by rotating the kV dial.
To exit the modification routine, press SETUP once.
Note:
Pressing the VALID key is effective ONLY when the entire value, including exponent, is displayed.
This means that if you correct a digit after making a mistake, as described above, you must press
the NEXT key repeatedly until the entire parameter value is visible before you can enter the new
value using the VALID key.
2-10. The DISP Parameter
• The DISP (display) parameter is provided to aid service procedures; it controls the display of certain
parameters after an exposure. When it is set to YES, the two groups of exposure parameters shown
in the table below are displayed alternately
:
Composition Breast composition
Thickness: Breast thickness in equivalent cm of plexiglass.
Prep dose: Dose for pre-shot (returned by IDC)
Prep kV: kV used for pre-shot
Prep track: Track used for pre-shot
Prep filter: Filter used for pre-shot
Delta grid: Delta time (ms) from end of X-ray emission to grid U-turn
To change the DISP parameter, access SETUP/GENE/INSTAL/GENE/CONFIG/DISP , and select YES or
NO. A CKSUM is NOT necessary because this parameter is not a saved parameter. When power is
turned off, the value of DISP is lost and reverts to NO at the next power-up.
Group 1: kV mAs mA Time Ih
Group 2:Composition Thickness Prep dose Prep kV Prep track Prep filter Delta grid

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 002 - Gantry and Generator Configuration
CFG002.FM Page no. 109 Chapter 3
Job Card CFG 002 - Gantry and Generator Configuration Chapter 3
1. INTRODUCTION
This module provides information for initial configuration of a Senographe 2000 D system. It contains two
sections:
•
Configuration procedures. Procedures which are required to set up the system for local site condi-
tions.
•
Configuration check information. For use only if you suspect that certain important settings have
been changed; most settings are made in the factory and do not need to be changed on site.
2. CONFIGURATION PROCEDURES
2-1. Prerequisites
The system must have powered up and reached a stable state with no checksum error displayed on the
console.
• If, on the first power up, CHECKSUM ERROR (GEN) or ARM CHECKSUM ERROR is displayed, pro-
ceed as follows:
Reload the Generator or Gantry (or both) saved parameters, using the diskette (p/n 2152250) deliv-
ered in the accessories kit. This diskette is labelled with the same product locator number as your
Senographe 2000 D. Files are identified as Gxxxx.DAT (Generator) and Axxxx.DAT (Gantry/Arm),
where xxxx is the factory serial number. There should be no checksum error after correct reloading of
these parameters.
• Enable access to installation menus by changing the position of the installation menu enable switch
(switch B1-8 on generator CPU board 400-PL3) as described in the Job Card
The Control Console.
2-2. Set Language
• If necessary, change the console display language by selecting: SETUP/LANGUAGE/ language ,
where language is the desired language (FRANC, ENGL, DEUTS, ESPAN, ITAL, or PORTU).
Note:
For compatibility with English language documentation, you are recommended to select ENGL
during installation of the system; change to the local language when installation is complete.
2-3. Set Maximum Generator Output Power
• Select SETUP/GENE/INSTAL/GENE/CONFIG/CONFIG/PMAX/ 3.0, to set the maximum generator
power to 3.0 kW.
2-4. Set Maximum X-ray Tube mAs
• Select SETUP/GENE/INSTAL/GENE/CONFIG/CONFIG/MASMAX . Enter the maximum mAs value
according to local regulations, using the NEXT and VALID keys and rotating the kV dial.
Note:
The default (and maximum possible) value is +6.000E+2 (600) mAs.

Chapter 3 Page no. 110 CFG002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 002 - Gantry and Generator Configuration
2-5. Set Maximum X-ray Exposure Time
•Select SETUP//GENE/INSTAL/GENE/CONFIG/CONFIG/T_MAX . Enter the maximum X-ray expo-
sure time according to local regulations.
Note:
The default value is +6.000E+0 (6) seconds; the maximum is +2.000E+1 (20) seconds.
2-6. Set Collimator Front Blade Presence
•Select SETUP/ARM/VALID/INSTAL/CONFIG/COLLIM/FRONT_B/YES.
The default setting is YES (collimator V4 with front blade control).
2-7. Set X-ray Beam Margins
Set the lateral, rear, and front beam margins as required to meet any specific local regulations and rec-
ommendations; note the current setting values for future reference. The tables below give examples of
margin settings for some of the main standards currently applicable.
1. Lateral margin: select SETUP/ARM/VALID/INSTAL/CONFIG/COLLIM/MARG_L
2. Rear Margin: select SETUP/ARM/VALID/INSTAL/CONFIG/COLLIM/MARG_R .
3. Front Margin: select SETUP/ARM/VALID/INSTAL/CONFIG/COLLIM/MARG_F .
4. Perform CKSUM; return to application mode
Standard Lateral margin (mm)
CDRH/CEI/MQSA/Factory default +5
UK +2
AUSTRALIA -4
Current setting ...
Standard Rear margin (mm)
CDRH/CEI/MQSA/Factory default +2
UK +2
AUSTRALIA -4
Current setting ...
Standard Front margin (mm)
CDRH/CEI/MQSA/Factory default +2
UK +2
AUSTRALIA +2
Current setting ...

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 002 - Gantry and Generator Configuration
CFG002.FM Page no. 111 Chapter 3
.
2-8. Set Gantry Medical Parameters
1. Ask the doctor for the desired paddle compression speed (fast or slow). From application mode,
select SETUP/MEDICAL/SPEED . Select FAST or SLOW as required.
2. Ask the doctor for the desired maximum compression force in daN. Select SETUP/FORCE. Press +
or - on the console repeatedly until the desired force value is reached (possible values range from 4
to 20 daN in steps of 1 daN). Select VAL on the console to confirm the desired value.
Note:
The ACR recommendation is 18daN (40lbs).
3. Ask the doctor if automatic decompression (i.e. return of the compression paddle to its home posi-
tion) following an exposure is desired. Select SETUP/DECOMP. Select YES or NO as required.
4. Ask the doctor if a beep tone is desired. Select SETUP/BEEP. Select YES or NO as required.
5. Ask the doctor for the desired maximum height during automatic decompression (between 10 and 38
cm). Select SETUP/HEIGHT. Press + or - on the console repeatedly until the desired maximum
height is displayed.
3. CONFIGURATION CHECK INFORMATION
Note:
It is not normally necessary to set the configuration parameters listed here. They are set to their
correct values in the factory, or by the link from the AWS. They are given here for convenience in
the event that you suspect that they have been changed, perhaps during trouble-shooting.
3-1. Set Date and Time
• The Generator clock is updated automatically by the AWS each time the DMR subsystem is powered
up (either alone or when the complete system is powered up). It is not normally necessary to change
it manually.
Note:
After a Generator CLEAR operation, the current date and the date of the last installation mode ac-
cess are both set to the Generator software release date; the time is set to 0:00:00. The current
date and time are updated at the next power up.
• The clock is used for the following functions:
- Thermal protection cooling algorithms, whether or not the generator is turned on.
- Recording dates and times in the error log.
- Recording the date of last access to installation mode.
• If it is necessary to change the date or time on the Gantry manually, note the following points:
- Dates (menu path: SETUP/GENE/INSTAL/GENE/CONFIG/DATE ) are entered in the form DD/
MM/YY; e.g., for February 14th, 2001, you must enter 14/02/01.
- Time (menu path: SETUP/GENE/INSTAL/GENE/CONFIG/HOUR ) is entered in 24-hour format;
e.g., for 3:13:27 pm you must enter 15:13:27.
3-2. Check X-ray Tube Housing Configuration.
• Select SETUP/ARM/VALID/INSTAL/CONFIG/TUBE/YES. (the X-ray tube housing is inclinable on all
Gantries).
Perform a gantry CKSUM and return to application mode. Check that NONE of the following error
messages appears on the console display:

Chapter 3 Page no. 112 CFG002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 002 - Gantry and Generator Configuration S53 105/080 (incompatible housing configuration)
S53 105/081 (incompatible housing configuration)
S54 155/082 (incompatible tube type)
S54 155/083 (incompatible tube type)
3-3. Check X-ray Tube Type
1. Select SETUP/GENE/INSTAL/GENE/CONFIG/CONFIG/TUBE .
Verify that the console display shows the correct tube type:
2foc/2met/1ang (dual focus, bi-metal, sin-
gle anode angle or slope; This corresponds to tube type GS412-49).
If the configuration is not correct, proceed with steps 2 and 3 below.
2. If necessary, set the correct X-ray tube type:
Select FOCUS/DUAL on the console.
Select SETUP/ANODE/DM on the console.
3. Press SETUP; verify that the console display now shows the correct tube type (above).
3-4. Check Presence parameters
The following equipment presence parameters should normally be set to YES:
• Gantry (Arm) presence; SETUP/GENE/INSTAL/GENE/GCONFIG/CONFIG/PRES/PRES_B
• Digital System presence; SETUP/GENE/INSTAL/GENE/CONFIG/CONFIG/PRES/PRES_DS
• IDC presence; SETUP/GENE/INSTAL/GENE/CONFIG/CONFIG/PRES/PRES_IDC
3-5. Check Paddle Deformation Parameters
• Check the values of paddle deformation parameters against the default factory settings given below;.
These parameters alter the X-Ray exposure behaviour in AOP mode...
• After any modification, perform a CKSUM and return to Application mode
Paddle Offset Value Gain Value
19 x 23 paddle in contact mode: access
SETUP/GENE/INSTAL/GENE/AOP/PADD/18x24
Standard metallic arm
paddle for
Senograpghe.2000 D (*)
+1.000 E + 0 +6.000 E - 1
Magnification paddles: access
SETUP/GENE/INSTAL/GENE/AOP/PADD/MAGNIF
All magnification
paddles(*)
+2.000 E + 0 +2.000 E - 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 003 - Senographe Jumpers and Switches
CFG003.FM Page no. 113 Chapter 3
Job Card CFG 003 - Senographe Jumpers and Switches Chapter 3
1. SUPPLIES
None
2. TOOLS
None
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
None
5. JUMPERS AND SWITCHES
The information in this Job Card is provided only to allow verification of the position of jumpers and
switches if necessary. Tthey should normally be factory-set.
Jumpers and switches are divided into three groups: Generator, Mains Distribution Rack, and Gantry.
5-1. Generator jumpers and switches.
1. Check (visually, do not remove it!) that the jumper on the Generator CPU board 400PL3 is installed in
position X2 as shown in Illustration 1.
ILLUSTRATION 1 - GENERATOR AND GANTRY CPU BOARD JUMPER X2
CAUTION
This jumper must NEVER be removed from position X2 and must NEVER be placed in posi- tion TP1. If this jumper is either removed or placed in position TP1, the generator CPU RAMs are totally erased, necessitating a TOTAL RE-CALIBRATION of the generator and
AEC.
2. Check that jumpers JP1 and JP2 are installed on generator command board 400PL1 (they must NOT
be installed in the parking position JP3 and JP4 at this point). See illustration 2.
Note:
JP1 and JP2 are connected to the room relays on 200PL4. See Installation Job card
Room Lamps
and Room Door Connection
for more information.
GENE CPU BOARD 400-PL3 or
GANTRY CPU BOARD 800-PL3
P2P1
B41
B46
BT1
X2
! !

Chapter 3 Page no. 114 CFG003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 003 - Senographe Jumpers and Switches
ILLUSTRATION 2 - GENERATOR COMMAND BOARD JUMPERS
5-2. Mains Distribution Rack jumpers and switches.
• Check that the jumpers on the Mains Distribution Board are installed as follows:
5-3. Gantry jumpers and switches.
1. Check (visually, do not remove it!) that the jumper on the Gantry CPU board 800PL3 is installed in
position X2 as shown in Illustration 1.
CAUTION
This jumper must NEVER be removed from position X2 and must NEVER be placed in posi-
tion TP1. If this jumper is either removed or placed in position TP1, the gantry CPU RAMs
are totally erased, necessitating a TOTAL RECALIBRATION of the gantry.
2. Check that the jumpers on the Gantry board
800PL1 are correctly set for the bucky motor
type. They should be set as follows
Jumper Present? Where?
TP1 NO Between K2 and K6
TP2 NO Beside K8
TP3 YES Beside K10
Jumper Present?
TP104 NO
TP105 YES
TP106 NO
TP107 YES
TP102 YES
TP103 NO
XJ21
TP105
TP104
TP107 TP106
TP102
TP103

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 004 - Check Supply Voltages
Cfg004.fm Page no. 115 Chapter 3
Job Card CFG 004 - Check Supply Voltages Chapter 3
1. SUPPLIES
None
2. TOOLS
• Accurate digital voltmeter (Fluke 8060 or equivalent)
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
The procedures described in Installation Job Card AC Connection must have been completed correctly.
5. FIRST POWER-ON CHECK
The objective in this section is to detect incorrect mains supply configuration - before it is too late!
5-1. Check keep-alive supply voltage.
1. Check that generator Mains Supply Switch 200-S1 is open
(i.e. in the ”0” position) and that the Mains Distribution Rack
circuit breaker is open (i.e. in the ”0” position ).
2. Close the main hospital supply circuit breaker.
3. Connect the voltmeter (set to 20 VDC scale) to generator
mains power board 200-PL1 between test points PT15 (+V
E)
and PT12 (0V
E
). See illustration.
4. Close generator Mains Supply Switch 200-S1 (i.e., turn it to
the ”1” position) and read the voltmeter immediately. It should
read +8 ±2 VDC.
CAUTION
If the voltmeter reading is outside this range, turn mains source OFF IMMEDIATELY and
check configuration of voltage selection board IN2 on generator mains distribution board
200-PL1 (see Installation Job Card
AC Connection).
5. Move the voltmeter from PT12 to PT14 (-VE). The voltmeter should read -8V ± 2VDC
6. Connect the voltmeter (set to 500 VAC scale) to generator mains power board 200-PL1 between ter-
minals 1 (0V) and 4 (220V).
7. Turn console on and read the voltmeter immediately. It should read 220 ± 20 VAC.
CAUTION
If the voltmeter reading is outside this range, open generator Mains Supply Switch 200-S1
IMMEDIATELY (turn it to the "0” position) and check mains voltage selection wire SEL on
generator mains distribution board 200-PL1
(see Installation Job Card AC Connection).
PT14 PT15
X1
TR2
R1
R4
PT12
- VE + VE
0V
E
+V
E
POWER BOARD 200-PL1

Chapter 3 Page no. 116 Cfg004.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 004 - Check Supply Voltages 6. CHECK GENERATOR AND GANTRY DC SUPPLY VOLTAGES
6-1. Check generator DC supply voltages.
1. On generator CPU board 400-PL3, measure the DC voltage between TP110 and TP111 (GND). It
should be in the range +5.1±0.05 V. If necessary, adjust the voltage:
WARNING
The adjustment is made using the potentiometer located on the low voltage power supply unit (on the rear of the board mounting plate). To avoid any risk of electric shock, DO NOT TOUCH any of the other components on the supply (later models are protected by a transparent cover).
2. On generator interface board 400-PL2, measure DC voltage (+12 V nominal) between PT3 and PT4.
The voltage should be in the range +11.5 to +12.5 VDC.
3. On generator interface board 400-PL2, measure DC voltage (-12 V nominal) between PT5 and PT4.
The voltage should be in the range -11.5 to -12.5 VDC.
Note:
Voltages measured in steps 2 and 3 above depend on the +5 V adjustment mentioned in step 1.
6-2. Check gantry DC supply voltages.
1. On gantry power board 800-PL1, measure DC voltage (+27 V nominal) between terminals 3 and 1
(GND) of connector TB1. The voltage should be in the range +26 to +34 VDC.
2. On gantry CPU board 800-PL3, measure DC voltage (+5 V nominal) between TP110 and TP111
(GND). The voltage should be in the range +4.75 to +5.25 VDC.
Note:
Unlike the +5 V supply on the generator CPU board, the +5 V supply on the gantry CPU board can-
not be adjusted.
7. CHECK MAINS DISTRIBUTION RACK INPUT
If the above check for generator voltages was correct, this confirms that the correct voltage has been
selected for the generator input voltage.
Before switching power to the Mains Distribution Rack, check that the same voltage has been
selected for the Mains Distribution Rack transformer input tap.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 005 - AWS Configuration
CFG005.FM Page no. 117 Chapter 3
Job Card CFG 005 - AWS Configuration Chapter 3
1. SUPPLIES
None.
2. TOOLS
None.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Display the Configuration/System Overview Window
1. From the application browser screen, select Toolkit/Service Desktop/Configuration.
2. The System Overview window shows current site and communications settings.
ILLUSTRATION 1 - SYSTEM OVERVIEW WINDOW

Chapter 3 Page no. 118 CFG005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 005 - AWS Configuration 5-2. Set Site Information
1. Click the Site button under Configuration to display the Site Information Screen (Illustration 2).
2. To change any of the displayed items, enter the new information and confirm the entry by clicking the
Submit button.
ILLUSTRATION 2 - SITE INFORMATION SCREEN
Note:
1. The entry
Station Name is the name that will be used for the system when interacting with the
RIS server. It is recommended that this name be the same as that used for TCP/IP networking con-
figuration (see the CFG Job Card
Networking Connection and Configuration).
2. If there is more than one Senographe 2000 D system on the hospital site, the names chosen
must clearly differentiate between them (for example, by including the unit number or name).

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 005 - AWS Configuration
CFG005.FM Page no. 119 Chapter 3
5-3. Set General Configuration Parameters
1. Click the General button under Configuration to display the General Configuration screen (Illustration
3).
2. To change any of the displayed items, enter the new information and confirm the entry by clicking the
Submit button.
ILLUSTRATION 3 - GENERAL CONFIGURATION SCREEN
3.Auto Delete. By default, the auto-delete feature is not active. To activate it on request from the user,
select
Toolkit/Medical application preferences/Auto delete. When the disk usage exceeds the config-
ured
Upper Threshold value, images are automatically deleted in chronological order, oldest first,
until the disk space reaches the
Lower Threshold. Images which have been on the disk for less than
the
Minimum Image Storage time are not deleted.
Notice:!
Do not activate Auto Delete or change Auto Delete settings without agreement from the user.
4.QAP Time-out. Sets the time at which a warning that QAP must be carried out will be displayed
5.System Clock. Sets the current date and time.
Notice:!
After changing the clock the system must be shut down and restarted. If this is not done, system
behaviour may be unpredictable, because several interacting processes use the clock.
6.Language. Check the button beside the required application language
Notice:!
If you change the application language, you must perform a shutdown and restart before the
change will take effect.

Chapter 3 Page no. 120 CFG005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 005 - AWS Configuration This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 006 - Declare Network Hosts
CFG006.FM Page no. 121 Chapter 3
Job Card CFG 006 - Declare Network Hosts Chapter 3
1. SUPPLIES
None.
2. TOOLS
None.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
Obtain full details of the addresses, names and characteristics of the host(s) to be declared, from the
network administrator concerned and/or Remote Host Conformance statement documents.
5. PROCEDURE
Note:
Hosts may be declared before they are physically connected.
5-1. Remote Host Selection Window
• Click on the Tools icon in the upper right corner of the Browser to display the Tools menu; select
Network management in the menu to display the
Remote host selection window. This window dis-
plays a list of all currently declared hosts. Click on a host name to select it. Click the appropriate but-
ton at the bottom of the window to Add or Remove a host. Click Update to apply the changes, or
Cancel to close the window without applying changes.
.
Remote Host Selection
List of Registered Hosts
Your choice :
Update Add Remove Cancel

Chapter 3 Page no. 122 CFG006.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 006 - Declare Network Hosts
5-2. Declaration of a new host
To add a host to the List of Registered Hosts, click the Add button in the Remote host selection window.
The
Remote host parameters window is displayed:
:
DICOM 3.0
Host Label
Hostname
Application Entity Title
Host Label
Port number
Query Retrieve
Storage Commitment
Message send
Network protocol
Application Entity Title
Network address
Port number
Comment
Save Clear Cancel
Remote Host Parameters

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 006 - Declare Network Hosts
CFG006.FM Page no. 123 Chapter 3
Complete the fields in this window according to the information received from the network administrator
concerned and the remote host device DICOM 3.0 Conformance Statement:
1. The following fields must be completed for all types of remote host devices (Review Station, Non-
Storage Commitment capable Archival device, and Storage Commitment capable Archival device):
- HOST LABEL: Enter the name you want to give to the icon that will represent the remote host
- HOSTNAME: Enter the host name provided by the hospital network administrator.
- APPLICATION ENTITY TITLE: Enter the Application Entity Title
(A.E.T.)
- NETWORK ADDRESS: The network address may be local (see Job Card Networking Connection
and Configuration
), or provided by the hospital network administrator.
- NETWORK PROTOCOL: Always select
DICOM 3.0.
- PORT NUMBER: Refer to the device DICOM 3.0 Conformance Statement; normal value is 4006.
2. The following field must be completed if the remote host device is a Review Station:
- QUERY RETRIEVE: Select
Patient, Study, or No Provider; (refer to the Review Station DICOM
3.0 Conformance Statement).
3. The following field must be completed if the remote host device is an Archival device:
- STORAGE COMMITMENT: Set to
Yes if the the Archival device DICOM 3.0 Conformance State-
men lists this feature; set to
No if the Statement does not list the feature.
• If STORAGE COMMITMENT is set to
No; set MESSAGE SEND to Yes if you want an S to appear
in the Report Status column of the Study or Exam list when a study has been sent to the Archival
device.
• If STORAGE COMMITMENT is set to
Yes; the fields Application Entity Title, Network Address,
and
Port Number in the lower part of the window are highlighted. You must complete these fields,
even if the information is the same as that already entered (this is necessary to allow the possibil-
ity of images being sent to one remote host and committed to another).
4. On completion, click the
Save button to update and close the window.
Note:
Network Remote Hosts are not updated in the
/etc/inet/hosts file, but are found in the file /export/
home/sdc/Prefs/SdCRHosts
. For this reason, it is not possible to use ping, telnet, spray and sim-
ilar commands using the hostname. The internet address must be used instead. For example, to
ping the MR operator's console from the workstation, do not use /usr/sbin/ping MR01_OC0. In-
stead, use the internet address, e.g., /usr/sbin/ping 192.9.200.1.
Therefore, though it is not mandatory to declare hosts in the /etc/hosts file when using the SdC
NET protocol, we recommend you to do it in any case.
5-2-1. Declaring a Dicom IC (independent console)
A new protocol has been implemented in order to solve some problems with the IC current software
Release 5.5 (echo, delay, repetition and inversion times display being multiplied by 1000).
Declare the host as any other Dicom3 host, but select the IC Dicom3.0 protocol.

Chapter 3 Page no. 124 CFG006.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 006 - Declare Network Hosts This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
CFG008.FM Page no. 125 Chapter 3
Job Card CFG 008 - Networking Connection and Configuration Chapter 3
1. SUPPLIES
• Senographe 2000 D Networking kit (M80061NK) containing six 60 m Ethernet cables and a Network
Hub.
2. TOOLS
• None.
3. SAFETY PRECAUTIONS
• None.
4. PREREQUISITES
• None.
5. PROCEDURE
5-1. Introduction
• The Senographe 2000 D must be configured in a local network with associated devices, e.g., Review
Workstation, Archive devices, Printers, and/or other Senographes. A typical installation is shown in
the diagram below.
• The networking kit containing six cables and a network hub is provided with the system for this pur-
pose.
• For HIS/RIS connection, a special optional HIS/RIS networking kit is required. Contact your GE HNS
representative for details.
• For connection to a hospital network, a special study is required. Contact your GE HNS representa-
tive for details.
Hub
Mass
RWS
AWS Cart
ReproRIS
Archiver
60 m (197 ft)
60 m (197 ft)
60 m (197 ft)
60 m (197 ft)
3 m (10 ft)
a.c. power (16 A)
3 m (10 ft)
DMPU
(CAD)

Chapter 3 Page no. 126 CFG008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
5-2. Ethernet Connection between AWS and Local Network
1. Route and connect the Ethernet cable from the network hub (provided in the Senographe 2000 D
networking kit) to the Ethernet network socket (C) on the AWS Cart patch panel. See illustration
below.
.
2. Connect the other local network devices to the hub. The hub connector positions used are not impor-
tant (except for special-purpose connector 8X, which must be left empty). However, you are recom-
mended to label the cables at the hub, to aid future trouble-shooting.
5-3. IP Addresses
This section explains how to change the Senographe 2000 D Hostname, IP address, subnet mask, and
other parameters
Each Senographe 2000 D system is given a default IP address during the factory check process. To
avoid conflicts (such as IP packet colliision) inside the customer network, this Senographe 2000 D IP
address should normally be changed at installation time to the recommended address given below.
For the RWS, printer(s), archiving device, etc., refer to their respective documentation.
Notice:!
If the system is connected to an external network (HIS/RIS, hospital network, etc.) the standard
recommended IP addresses may not be applicable. Suitable addresses will be defined by the GE
HNS consultant and the hospital network administrator.
Standard recommended IP addresses:
Senographe 2000 D: 192.1.1.100
RWS 192.1.1.110
Printer 192.1.1.120
Archiver 192.1.1.130
DMPU (CAD) 192.1.1.140
If more than one S2000 D is connected to the network, use adresses 192.1.1.101, 192.1.1.102, etc. Sim-
ilarly, if other devices are duplicated, add 1 to the recommended base address for each additional device.
Subnet mask:
ALL devices: 255.255.255.0
C

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
CFG008.FM Page no. 127 Chapter 3
6. NETWORK PARAMETER CONFIGURATION
6-1. Enter configuration mode
1. Open a command window.
2. Log in as
root on the AWS:
At the prompt type: su -root<Return>
At the password prompt type: operator <Return>
(factory loaded default root password)
3. Type the following command:
> sys-unconfig<Return>
A warning mesage is displayed; answer Yes to continue.
4. The AWS shuts down. Wait until AWS power has been switched off by the UPS (AWS shut down,
power light on monitor goes out), then press the UPS ON switch on the Generator Cabinet to power
up the AWS.
5. The AWS starts up in configuration mode.
6. When asked to select the system language, select
English (type 0<Return>). This is mandatory.
7. When asked to select the system character set, select
USA ASCII (type 0<Return>). This is manda-
tory.
8. The system now displays a succession of screens in which you must enter the network parameters,
starting with the Hostname.
Enter parameters as outlined in the following sections. Where necessary, use the arrow keys to move
the cursor, and use the space bar to select an item.
6-2. Hostname
• Enter the Hostname to be used for the workstation; press <Return> when done..
Note:
The Hostname will automatically be used as the Dicom A.E. Title for the workstation. If there is
more than one Senographe 2000 D system on the hospital site, the names chosen must clearly
differentiate between them (for example, by including the unit number or name).
Do not use the following characters in the Hostname:
! $ ” ‘ ’ { } [ ] * @ # ? ^ _ / etc.
ONLY USE ALPHANUMERIC CHARACTERS WITHOUT SPACES.
It is also possible to use the characters - (dash) and . (dot).
• Press the <F2> key to continue with the Network Connectivity screen
- Hostname - - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
A host name must be .............................
.................................................
Hostname :

Chapter 3 Page no. 128 CFG008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration6-3. Network Connectivity
WARNING
MAKE SURE that Yes (networked) is selected even though you do not intend to connect
the workstation immediately to a network.
• Press the <F2> key to continue.
6-4. Primary Network Interface
• Make sure that hme1 is selected. Press the <Return> key when done.
• Press the <F2> key to continue.
- Network Connectivity - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
> To make a selection .............................
.................................................
Networked :
X Yes
No
- Primary Network Interface - - - -- - - - - - - - - - - - - - - -
On this screen .................................
Primary Network Interface
hme0
hme1X

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
CFG008.FM Page no. 129 Chapter 3
6-5. IP Address
• Enter the Network IP address to be used for the workstation. Press the <Return> key when done.
• Press the <F2> key to continue.
6-6. Confirmation Screen
• Check that the correct information is displayed..
• Press the <F2> key to continue.
• A screen indicating a delay may be displayed:
- IP Address - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
IP addresses must .............................
.................................................
IP address :
- Confirm information - - - - - - - - - - - - - - - - - - - - - -
> Confirm the following ...........................
.................................................
Host name : XXXXX
Networked : Yes
Primary Network Interface : hme1
IP address : X.X.X.X
Just a moment

Chapter 3 Page no. 130 CFG008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration• Name Service.
• Use the Down arrow key to move to
None then press <Return> to select
WARNING
DO NOT SELECT NIS+ or NIS Client services. None is the primary selection
• Press the <F2> key to continue
6-7. Confirm Name Service screen
• Check that the correct information is displayed..
• Press the <F2> key to continue.
- Name service - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
> To make a selection .............................
.................................................
Name Service :
X
NIS +
NIS (formerly yp)
Other
None
- Confirm information - - - - - - - - - - - - - - - - - - - - - -
> Confirm the following ...........................
.................................................
Name service : None

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
CFG008.FM Page no. 131 Chapter 3
6-8. Subnets (NetMask) Information.
• Select the appropriate response Yes or No. The normal response is No.
Notice:!
Select
Yes only when your Network Administrator explicitly requires a tree-like network
with netmasks. In this case only, the screen below is displayed. The Netmask value will be
given by the Network Administator. Erase the displayed default Netmask by back-spacing
over it and type in the new address..
• Press the <F2> key to continue.
- Subnets - - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
> To make a selection .............................
.................................................
System part of a subnet :
X
Yes
No
- Netmask - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
Netmask : 255.255.255.0

Chapter 3 Page no. 132 CFG008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
6-9. Time Zone (Regions)
• Use the Up or Down arrow keys to select the Other region, then press <Return> to select.
• Press the <F2> key to continue.
6-10. Time Zone Selection
• Use the Up or Down arrow keys to select the GMT Time Zone, then press <Return> to select.
•Select
Offset = 0.
• Press the <F2> key to continue.
- Time Zone - - - - - - - - - - - - - - - - - - - - - - - - - - -
On this screen .................................
> To make a selection .............................
.................................................
Regions :
X
Africa
Asia, Eastern
United States
- - -
Other
- Time Zone - - - - - - - - - - - - - - - - - - - - - - - - - - -
> To match a selection .............................
.................................................
Time Zone :
X
- - - - - - -
GMT
- - - - - - -
- - - - - - -
- - - - - - -

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration
CFG008.FM Page no. 133 Chapter 3
6-11. Date and Time
• Use the Up or Down arrow keys to move to the field to be modified (as applicable), type in the correct
value then press <Return> to move to the next field.
• Press the <F2> key to continue
6-12. Confirmation Screen
• Check that the correct information is displayed..
• Press the <F2> key to continue.
PRESS ON THE <F2> key to continue.
- Date and Time - - - - - - - - - - - - - - - - - - - - - - - - -
> Accept the default .............................
.................................................
Date and time : mm/dd/yy hh:mm
Year (4 digits) : 1995
Month (1-12) : XX
Day (1-31) : XX
Hour (0-23) : XX
Minute (0-59) : XX
- Confirm information - - - - - - - - - - - - - - - - - - - - - -
> Confirm the following ...........................
.................................................
System part of a subnet : yes / no
Netmask : XXX.XXX.XXX.XXX
Time Zone : XXX
Date and time : mm/dd/yy hh:mm

Chapter 3 Page no. 134 CFG008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 008 - Networking Connection and Configuration6-13. Root Password Screen
• This screen allows you to change the Root password for the workstation. If you want to change it,
type in the new password and press the <Return> key when done. The password you type in will not
be displayed. The standard password for root is
operator.
Notice:!
Make sure that the caps-lock key is not engaged before entering the password.
• If you change the password, the system asks you to re-enter the new password. If there is a differ-
ence between the two entries, it will request it again.
• When the password has been correctly changed, or if you press <Return> without changing the
password, the system reboots.
6-14. Login
• The configuration process is now complete.
• On completion of the boot sequence, the prompt
login: is displayed.
• Login with the user name sdc, and the password adw3.1, to restart the system.
On this screen, you can create a root password .....
A root password .................................
> If you do not want ...........................
.................................................
Root password :
Press Return to continue.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 009 - Declare Printers (Filming Devices)
CFG009.FM Page no. 135 Chapter 3
Job Card CFG 009 - Declare Printers (Filming Devices) Chapter 3
1. SUPPLIES
None
2. TOOLS REQUIRED
None.
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
The AWS must have been correctly installed and configured.
The local Network must have been correctly installed, and the Network Printers to be declared must have
been correctly installed and connected to the Network.
Make sure the following information is available before starting to install your printer:
• Internet address and Hostname of Network printer(s). This information can be obtained either from
the Network Administrator of the Hospital, or from the Needs assesment Form completed by the Net-
work Champions.
• Application Entity Title (A.E.T.) of Network printer(s). This information can be obtained from the
Printer Supplier.
• Dicom Port number of Network printer(s). This information can be obtained from the Printer Supplier.
5. PROCEDURE
5-1. Test the network connection to the printer
Once a network printer is connected to the network and set with an IP address, it should respond to
TCP/IP messages from another Host on the network (e.g., the AWS). Before declaring a printer on the
AWS, check the following:
• Log in as root on the AWS:
console login: root <Return>
Password: operator <Return> (factory loaded default root password)
• Type in the following command:
/usr/sbin/ping <IP_address_of_network_printer> <Return>
e.g.: /usr/sbin/ping 192.100.9.5 <Return>
<IP address of Network printer> is alive
Note:
Do not enter leading zeros in any of the four numeric fields separated by dots of the IP address.
e.g., do not type in 192.100.009.005, but type 192.100.9.5 instead.

Chapter 3 Page no. 136 CFG009.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 009 - Declare Printers (Filming Devices)The "alive" answer from the system means that the Network printer can be reached from the Acquisition
Workstation system through the network.
If the system displays "unknown host", your printer may not be correctly declared on the network.
If the system displays "Network unreachable", you probably have given the printer an IP address incom-
patible with your Review Workstation's IP address. Check that they belong to the same network,
192.100.9.X, for example.
• Log out
5-2. Declare the Dicom printer(s) to the AWS
• Click on the Tools Menu icon in the Browser to open a Menu displaying available applications.
• Click on
Printer Management on the just displayed menu to display the Dicom Printer Selection win-
dow:.
• Click the
Add button to display the Dicom Printer parameters window (Illustration 1):
Complete the different fields according to the information given by the Dicom Printer manufacturer .
•Click on
Clear to clear all fields, Save to store the entered information, and Done to exit and return to
the
Printer Selection window.
Refer to section 6 for information on entries for Kodak 8600 and Agfa LR5200 printers.
Dicom Printer Selection
List of Registered Dicom Printers
Your choice :
Update Add Remove Cancel

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 009 - Declare Printers (Filming Devices)
CFG009.FM Page no. 137 Chapter 3
ILLUSTRATION 1 - THE DICOM PRINTER PARAMETERS WINDOW
DICOM printer parameters
DICOM printer label :
Hostname :
Application Entity Title (AET) :
Network address :
Port number :
Layouts :
Slide formats : 35mm 40mm
Pixel depth : 10 12
Film size :
8 in x 10 in W H
10 in x 12 in W H
10 in x 14 in W H
11 in x 14 in W H
14 in x 14 in W H
14 in x 17 in W H
24 cm x 24 cm W H
24 cm x 30 cm W H
Printer pixel size (micron) :
Configuration information :
Density : Min Max
Magnification type : Cubic
Smoothing factor :
Trim : No
Save Clear Done

Chapter 3 Page no. 138 CFG009.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 009 - Declare Printers (Filming Devices)
6. KODAK 8600/8610 AND AGFA LR5200 PRINTERS
This section gives information on entries to be made in the Dicom Parameters window for Kodak 8600,
Kodak 8610, and Agfa LR5200 printers. These are currently the only printers qualified for use with the
Senographe 2000 D system.
6-1. Kodak DryView 8600 and 8610 Dicom printers
Refer to Illustration 2.
• DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g., Kodak 8610.
• Hostname: Use printer Hostname if any, or choose a name unique to your Network,e.g.,
kodak1printer.
• Network address: A unique IP address given by the Network Administrator of the hospital, e.g.,
3.48.16.135.
• Application Entity Title (AET):
Must be IMN_9410
• Port number:
Must be 1024
• Layouts:
Must be 1 x 1, 1 x 2*, 2 x 1*, 2 x 2*
•Film size:
Must be 8 in x 10 in W = 5192* H = 6201*
•Pixel size 38.75*
• Configuration information
Must be LUT=0,3
• Magnification type:
Must be Cubic
•Trim:
Must be NO
6-2. Agfa LR5200 Dicom printer
Refer to Illustration 3.
• DICOM printer label: Any name you wish to see appearing on the Film Composer e.g., Agfa LR5200.
• Hostname: Use printer Hostname if any, or choose a name unique to your Network e.g., agfa1LR.
• Network address: A unique IP address given by the Network Administrator of the hospital, e.g.,
3.48.16.136.
• Application Entity Title (AET):
To be specified by Agfa Field Engineer
• Port number:
Must be 104
• Layouts:
Must be 1 x 1, 1 x 2*, 2 x 1*, 2 x 2*
•Film size:
Must be 8 in x 10 in W = 4776* H = 5944*
•Pixel size 40*
• Configuration information:
Must be PERCEPTION_LUT=150
• Magnification type:
Must be Cubic
• Smoothing factor:
Must be 140
•Trim:
Must be NO
Notice:!
* The values marked with an asterisk can be left blank for normal printing, but must be set as spec-
ified for successful use with the
ScrapBook function.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 009 - Declare Printers (Filming Devices)
CFG009.FM Page no. 139 Chapter 3
ILLUSTRATION 2 - TYPICAL KODAK DRYVIEW 8600/8610 ENTRIES
DICOM printer parameters
DICOM printer label :
Hostname :
Application Entity Title (AET) :
Network address :
Port number :
Layouts :
Slide formats : 35mm 40mm
Pixel depth : 10 12
Film size :
8 in x 10 in W H
10 in x 12 in W H
10 in x 14 in W H
11 in x 14 in W H
14 in x 14 in W H
14 in x 17 in W H
24 cm x 24 cm W H
24 cm x 30 cm W H
Printer pixel size (micron) :
Configuration information :
Density : Min Max
Magnification type : Cubic
Smoothing factor :
Trim : No
Save Clear Done
IMN_9410
1024
X
X
CHECK HERE
IMN_9410
1024
X
LUT=0,3

Chapter 3 Page no. 140 CFG009.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CFG 009 - Declare Printers (Filming Devices)
ILLUSTRATION 3 - TYPICAL AGFA LR5200 ENTRIES
DICOM printer parameters
DICOM printer label :
Hostname :
Application Entity Title (AET) :
Network address :
Port number :
Layouts :
Slide formats : 35mm 40mm
Pixel depth : 10 12
Film size :
8 in x 10 in W H
10 in x 12 in W H
10 in x 14 in W H
11 in x 14 in W H
14 in x 14 in W H
14 in x 17 in W H
24 cm x 24 cm W H
24 cm x 30 cm W H
Printer pixel size (micron) :
Configuration information :
Density : Min Max
Magnification type : Cubic
Smoothing factor :
Trim : No
Save Clear Done
X
104
X
CHECK HERE
TO BE SPECIFIED BY AGFA FIELD ENGINEER
PERCEPTION_LUT=150
140

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
CALIB.FM Page no. 141
CHAPTER 4 CALIBRATION
This chapter contains information for calibration of Senographe 2000 D functions.

Page no. 142 CALIB.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 000 - Calibration Interactions
Cal000.fm Page no. 143 Chapter 4
Job Card CAL 000 - Calibration Interactions Chapter 4
1. CALIBRATION INTERACTIONS
• Many calibration procedures function correctly only if other related calibrations have already been
successfully completed. These are specified as
Prerequisite Calibrations in the table below.
• Settings made during calibration procedures may interact with others. It may therefore be necessary
to carry out other procedures (calibrations or checks) after completing a calibration procedure. These
are specified as
Post-Completion Requirements in the table below.
Calibration Prerequisite Calibrations Post-Completion Requirements
CAL 001 Heater Scale Factor None None
CAL 002 X-ray Bias Scale Factor CAL 001 Heater Scale Factor None
CAL 003 Tube heater Current CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
None
CAL 004 Tube mA Measurement None None
CAL 005 kV Scale Factor CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 004 Tube mA Measurement
CAL 003 Tube heater Current
CAL 006 Comp. Force Detector None None
CAL 007 Breast Thickness CAL 006 Comp. Force Detector None
CAL 008 Minimum Force CAL 006 Comp. Force Detector None
CAL 009 Compression Arm Angle None None
CAL 010 Tube Tilt CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 011 Beam Alignment
CAL 012 Collimator Format
CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 015 Conversion Factor
CAL 017 AOP
CAL 020 mAs Non-linearity
CAL 011 Beam Alignment CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 010 Tube Tilt
VF 008 Flat Field Test

Chapter 4 Page no. 144 Cal000.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 000 - Calibration Interactions
CAL 012 Collimator Format CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 010 Tube Tilt
CAL 011 Beam Alignment
VF 008 Flat Field Test
CAL 013 Bad Pixel CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 014 Detector Gain
CAL 015 Conversion Factor
CAL 017 AOP
CAL 020 mAs Non-linearity
VF 008 Flat Field Test
CAL 014 Detector Gain CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 013 Bad Pixel
If Rh/Rh/LF/nogrid configuration
has been calibrated, do:
CAL 015 Conversion Factor
CAL 017 AOP
For all configurations, do:
VF 008 Flat Field Test
CAL 015 Conversion Factor CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 012 Collimator Format
CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 017 AOP
VF 010 Acquisition in AOP Mode
CAL 016 Half-Value Layer CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 012 Collimator Format
None
Calibration Prerequisite Calibrations Post-Completion Requirements

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 000 - Calibration Interactions
Cal000.fm Page no. 145 Chapter 4
CAL 017 AOP CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 006 Comp. Force Detector
CAL 007 Breast Thickness
CAL 009 Compression Arm Angle
CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 012 Collimator Format
CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 015 Conversion Factor
VF 010 Acquisition in AOP Mode
CAL 018 ISC Monitor None None
CAL 019 Light Beam CAL 011 Beam Alignment
CAL 012 Collimator Format
None
CAL 020 mAs Non-linearity CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale Factor
CAL 003 Tube heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 013 Bad Pixel
CAL 014 Detector Gain
None
Calibration Prerequisite Calibrations Post-Completion Requirements

Chapter 4 Page no. 146 Cal000.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 000 - Calibration Interactions This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 001 - Heater Current Scale Factor.
CAL001.FM Page no. 147 Chapter 4
Job Card CAL 001 - Heater Current Scale Factor. Chapter 4
1. SUPPLIES
None
2. TOOLS
• True R.M.S. AC ammeter, e.g.: Fluke 8060, Fluke 87, Fluke 45. HP 3457A, Tektronix DMM 254.
• HV unit/cable adaptor, Part number 36003399.
Notice:!
IMPORTANT NOTE: To ensure accurate measurement:
• For ALL Digital Multimeters: The Autorange function MUST NOT be used, as it can lead to severe
measurement errors (>40% !!).
• For ALL Digital Multimeters: The Relative Measurement function MUST NOT be used to reset to zero
the natural offset displayed when there is no current measured. The offset is automatically reset to
zero by the principle of the rms measurement.
• Measurement configurations to be used are as follows:
- Fluke 8060: used with 20 A shunt in the 200 mV AC range (shunt reference: 80J10).
- Fluke 87: no shunt, in the 10 A AC range.
3. SAFETY PRECAUTIONS
CAUTION
In case of equipment breakdown or service engineer error during this calibration proce-
dure, the resulting parameter values might become so unreasonable as to render the X-ray
tube filament supply circuit unusable. This can show up as an error message during power
up, and thus inhibit use of the calibration software. If this happens, the only way to cancel
the error is to manually re-enter the default parameter values (see section 10 for default
parameter values) and switch the Senographe DMR off and back on again before again
attempting a calibration.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF HEATER CURRENT SCALE FACTOR
The objective is to determine the scale factor K and fixed offset VOF between the X-ray tube filament cur-
rent command from the software and the real filament current value obtained. Each tube track has its
own filament supply circuit, so this calibration procedure is to be performed for each of the two tracks.
DEFAULT PARAMETER VALUES ARE K = +9.450E+0 (9.45) AND VOF = +0.000E+0 (0.0).
5-1. Preparation
1. Turn the Senographe DMR OFF.
2. Open the HV converter control signal inputs by disconnecting flat cable WG8 from generator com-
mand board 400-PL1 connector XJ7. This eliminates any possibility of energy build-up in the HV unit

Chapter 4 Page no. 148 CAL001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 001 - Heater Current Scale Factor. during this calibration procedure.
3. Disconnect the DMR HV cable at the HV tank end.
4. Connect the HV cable to the HV unit HV cable adaptor (36003399).
5. Connect the HV unit HV cable adaptor to the DMR HV tank.
6. Connect the true r.m.s. AC ammeter in series with the filament being calibrated on the HV unit/cable
adaptor; and place a jumper on the other track (see illustrations 1 and 2).
Note:
1. See “important note” in section 2 to ensure an accurate measurement.
2. Keep the other filament connected in order to avoid error messages when the Senographe DMR
is switched on.
ILLUSTRATION 1 - HV UNIT HV CABLE ADAPTOR 36003399
Note:
For safety reasons, -HT is permanently connected to ground
ILLUSTRATION 2 - HV CABLE CONNECTOR PIN CONNECTION DIAGRAM
The HV cable connector is marked with the standard X-ray industry markings (L, S, C, and G). However,
the functions of the corresponding conductors in the Senographe DMR are as follows:
L = track 1 (Mo) filament supply
S = track 2 (Rh) filament supply
C = filament supply return and X-ray tube cathode HV
G = focal bias voltage supply
7. Switch on the Senographe DMR.
L
C
S
G
S
G
L
C
Track 2 (Rh): jumper or ammeter
Bias measurement
Ground (+HV)
BIAS
TRACK 1 (Mo)
TRACK 2 (Rh)
-HT
Ground (+HV)
BIAS
TRACK 1 (Mo)
TRACK 2 (Rh)
-HT
From HV tank To X Ray tube
HV cable
shield
Track 1 (Mo): jumper or ammeter
CG
L
C
S
G
POLARIZING TAB
NOTE: THIS IS A HEAD-ON VIEW
OF THE HV CABLE CONNECTOR
AFTER BEING DISCONNECTED
FROM THE HV UNIT.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 001 - Heater Current Scale Factor.
CAL001.FM Page no. 149 Chapter 4
Note:
ERROR 007/024 is normal with 400 PL1 XJ7 disconnected.
8. Change the position of the installation menu enable switch (switch B1 on generator CPU board 400-
PL3, see
Accessing the Different Installation Menus in the Tree Structure).
5-2. Perform the calibration procedure for the desired X-ray tube track.
1. Select SETUP/GENE/INSTAL/GENE/GENE/HTR_SCL/TRACK1 ( or TRACK2) /CALIB/1st pt/
CALIB
on the console for the first calibration point.
Note:
The default filament current command value for this first point is +4.000E+00 (4.0) A.
2. Press the 2nd trigger button (no X-rays are produced) and hold it until the ammeter reading stabilizes
(this may take a few seconds).
3. Note the filament current value measured from the ammeter.
4. Release the 2nd trigger button.
5. Select SETUP/MeasI on the console.
6. Enter and confirm the measured filament current value from the ammeter for this first calibration
point.
7. Select SETUP/SETUP/2nd pt/CALIB on the console for the second calibration point.
Note:
The default filament current command value for this second point is +5.000E+00 (5.0) A.
8. Press the 2nd trigger button (no X-rays are produced); hold it until the ammeter reading stabilizes
(this may take a few seconds).
9. Note the filament current value measured by the ammeter.
10. Release the 2nd trigger button.
11. Select SETUP/MeasI on the console.
12. Enter and confirm the measured filament current value from the ammeter for this second calibration
point.
13. Select SETUP/SETUP/calcul/VALID on the console. This executes the calculation of the scale factor
and fixed offset of the filament current command from the software.
5-3. Check the calibration accuracy.
1. Select SETUP/1st pt/CALIB on the console.
Note:
Again, the default filament current command value for this first point is +4.000E+00 (4.0) A.
2. Press the 2nd trigger button (no X-rays are produced) and hold it until the ammeter reading stabilizes
(this may take a few seconds).
3. Note the filament current value measured from the ammeter.
4. Release the 2nd trigger button.
5. Select SETUP/SETUP/2nd pt/CALIB on the console.
Note:
Again, the default filament current command value for this second point is +5.000E+00 (5.0) A.
6. Press the 2nd trigger button (no X-rays are produced) and hold it until the ammeter reading stabilizes
(this may take a few seconds).
7. Note the filament current value measured from the ammeter.

Chapter 4 Page no. 150 CAL001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 001 - Heater Current Scale Factor. 8. Release the 2nd trigger button.
9. The expected results are 4.0 ±0.02 A for the first point and 5.0 ±0.02 A for the second point. If the
results are within these tolerances, the calibration for this X-ray tube track is finished. If the results are
out of tolerance, repeat the calibration procedure in section 5-2 and the check procedure in section 5-
3 for the same X-ray tube track.
Note:
Performing the calibration procedure once is normally sufficient for correct calibration.
If it is necessary to perform the calibration procedure more than once, the measured values of cur-
rent for the two points (4.0 and 5.0 A) should become increasingly accurate. If not, the procedure
is probably not being followed correctly.
10. Select SETUP/SETUP/SETUP/PARAM on the console. Note down the calculated values of scale
factor K and offset VOF. The two values can be displayed alternately by rotating the console kV dial.
CAUTION
If the value of K is less than +8.000E+0 (8.0), either the calibration procedure was done
incorrectly or there is a hardware defect.
11. Repeat the calibration and check procedures on the other tube track, if necessary.
12. Perform a ”CKSUM”, then turn the system off.
13. Reconnect flat cable WG8 to generator command board 400-PL1 connector XJ7; Remove the high
voltage cable adapter. Remember to secure the HV ring Allen screw, to insure proper ground continu-
ity.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 002 - X-Ray Tube Focal Bias Voltage Scale Factor
CAL002.FM Page no. 151 Chapter 4
Job Card CAL 002 - X-Ray Tube Focal Bias Voltage Scale FactorChapter 4
1. SUPPLIES
None
2. TOOLS
• Voltmeter
• HV unit / cable adaptor, part number 36003399
3. SAFETY PRECAUTIONS
CAUTION
In case of equipment breakdown or service engineer error during this calibration proce-
dure, the resulting parameter values might become so unreasonable as to render the X-ray
tube focal bias voltage supply circuit unusable. This can show up as an error message
during power up, and thus inhibit use of the calibration software. If this happens, the only
way to cancel the error is to manually re-enter the default parameter values (see section 5
for these values) and switch the DMR off and back on again before attempting to repeat the
calibration.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF X-RAY TUBE BIAS VOLTAGE SCALE FACTOR
The objective is to determine the scale factor alpha and offset beta between the X-ray tube focal bias
voltage command from the software and the real focal bias voltage value obtained. The focal bias voltage
applied to the X-ray tube electron beam concentration device establishes the focal spot width.
DEFAULT PARAMETER VALUES ARE ALPHA = +9.389E+1 (93.89) AND BETA = +3.801E-2 (0.03801).

Chapter 4 Page no. 152 CAL002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 002 - X-Ray Tube Focal Bias Voltage Scale Factor
5-1. Preparation
1. Turn the Senographe DMR OFF.
2. Open the HV converter control signal inputs by disconnecting flat cable WG8 from generator com-
mand board 400-PL1 connector XJ7. This eliminates any possibility of energy build-up in the HV unit
during this calibration procedure.
3. Disconnect the DMR HV cable at the HV tank end.
4. Connect the HV cable to the HV unit / cable adaptor (36003399).
5. Connect the HV unit / cable adaptor to the DMR HV tank.
6. Connect a DC voltmeter between common (C) and the focal bias voltage wire (G) (see illustrations 1
and 2).
ILLUSTRATION 1 - HV UNIT HV CABLE ADAPTOR 36003399
Note:
For safety reasons, -HT is permanently connected to ground
ILLUSTRATION 2 - HV CABLE CONNECTOR PIN CONNECTION DIAGRAM
The HV cable connector is marked with the standard X-ray industry markings (L, S, C and G). However,
the functions of the corresponding conductors in the Senographe DMR are as follows:
L = track 1 (Mo) filament supply S = track 2 (Rh) filament supply
C = filament supply return and X-ray tube cathode HVG = focal bias voltage supply
L
C
S
G
S
G
L
C
Bias measurement
Ground (+HV)
BIAS
TRACK 1 (Mo)
TRACK 2 (Rh)
-HT
Ground (+HV)
BIAS
TRACK 1 (Mo)
TRACK 2 (Rh)
-HT
From HV tank To X Ray tube
HV cable
shield
Track 2 (Rh): jumper or ammeter
Track 1 (Mo): jumper or ammeter
CG
L
C
S
G
POLARIZING TAB
NOTE: THIS IS A HEAD-ON VIEW
OF THE HV CABLE CONNECTOR
AFTER BEING DISCONNECTED
FROM THE HV UNIT.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 002 - X-Ray Tube Focal Bias Voltage Scale Factor
CAL002.FM Page no. 153 Chapter 4
5-2. Check focal bias command values
1. Switch on the Senographe DMR.
ERROR 007/024 is normal with 400 PL1 XJ7 disconnected.
2. Change the position of the installation menu enable switch; switch B1 on generator CPU board 400-
PL3, see Accessing the Different Installation Menus in the Tree Structure.
3. Select
setup/gene/INSTAL/gene/gene/BIAS/voltage on the console.
4. Rotate the kV dial in order to check that all four focal bias voltage command values are the same as
the values specified on the label on the X–ray tube housing:
MO/LF =
(typical value is +70v)
RH/LF = (typical value is +70v)
MO/SF = (typical value is +250v)
RH/SF = (typical value is +250v)
Modify any values that appear different from these on the tube label by using the CHANGE key, rotat-
ing the kV dial and using the NEXT, VALID and SETUP keys. See ”Visual Display of Parameter Val-
ues” and ”Modification of a Parameter Value or Entry of a Measurement” in chapter xxxxx.
Note:
The bias voltages are repeated on a label inside the generator, above the CPU board.
5-3. Calibration procedure
1. Open the work list and declare a new Patient called "X-ray Tube BIAS cal",
2. First calibration point:
a. Select SETUP/CALIB/1st pt/CALIB
on the console; press the 2nd trigger button (no X-rays are
produced) and hold it until the voltmeter reading stabilizes.
b. Note the focal bias voltage value measured from the voltmeter; release the 2nd trigger button.
c. Select SETUP/MEASV on the console.
d. Enter and confirm the measured focal bias voltage value (see
Entry of an Alpha-Numeric Value in
the Configuration Job Card
The Control Console for information on entering alpha-numeric val-
ues).
3. Second calibration point:
a. Select
setup/setup/2nd pt/calib on the console; press the 2nd trigger button (no X-rays are pro-
duced) and hold it until the voltmeter reading stabilizes.
b. Note the focal bias voltage value measured from the voltmeter; release the 2nd trigger button.
c. Select
setup/MeasV on the console.
d. Enter and confirm the measured focal bias voltage value.
4. Select
setup/setup/calcul/valid on the console. This calculates of the parameters alpha and beta.
The message CALCUL DONE appears on the console display.

Chapter 4 Page no. 154 CAL002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 002 - X-Ray Tube Focal Bias Voltage Scale Factor
5-4. Calibration accuracy check.
1. First calibration point:
a. Select
setup/1st pt/calib on the console; press the 2nd trigger button (no X-rays are produced)
and hold it until the voltmeter reading stabilizes.
b. Note the measured focal bias voltage value; release the 2nd trigger button. DO NOT ENTER
THIS MEASURED VALUE.
c. The expected result is the value of MO/LF ±5% (see above).
2. Second calibration point:
a. Select
setup/setup/2nd pt/CALIB on the console; press the 2nd trigger button (no X-rays are
produced) and hold it until the voltmeter reading stabilizes.
b. Note the measured focal bias voltage value; release the 2nd trigger button. DO NOT ENTER
THIS MEASURED VALUE.
c. The expected result is the value of MO/SF ±5% (see above).
3. If the results are within the tolerances given, the focal bias voltage calibration is finished. If the results
are out of tolerance, repeat the calibration and check procedures.
Note:
Performing the calibration procedure once is normally sufficient for correct calibration.
If it is necessary to perform the calibration procedure more than once, the measured values of focal
bias voltage for the two points should become increasingly accurate. If not, check that you are fol-
lowing the procedure correctly.
4. Select
setup/setup/setup/param on the console. Note down the calculated values of parameters
alpha and beta. Each of these two values can be displayed alternately by rotating the kV dial on the
console,
5. Delete the previously created patient called "X-ray Tube BIAS cal".
6. Perform a ”CKSUM”, then turn the system OFF.
7. Re-connect flat cable WG8 to generator command board 400-PL1 connector XJ7.
8. Remove the high voltage cable adapter. Remember to secure the HV ring correctly using the Allen
screw, to insure proper ground continuity.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 003 - X-Ray Tube Heater Current
CAL003.FM Page no. 155 Chapter 4
Job Card CAL 003 - X-Ray Tube Heater Current Chapter 4
1. SUPPLIES
None
2. TOOLS
Detector protection steel cover.
3. SAFETY PRECAUTIONSWARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF X-RAY TUBE HEATER CURRENT
The objective is to determine the X-ray tube heater current required to produce the desired X-ray tube
anode current at a given kV value.
The calibration procedure must be performed on each of the four focal size/track combinations. Four rep-
etitions of the calibration procedure are therefore necessary.
Three exposures are taken automatically at different kV values. Using the results, the software calculates
the necessary parameters to determine the X-ray tube heater current needed to produce the desired X-
ray tube anode current for all possible kV values.
5-1. Preparation
• Place a lead cover on the top of the image receptor to protect the detector panel from excessive X-
ray doses.
• Enter installation mode on the console (change the position of installation menu enable switch B1 on
Generator CPU board 400-PL3 if the DMR has been switched off since the last access to install
menus,)
5-2. Calibration procedure.
1. Open the work list and declare a new Patient called "X-ray Tube Heater current",
2. Select: SETUP/GENE/INSTAL/GENE/GENE/TUBE/HEATER/
focal size/track/CALIB.
Perform the calibration procedure first on the LF/MO focal size/track combination, then on the others
in any order (SF/MO, LF/RH, SF/RH).

Chapter 4 Page no. 156 CAL003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 003 - X-Ray Tube Heater Current
CAUTION
Before proceeding to make exposures, be sure that the detector panel has been protected
from excessive X–ray doses as instructed (by placing a lead cover on the top of the image
receptor).
3. Press the 2nd trigger button and hold it down. The messages EXPOSURE NO 1, EXPOSURE NO 2,
EXPOSURE NO 3, and CALIBRATION END appear in sequence on the console display. The soft-
ware automatically performs the three exposures and the calculations.
Note:
1.Calibration time varies according to tube temperature. The message TOO WARM appears on
the console display when tube temperature is too high for taking the next exposure. It is not nec-
essary to hold the 2nd trigger button down while waiting for the tube to cool. However, do not exit
the calibration menu while waiting. As soon as the TOO WARM message disappears, continue the
calibration by pressing and holding the 2nd trigger button.
2. In case of arcing, the software re-attempts the same exposure automatically.
3. The calibration procedure includes an exposure at 40 kV. If for some reason this exposure is
impossible to perform (e.g. excessive arcing), tube warm-up may be necessary.
4. Repeat steps 2 and 3 for each of the remaining focal size/track combinations.
5. Carry out the following steps only if the warning message
Limited Ich (limited heater current) is dis-
played on the console at the end of the tube current calibration. This message indicates that one or
more filament heating current values were calculated above 5.6 A (but internally limited to 5.600 A
when stored in memory); the message is only a warning and is not associated with an error code.
Proceed as follows:
a. Access the DMR installation menu to read the heater parameters with the LF/MO combination
(the path is: SETUP/GENE/INSTAL/GENE/GENE/TUBE/HEATER/LF/MO/Param/I-Value ).
b. Explore the 28 kV/A combinations; note the kV values showing a current value equal to 5.600 A.
c. Repeat steps a and b for the LF/RH combination (there is a very low probability of current limita-
tion occurring in small focus).
d. Return to Application Mode (press SETUP).
e. Set the DMR for 2-point exposure technique at 60 mAs. Make an exposure for each kV point
detected at 5.600 A (LF/MO and LF/RH). Allow at least 60 seconds between each exposure.
f. Repeat steps a and b: If no more kV points are set to 5.600 A, the problem has been resolved. If
one or more points still show a current value of 5.600 A, repeat Calibration Job Card
Heater Cur-
rent Scale Factor
, then restart this procedure from step a above. If the problem recurs, replace the
X-ray tube

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 003 - X-Ray Tube Heater Current
CAL003.FM Page no. 157 Chapter 4
5-3. Check results.
1. Be sure that the gantry presence parameters PRES_A, PRES_IDC, and PRES_ADC are set to yes
before proceeding
.
2. Activate the display of parameters following an exposure (select SETUP/GENE/INSTAL/GENE/CON-
FIG/DISP/YES).
3. Perform a CKSUM and return to application mode.
4. Set up the DMR for a manual (2-point) exposure at 30 kV and 60 mAs. For each of the four focal size/
track combination, take an exposure following a 3–second preparation time. Note the resulting X-ray
tube anode currents displayed on the console display.
Note:
A lower mAs value would decrease the measurement accuracy.
5. Expected X-ray tube anode currents for each focal size/track combination are:
LF/MO - 100 mA ± 5% (± DVM accuracy)
SF/MO - 40 mA ± 5% (± DVM accuracy)
LF/RH - 75 mA ± 5% (± DVM accuracy)
SF/RH - 40 mA ± 5% (± DVM accuracy)
6. Delete the prevously created patient called "X-ray Tube Heater current" .
If you are handing the machine back to the user at this point, switch the DMR off, then on again.

Chapter 4 Page no. 158 CAL003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 003 - X-Ray Tube Heater Current This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 004 - X-Ray Tube mA Measurement
CAL004.FM Page no. 159 Chapter 4
Job Card CAL 004 - X-Ray Tube mA Measurement Chapter 4
1. SUPPLIES
None
2. TOOLS
• Milliammeter
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF X-RAY TUBE MA MEASUREMENT
The objective is to determine the scale factor F/mA between the X-ray tube mA measurement input to
the software and the real X-ray tube mA value.
The default parameter value is F/mA = +1.000E+3 (1000) Hz/mA
CAUTION
DO NOT touch potentiometer P1 on generator kV/mA board 701-PL1 at any point in this
procedure.
5-1. Preparation
1. Turn off the Senographe DMR.
2. Remove strap between terminal lugs
MES_M_NEG and MES_M_POS of generator kV/mA board
701-PL1.
3. Connect the milliammeter between test points
MES_M_NEG and MES_M_POS of generator kV/mA
board 701-PL1. Set the milliammeter to its 100 mA DC scale.
4. Turn on the Senographe DMR.
5. Change the position of the installation menu enable switch (switch B1 on generator CPU board 400-
PL3; see Accessing the Generator or Gantry Installation Mode from the Console in Chapter xxx
5-2. Perform the calibration procedure.
1. Starting from application mode, select SETUP/GENE/INSTAL/GENE/GENE/mA_meas/CALIB/CALIB
on the console. Press pushbutton SW2 on generator kV/mA board 701-PL1 and HOLD IT DOWN.
Read and note down the value measured on the milliammeter. While still holding SW2 down, select
MEASUR on the console (this request for a measurement of X-ray tube current is needed by the cal-
ibration software, but the value displayed on the console at this point is of no importance to the ser-
vice engineer). Release pushbutton SW2.
2. Select SETUP/I_MEAS on the console. Enter the value of current (in milliamperes) read from the mil-
liammeter by using the NEXT and VALID keys and rotating the kV dial. See Visual Display of Param-
eter Values and Modification of a Parameter Value or Entry of a Measurement in Job Card CFG "The

Chapter 4 Page no. 160 CAL004.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 004 - X-Ray Tube mA Measurement Control Console".
Note:
Only values between +4.000E+1 (40) and +6.000E+1 (60) mA are accepted by the calibration soft-
ware. An error message appears on the console display if a value outside of this range is entered.
3. Select SETUP/CALCUL/VALID on the console to execute the calculation. If the calculated F/mA
value is between 850 and 1150 mA
calib END appears on the console display - if it is not, then an
error message appears.
5-3. Check results.
1. Select SETUP/CALIB on the console. Press pushbutton SW2 on generator kV/mA board 701-PL1
and HOLD IT DOWN. Read and note down the value measured on the milliammeter. While still hold-
ing SW2 down, select MEASUR on the console. Verify that the measurement on the console display
now corresponds to the value read on the milliammeter ±400 µA ± milliammeter error. Release push-
button SW2. DO NOT ENTER THIS MEASURED VALUE.
2. Change the milliammeter scale to 200 mA DC. Press pushbutton SW1 on generator kV/mA board
701-PL1 and HOLD IT DOWN. Read and note down the value measured on the milliammeter (the
current should be approximately 120 mA). While still holding SW1 down, select MEASUR on the con-
sole. Verify that the difference between the measurement on the console display and the value read
on the milliammeter is less than 2.4 mA. Release pushbutton SW1. DO NOT ENTER THIS MEA-
SURED VALUE.
3. Select SETUP/SETUP/PARAM on the console and note down the displayed value of F/mA.
4. Perform a CKSUM and return to application mode.
5. Turn off the Senographe DMR.
6. Disconnect the milliammeter and replace the strap between terminal lugs MES_M_NEG and
MES_M_POS of generator kV/mA board 701-PL1.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 005 - Calibration of kV Scale Factor
CAL005.FM Page no. 161 Chapter 4
Job Card CAL 005 - Calibration of kV Scale Factor Chapter 4
1. SUPPLIES
None
2. TOOLS
• RMI 232 kV peak meter is recommended. Keithley 35080 may be used when RMI 232 is not available
on site.
Note:
The calibration curve attached to the meter (a curve obtained during periodic recalibration of the
meter) must be used to correct the meter reading throughout this procedure.
• Lead sheet.
3. SAFETY PRECAUTIONS
CAUTION
These procedures produce X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF KV SCALE FACTOR
The objective is to determine the scale factor G and offset VOF between the X-ray tube kV command
from the software and the real kV value obtained.
DEFAULT PARAMETER VALUES ARE G = +6.554E+1 (65.54) and VOF = +0.000E+0 (0.0)
CAUTION
DO NOT touch potentiometer P1 on generator kV/mA board 701-PL1 at any point in this
Procedure
.
5-1. Preparation
1. Set up the DMR in the following configuration:
- no compression paddle
- maximum magnification mode (the purpose here is simply to reduce the distance between the kV
peak meter and the X-ray source - see applicable documentation for minimum acceptable dis-
tance from X-ray source, if any).
2. Install a lead sheet on the top of the image receptor (to protect the detector from excessive exposure
to X-rays).
3. Place the kV peak meter on the magnification device. Turn it on and set it to the MO/MO position. Set
the waveform selector to CP (constant potential).
4. Use the Senographe DMR light centering device to center the kV peak meter target with the X-ray
zone.
5. Starting from application mode, change the position of the installation menu enable switch (switch B1

Chapter 4 Page no. 162 CAL005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 005 - Calibration of kV Scale Factor on generator CPU board 400-PL3).
Note:
This must only be done if the DMR has been switched off since the last access to SETUP/GENE/
INSTAL or SETUP/ARM/VALID/INSTAL
5-2. Perform the calibration procedure.
1. Open the work list and declare a new Patient called "X-ray Tube kV scale factor"
2. Select SETUP/GENE/INSTAL/GENE/GENE/KV/CALIB/1st pt/CALIB on the console.
Note:
By default, the kV command value sent by the software for this 1st point is 25 kV.
3. Press the 2nd trigger button and hold it down. A single 1-second exposure is taken.
CALIBRATION
END
appears on the console display. Release the 2nd trigger button. Note down the kV value dis-
played on the kV peak meter.
Note:
If for some reason this exposure has to be repeated, select SETUP/CALIB on the console before
pressing the 2nd trigger button again.
4. Select SETUP/kV_M on the console. Enter the kV value read from the kV peak meter by using the
NEXT and VALID keys and rotating the kV dial. See Visual Display of Parameter Values and Modifi-
cation of a Parameter Value or Entry of a Measurement in Job Card CFG "The Control Console".
5. Select SETUP/SETUP/2nd pt/CALIB on the console.
Note:
By default, the kV command value sent by the software for this 2nd point is 40 kV. Due to DMR
software, the Mo filter remains selected.
6. Press the 2nd trigger button and hold it down. A single 1-second exposure is taken.
CALIBRATION
END
appears on the console display. Release the 2nd trigger button. Note down the kV value dis-
played on the kV peak meter.
Note:
If for some reason this exposure has to be repeated, select SETUP/CALIB on the console before
pressing the 2nd trigger button again.
Note:
Select SETUP/kV_M on the console. Enter the kV value read from the kV peak meter by using the
NEXT and VALID keys and rotating the kV dial. See
Visual Display of Parameter Values and Mod-
ification of a Parameter Value or Entry of a Measurement
in Configuration Job Card The Control
Console
.
7. Select SETUP/SETUP/calcul/VALID on the console to execute the calculation. Normally,
KV CALIB
END
appears on the console display. The message OUT OF RANGE the console display indicates
either faulty kV measurement(s) in the steps above, or a hardware failure.
5-3. Check results.
1. Leave the Senographe DMR set up as given in section 5.1.1 , with the kV peak meter on the cassette
holder, and select SETUP/1st pt/CALIB on the console.
2. Press the 2nd trigger button and hold it down. A single 1-second exposure is taken. ”CALIBRATION
END” appears on the console display. Release the 2nd trigger button. Note down the kV value dis-
played on the kV peak meter. DO NOT ENTER THIS MEASURED VALUE.
3. The kV value read on the kV peak meter should be 25 kV ± 0.3 kV.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 005 - Calibration of kV Scale Factor
CAL005.FM Page no. 163 Chapter 4
Note:
The generator kVp accuracy is ±1% (one per cent) when measured just after a calibration.
4. Select SETUP/SETUP/2nd pt/CALIB on the console.
CAUTION
Before proceeding to make exposures, be sure that the detector has been protected from
excessive X–ray doses as instructed (by placing a lead sheet on the top of the image
receptor).
5. Press the 2nd trigger button and hold it down. A single 1-second exposure is taken. ”CALIBRATION
END” appears on the console display. Release the 2nd trigger button. Note down the kV value dis-
played on the kV peak meter. DO NOT ENTER THIS MEASURED VALUE.
6. The kV value read on the RMI kV peak meter should be 40 kV ± 0.4 kV.
Note:
The generator kVp accuracy is ±1% (one per cent) when measured just after a calibration.
Note:
If the need should arise to check this point in application mode, it is not possible to do so because
exposures at 40 kV are prohibited with the Mo filter in application mode.
7. Select SETUP/SETUP/SETUP/PARAM on the console. Note down the displayed values of G and
VOF by rotating the kV dial.
8. Perform a CKSUM and return to application mode,
9. Delete the previously created Patient called "X-ray Tube kV scale factor".
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

Chapter 4 Page no. 164 CAL005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 005 - Calibration of kV Scale Factor This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 006 - Calibration of Compression Force Detector
CAL006.FM Page no. 165 Chapter 4
Job Card CAL 006 - Calibration of Compression Force DetectorChapter 4
1. SUPPLIES
None
2. TOOLS
• Calibrated force detector capable of measuring compression force between the compression paddle
and the image receptor; required accuracy ±0.5 daN. The Mammo compression scale Part No. 46–
194427P407 is recommended.
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF COMPRESSION FORCE DETECTOR
The objective is to calibrate the compression force detector.
WARNING
If a power line failure occurs during the calibration (before a checksum is done), it is man-
datory to repeat this calibration from the beginning.
5-1. Preparation
Starting from application mode, change the position of the installation menu enable switch (switch B1 on
generator CPU board 400-PL3).
Note:
This must only be done if the DMR has been switched off since the last access to SETUP/GENE/
INSTAL or SETUP/ARM/VALID/INSTAL
5-2. Perform the calibration.
1. Set up the DMR in the following configuration:
- Arm rotation at 0°
- Contact mode
- Any compression paddle installed
2. Raise the compression paddle and place the force detector on the image receptor.
3. Starting from application mode, select SETUP/ARM/VALID/INSTAL/CALIB/COMP/FORCE/CALIB/
MEASURE on the console.
4. With the compression paddle completely raised and out of contact (i.e. in its home position), press
the
0daN key on the console.
5. Lower the compression paddle into contact with the force detector. Using the compression fine-
adjustment knob, adjust the compression to exactly 3 daN on the force detector.
6. Press the
3daN key on the console.

Chapter 4 Page no. 166 CAL006.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 006 - Calibration of Compression Force Detector7. Still using the compression fine-adjustment knob, increase the compression to exactly 6 daN on the
force detector.
8. Press the
6daN key on the console.
9. Continue this process for 9 and 12 daN, then press the
1 ->2 key on the console and proceed with 15,
18, 21, 24, and 27 daN.
10. Select SETUP/CALCUL on the console. This executes the calculation of the new force detector cali-
bration parameters.
11. Perform a CKSUM and return to application mode.
12. Request a total decompression of the compression paddle (the paddle must return to its home posi-
tion).
5-3. Check results.
1. With the arm at 0° (i.e. X-ray tube above the image receptor), bring the compression paddle into con-
tact once again with the force detector. Increase the compression until maximum force is reached
(i.e. compression cannot be further increased). Read the force detector measurement and check that
it is within ±1 daN of the maximum compression force chosen in the SETUP/MEDICAL/FORCE
menu.
2. Repeat the above check with the arm rotated to 180° (i.e. X-ray tube below the image receptor).
Note:
The maximum compression force must never exceed 30 daN. This limit must always be checked
after any maintenance (preventive or corrective) of the compression assembly to insure patient
safety.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 007 - Calibration of Breast Thickness Measurement
CAL007.FM Page no. 167 Chapter 4
Job Card CAL 007 - Calibration of Breast Thickness MeasurementChapter 4
1. SUPPLIES
None
2. TOOLS
• Plexiglass in thickness increments of 5 mm (minimum plexiglass dimensions 200 x 200 mm)
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. CALIBRATION OF BREAST THICKNESS MEASUREMENT
The objective is to calibrate the breast thickness measurement. Since the compression paddle and its
drive train, as well as the image receptor or magnification device undergo deflection when under com-
pression, a compensation factor must be calculated to correct for changes in compression force at a
fixed breast thickness.
Two calibrations must be done, one for contact mode and one for magnification mode.
5-1. Prepare the DMR for the calibration.
Starting from application mode, change the position of the installation menu enable switch (switch B1 on
generator CPU board 400-PL3).
Note:
This must be done only if the DMR has been switched off since the last access to SETUP/GENE/
INSTAL or SETUP/ARM/VALID/INSTAL
5-2. Perform contact mode calibration.
1. Set up the DMR in contact mode with the 19 x 23 compression paddle installed.
2. Request a total decompression of the compression paddle to its home position.
Note:
This MUST be done before entering the gantry installation menus.
3. Starting from application mode, select SETUP/ARM/VALID/INSTAL/CALIB/THICK/CONTACT/
CALIB on the console.
4. Lower the compression paddle into contact with the image receptor. Using the compression fine-
adjustment knob, adjust the compression to between 3 and 5 daN as read on the gantry display
panel. Press the
<5daN key on the console.
5. Increase the compression to between 15 and 20 daN as read on the gantry display panel. Press the
>15daN key on the console. The Calcul key appears on the console display.
6. Press the
Calcul key on the console. This executes the calibration calculation for contact mode.

Chapter 4 Page no. 168 CAL007.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 007 - Calibration of Breast Thickness Measurement5-3. Check the contact mode calibration results.
1. Request a total decompression of the compression paddle.
2. Keeping the
DMR in contact mode, place 40 mm of plexiglass in the center of the bucky.
3. Lower the compression paddle into contact with the plexiglass.
4. Verify for any compression force between zero and 20 daN of compression (as displayed on the gan-
try display panel), that the breast thickness displayed on the gantry display panel does not exceed
the range of 35 to 45 mm (i.e. plexiglass thickness of 40 mm, ± 5 mm).
5-4. Perform the magnification mode calibration.
1. Install the magnification device and the 19 x 23 compression paddle.
2. Select SETUP/SETUP/MAGNIF/CALIB on the console.
3. Request a total decompression of the compression paddle to its home position, using the appropriate
pedal.
4. Lower the compression paddle into contact with the magnification device. Using the compression
fine-adjustment knob, adjust the compression to between 3 and 5 daN as read on the gantry display
panel. Press the
<5daN key on the console.
5. Increase the compression to between 15 and 20 daN as read on the gantry display panel. Press the
>15daN key on the console. The Calcul key appears on the console display.
6. Press the
Calcul key on the console. This executes the calibration calculation for magnification mode.
5-5. Check the magnification mode calibration results.
1. Request a total decompression of the compression paddle.
2. Place 40 mm of plexiglass on the magnification device.
3. Lower the compression paddle into contact with the plexiglass.
4. Verify for any compression force between zero and 20 daN of compression (as displayed on the gan-
try display panel), that the breast thickness displayed on the gantry display panel does not vary out-
side the range of 37 to 43 mm (i.e. plexiglass thickness of 40 mm, ± 3 mm).
5-6. Return the DMR to its normal state.
1. Perform a CKSUM and return to application mode.
2. If you are handing the system back to the user at this point, turn the
DMR off, then back on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 008 - Compression/Decompression Minimum Force
CAL008.FM Page no. 169 Chapter 4
Job Card CAL 008 - Compression/Decompression Minimum Force Chapter 4
1. SUPPLIES
None
2. TOOLS
• A 3 kg weight suitably sized to allow it to be placed on top of the compression paddle.
3. SAFETY PRECAUTIONS
None.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. MINIMUM COMPRESSION/DECOMPRESSION FORCE TESTS
The objectives are:
• Check that a compression force (limit IC: In Compression) of at least 22 daN can be reached when
the arm is rotated by 180° with the paddle drive motor current limit value set between 50% and 70%
(motor’s maximum current rating).
• Check that a decompression force (limit OC: Out of Compression) of at least 3 daN can be reached
when the arm is at 0° with the paddle drive motor current limit value set to 65% of the motor’s maxi-
mum current rating (this ensures that the compression paddle will rise and stay up when decompres-
sion is requested).
5-1. Prepare the DMR
1. Starting from application mode, change the position of the installation menu enable switch (switch B1
on generator CPU board 400-PL3).
Note:
This must be done only if the DMR has not been switched off since the last access to
SETUP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL.
2. Set up the DMR in contact mode with a compression paddle installed.
3. Inhibit the cassette detection in medical menu: starting from application mode, select: SETUP/MEDI-
CAL/FILM/NO
5-2. Check and adjust the limit IC
1. Starting from application mode, select:
SETUP/ARM/VALID/INSTAL/CALIB/COMP/LIMIT/CALIB/LIMIT_IC on the console. Check that the
limit IC is set between 50% and 70% for the motor current value and perform the test.
2. Rotate the arm to 180° (i.e., X-ray tube below the image receptor).
3. Bring the compression paddle into contact with the image receptor.
Note:
Be ready to read the compression force displayed on the gantry readout unit. The paddle should
release and fall towards the tube housing for a compression close to 22 daN.

Chapter 4 Page no. 170 CAL008.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 008 - Compression/Decompression Minimum Force4. Increase compression to 22 daN as displayed on the gantry display panel. You must be able to reach
22 daN without the compression paddle releasing. If the compression paddle releases for a force
lower or higher than 22 daN, perform the following adjustment ; otherwise go straight to section 5-3.
a. Set up the limit IC to its minimum: Starting from application mode, select
SETUP/ARM/VALID/INSTAL/CALIB/COMP/LIMIT/CALIB/LIMIT_IC on the console.Use the + and
– key on the console to set the motor current value to 50%. Enter this values with the VAL key on
the console.
b. Increase compression to 22 daN as displayed on the gantry display panel. You must be able to
reach 22 daN without the compression paddle releasing.
c. If the compression paddle releases before a force of 22 daN is reached, increase the compres-
sion motor current value by 5%, then try again to reach 22 daN. Continue increasing the maxi-
mum in compression motor current value in steps of 5%, if necessary, until 22 daN can be
reached.
Note:
Between each compression, request a total decompression of the compression paddle, using the
appropriate footswitch.
5-3. Check the limit OC.
1. Rotate the arm to 0° (i.e., X-ray tube above the cassette holder).
2. Bring the compression paddle into contact with the cassette holder.
3. Select SETUP/LIMIT_OC on the console. The maximum motor current value (in decompression
mode) should be set to 65% otherwuise use the + and – key on the console to set this value. Enter
this value with the VAL key on the console and return to the application menu.
4. Place the 3 kg weight on top of the compression paddle.
5. Request a total decompression of the compression paddle using the appropriate pedal. The com-
pression paddle must rise smoothly to its home position.
6. If the compression paddle slips or releases, check that the compression paddle drive belt is not slip-
ping. If it is slipping, adjust or replace the belt.
Note:
The maximum out-of-compression motor current adjustment range is limited to 65%.
7. Perform a CKSUM and return to application mode.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 009 - Compression Arm Angle Calibration
CAL009.FM Page no. 171 Chapter 4
Job Card CAL 009 - Compression Arm Angle Calibration Chapter 4
1. SUPPLIES
• None
2. TOOLS
• None
3. SAFETY PRECAUTIONS
• None (Calibration is performed on DMR console (no X-rays required).
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. PROCEDURE
Note:
The compression arm angle must be calibrated for automatic selection of view names.
Note:
The sign convention for Senographe 2000 D applications is illus-
trated here, shown looking at the tube head or chest wall of the
DMR.
1. Turn the Compression arm clockwise to the mechanical stop (-178°).
2. Set MIN compression arm rotation value:
arm/valid/instal/calib/angle/calib/MIN
3. Turn the Compression arm counterclockwise to the mechanical stop (+178°)
4. Set MAX compression arm rotation value:
Select:MAX
5. Perform calculation:
Select:CALCUL
6. Go up the menu and perform cksum
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.
+ -

180°

Chapter 4 Page no. 172 CAL009.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 009 - Compression Arm Angle Calibration This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 010 - Tube Tilt Calibration
CAL010.FM Page no. 173 Chapter 4
Job Card CAL 010 - Tube Tilt Calibration Chapter 4
1. SUPPLIES
• None.
2. TOOLS
• None.
3. SAFETY PRECAUTIONSWARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
• All detector calibrations must have been cleared (select
Service desktop/Calibration/Detector; then
select
Clear Calibrations and perform the clear procedure as indicated on the screen).
5. CALIBRATION OF X-RAY TUBE TILT
The objective is to determine the X-ray tube angle when in its reference position. This calibration is done
by taking an exposure with the field coverage depth limited by the anode geometry (and not by the rear
collimator blade), measuring this field coverage depth, and entering this depth into the DMR, which can
then calculate the X-ray tube angle when in its reference position.
Notice:!
DO NOT perform Tube Tilt Calibration unless it is strictly necessary (e.g., after tube replace-
ment or when there is clear evidence of bad calibration parameters).
BEFORE performing Tube Tilt Calibration, clear all detector calibrations (select
Service
desktop/Calibration/Detector
; then selectClear Calibrations and perform the clear procedure as in-
dicated on the screen).
AFTER performing Tube Tilt Calibration, it is essential to perform the following CAL Job Cards:
- Bad Pixel Calibration.
- Detector Gain Calibration (for all techniques).
- Conversion Factor Measurement.
- AOP Calibration.
5-1. Procedure
• Before performing this calibration, make an exposure in normal application mode; check that the red
X-ray inhibition light is not on, and no errors are reported. The exposure required by the calibration is
not made correctly if the detector temperature is out of range; the pre-exposure is aborted and
relaunched as long as the exposure button is pressed, but no exposure is possible. Error messages

Chapter 4 Page no. 174 CAL010.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 010 - Tube Tilt Calibration may be confusing or non-existent.
1. Select Service Desktop from the Browser Toolkit menu.
2. Click the
Calibrations button, then the Tube Tilt button.
3. A full screen appears on the AWS monitor.
4. On the DMR console, parameters are automatically set to HOUSING ANGULATION CALIB.
5. Remove Bucky and compression paddle. No phantom should be present.
6. Press EXPOSE on DMR console.
7. Verify that the mAs are set to 25, press VALID/VALID.
8. When the message "ADJUST EXPOSURE" is displayed, press and hold the exposure button (no
PREP) while the calibration exposure is taken.
Be sure to hold the button until the message "Calibration End" is displayed.
The screen displays the measured chest wall to anode heel distance (for information only).
Note:
Do not try to make another exposure at this point without terminating the calibration. If you want to
re-calibrate, first end the current calibration by performing the next step (press SETUP/SETUP),
then restart the calibration from step 2.
9. Press SETUP/SETUP. The message "HOUSING ANGULATION CALIB calibration OK". should be
displayed. If an error is detected, repeat the calibration.
10. To exit the calibration, click
Quit on the screen, then press SETUP on the console.
11. Click the
Refresh button on the CAL SUMMARY page to update the calibration date.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 011 - X-ray Beam Alignment Calibration
CAL011.FM Page no. 175 Chapter 4
Job Card CAL 011 - X-ray Beam Alignment Calibration Chapter 4
1. SUPPLIES
• None.
2. TOOLS
• Screw driver to open cover.
• Set of metric Allen wrenches.
• 8 mm hollow offset tubular socket wrench
3. SAFETY PRECAUTIONS
WARNING
These procedures produce X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
Note:
Before performing this calibration, make an exposure in normal application mode. Check that the
red X-ray inhibition light is not on, and no errors are reported. The exposures required by the cal-
ibration are not made correctly if the detector temperature is out of range; the pre-exposure is
aborted and relaunched as long as the exposure button is pressed, but no exposure is possible.
Error messages may be confusing or non-existent.
5. DESCRIPTION
The calibration function automatically calculates and displays measurements indicating the position of
the X-ray beam in relation to the detector. It uses these measurements to determine whether the calibra-
tion was successful. If so, no further action is required.
If an error is signalled, you must use the displayed values to determine the source of the problem and
make appropriate adjustments before repeating the calibration. Two checks are required:
• Collimator Blade Parallelism. Use the displayed angles to check for correct parallelism of the two
pairs of collimator blades; adjust the collimator position if necessary.
• Lateral Centering of X-ray Beam. Use the displayed values to check for correct positioning of the
beam on the detector; adjust the beam position if necessary.
Notice:!
The
Collimator Blade Parallelism check must be completed, and any errors corrected, before mak-
ing the
Lateral Centering of X-ray Beam check.

Chapter 4 Page no. 176 CAL011.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 011 - X-ray Beam Alignment Calibration
6. PROCEDURE
1. Remove the bucky or breast holder and remove the compression paddle. No phantom should be
present
2. From the Service Desktop, click the Calibration button and select X-ray Alignment. Exposure param-
eters are automatically set by the system.
3. Press EXPOSE on the control console
4. Verify that mAs is set to 25 and press VALID/VALID
5. When the message ADJUST EXPOSURE is displayed, press and hold the exposure button (no
PREP) until an exposure has been taken and the message "calibration end" is displayed (about 10
seconds).
6. The system uses the acquired image to assess the position of the collimator and displays a list of
measurements. Disregard these measurements at this time.
7. Press SETUP/SETUP. The message "calibration OK" should be displayed.
8. If an error is signalled, follow the instructions given in the section "Error Actions".
9. To exit the calibration, click "Quit" on the screen, then press SETUP on the console.
7. ERROR ACTIONS
7-1. Measurements
If an error is signalled, first note the values displayed on the screen and use them in the calculations
described below:
Displayed Measurement Value Units Comment
Distance from detector center to front blade mm
Not used for service
Distance from detector center to rear blade mm
Distance from detector center to left blade mm
Distance from detector center to right blade mm
Angle between detector column and front blade radians
Use for collimator
blade
parallelism
Angle between detector column and rear blade radians
Angle between detector row and left blade radians
Angle between detector row and right blade radians
Distance from detector edge to left blade mm
Use for X-ray beam
lateral centering
Distance from detector edge to right blade mm
Lateral offset mm

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 011 - X-ray Beam Alignment Calibration
CAL011.FM Page no. 177 Chapter 4
7-2. Collimator Blade Parallelism check
Perform the instructions of this section only if an error is signalled. They must be completed, and adjust-
ments made if necessary, before continuing with the Lateral Centering check.
1. After performing the calibration measurement as described above, note the displayed angular values
(shown in the table below with references a1 through a4):
2. Calculate the differences |a1 - a2| and |a3 - a4|.
3. Check the displayed and calculated angles against the maximum values given in the table::
4. If all values are less than the maxima given in the table, no action is necessary; the check is com-
plete, and you can proceed to check lateral centering of the X-ray beam.
5. If any value exceeds that given in the table, proceed as follows:
•If the value |a1 - a2| is greater than 0.018 radians,
or the value |a3 - a4| is greater than 0.016 radians,
Change the collimator. No field adjustment is possible.
•If the displayed value for |a1| exceeds 0.007 radians:
- Loosen the collimator assembly mounting slightly; adjust the front blade position as required and
secure the assembly
- Press SETUP once; repeat the calibration measurement to check the new position.
- Repeat as necessary until the criterion for a1 is met.
•If the displayed value for |a2|, |a3|, or -a4| exceeds the value given:
Change the collimator. No field adjustment is possible.
Displayed and calculated angles Ref. Value (radians)
Actual Maximum
Angle between detector column and front blade |a1| 0.007
Angle between detector column and rear blade |a2| 0.007
Angle between detector row and left blade |a3| 0.008
Angle between detector row and right blade |a4| 0.008
Calculate the difference between a1 and a2| a1 - a2| 0.018
Calculate the difference between a3 and a4| a3 - a4| 0.016

Chapter 4 Page no. 178 CAL011.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 011 - X-ray Beam Alignment Calibration
7-3. Lateral Centering of X-ray Beam check
Carry out this procedure only if an error has been signalled and all angles calculated in the Collimator
Blade Parallelism procedure are within the specified limits.
1. After performing the calibration measurement as described above, note the displayed
Lateral offset,
which is the measurement t2 - t1 (see diagram).

2. If offset½t2 -t1½ is equal to or less than 1 mm, no adjustment is required; the procedure is complete.
3. If offset½t2 -t1½is greater than 1 mm, the beam position must be adjusted. Proceed as follows:
7-3-1. Adjustment
•If the offset ½t2 -t1½is greater than 1 mm adjust the beam position as follows:
- If t2 -t1 is POSITIVE, adjust from the LEFT side of housing.
If t2 -t1 is NEGATIVE, adjust from the RIGHT side of housing
1. Place the 8 mm hollow offset tubular socket around the adjustment screw locknut on the side of the
collimator housing (right or left as determined above).
2. Holding the adjustment screw in the center of the locknut in place with the 2.5 mm Allen wrench,
loosen the locknut.
3. Use the 2.5 mm Allen wrench to UNSCREW (turn COUNTER-clockwise) the adjustment screw
according to the formula: NUMBER OF TURNS = 0.2 x(t2 -t1).
4. While holding the adjustment screw in place, tighten the locknut.
5. On the OPPOSITE side; loosen the adjustment screw locknut. TIGHTEN (turn CLOCKWISE) the
adjustment screw until snug, then re-tighten the locknut.
6. After adjustment, press SETUP once and repeat the Beam Position Measurement procedure.
7. Repeat the measurement and adjustment until the offset ½t2 -t1½ < 1 mm.
7-3-2. After adjustment
• After making any adjustments, repeat the calibration. When all required actions have been com-
pleted, click the "Quit" button on the screen, and click the "Refresh" button on the CAL SUMMARY
page to update the calibration date.
• If any adjustments were required to correct the X-ray beam alignment, carry out the procedures given
in Calibration Job Card
Collimator Format Calibration.
8. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.
Chest Wall
Detector FOV
Collimator
t1 t2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 012 - Collimator Format Calibration
CAL012.FM Page no. 179 Chapter 4
Job Card CAL 012 - Collimator Format Calibration Chapter 4
1. SUPPLIES
• None.
2. TOOLS
• None.
3. SAFETY PRECAUTIONSWARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. PROCEDURE
5-1. Presentation
The collimator blade movement must be calibrated in order to produce the different fields of view (large,
medium, or small) on the detector. The exact magnification of the collimator blades must be calibrated.

Chapter 4 Page no. 180 CAL012.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 012 - Collimator Format Calibration
5-2. Description
1. Select Service Desktop from the Browser Toolkit menu; click the “Calibrations” button, then the “Colli-
mator Format” button.
2. A full screen appears on AWS monitor.
3. On DMR console, parameters are automatically set to COLLIMATOR FORMAT CALIB.
4. Remove Bucky and compression paddle. No phantom should be present.
5. Press EXPOSE on DMR console.
6. Verify that the mAs are set to 25; press VALID/VALID.
7. When the message “ADJUST EXPOSURE” is displayed, press and hold the Exposure button (no
PREP) while three exposures are taken. Be sure to hold the button until the “Calibration end” mes-
sage is displayed (about 45 seconds). A table of measurements is displayed; they have no signifi-
cance for this procedure.
8. Press SETUP/SETUP. You should have the message
Calibration OK.
If a
Calibration error message is displayed, repeat the calibration.
9. To exit the calibration, click “Quit” on the screen.
10. Click the “Refresh” button on the CAL SUMMARY page to update the calibration date.
5-3. Error messages
•If Calibration errors persist, first check that the X-ray Beam Alignment Calibration has been correctly
carried out. Then verify the status of the collimator blades and motor, and change them if necessary.
•
Collimator blades are not detected within detector FOV. This error occurs when the collimator blades
do not enter the field of view for medium and small format. It usually indicates a failure in the step
motor.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 013 - Bad Pixel Calibration
CAL013.FM Page no. 181 Chapter 4
Job Card CAL 013 - Bad Pixel Calibration Chapter 4
1. SUPPLIES
• None.
2. TOOLS
- Standard toolkit with metric wrenches.
- Flat Field phantom (25 mm lucite).
Notice:The lucite must be clean and free from scratches. This is essential to avoid incorrect detec-
tion of dead pixels.
3. SAFETY PRECAUTIONS
WARNING
This procedure requires the use of X–rays; take all due precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
• It is strongly recommended to perform bad pixel calibration when the detector has not been used for
a while (typically on a Monday morning if the detector is not used over the weekend).
• The temperature of the detector must be stabilized and within the specified range before commenc-
ing and during bad pixel calibration.
• The Senographe DMR must be in the Medical Application menu.
5. BAD PIXEL CALIBRATION
5-1. Description
A Bad Pixel is an Image Receptor cell which has a response curve very different from that of other cells
subject to the same dose, even after flat field correction. It thus provides no useful information, and
appears as a white or black pixel on the image. Bad Pixel Calibration is a process which uses a number
of acquisitions to identify such cells and store their positions in a Bad Pixel Map. The bad pixel correction
process, applied to acquired images, uses the map to replace the level read from each bad pixel with the
mean level of neighboring cells.
Notice:!
It is important that the phantom used in these procedures, and all surfaces in the beam (detector,
tube, etc.) are very clean and free from imperfections, which can affect the calibration.

Chapter 4 Page no. 182 CAL013.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 013 - Bad Pixel Calibration
5-2. Procedure
Notice:!
Before starting the calibration, check that the Conditioner temperature has stabilized at 20°C ±2°C.
1. Remove the Bucky or Breast Holder and remove the compression paddle. Install the Flat Field phan-
tom (25 mm of lucite) on the Image Detector.
2. Open the Service Desktop; click the Calibration button and select
Detector Calibration.
3. Select
Bad pixel and Start; follow the on-screen instructions to run the calibration.
4. The IDC automatically acquires 16 offset images (no X-rays) and computes results.
5. When the offset acquisitions and calculations are complete, you are requested to install the Flat Field
phantom and make four exposures. The parameters are set automatically (Rh/Rh; 30 kV; 56 mAs).
Release the Prep and Expose buttons between each exposure.
6. You are requested to make a low dose exposure. The parameters are set automatically (Rh/Rh; 30
kV; 5.6 mAs).
7. You are requested to make a high dose exposure using the following parameters: (Rh/Rh; 30 kV; 80
mAs).
8. The IDC uses the acquired images to create the bad pixel map which will be used for subsequent
image correction. The bad pixel results are displayed on the screen in the form of a table indicating
Pass or Fail for all tested specifications.
5-3. Errors
•Exposure abort during calibration. If the X-ray exposure is aborted during bad pixel calibration, the
buzzer sounds and the calibration is terminated at the end of the current set of exposures. The cali-
bration must be re-started.
• If any failure is reported, check to see if the collimator blades are within the detector FOV. If so, cor-
rect the problem and try again.
•If an
Isolated bad pixels failure is reported, check detector temperature and repeat the calibration
when the temperature is correct.
•If a
Cluster or Bad Columns failure is reported, check the Flat Field phantom for imperfections such
as scratches or dust, and repeat the calibration after correction.
When a
Bad Columns failure is reported, also check that the floppy disk supplied with the detector
was correctly loaded when the Image Receptor was installed, as instructed in the D/R Job Card
Image Receptor. It contains bad column information specific to the detector.
5-4. Gain Calibration
• .After completing the Bad Pixel calibration, Gain Calibration must be performed for all techniques
(perform CAL Job Card
Detector Gain Calibration).
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 014 - Detector Gain Calibration
CAL014.FM Page no. 183 Chapter 4
Job Card CAL 014 - Detector Gain Calibration Chapter 4
1. SUPPLIES
• None.
2. TOOLS
• Flat Field phantom (25 mm lucite).
3. SAFETY PRECAUTIONSWARNING
These procedures produce X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
• It is strongly recommended to perform gain calibration when the detector has not been used for a
while (typically on a Monday morning if the detector is not used over the weekend).
• The Senographe DMR must be in the Medical Application menu.
5. DETECTOR GAIN CALIBRATION
5-1. Presentation
The goal of gain calibration is to compensate for differences in pixel responses and differences in elec-
tronic gain. This calibration makes the image of a flat uniform object look flat (flat field uniformity).
5-2. Procedure
1. Open the Service Desktop; click the Calibration button and select Detector Gain Calibration.
2. Before starting the calibration, follow the on-screen instructions for pre-requisites.
3. Follow the on-screen instructions to set up the system: remove the Bucky or Breast Holder and
remove the compression paddle. Install the flat field phantom (25 mm of lucite) on the Image Recep-
tor.
4. You are requested to check that the collimator is in full FOV, and then to make an exposure for an
image quality check. The IQ check technique and parameters are selected automatically).
5. If the check is successful you are requested to make four exposures for gain calibration. The system
selects the calibration technique and parameters, as beforerh/rh but with a higher mAs value.
6. The system uses information from the four exposures to update the gain map; a message is dis-
played to indicate successful completion.
7. Repeat the procedure for each of the listed techniques.

Chapter 4 Page no. 184 CAL014.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 014 - Detector Gain Calibration
The following table lists the techniques to be calibrated, with specified parameters and configurations.
6. ERROR ACTIONS
If an error is found, an error message is displayed, with a request for further action. Possible error mes-
sages are:
1.
IQ check failed - Detector not ready. You are informed that the IQ check failed because the check for
detector lag failed, or because the detector temperature was outside limits. You are then requested to
try the test again after a 15 minute delay or after the detector temperature has stabilized.
2.
IQ check failed - system configuration is not OK. You are asked to repeat the IQ check after checking
the system configuration (phantom position, removal of paddle, collimator in full field of view).
7. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.
Technique for calibration IQ check technique;
Calibration technique
Configuration
Mo/Mo/LF/grid Mo/Mo 26 kV 32 mAs, LF;
Mo/Mo 26 kV 320 mAs, LF
Bucky installed, no paddle.
Phantom on bucky.
Mo/Mo/LF/nogrid Mo/Mo 26 kV 16 mAs, LF;
Mo/Mo 26 kV 160 mAs, LF
No bucky, no paddle.
Phantom on image receptor.
Rh/Rh/LF/grid Rh/Rh 30 kV 11 mAs, LF;
Rh/Rh 30 kV 110 mAs, LF
Bucky installed, no paddle.
Phantom on bucky
Rh/Rh/LF/nogrid Rh/Rh 30 kV 5.6 mAs, LF;
Rh/Rh 30 kV 56 mAs, LF
No bucky, no paddle.
Phantom on image receptor.
Mo/Mo/SF/nogrid/mag Mo/Mo 26 kV 16 mAs, SF;
Mo/Mo 26 kV 160 mAs, SF
No bucky, no paddle.
Phantom on mag. stand (1.8; mid-
dle hole).
Rh/Rh/SF/nogrid/mag Rh/Rh 30 kV 10 mAs, SF;
Rh/Rh 30 kV 100 mAs, SF
No bucky, no paddle.
Phantom on mag. stand (1.8).

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 015 - Conversion Factor Measurement
CAL015.FM Page no. 185 Chapter 4
Job Card CAL 015 - Conversion Factor Measurement Chapter 4
1. SUPPLIES
• None
2. TOOLS
• RADCAL Dosimeter (6 cc chamber).
• Dosimeter support with steel protection sheet.
• 5 cm of lucite.
3. SAFETY PRECAUTIONS
WARNING
This procedure requires the use of X–rays; take all due precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
• The Senographe DMR must be in the Medical Application menu.
5. CONVERSION FACTOR MEASUREMENT
5-1. Description
The Conversion Factor is required for AOP (Automatic Optimization of Parameters) calculations. It is
determined by using a dosimeter to measure the X-ray dose while acquiring an image. This value is com-
pared with the brightness of the corrected image to give a Conversion Factor, used by the system to con-
vert brightness to dose for all AOP-related calculations. The brightness and dose values are also sent to
the AWS for the image quality tests.
5-2. Procedure
1. Open the Service Desktop; click the Calibration button and select Detector/CF Measurement.
2. Before starting the calibration, follow the on-screen instructions for pre-requisites.
3. Follow the on-screen instructions to run the calibration. You are asked to:
- Remove the bucky or breast holder and remove the compression paddle.
- Install the dosimeter support.
- Install steel sheet for detector protection.
- Place 5 cm of lucite on the dosimeter support and the dosimeter in the low position (Source-
Dosimeter distance = 622.5 mm).
- Take an exposure (parameters set automatically: Rh/Rh; 30 kV; 56 mAs).
- Read the dosimeter and enter the dose value (in microGray or mR).
- Remove the dosimeter and steel sheet; take a second exposure with the same technique.
4. The IDC calculates the brightness level in the pre-defined region of interest.

Chapter 4 Page no. 186 CAL015.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 015 - Conversion Factor Measurement 5. The IDC computes the counts/microGray value and checks the result. If good, it displays the mes-
sage "Calibration successful".
5-3. Error codes
• Calibration calculation failed.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 016 - Half-Value Layer Measurement
CAL016.FM Page no. 187 Chapter 4
Job Card CAL 016 - Half-Value Layer Measurement Chapter 4
1. SUPPLIES
• None
2. TOOLS
• RADCAL dosimeter and 6M ionization chamber (6cc) calibrated at mammographic X ray beam ener-
gies,
• Dosimeter support with steel protection sheet,
• Two sheets of 0.2 and 0.5mm thickness of 99.9% pure aluminum, of length and width sufficient to
fully cover the ionization chamber. The stated thickness should be accurate within ±1%,
• 19x23 compression paddle.
3. SAFETY PRECAUTIONS
CAUTION
This procedure produces X Rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
Notice:!
The Detector must be protected with the steel protection sheet; it is not required for this measure-
ment
5. INTRODUCTION
The purpose of this procedure is to verify that the Half-value layer of the X-ray beam is adequate to min-
imize patient breast dose, while not so excessive that contrast is lost in the resultant image.
The tube yield is measured for all track/filter/focal spot combinations and the values used for skin dose
computation.
For each spectrum the HVL and tube Yield are determined and stored on the DMR, for use in the skin
dose and Mean Glandular Dose evaluation.
• The following techniques will be set automatically by the DMR Senographe:
• The calibration of each configuration (focal spot, track/filter) is independent of the results obtained in
other configurations.
Configuration Technique
Mo/Mo 26kV/63mAs/LF
Mo/Rh 28kV/56mAs/LF
Rh/Rh 30kV/45mAs/LF

Chapter 4 Page no. 188 CAL016.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 016 - Half-Value Layer Measurement 6. PROCEDURE
Notice:!
The dose values to be entered in the DMR menu must be expressed in mGy.
- If your dosimeter displays dose values in mR, divide the reading by 114.2 to obtain mGy.
- If your dosimeter displays dose values in mGy, divide the reading by 1000 to obtain mGy.
Note that you must use the format x.xxx E+0 to enter a dose value
x.xxx.
1. Ensure that paddle decompression at end of exposure is set to “NO”. In the DMR menu, set:
MEDICAL/DECOMP/NO
2. Place the 19 cm x 23 cm compression paddle
upside down as close as possible to the X-ray tube.
3. Remove the bucky or breast support; place the dosimeter support with steel protection plate on the
image receptor.
4. Place the dosimeter with 6M chamber on
the high position of the dosimeter support (the source-to-
dosimeter distance SDD = 592mm).
5. Select the 9x9 collimator format; verify using the light beam that the 6M ionization chamber is fully in
the beam.
6. In the HVL/Yield calibration menu on the DMR console, enter:
SET UP/GENE/INSTAL/GENE/DOSE /LF/MO/CALIB/0/CALIB
7. The DMR automatically selects the acquisition parameters (see table above).
8. Take the calibration exposure (press DMR exposure button) with no aluminum sheet on paddle, and
read the dose value on the dosimeter.
9. Press Set Up once on DMR console, select Meas.D, enter the dose value E
0
in mGy (see Notice
above) and press Valid.
10. Press Set Up twice, and select 0.2.
11. Place 0.2 mm aluminum on top of the paddle; verify using the light beam that the aluminum sheet
fully covers the ionization chamber field of view.
12. Repeat steps 8 and 9 with 0.2 mm aluminum
13. Press Set Up twice, and select 0.5
14. Place 0.5 mm aluminum on top of the paddle. Verify using light beam that the aluminum sheet fully
covers the ionization chamber field of view.
15. Repeat steps 8 and 9 with 0.5 mm aluminum
16. Press Set Up twice, and press Calcul
17. Press Set Up, press Param, verify that the value is within the limits given below (
Acceptable Values)
18. If the values are within limits, the DMR stores HVL and Yield values.
If not, check the following and repeat the calibration:
• Paddle is present,
• Dosimeter is properly calibrated,
• 6M probe is correctly placed in the high position of the dosimeter holder tool
• Tube tilt and format are correctly calibrated, and probe is in the 9x9 field of view
• Generator kV and mA are properly calibrated.
If the calibration still fails, the tube must be replaced.
19. Repeat steps 6 through 18 for the other track/filter/focal spot combinations: LF/MoRh, LF/RhRh, SF/
MoMo, SF/MoRh and SF/RhRh techniques

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 016 - Half-Value Layer Measurement
CAL016.FM Page no. 189 Chapter 4
6-1. Acceptable values:
The specification limits for HVL and Yield values measured under these conditions are:
6-2. Information:
The DMR automatically computes the HVL by interpolation using the following formula:
(the units are mm of Al)
Where E
0
, E
0.2
, E
0.5
, are dose values measured behind 0, 0.2 mm and 0.5 mm aluminum respectively.
The DMR calculates Tube Yield as E
0
/ mAs (unit: mGy/mAs) for this technique.
7. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.
Configuration HVL (mm Al) Yield ( mGy/mAs)
MoMo 26kV 0.29 - 0.38 35 – 130
MoRh 28kV 0.31 - 0.47 35 – 130
RhRh 30kV 0.33 - 0.52 35 – 130





ln
E E lnE E ln(2*0.5
HVL
00.500.2

Chapter 4 Page no. 190 CAL016.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 016 - Half-Value Layer Measurement This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 017 - AOP Calibration
CAL017.FM Page no. 191 Chapter 4
Job Card CAL 017 - AOP Calibration Chapter 4
1. SUPPLIES
• None
2. TOOLS
• Flat Field Phantom (25 mm Lucite).
3. SAFETY PRECAUTIONSWARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. PROCEDURE
5-1. Presentation
This procedure will be done for the three following techniques:
• With Grid and Large Focal Spot.
• With No Grid and Large Focal Spot.
• With No Grid and Small Focal Spot.
Note:
AOP Calibration is required to adapt AOP parameters to tube yield (mGy/mAs).
5-2. Calibration with Grid/Large Focal Spot configuration
1. Close the exam window if opened.
2. Insert the bucky.
3. Place the Flat Field Phantom on the bucky.
4. Compress the phantom.
5. Select Large Focal Spot from DMR console.
6. From DMR console, enter GENE/INSTAL/GENE/AOP/APOLLO/CALIB/GRIDLF menu.
7. Press CALIB
8. Press the acquisition button of DMR console (not PREP button) and hold it down until the system has
performed 3 acquisitions (about 20 seconds).
9. A message is displayed to indicate that the calibration was successful.
10. Perform a checksum from DMR console: GENE/INSTALL/CKSUM.

Chapter 4 Page no. 192 CAL017.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 017 - AOP Calibration
5-3. Calibration with No Grid/Large Focal Spot configuration
Note:
This calibration is applicable only if the Breast Holder option is present.
1. Remove the bucky and insert the Breast Holder.
2. Place the Flat Field Phantom on the Breast Holder.
3. Compress the lucite.
4. Select Large Focal Spot from DMR console.
5. From DMR console, enter GENE/INSTAL/GENE/AOP/APOLLO/CALIB/NOGRIDLF menu.
6. Press CALIB
7. Press the acquisition button of DMR console (not PREP button) and hold it down until the system has
performed 3 acquisitions (about 20 seconds).
8. A message is displayed to indicate that the calibration was successful.
9. Perform a checksum from DMR console: GENE/INSTALL/CKSUM.
5-4. Calibration with No Grid/Small Focal Spot configuration
1. Remove the bucky.
2. Install the magnification stand
3. Place the Flat Field Phantom on the magnifiaction stand.
4. Compress the phantom.
5. Select Small Focal Spot from DMR console.
6. From DMR console, enter GENE/INSTAL/GENE/AOP/APOLLO/CALIB/NOGRIDSF menu.
7. Press CALIB.
8. Press the acquisition button of DMR console (not PREP button) and hold it down until the system has
performed 3 acquisitions (about 20 seconds).
9. A message is displayed to indicate that the calibration was successful.
10. Perform a checksum from DMR console: GENE/INSTALL/CKSUM.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
CAL018.FM Page no. 193 Chapter 4
Job Card CAL 018 - ISC Monitor Adjustments Chapter 4
Notice:!
This Job Card applies only to monitors manufactured by Image Systems Corporation (refer to the
label on the rear panel if in doubt).
For other monitors, consult the appropriate documents.
1. SUPPLIES
None
2. TOOLS
- Image Systems’ Remote Commander (RC1), p/n 2190157.
- Gray reference card 18% p/n 2110360 (or equivalent).
3. SAFETY PRECAUTIONS
None
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. USING THE REMOTE COMMANDER (RC1)
Adjustments to Image Systems monitor are made using a hand-
held digital controller, the Remote Commander (RC1), illustrated
here.
When connected to an Image Systems monitor, the RC1
accesses microprocessors in the monitor electronics. Internal
parameters may be monitored and adjusted, using the keyboard
and display. The RC1 also functions as a light meter.
-+
123
456
789
0#
The Calibrator
image
SYSTEMS
Light meter
LCD
display
To connector A

Chapter 4 Page no. 194 CAL018.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments 1. Connect the 3.5 mm jack plug on the RC1 lead to the connector
(A) at the rear of the Image Systems monitor (remove the rear
panel for access).
2. Select the required internal processor unit (see Table 1 for a list
of available processor units):
a. Press the # key; the RC1 responds by displaying:
Proces-
sor?
b. Select the processor unit by pressing the corresponding
number (Table 1). the RC1 displays the name and software
revision of the processor unit selected.
3. Select the required function (see Tables 2 and 3 for lists of available functions).
Selection can be made in two ways:
a. By direct access: press the * key; the RC1 responds by displaying:
Function? Then press the two-
digit number given in the Code column of the function table for the required function.
b. By scrolling: press the
¿ key to access the next function in the menu (in the order given in the
function table). Press it as many times as required to display the required function.
Note that when you reach the last function, the next press will return you to the first function which
has a number in the Scroll column of the function table.
4. After the name of the function, the RC1 displays the value of the associated parameter (see table).
5. To alter the value of the parameter (where this is possible it is indicated by:
Adjustable? = Y in the
table), press the + key to increase the value, or the - key to decrease the value.
TABLE 1 - PROCESSOR UNIT SELECTION
Selection key Unit Processor
1 Unit 1 I/O Interface Board
9 Unit 9 Remote Commander (RC1)
A

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
CAL018.FM Page no. 195 Chapter 4
TABLE 2 - UNIT 1: I/O INTERFACE BOARD FUNCTION
TABLE 3 - UNIT 9: REMOTE COMMANDER FUNCTIONS
* Direct access only
Code Function Parameter Value Adjust-
able?
Scroll
00 Unit 1: Name & software rev. IP3–C N Ø
01 Date: Software rev. date month/day/year N Ø
02 Hours: Total hours of operation 0-9999 N Ø
03 Cycles: Total no. of ON/OFF cycles 0-9999 N Ø
10 Vert Size Vertical size -128 to +127 Y
® 1 ¯
11 Vert Pos Vertical position -128 to +127 Y 2
¯
12 Horiz Size Horizontal size -128 to +127 Y 3
¯
13 Horiz Pos Horizontal position -128 to +127 Y 4
¯
20 White level Video gain brightness intesity -128 to +127 Y 5
¯
21 Black Level Raster brightness intensity -128 to +127 Y 6

¯
22 Rotation Magnetic field correction -128 to +127 Y 7

¯
30 Hsync: Horizontal input frequency 50-150 kHz N 8

¯
31 Vsync: Vertical input frequency 45-255 Hz N 9

¯
40 B/W Mode: A/B brightness mode switch A or B Y ­

10
Code Function Parameter Value Adjusta-
ble?
Scroll
00 Unit 9: Name & software rev. RC1–EX N Ø
01 Date: Software rev. date month/day/year N Ø
10 Lambda: Light meter function
0-500
cd/m2 N ® 1 ¯
11 Black Display black value setting
value cd/m2 Y 2

¯
12 White Display white value setting
value cd/m2 Y ­

3
21 cal black Set black value (use +/- keys)
value
cd/m2 Y*
22 cal white Set white value (use +/- keys)
value
cd/m2 Y*
24 fl/Cd mode
Toggles units:
cd/m2
or ftLamberts
Cd or fL Y *

Chapter 4 Page no. 196 CAL018.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
6. MONITOR GEOMETRY AND LINEARITY
Notice:!
Before beginning any check or calibration, be sure that the monitor has been powered on for at
least 20 minutes. The viewing position can impact checks and adjustments; all observations must
be made while directly facing the center of the screen.
1. Connect the RC1 at the rear of the monitor as described above.
2. Display the Monitor Pattern:in the Service Calibration screen, click on the
AWS Monitor button.
Click
Start to display the Monitor Pattern; shown below with reference points for linearity and bright-
ness checks. The pattern (
BDLJ) is shown in a full screen window (AMOI).
6-1. Geometry Check and Adjustment
Check/adjust monitor geometry. Observe the full screen displayed image:
1. Measure the overall horizontal width of the displayed image (EN) with a flexible rule at the top, bottom
and center of the screen: it must be 350 mm +
3 mm (13.8” +0.12”) at all points. If necessary, use the
RC1 to adjust the width. Note the result in the table at the end of the Job Card.
2. Measure the height of the pattern (CK) with a flexible rule at the left, right, and center of the screen: it
must be 280 mm +3 mm (11” +0.12”) at all points. If necessary, use the RC1 to adjust the height.Note
the result in the table at the end of the Job Card.
3. Verify that the image is not rotated. If necessary, use the RC1 to make the pattern horizontaL Note
the result (OK, or the deviation from horizontal) in the table at the end of the Job Card.
4. Verify that the image is well centered vertically and horizontally with the monitor fascia (front cover). If
necessary, use the RC1 to adjust horizontal and vertical centering. Note the results in the table at the
end of the Job Card.
AD
IL
B
EH FG
C
JK
Ref. 4
Ref. 3
Ref. 2
Ref. 1
M
O
N

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
CAL018.FM Page no. 197 Chapter 4
6-2. Linearity Check
1. Verify screen linearity by measuring and noting the sizes of the various monitor pattern line segments
listed in the tables below, as defined in the Illustrations.
2. Calculate the difference between segments as shown in the tables below (e.g. BC - CD). For FG-GH,
and for CG-GK, the difference must not exceed +1.5 mm. For the other checks, the difference must
not exceed +3 mm.
• No linearity adjustment is possible, but note that apparent linearity errors may be caused by incorrect
geometry adjustment. Check geometry adjustment carefully if linearity errors are observed.
Horizontal linearity checks:
Vertical linearity checks:
BC CD BC - CD
FG GH FG - GH
JK KL JK - KL
BF FJ BF - FJ
CG GK CG - GK
DH HL DH -HL
AD
IL
B
EH FG
C
JK
M
O
N
CG
GK

Chapter 4 Page no. 198 CAL018.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
7. AMBIENT LIGHT ESTIMATION
Conditions for ambient light estimation should be as close as possible to normal working conditions. Ver-
ify that the cart and monitor are oriented as they will be in normal use, and verify that the ambient light is
at the level desired by the customer. If, as recommended, the room has a a light dimmer, use it to adjust
the level as required.
For more accurate measurement, a gray reference card should be used; if this is not available, the side of
the monitor rear cover may be used as a substitute, as described below.
1. Set the RC1 for use as a light meter:
a. Press # 9 on the RC1 to select the RC1 processor.
b. Enter *10 to select the light measurement function.
2. Check that the light measurement units are set to cd/m
2
(this is the ISO unit of luminance, and should
be used for all procedures in this Job Card). To change the units proceed as follows:
a. Press # 9 on the RC1 to select the RC1 processor.
b. Enter *24 for direct access to the unit select function.
c. Enter + or - to toggle the selected unit between cd/m
2
(cd displayed) and foot lamberts (fL dis-
played). Note that 1 fL = 3.4 cd/m
2
.
d. Enter *10 to return to the measurement function.
3. Position the reference surface (from which the ambient light will be measured) as closely as possible
to the normal screen posiiton:
• If the gray reference card p/n 2110360 is available, place it against (and parallel to) the monitor
screen.
• If no gray reference card is available, rotate the monitor left or right to 90
°. The side of the monitor
rear cover will be used as the reference surface.
4. Position the RC1 toward the reference card or the monitor cover, at about 300 mm distance , Make
sure that no shadow (from yourself for example) can affect the measurement; record the measured
light value ML (in cd/m
2
) shown on the RC1.
5. Calculate the ambient light level IA
(in lux): IA = p x ML/RR, where RR is the Reflection Ratio of
the surface used. The table shows RR and values for the reference card and the monitor cover.

Notice:!
1. If the ambient light level is over 50 lux, image quality cannot be guaranteed; the customer must
be warned of the problem.
2. Screen luminance cannot exceed 275 Cd/m
2
(80 fL) without loss of resolution!
6. Note the result in the table at the end of the Job Card.
7. Remove the gray reference card or rotate the monitor to its normal posiiton.
Parameter Gray reference card Monitor cover
Reflection Ratio RR 0.18 0.75
Ambient Light IA (lux) 17.5 x ML 4.2 x ML
Measured value ML for 50 lux IA
2.8
cd/m
2
12 cd/m
2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
CAL018.FM Page no. 199 Chapter 4
8. BRIGHTNESS AND CONTRAST
8-1. Select RC1 Light Measurement Units
Check that the the light measurement units are set to cd/m
2
. If necessary, change the measurement
units as described in the previous section.
8-2. Prepare room and monitor for adjustment
• The monitor must be displaying the Monitor Pattern (as described in section 6)
• The ambient light (as measured in section 7) must be at a low value (about 50 lux).
• There must be no direct or reflected light shining on the screen.
8-3. Check the displayed black and white levels
1. Set the RC1 for use as a light meter:
a. Press # 9 on the RC1 to select the RC1 processor.
b. Enter *10 to select the light measurement function.
2. Measure the brightness of the central black box of the Monitor Pattern; use the RC1, touching the
screen at the center of the box. Wait until the displayed Black Level light output value is steady.
The Black Level value B should be in the range 1.0 to 1.5 cd/m
2
.
3. Measure the brightness of the central white box of the Monitor Pattern; use the RC1, touching the
screen at the center of the box. Wait until the displayed White Level light output value is steady.
The White Level value should be in the range 165 to 175 cd/m
2
.
4. If either of these values is outside the specified range, adjust the value(s) as described below (section
8-6).
5. Note the values (after adjustment if necessary) in the table at the end of the Job Card.
8-4. Check gray level visibility
1. Check that each of the following items can be seen clearly against adjacent parts of the pattern (refer
to the illustration of the Monitor Pattern):
a. The 5% black square within the 0% black square (Ref. 1).
a. the 95% white square within the 100% white square (Ref. 2).
b. Each of the 10% percent step levels (Ref. 3; 10% through 100%).
c. Each of the step levels in the gray-scale map (Ref. 4).
2. If any of these gray levels is not clearly visible, check that the ambient light level and the displayed
black and white levels have been correctly set. Adjust the levels if necessary.
8-5. Adjust the monitor black and white levels
1. Before adjusting these levels, check that the ambient light level at the screen is below 50 lux.
2. If the value of the displayed black or white level is out of the ranges specified above, or gray level vis-
ibility is not correct, the Black and/or White level settings must be adjusted. In case of difficulty, it may
be advisable to return the settings to their nominal values before repeating checks and adjustments.
The nominal values of the Black and White Levels are: black = 40; white = 80.
3. To change Black and White Levels:
a. Press # 1 on the RC1 to select the I/O function.l
b. Press the
¿ key to access the Black Level function (or enter *21 for direct access). Enter + or - to

Chapter 4 Page no. 200 CAL018.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments adjust the level as required.
c. Press the
¿ key as often as required to access the White Level function (or enter *20 for direct
access). Enter + or - to adjust the level as required .
The following points may be helpful:
• Always adjust the black level first.
• Always check both levels, black and white, after adjustment, even if only one has been changed.
• If the displayed black level is below 1.0 cd/m
2
, increase the level. If it is above 1.5 cd/m
2
, reduce
the level.
• If the displayed white level is below 165 cd/m
2
, increase the level. If it is above 175 cd/m
2
, reduce
the level.
4. Repeat the checks and adjustments for black and white levels and gray level visibility as required.
Notice:!
Under certain site conditions (where the ambient light cannot be reduced to 50 lux), or if requested
by the customer, it may be necessary to readjust the Black Level value in the range 0.5 to 1 cd/m
2
.
Black levels above 1.5 cd/m
2
will affect image quality.
9. RESOLUTION CHECK
1. Observe the resolution patterns located in the four corners of the screen. They must all be equally
visible, sharp, and well contrasted.
2. Note your evaluation in the table at the end of the Job Card.
Note that no resolution adjustment is possible.
10. RETURN TO CALIBRATION SCREEN
When all measurements are satisfactorily completed, return to the Calibration screen by clicking the
close button at the top left corner of the Monitor Pattern screen.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments
CAL018.FM Page no. 201 Chapter 4
11. RECORD OF MONITOR ADJUSTMENTS
Use the table below to keep a record of monitor measurements and adjustments.
When recording the values shown on the RC1, be sure to note the sign (+/-).
12. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.
Adjustment Date: Date: Date: Date: Date: Date:
1 H position
2H size
3 V position
4V size
5Rotation
6 Black level
7 White level
8 Measured ambient value
9 Resolution check
(OK/Not OK)
10
Remarks

Chapter 4 Page no. 202 CAL018.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 018 - ISC Monitor Adjustments This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 019 - Light Beam Centering
CAL019.FM Page no. 203 Chapter 4
Job Card CAL 019 - Light Beam Centering Chapter 4
1. SUPPLIES
• None.
2. TOOLS
• White paper and adhesive tape.
•4 coins.
3. SAFETY PRECAUTIONS
WARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. PROCEDURE
5-1. Settings
1. Configure the collimator with following margins
ARM/VALID/INSTAL/CONFIG/COLLIM /MARG_L = -5
ARM/VALID/INSTAL/CONFIG/COLLIM /MARG_R = -5
ARM/VALID/INSTAL/CONFIG/COLLIM /MARG_F = -5
2. Go up the menu and press CKSUM twice.
3. Set technique to large focal spot, MoMo, 25kV, 25mAs.
4. Install bucky with no additional filtration, select full FOV.
5. Use adhesive tape to attach white paper on the bucky, covering the full FOV.
6. Press the light beam button and place a coin tangential to each internal border of the light beam:
Bucky
Light beam borders
Coin
White paper

Chapter 4 Page no. 204 CAL019.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 019 - Light Beam Centering 5-2. Mechanical adjustment of light position
1. Open the work list and declare a new Patient called "Light Beam centering".
2. Take an exposure, select zoom = fit to screen;
3. Check that the 4 coins are tangential to X-ray field of view;
4. If not, move the light FOV as described in the following sections (5-3 and 5-4). Use the software
adjustment procedure for fine adjustment when coincidence has been almost achieved using the
mechanical procedure.
5. Correct the position of the coins for the new light beam field of view.
6. Repeat the procedure until coincidence is reached between the light beam and the X-ray beam.
7. Reconfigure the collimator to margins 0, 0, 0.
8. Go up the menu and press CKSUM twice.
9. Delete the previously created Patient called "Light Beam centering".
5-2-1. Mechanical Adjustment of Lamp
Refer to the illustration below.
1. To move in Rear-Front direction, use the Mirror adjusting screw (1).
2. To move in Lateral direction, use the Light Box screw (2).
TUBE HOUSING ARM
MIRROR FRAME
MIRROR MOUNTING SC REWS
AND LOCKNUTS
MIRROR ANGLE ADJUS TMENT SCREW
AND SPRING
(2)
(1)

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 019 - Light Beam Centering
CAL019.FM Page no. 205 Chapter 4
5-3. Software adjustment of blade position
1. For a finer adjustment of front blade:
- Select SETUP/ARM/VALID/INSTAL/CALIB/FORMAT/CALIB/LIGHT/ALIGN.
- Adjust the front blade position by rotating the console kV knob as described in the Configuration
Job Card
The Control Console.
2. Fora finer adjustment of lateral and rear blades:
- Select SETUP/ARM/VALID/INSTAL/CALIB/FORMAT/CALIB/LIGHT/COVERAGE.
- Adjust lateral and rear blades by rotating the console kV knob as described in the Configuration
Job Card
The Control Console.
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

Chapter 4 Page no. 206 CAL019.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 019 - Light Beam Centering This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 020 - mAs Non-linearity Calibration (Automatic)
cal020.fm Page no. 207 Chapter 4
Job Card CAL 020 - mAs Non-linearity Calibration (Automatic)Chapter 4
1. SUPPLIES
• None
2. TOOLS
• Flat Field Phantom (25 mm Lucite).
3. SAFETY PRECAUTIONSWARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• If not already done, carry out any procedure(s) specified as Prerequisites for this Job Card in Job
Card
Calibration Interactions.
5. PROCEDURE
5-1. Presentation
The object of mAs non-linearity calibration is to compensate for the non-linearity of exposure delivered
by the Generator at low mAs values, as used in AOP pre-exposure operations.
This procedure measures pre-exposure linearity deviation, and stores a correction factor in the DMR
menu, to be applied to the mAs value of all AOP exposures.
The correction factor is automatically calculated as:
MASPRE CORR = S
100
/ (25 x S
4)
, where S
100
and
S
4
are the signals measured in the 100 mAs image and the 4 mAs image respectively.
5-2. Procedure
1. Remove the compression paddle and install the Flat Field Phantom on the Bucky.
2. From the DMR console menu, enter:
SET UP/GENE/INSTAL/GENE/AOP/APOLLO/CALIB/mAsPRE/CALIB.
3. Press CALIB. The 9x9 collimator format is automatically selected.
4. Press the DMR console acquisition button (not PREP button) and hold it down until the system has
performed two acquisitions (about 20 seconds). Parameters for the two exposures are automatically
set to MoMo/28 kV/4 mAs and MoMo/28 kV/100 mAs.
5. Press SETUP once, then press PARAM.
6. Verify that the value displayed for MASPRE CORR is within the limits given below (
Acceptable val-
ues
).
7. Perform a checksum from DMR console: GENE/INSTALL/CKSUM.
5-3. Acceptable Values
• The value displayed for MASPRE CORR should be within the limits 1.00 and 1.20.
• If it is outside this range, check the Generator calibration and repeat the calibration.

Chapter 4 Page no. 208 cal020.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card CAL 020 - mAs Non-linearity Calibration (Automatic)
6. COMPLETION
• Carry out any procedure(s) specified as Post Completion Requirements for this Job Card in Job Card
Calibration Interactions.
• If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
VF.FM Page no. 209
CHAPTER 5 FUNCTIONAL CHECKS
This chapter contains information for verification of Senographe 2000 D functions.

Page no. 210 VF.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
VF001.FM Page no. 211 Chapter 5
Job Card VF 001 - Boot/Init and Shutdown Chapter 5
1. SUPPLIES
None.
2. TOOLS
None.
3. SAFETY PRECAUTIONS.
•To remove power from the complete system, switch off the external circuit breaker which con-
nects the system to the hospital supply. There are two circuit breakers inside the Generator Cabi-
net, one for the Gantry and Generator, the other for the Cart and imaging equipment. Switching off
one of these breakers removes power from only part of the system.
• Remember that the UPS will continue to supply power to parts of the Cart and imaging equipment
until it is shut down.
• NEVER switch off the UPS manually (risk of data loss).
• NEVER switch off power to the Conditioner except in emergency (risk of damage to the Detector if
the correct conditions of temperature and humidity are not maintained).
WARNING
In normal operation (circuit breakers closed):
Power is present in the Generator Cabinet, MDR, and AWS Cart at all times, even after the
AWS, UPS, and other components have shut down. Cart components remain powered by
the UPS for about two minutes after switch off. The 230 V supply to the Conditioner the
UPS, and the Light Box is maintained at all times.
4. PREREQUISITES
None.

Chapter 5 Page no. 212 VF001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
5. PROCEDURE
This Job Card provides check lists for confirmation of the correct operation of power up and power down
sequences, and for the steady state conditions to be expected with power on or off.
• Refer to the Power Supply Schematic diagram overleaf for an overview of power connections.
• In normal operation, system power switches are used as follows:
Power comes from the hospital supply circuit breaker to the Generator Circuit Breaker and to a circuit
breaker in the MDR (the Main Distribution Rack; at the top of the Generator Cabinet), in parallel.
Both circuit breakers are normally left ON.
All subsystem power switches (items in the Cart, Conditioner, etc.) are normally left ON.
The system is normally switched ON using the ON switch (I) on the Gantry Console.
The system is normally switched OFF by clicking the System Off button in the AWS Browser
window. When switching off, wait for complete system and UPS shut down (about two minutes).
The Gantry and Generator can be switched OFF without affecting components in the AWS cart by
pressing the OFF switch (STOP) on the Gantry Console.

MDR Breaker
ON/OFF
Conditioner
Detector supply
Cart
UPS
Generator
Breaker
pushbuttons
:
Conditioner
and
MDR
Generator

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
VF001.FM Page no. 213 Chapter 5
ILLUSTRATION 1 - POWER SUPPLY SCHEMATIC
Generator
Conditioner
Gantry
UPS
IDC
AWS
UPS off
MDR (Main
Distribution Rack)
Detector
Power Supply
Detector
Conditioner, Detector,
and Cart on/off
Control
console
AWS monitor
CD writerModem
Operator
FE
Operator
Hospital Supply
circuit breaker
a.c. mains
ON/ OFF
All system power
OFF
breaker
breaker
Light Box
Emergency
Stop
UPS on
HV & LV
System ON;
Gantry & Gene. OFF
Conditioner
ON/ OFF
Detector PS
ON/OFF
AWS Cart
ON/OFF
UPS
ON
Legend:Control signals
Manual operations
Power supplies
Detector 0n/
0ff
Detector output and control
Conditioner state (chiller, dryer, on/off)
Gantry
AWS Cart
Generator Cabinet
Gene. off
HV & LV
Gantry/Gene. off
Login;
System OFF

Chapter 5 Page no. 214 VF001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown 6. POWER OFF STATE CHECKS
Remove Generator Cabinet covers as required for access to the main circuit breaker in the MDR, the
Generator circuit breaker, the Conditioner, and to the pushbuttons controlling power to the Conditioner
and the Detector.
6-1. Normal power OFF state checks
For these power OFF checks, the MDR and Generator circuit breakers should be
closed (ON), and the system should have been switched off using the normal
power off sequence (by clicking the System Off button in the AWS Browser window). Check:
• No display on the Senographe Console and STOP pushbutton is lit.
• No display on the AWS monitor and no monitor LED lit.
• None of the UPS front panel LEDs are lit (visible through the plexiglass
cover at the base of the Cart).
• The Cart and Conditioner pushbuttons ON (inside the MDR) are lit.
• The Conditioner is powered up and operating normally.
6-2. Complete power OFF state checks
For these checks, all items, including the UPS and conditioner, must first be switched off. Starting from
the normal power off state as described above, proceed as follows:
1. Switch the MDR circuit breaker OFF
2. Switch the Generator circuit breaker OFF
Remember that mains power is still present in the cabinet, at the input to the breakers.
3. Check the following:
• No LED lit in the Generator Cabinet
• No LED lit in the MDR.
• None of the internal and external pushbuttons in/on the MDR are lit.
• The Conditioner is not running (no LCD display, no sound from the pump)
NOT OKOK

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
VF001.FM Page no. 215 Chapter 5
7. POWER UP SEQUENCE CHECKS
Remove Generator Cabinet covers as required for access to the main circuit breaker in the MDR, the
Generator circuit breaker, the Conditioner, and to the pushbuttons controlling power to the Conditioner
and the Detector.
7-1. Preliminary Power-Up sequence check
Start this sequence with the MDR and Generator circuit breakers switched OFF (as described for the
Complete power OFF state checks above).
1. Switch the Generator circuit breaker ON.
2. Switch the MDR circuit breaker ON.
3. Press the CART ON button on the Generator Cabinet; The button lights and
power is suplied to the Cart; check by switching the Light Box on.
4. Press the CONDITIONER ON pusbbutton at the top of the MDR; the button
lights and the Conditioner starts.
5. Press the UPS ON pushbutton on the Generator Cabinet for at least one sec-
ond; the AWS starts up (see VF Job Card
UPS Battery if you suspect a prob-
lem with the UPS).
6. The
Login prompt appears on the AWS monitor within 90 seconds of switch-
on. Enter sdc<Enter>. At the
Password prompt, enter adw3.1<Enter>. The
browser application begins to load.
Notice:!
At this point of the sequence, communication errors bertween AWS and the DMR will appear be-
cause the DMR is not yet switched ON. This is normal and does not imply any problem.
7. Press the GANTRY ON pushbutton in the MDR; the button lights.
8. Press the GANTRY OFF pushbutton in the MDR; the light goes out.
9. Shut the system down by clicking the System Off button on the Browser window; wait for com-
plete shut down.
The system is now in the normal switched-off state, ready for the normal power up sequence (below).
NOT OKOK

Chapter 5 Page no. 216 VF001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
7-2. Normal power UP sequence check
Start this sequence with the system in the normal switched-off state (above).
1. Press the power ON pushbutton I on the Senographe Console. Power is
applied to all parts of the system:
2. .The Gantry Console central LCD panel displays default settings for kV and
mAs, and displays a message indicationg temporary communications errors.
3. The AWS Monitor LED lights. On the UPS the orange LEDs flash during the
selftest process, then the On-line LED (green) is lit. No red LED should be
lit and no LEDs should continue flashing.
4. The
Login prompt appears on the AWS monitor within 90 seconds of switch-
on. Log in as described above.
5. The popup message
All devices are OK appears on the AWS monitor.
6. The Gantry Console displays the two following messages:
S70 Wrong applica-
tion selected
and S34 Select Laterality; these messages are normal, because
an exam has not been started and no laterality has been selected.
7. The Browser screen is displayed on the AWS monitor
. Within about 5 minutes
of switching on, it should be possible to use the
Start Exam button in the
Worklist window (see Operator Manual).
8. POWER ON STATE CHECKS
These checks confirm normal operation after the normal power up sequence (above);
1. The Browser screen is displayed on the monitor.
2. None of the UPS front panel LEDs are flashing; the On–line LED (green)
is lit.
3. The Conditioner is running (pump audible and no conditioner error messages
displayed).
NOT OKOK

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
VF001.FM Page no. 217 Chapter 5
9. NORMAL POWER OFF SEQUENCE CHECKS
1. Click the System Off button in the AWS Browser window.
2. A message appears, requesting confirmation of the shutdown request. con-
firm by clicking OK.
3. After a few minutes, the Console display goes blank.
4. The Cart and AWS system perform the shut down sequence, which takes
approximately two minutes. The normal shut-down message is displayed on
the monitor:
System is down. System will power off within 30 seconds; do NOT reboot now! (if system does not
turn off, switch off UPS manually
.
If at any time the Ups does not shut down after the System Off request, the UPS should be turned off
by pressing the CART OFF button at the front of the Generator Cabinet.
5. Within about two minutes of confirmation of the shutdown request, the Cart
system is switched off and the Gantry/Generator system is switched off (check
again at the Gantry Control Console).
6. No LED lit or flashing on the UPS.
7. Check that the conditioner continues to operate; check that the
CART ON but-
ton (at the front of the Generator Cabinet), and the
CONDITIONER ON button
(inside the MDR) are still lit.
10. THERMAL CUTOUTS
Two thermal cutouts are fitted in the Generator Cabinet, one in the autotransformer, and the other on the
200PL1 board.
If either of these cutouts operates, the effect is the same as if the Gantry Emergency Stop button had
been pressed, except that it is impossible to restart the system; the Console ON pushbutton has no
effect. The system will not start until it has cooled and the cutout has recovered.
• Autotransformer cutout. Operates at 120°C; the OFF button on the Main Supply Rack remains lit, but
it is impossible to restart the system. Recovery time is one to two hours.
• 200PL1 board cutout. Operates at 110°C; the OFF button on the Main Supply Rack is not lit; it is
impossible to restart the system. Recovery time is about half an hour.
NOT OKOK

Chapter 5 Page no. 218 VF001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 001 - Boot/Init and Shutdown
11. SYSTEM COMPONENT RESET
It may occasionally be necessary to reset one or more components of the system. This usually involves
switching the component(s) off and on again. The following table shows suggested switch-off points,
classified for use by operating personnel or by F.E.s (i.e., trained personnel with access to system cabi-
nets).
TABLE 1 - RESET OPERATIONS
CAUTION
It is recommended that the IDC should not be booted or reset by itself without a complete
system shutdown/power on.
If the IDC is rebooted, a minimum interval of two (2) minutes must be left before booting it
again. Booting the IDC twice within two minutes may seriously degrade the detector.
Component(s) Operator F.E. Comments
Whole system Hospital supply ON/OFF
switch
Hospital supply ON/OFF
switch
Conditioner Hospital supply ON/OFF
switch
Conditioner OFF button
inside MDR
Gantry and Generator OFF/ON at Gantry con-
sole
Generator Circuit breaker
in Generator Cabinet
Also resets the Detector
and its power supply
Detector and its power
supply
OFF/ON at Gantry con-
sole
Detector OFF button
inside MDR
OFF/ON at Gantry con-
sole also resets the Gan-
try and Generator
Cart and all contents
(AWS system, monitor,
IDC, modem, CD-R/W,
light box)
Click on System Off but-
ton in Browser. Wait two
minutes for UPS shut-
down
Cart OFF button on side
of Generator cabinet
System Shutdown
includes shutdown of
Gantry and Generator.
UPS Click on System Off but-
ton in Browser. Wait two
minutes for UPS shut-
down
Cart OFF button on side
of Generator cabinet
System Shutdown
includes shutdown of
Gantry and Generator.
IDC Click on System Off but-
ton in Browser. Wait two
minutes for UPS shut-
down
IDC reset button at rear
of the IDC.
See Caution below.
System Shutdown
includes shutdown of
Gantry and Generator.
AWS Click on System Off but-
ton in Toolkit menu.
Reboot
Open a command win-
dow; type halt. Then type
reset.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 002 - Cart Check
VF002.FM Page no. 219 Chapter 5
Job Card VF 002 - Cart Check Chapter 5
1. SUPPLIES
None
2. TOOLS
None.
3. SAFETY PRECAUTIONS.
None.
4. PREREQUISITES
None.
5. PROCEDURE
Inspect the FFDM Cart for any signs of damage or impairment. In particular, check the following points:
1. Wheels and brakes. Check that the Cart can be easily moved, with no indication of noise or binding
on any of the wheels. Check that the brakes on the front wheels can be easily operated, and that they
prevent movement of the Cart when operated.
2. Centre table. Check that the centre table (keyboard support) can be easily raised from the normal
horizontal position to +60
°, and locks correctly in both positions.
3. Side tables. Check that each of the side tables (mouse pad and light box) can be easily lowered from
the normal horizontal position to -90
°, and locks correctly in the horizontal position.
4. Sliding doors. Check that all sliding access panels open and close smoothly.
5. Check the operation of the light box tube.

Chapter 5 Page no. 220 VF002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 002 - Cart Check This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 003 - UPS Batteries
VF003.FM Page no. 221 Chapter 5
Job Card VF 003 - UPS Batteries Chapter 5
1. SUPPLIES
None
2. TOOLS
None
3. SAFETY PRECAUTIONS.
.None
4. PREREQUISITES
None
5. UPS FUNCTION CHECK
The LEDs and control buttons of the UPS front panel may be observed and accessed through the plexi-
glass panel at the front of the AWS Cart.
Notice:!
If you intend to re-calibrate the battery capacity indicator (procedure described below), you should
do this before the function check, to avoid an additional delay while the battery recharges.
Proceed as follows to carry out the function check:
1. With the system powered on as normal, observe the state of the front panel LEDs:
a. The On Line LED (B) should be lit.
b. If any of the remaining five status LEDs (A, C, D, E, F, G) are lit, the cause should be investigated.
c. The Load Indicator LEDs indicate the power being drawn from the UPS as a percentage of the
rated capacity. Normally, four LEDs or less should be lit.
d. The Charge/Line Voltage LEDs indicate the current state of battery charge. With all five LEDs lit,
the battery is fully charged. When the lowest LED flashes, the battery can supply less than two
minutes of run time; you must wait for the battery to recharge before using the system.
A
B
C
D
E
F
ON/TEST BUTTON
LOAD INDICATOR
OFF BUTTON
CHARGE/LINE VOLTAGE
INDICATOR (5 LEDS)
A - SMART TRIM (INPUT HIGH)
B - ON LINE
D - OVERLOAD
E - ON BATTERY
F - REPLACE BATTERY
STATUS LEDS:
C - SMART BOOST (INPUT LOW)
(5 LEDS)

Chapter 5 Page no. 222 VF003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 003 - UPS Batteries 2. Start a UPS self test; press and hold the ON/TEST button until a beep is heard and the On Line LED
(B) flashes.
During the test the UPS briefly supplies the load from the battery; the On Battery LED (E) is lit. If
the test passes, the On Battery LED goes out and the On Line LED (B) again lights steadily.
If the test fails, the UPS returns to on-line status, but the Replace Battery LED (F) is lit. The
cause should be investigated.
3. To check the input voltage (hospital supply), press and hold the ON/TEST button for at least four sec-
onds (this will start a self test as above):
The Charge/Line Voltage LEDs now indicate the input voltage status; normally two, three, or four
LEDs should light.
If all five LEDs are lit, the input is more than 247 V; if one LED is lit or no LEDs are lit, the input is less
than 213 V.
6. BATTERY CAPACITY RE-CALIBRATION
About once a year, it may be desirable to re-calibrate the battery capacity indicator.
The procedure, including battery recharge, takes about one hour
Notice:!
It may not be possible to start this procedure immediately after the function check described above,
because it requires the battery charge to be 100% of capacity.
.Proceed as follows:
1. From the Service Desktop, click the
Diagnostics button, then the UPS button.
2. To prevent the system from being shut down during the calibration process, select
Configuration/
Event Actions, and then the UPS Battery is Discharged event; remove the check mark from the Shut-
down Server
box.
3. Note that the calibration process will not start if the battery charge is less than 100%. To display
Bat-
tery Capacity
(in place of Run Time Remaining), select Configuration/Monitoring Preferences, and
select the parameter to be displayed in the left hand graphical bar of the PowerChute interface.
4. Start the calibration process by selecting
Diagnostics/Initiate run Time Calibration.
5. Once started, the calibration process continues until battery capacity falls to 25% (the time from
100% to 25% is about 10 minutes).
With the AWS system powered off and power applied to the UPS, the battery takes 45-60 minutes to
recharge.
6. Before returning the system to normal operation, you must ensure that it will be shut down if battery
capacity is low:
- EITHER: In the PowerChute interface menu, select
Configuration/Event Actions, and select the
UPS Battery is Discharged event; click the Shutdown Server box to restore the check mark.
- OR: Log in as root in a command window; execute the command ups stop, then ups start.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 005 - Senographe Functional Checks
VF005.FM Page no. 223 Chapter 5
Job Card VF 005 - Senographe Functional Checks Chapter 5
1. SUPPLIES
None
2. TOOLS
- Digital mAs meter (GSPONER type) able to display tube current
- Compression measurement tool: GAMMEX RMI model 817 or Mammo compression scale # 46-
194427P407 (digital display)
- Plexiglass in thickness increments of 0.5 cm
3. SAFETY PRECAUTIONS
These procedures produce X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
None
5. GENERATOR TESTING
The objective is to check quickly the generator kV, mA and mAs calibrations, which have already been
done in the factory.
5-1. Set up generator for testing.
1. Configure the DMR for the following exposure:
- Mode manual (2-point)
- Focal spot: large
-Track: Mo
-Filter: Mo
-mAs: 60 (lower mAs value may give inaccurate measurements)
-HV: 22 kV
2. Activate display of parameters following an exposure by selecting SETUP/GENE/INSTAL/GENE/
CONFIG/DISP/YES on the console (see Job Card CFG 001 ). Return to application mode.
3. Connect the digital mAs meter to kV mA board 701-PL1 (on HV tank unit 701) between terminals
MES_MA_POS and MES_MA_NEG. Disconnect the jumper that is normally between these same
two terminals.
4. Set the mAs meter to current (mA) measurement.
5. Protect the detector with 2.5 cm of plexiglass.
6. Open
New Patient in the Browser, and take 5 exposures at approximately 20 second intervals, incre-
menting the HV by 2 kV each time (i.e., 22 - 24 - 26 - 28 - 30 kV).
7. Take 5 more exposures at the same interval, leaving the HV set at 30 kV.
8. The X-ray tube current value displayed on the console (I
display
) must be 100 mA ± 5 mA

AND should
agree with the value read on the mAs meter (Imeasure) ± 5%.
If I
display
is not equal to I
measure
then perform Calibration Job Card X-ray Tube mA Measurement, fol-
lowed by Job Card
X-ray Tube Heater Current.

Chapter 5 Page no. 224 VF005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 005 - Senographe Functional Checks 9. Take 2 exposures at 40 kV and 50 mAs at an interval of 30 seconds. No arcing should be detected. If
arcing is detected, perform the procedure given in the IST Job Card
Automatic X-ray Tube Warm-up.
10. Reconfigure the
Senographe DMR to 30 kV and 25 mAs. Take one exposure for each of the three
following focal spot/track combinations and verify the expected X-ray tube current reading on the con-
sole display:
SMALL FOCAL SPOT/MO TRACK => I = 40 mA ± 2 mA
SMALL FOCAL SPOT/RH TRACK => I = 40 mA ± 2 mA
LARGE FOCAL SPOT/RH TRACK => I = 75 mA ± 4 mA
11. Disconnect the mAs meter from the
DMR and reconnect the jumper on kV mA board 701-PL1.
12. Verify kVp using a non-invasive kVp meter and, if needed, perform Job Card CAL 005 (Calibration of
kV Scale Factor). This step may be mandatory, when required by local regulations.
6. GANTRY TESTING
The objective is to quickly check the basic functions of the gantry.
Note that the following functions are tested by the software during each power-up sequence:
- Filter holder rotation
- Lateral, front and rear collimator blade movement
- X-ray tube housing movement
- Compression paddle movement to home position
- Column movement limit stops
- Coherency between focal spot size, magnification and X-ray tube housing angulation
- Grid movement
- Generator-gantry communications link
- Gantry CPU board
6-1. Check gantry readout.
Immediately following the power-up sequence, the gantry readout should indicate the X-ray format.
6-2. Check column movement.
1. With the compression paddle out of compression, check that all four sets of column up/down control
buttons are operational (check each button for up and down movement of column).
2. Move column to its upper and lower limits. Optical sensors should stop movement. If optical sensors
have failed, mechanical safety stops halt movement. Total one-way movement time (going either up
or down) between upper and lower limits should be 16 ±2 seconds. This time includes initial acceler-
ation, which lasts 3 seconds.
6-3. Check arm rotation.
1. With the compression paddle out of compression, check that all four arm rotation control buttons are
operational (check each button for release and re-locking of arm rotation).
2. Rotate arm through its full range of ±180°. Check in several positions that electro-brake is sufficient,
by manual action on the handles.
3. Check alignment of angle markings at 0°.
6-4. Check light centering device.
Press one of the light centering device activation buttons. The light centering device should light for 30

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 005 - Senographe Functional Checks
VF005.FM Page no. 225 Chapter 5
seconds.
6-5. Check format control.
Check that repeated pressing of each of the two format control pushbuttons, gives the sequence:
1) light centering device lights; 2) first format reduction; 3) second format reduction; 4) full format.
6-6. Check magnification sensors.
1. With magnification device removed, the gantry readout should show a magnification of 1.0.
2. Install the magnification device into the lower, middle and upper holes successively. The gantry read-
out should show a magnification of 1.5, 1.8, and 1.9, respectively.
6-7. Check tube housing arm detents.
1. Activate the tube housing arm locking lever and rotate the tube housing arm to its ± 15° and ± 27°
detents. When tube housing arm is locked in each of these four positions, the error message
Incor-
rect X_Ray Incidence
should appear on the console display.
2. Rotate the tube housing arm so that it is "floating" between detents. After five seconds, the error mes-
sage
Tube Housing Arm Unlocked should appear on the console display and Arm Unlocked should
appear on the gantry readout.
6-8. Check paddle compression.
1. Verify that both sets of compression/decompression pedals and the manual compression knobs
located on the paddle holder all function correctly (i.e., the paddle moves up or down in response to
each command).
2. Use the mammo compression scale placed between the cassette holder and the compression paddle
to check calibration of the compression force display on the gantry readout. Measure one point at 5
daN and one point at 15 daN. The value on the gantry readout must agree within ±1 daN. If it does
not, perform Job Card CAL 006 (Calibration of Compression Force Detector).
Note:
If the recommended compression measurement tool is not available, this verification can be done
by lowering the paddle close to the top surface of the cassette holder, installing the magnification
device in the upper pair of holes, and connecting a spring dynamometer (calibrated from 0 to
15 kg) between the compression paddle and the magnification device.
6-9. Check breast thickness measurement.
1. Install the bucky and place 10 mm of plexiglass in the center of the bucky.
2. Lower the compression paddle into contact with the plexiglass.
3. Slowly increase compression force on the plexiglass while watching the breast thickness display on
the gantry readout. The displayed value should remain in the range 10 mm ±3 mm for any value of
compression force (as displayed on the gantry readout) between 5 and 15 kg.
4. Replace 10 mm of plexiglass with 50 mm of plexiglass and repeat the above check. Now the dis-
played value should remain in the range 50 mm ±3 mm for any value of compression force (as dis-
played on the gantry readout) between 5 and 15 kg.
5. If either or both of the above checks does not give a satisfactory result, perform Job Card CAL 007
(Calibration of Breast Thickness Measurement).

Chapter 5 Page no. 226 VF005.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 005 - Senographe Functional Checks
6-10. Check grid presence sensor.
•Select ARM/VALID/INSTAL/CALIB/GRID/CALIB on the console.
- With the bucky present, the reponse should be CALIBRATION OK .
- If the bucky is not present or is not recognized, the response is GRID ABSENT.
6-11. Check emergency stop buttons.
Pressing either of the emergency stop buttons (located on either side of the gantry column base) should
result in a complete power-down of the generator and gantry (but not the AWS Cart).

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 008 – IQ Tools Flat Field Test
Vf008.fm Page no. 227 Chapter 5
Job Card VF 008 – IQ Tools Flat Field Test Chapter 5
1. SUPPLIES
•None
2. TOOLS
• Flat field phantom (25 mm lucite).
• ACR standard phantom (16 cells Nuclear Associates 18-220 or RMI-156).
3. SAFETY PRECAUTIONS.
• Take all normal precautions when operating with X–rays.
4. PREREQUISITES
• The system must have been correctly installed and calibrated.
5. FLATFIELD TESTS
5-1. Objective
The flatfield test measures the ability of the detector to reproduce the image of a flat and homogeneous
phantom.
The ideal response of the detector is a uniform image with no large or small defects, with no pixel having
a signal much higher or much weaker than its neighbours, and with a uniform signal-to-noise ratio. The
quantities to look for are thus:
-Brightness uniformity: the magnitude of low frequency structures (large shapes).
-High frequency modulation: the magnitude of high frequency structures (small defects).
-Number and density of bad pixels: the number of pixels having signals significantly different
from those of their neighboors (new bad pixels appeared since the latest bad pixel calibration),
and the density of these bad pixels.
-Signal-to-noise ratio uniformity: the variation of signal-to-noise ratio over the image.
A test failure indicates incorrect test conditions (phantom position, cleanliness), or an artifact causing
inconsistency between the current image and the associated gain map. Such an inconsistency means
that either the gain calibration needs to be updated, or the current image contains a defect which must
be identified (see section 7). Possible sources of artifacts include the tube output window, the mirror, the
Molybdnenum filter, the collimator blades, and the detector.

Chapter 5 Page no. 228 Vf008.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 008 – IQ Tools Flat Field Test
5-2. Method
The flatfield image is analyzed by measuring the mean
signal level, the noise, and the number of new bad pix-
els in a moving analysis box (a 20 mm x 20 mm region
of interest scanning the flatfield image with a 10 mm
overlap). The total number of new bad pixels is
summed over the whole image.
The following checks are performed:
-Brightness uniformity (BNU): compute maximum deviation of the mean signals measured in
each ROI.
-High frequency modulation (HFM): measure the minimum signal–to–noise from all ROIs
-New bad pixels: detect and count all bad pixels in the whole image
-New bad ROIs: count number of regions of interest containing more than 2 bad pixels
-Signal–to–noise ratio (SNR) uniformity: compute maximum deviation of the SNR measured in
each ROI.
6. PROCEDURE
1. From the Browser, click on the QAP button, then select Flat Field Test.
2. Follow the on–screen instructions. You are asked to:
e. Remove the bucky or breast holder and the compression paddle, and to place the flatfield phan-
tom on the image receptor.
f. Verify using the light beam that the collimator is in full field of view.
g. Verify that the tube arm angle is 0 degrees.
h. Make two exposures using the parameters set automatically (Large focal spot; Mo/Mo; 26 kV; 200
mAs).
3. Use the middle-button mouse function to adjust the image for optimum viewing conditions. Sug-
gested starting values are WC = 3000 and WL =4500.
4. The system calculates and displays test results in a format similar to that shown below. The LSL and
USL columns show the minimum acceptable values (Lower Specification Limit) and the maximum
acceptable values (Upper Specification Limit) for each test.
Test Measurement LSL USL Pass/Fail
Brightness Uniformity n/a 10.00
High Frequency Modulation n/a 0.80
Bad Pixel n/a 100.0
Bad ROI n/a 0.00
SNR n/a 40.00
Measurements made in
a moving analysis box

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 008 – IQ Tools Flat Field Test
Vf008.fm Page no. 229 Chapter 5
7. CORRECTIVE ACTIONS IN CASE OF FAILURE
7-1. Brightness Uniformity test fails
Typical causes are:
1.
The flatfield phantom does not cover the full FOV;
check the position of the flatfield phantom. The flat-
field phantom must exceed the detector field of view
in ever
y direction.
2.
A collimator blade is within the FOV; make an ac
qui-
sition in the Medical application in the same condi-
tions as the flatfield test (flatfield phantom, MoMo,
26 kV, 200 mAs) and verify in the raw image that the
X-ray beam field covers the detector field of view by
adjusting the contrast such that window width is less
than 200. If collimator blades are visible at the bor-
ders, check tube tilt, collimator centerin
g, parallelism
and format calibrations
3.
The gain map is no longer valid; calibrate gain in the MoMo/no grid configuration, move the phantom
slightly and run the test again.
7-2. High Frequency Modulation test fails
Typical causes are:
1. The detector temperature is not stabilized; check that the
s
ystem has not been powered on too recently.
2. The gain map is no longer valid; calibrate gain in the
MoMo/no grid configuration, and run the test again.
Brightness
non-uniformity:
Significant signal
variation detected
in the image
High frequency
modulation:
High noise detected
within a ROI

Chapter 5 Page no. 230 Vf008.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 008 – IQ Tools Flat Field Test
7-3. BadPixels or Bad ROI test fails
7-3-1. Typical causes are:
1.
Dust is present on the flatfield phantom or the
ima
ge receptor; ensure that the phantom and the
image receptor are completely clean and run the
test again.
2. The flatfield phantom contains inclusions or is
scratched;
- Run the flatfield test, select bad pixel graphi-
cal window, locate the bad ROIs or bad pixels.
- Move the phantom, run the flatfield test again
and note whether the bad ROIs or pixels have
moved.
If you have difficulty in observing a moving
spot, repeat the test in application mode. After making an exposure, adjust the viewing contrast to
make the spot visible. Then move the phantom slightly, make a second exposure, and switch the
display from one image to the other to observe any movement in the spot of interest.
If the bad pixels move, there are some inclusions inside the phantom. Obtain a new phantom and
run the test again.
3. The collimator is visible in the field of view;
- Run the flatfield test, select bad pixel graphical
window, locate the bad ROIs or bad pixels.
Check that they are not concentrated on the bor-
ders of the image. If they are, check tube tilt, cen-
tering, parallelism and format calibrations.
Collimator visible in
field of view

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 008 – IQ Tools Flat Field Test
Vf008.fm Page no. 231 Chapter 5
4. Detector response is unequal at some rows, columns,
or groups of pixels;

- Check that there is no visible bad row or bad column
in the flatfield image displayed in the IQ tool viewer
(use the mouse middle button to adjust contrast and
brightness such that window width WW is less than
200). If necessary, take an image in the Medical
Application and use the AWS display tools.
- If there is a distinct bad row or a bad column
(exactly parallel to the detector matrix), or a distinct
cluster, calibrate bad pixels.
If the bad row, bad column, or or cluster is detected
in the bad pixel calibration and is above the Upper Specification Limit, the detector must be
changed. If not, calibrate gain in the MoMo/no grid configuration, and run the test again.
If gain calibration succesfully corrects bad rows or columns, use the ACR phantom to check the
correction. Place the ACR phantom so as to partially cover the suspect rows or columns (e.g., to
the left or top of the detector) and make an acquisition at MoMo/26 kV/100 mAs/no grid, so that
the detector is saturated outside the area of the phantom. Verify in the raw image that the rows or
columns do not appear again in the phantom area. If they do appear, contact your OLC; it is likely
that the detector must be changed.
7-3-2. Other possible causes are :
1.
Durin
g the last gain calibration, dust was present (on the flatfield phantom, image receptor, mirror, or
filter ), or the flatfield phantom contained inclusions or scratches; verify that the bad pixels do not
move with the phantom and that they persist after careful cleaning of phantom, image receptor, mir-
ror, and Mo filter. Calibrate gain in the MoMo/no grid configuration, move the phantom slightly, and
run the test again.
2.
Dust is present on the light beam alignment mirror or on the Mo filter; clean the mirror and Mo filter
without direct contact and run the test again.
3.
Metallic particles have been deposited on the tube output window; tilt and move the tube and verify in
the image or in the bad pixel graphical window that the bad pixels have moved after the tube move-
ment. If you are sure that particles are present on the tube output window, change the tube and per-
form the resulting calibrations. before repeating the test.
Visible bad row or
bad column

Chapter 5 Page no. 232 Vf008.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 008 – IQ Tools Flat Field Test 7-4. Signal–to–Noise Ratio uniformity fails
Typical causes are:
1.
The X-ray beam is badly centered; carry out the
CAL Job Cards for Tube Tilt Calibration, X–ray
Beam Alignment Calibration, and Collimator For-
mat Calibration.
2.
The detector sensitivity is not uniform; locate in the
graphical window the area having a poor SNR (low
mean signal or high standard deviation). Calibrate
gain in the MoMo/no grid configuration, and run the
test again. If the SNR uniformity result is still above
the Upper Specification Limit, call your OLC for a
possible detector replacement.
7-5. Final checks
Notice:!
If a new gain calibration was done, you are recommended to recalibrate all other gain configura-
tions.
After investigating and correcting the suggested causes of error, repeat the test.
If the test still fails, call your OLC.
Significant
signal-to-noise variation
detected in the image

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 009 - Image Acquisition and ACR Score
VF009.FM Page no. 233 Chapter 5
Job Card VF 009 - Image Acquisition and ACR Score Chapter 5
1. SUPPLIES
None
2. TOOLS
• ACR standard phantom (16 cells Nuclear Associates 18-220 or RMI-156).
• Standard toolkit with metric wrenches.
3. SAFETY PRECAUTIONS.
• Take all normal precautions when operating with X–rays.
4. PREREQUISITES
• The system must have been correctly installed and calibrated.
5. TEST DESCRIPTION
5-1. ACR Phantom
The phantom used in this procedure is a block of plexiglass containing 16 cells. Each cell simulates an
anatomical structure which may be found in breast tissue.
The plexiglass gives attenuation equivalent to a breast of average density compressed to a thickness of
45 mm.
Starting from the top left hand corner, as shown in the illustration,
the cells are of three types:
1. Fibres.
Six cells represent fibrous calcifications; they contain fibers
with sections from 1.56 mm to 0.40 mm.
2. Specks (Calcifications).
Five cells represent groups of specks or micro–calcifications,
with sizes from 0.54 mm to 0.16 mm.
3. Masses.
Five cells represent tumors or masses, with sizes from 2 mm
to 0.25 mm; each has a different attenuation.
5-2. ACR Score
The ACR Score check program obtains and displays an image of the phantom. After suitable adjustment
of the viewing parameters, the user notes the visibility of the object in each cell, using a three-point scale
of 1, 0.5, or 0. For each cell, score:
-1 If the object (fiber, mass, or group of six specks) is completely visible.
-0.5If the object is partially visible, according to the following guidelines;
Fibres; More than half of the fibre is visible.
Masses; The mass does not appear circular, but is visible.
Specks; At least two specks are visible
-0 For all other cases,

Chapter 5 Page no. 234 VF009.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 009 - Image Acquisition and ACR Score6. PROCEDURE
1. Install the Bucky, with compression paddle.
2. Place the ACR phantom on the Bucky, against the chest wall side and laterally centered.
3. Lightly compress the phantom with the compression paddle.
4. Verify that the collimator is in full FOV and that the tube arm angle is 0 degrees.
5. Enter the Phantom IQ test. From the Browser, select: Tools Menu/Service Desktop/Image Quality/
New Acquisition/Phantom IQ.
6. The system applies suitable parameters. Make an exposure and display the image.
7. The image is displayed with standard parameters which may need adjusting for your system. Press
the right mouse button to display a two-item menu which allows you to select the middle button func-
tion. Use it either to control brightness and contrast (WL changes with vertical movement and WW
changes with horizontal movement) or to obtain the magnifying glass function. As a rough guide, it is
suggested that the contrast should be adjusted to obtain an on-screen displayed value for WW of less
than 200 for good viewing.
8. Obtain the best viewing conditions for an accurate assessment of the score (if necessary complete
the test and repeat it with the phantom in a better position) and observe the 16 cells in the phantom
image. Note the ACR score (1, 0.5, or 0, as described in the Test Description above) for each cell,
and total the scores for each type of cell (fibers, masses, speck groups).
9. A group of three on-screen sliders are displayed at the top right of the screen.For each type of cell,
move the on-screen slider (with the cursor on the slider, press and hold the mouse button while mov-
ing the mouse) until the correct score is displayed. Then click on OK to enter the score into the sys-
tem.
10. The results of the test are displayed in a format similar to that shown below. The LSL and USL col-
umns show the minimum acceptable score (Lower Specification Limit) and the maximum score
(Upper Specification Limit) for each item. The figures given here in the Measurement and Status col-
umns are for example only
.
11. The test is successful if the total score is 11 or more, with scores for masses of at least 3, for speck
groups of at least 3, and for fibers of at least 4.
In the example shown the total score is 11.5, but the test fails because the speck group score is less
than 3.
12. If the test fails, repeat the flatfield test and then try again. If the flatfield test passes successfully but
the ACR Score test fails, the cause must be identified and corrected before further patient examina-
tions are made.
.
Test Measurement LSL USL Status
Masses 4.00 3.00 5.00 Passed
Speck groups 2.50 3.00 5.00 Failed
Fibers 5.00 4.00 6.00 Passed
Score 11.50 11.00 16.00 Passed

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 010 - Acquisition in AOP Mode
VF010.FM Page no. 235 Chapter 5
Job Card VF 010 - Acquisition in AOP Mode Chapter 5
1. SUPPLIES
None
2. TOOLS
• Set of acrylic plates (PMMA) (minimum size 20 cm x 20 cm) allowing thicknesses of 25 mm, 40
mm and 60 mm to be used, with a tolerance of ±0.1 mm.
3. SAFETY PRECAUTIONS.
• Take all normal precautions when operating with X–rays.
4. PREREQUISITES
• The system must have been correctly installed and completely calibrated.
• Breast thickness calibration must have been carefully checked in contact mode.
5. TEST DESCRIPTION
The tests described here are designed to check the following aspects of system operation:
• Selection of parameters in AOP (Automatic Optimization of Parameters) mode.
• The level of SNR (Signal-to-Noise Ratio) in images acquired in AOP mode.
6. PROCEDURE
Notice:!
To avoid false results, the acrylic plates used must be clean and free from scratches.
1. Create an AOP patient file by selecting
New Patient on the Browser; complete the Medical Procedure
Card as required, with
AOP as the patient name.
2. Install the bucky on the image receptor and select Large field of view.
3.Carry out the following procedure three times, with different thicknesses of acrylic in the field of
view (25, 40, and 60 mm). The table below shows the values expected for each thickness:
a. Place acrylic plates of the chosen thickness on the bucky.
b. Apply 5daN compression to the plates
c. Check that the breast thickness indicated on the Console Display is equal to the expected value
(25, 40, or 60 mm) ±2 mm.
d. Select the AOP STD mode and make an exposure.
e. Record the exposure parameters (track/filter/kV/mAs) selected.

Chapter 5 Page no. 236 VF010.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 010 - Acquisition in AOP Mode f. Check that the exposure parameters selected are as shown in the table below:
4. On the AWS, close the exam, then open the three raw images for review.
5. For each of the three images in turn:
- Use the ellipse ROI tool (use the default size) to measure the mean value (MV)and standard devi-
ation (sd) of the image, in the region close to the chest wall and laterally centered.
- Calculate the SNR as MV/sd. The value of SNR should exceed 50.
Note that a calculator can be displayed on the AWS screen by obtaining the
root menu (in the
background) and selecting
calculator. Use the Front key to hide/display the calculator.
7. COMPLETION
• The check is successful if the exposure parameters are within the limits shown in the table above,
and if all SNR values exceed 50.
• If the check fails because of incorrect track/filter or kV selection, carry out AOP calibration in contact
mode with grid, and repeat the check.
• If the check fails because of incorrect mAs or SNR values, carry out Conversion Factor and HVL/Yield
calibrations, and repeat the check.
Acrylic
Thickness
(mm)
Exposure Parameters
Track/Filter mAs kV
25 Mo/Mo 20 - 60 27
40 Mo/Rh 50 - 100 28
60 Rh/Rh 50 - 100 32

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 011 - Light Beam Centering
VF011.FM Page no. 237 Chapter 5
Job Card VF 011 - Light Beam Centering Chapter 5
1. SUPPLIES
• None.
2. TOOLS
• White paper and adhesive tape.
•4 coins.
3. SAFETY PRECAUTIONS
WARNING
This procedure produces X-rays. Be sure to take appropriate precautions.
4. PREREQUISITES
• None.
5. PROCEDURE
5-1. Settings
1. Configure the collimator with the following margins (take note of the current values to re-introduce
them at the end of the verification:
ARM/VALID/INSTAL/CONFIG/COLLIM /MARG_L = -5
ARM/VALID/INSTAL/CONFIG/COLLIM /MARG_R = -5
ARM/VALID/INSTAL/CONFIG/COLLIM /MARG_F = -5
2. Go up the menu and press cksum.
3. Set technique to large focal spot, MoMo, 25kV, 25mAs.
4. Install bucky with no additional filtration, select full FOV.
5. Use adhesive tape to secure white paper onto the bucky, covering the full FOV.
6. Press the light beam button and place a coin on the paper tangential to each internal border of the
light beam. The coins must be placed at the edge between the completely lit area and the twilight
area as indicated below:
Bucky
Light beam borders
Coin
White paper
Twilight area

Chapter 5 Page no. 238 VF011.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 011 - Light Beam Centering 7. Take an exposure, select zoom = fit to screen
8. Check that the 4 coins are tangential to X-ray field of view
If not, perform the instructions of Calibration Job Card
Light Beam Centering.
9. Reconfigure the collimator to previous margins.
10. Go up the menu and perform a cksum

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 012 - Dicom Printer Adjustments and Checks
VF012.FM Page no. 239 Chapter 5
Job Card VF 012 - Dicom Printer Adjustments and Checks Chapter 5
1. SUPPLIES
• None.
2. TOOLS
• Densitometer
3. SAFETY PRECAUTIONS.
• None.
4. PREREQUISITES
• The printer must be correctly calibrated
5. PROCEDURE
5-1. Preparation and Print
• Install the SMPTE pattern (together with the Test_pattern pattern): from the root menu, select Ser-
vice Tools/Install SMPTE
.
• If the display WW and/or WL values have been changed, select
Auto to return them to the default val-
ues of WW =1500 and WL = 2000.
• Select the
FFDM_printer_pattern patient from the Browser menu.
-Select
Review Exam.
- Click on the
PRINT button.
- Select the DICOM printer to be checked in the Printing window; click on
Printing Properties.
- In the Properties window, select blue film, processor and portrait. Then
Save and Done.
- Cclick on the
PRINT button.
- The image is sent to the printer. Check the quality of the printed image as described below.
CAUTION
Make sure that the SMPTE pattern is displayed with WW =1500 and WL = 2000 for the
image quality tests.
CAUTION
Before you start to check the calibration of an Agfa LR5200 printer, make sure that it has been properly calibrated between OD max= 3.5 and OD min = 0.14. Refer to Agfa documen- tation.

Chapter 5 Page no. 240 VF012.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 012 - Dicom Printer Adjustments and Checks
5-2. Image Quality checks
• Examine the the printed image (illustrated below).
• Check the image quality according to the points listed.
- Verify continuity of the white diagonal lines [1]. Check that they do not show steps.
- In the two white squares [2] the vertical and horizontal lines must be well separated for good res-
olution.
- In the large light gray square, check that the arc [3] is not duplicated.
- The small black square [4] within the larger square must not be surrounded by a white line.
- The small white square [6] within the larger square, must not be surrounded by a black line.
- Check that the optical density is the same everywhere in the long grey rectangle (+7%) [5]
- Measure the density of each of the grayscale rectangles 1-21 [7] with a densitometer and com-
pare the measured values to the
Expected O.D. values in the table given on the next page. The
values must be accurate to +
8%.
1
11
12
21
2
3
4
5
6
7
8
9
10
13
14
15
16
17
18
19
20
Step
[1]
[2]
[3]
[4]
[5]
[6]
[7]
[7]

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 012 - Dicom Printer Adjustments and Checks
VF012.FM Page no. 241 Chapter 5

CAUTION
Before proceeding to OD measurement, reset the Zero of the Densitometer with no film.
Step Gray Level Expected O.D.
E
Agfa LR5200
Expected O.D.
E
Kodak
8600/8610
Measured O.D.
M
Difference (%)
(E - M)/E X 100
10 3.5 3.59
2 12 2.98 2.80
3 25 2.66 2.46
4 38 2.42 2.23
5 51 2.22 2.04
6 63 2.04 1.89
7 76 1.87 1.74
8 89 1.72 1.61
9 102 1.58 1.48
10 114 1.44 1.37
11 127 1.30 1.25
12 140 1.18 1.14
13 153 1.05 1.03
14 165 0.93 0.93
15 178 0.81 0.82
16 191 0.69 0.72
17 204 0.58 0.61
18 216 0.47 0.51
19 229 0.35 0.41
20 242 0.24 0.31
21 255 0.14 0.21

Chapter 5 Page no. 242 VF012.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 012 - Dicom Printer Adjustments and Checks This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 013 - Disk Full Check
Vf013.fm Page no. 243 Chapter 5
Job Card VF 013 - Disk Full Check Chapter 5
1. SUPPLIES
• None.
2. TOOLS
• None.
3. SAFETY PRECAUTIONS.
• None.
4. PREREQUISITES
• The system must be correctly installed and functional.
5. PROCEDURE
• A full disk may be caused by the storage of unnecessary temporary files. These can be created when
print requests are sent to a printer and the printer does not respond. They should be removed, but in
practice the deletion is not always successful.
5-1. Symptoms
• The disk remains full or almost full, even after deletion of patient files.
• Error logs may be truncated, the Workstation may appear to be running slowly, and/or functions or
image transfers may be slow.
5-2. Correction
1. Check that the previous exam is finished and that all images have been declared, pushed, and
printed as required.
2. Open a Command Window (right-click on screen background and select Service Tools/Command
Window).
3. Change directory to
/export/home/sdc by typing the command: cd<Return>.
4. Change directory to
/export/home/sdc/scripts by typing the command: cd scripts<Return>.
5. Launch the cleanup script by typing the command: iop_cleaner<Return>.
6. If you are asked if you wish to delete files in the printer queue, enter: y.
7. Wait for a messsage indicating successful completion of the iop cleaning script, then select Restart
Browser from the root menu.
8. Check disk spce and system performance.

Chapter 5 Page no. 244 Vf013.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card VF 013 - Disk Full Check This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
PM.FM Page no. 245
CHAPTER 6 PLANNED MAINTENANCE
1. INTRODUCTION
This chapter contains recommendations for planned maintenance of the Senographe 2000 D system.

Page no. 246 PM.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card PM 001 - Senographe 2000 D Maintenance
PM001.FM Page no. 247 Chapter 6
Job Card PM 001 - Senographe 2000 D Maintenance Chapter 6
1. SENOGRAPHE 2000 D PLANNED MAINTENANCE
Recommended periodic checks for the Senographe 2000 D system are listed below. The Job Cards
referred to are in the Sevice Manual.
Table 1 lists checks which should be carried out twice per year; Table 2 lists annual checks.
Table 3 lists items which must be replaced by new items every three years
TABLE 1 - SIX-MONTHLY CHECKS
Test Description Document
reference
System/power checks:
System: Boot/Init and
shutdown
Observe correct boot and initialization of system;
observe system behaviour and timing during shutdown.
VF 001
UPS: Battery Verify UPS operation, battery load and charge status. VF 003
DMR functional checks Check DMR mechanical operations (arm movements,
rotation, brakes), dust on fans, general cleanliness,
etc.)
VF 005
Conditioner checks:
Conditioner check and
Conditioner flush
Check water and flow, check pump noise, inspect
detector and tubes for leakage. Flush and refill Condi-
tioner
D/R 106
Image quality checks:
Light beam centering veri-
fication
Check field covered by centering light VF 011
X-ray beam parallelism
and centering
Check that field covered by X-ray beam is centered and
parallel to detector edges.
VF 006
X-ray format geometry
check
Check size and centering of FOV for each focal spot. VF 007
Flat Field Validation Perform Flat Field Test;
if it fails, repeat Detector Gain, Correction factor meas-
urement, and/or Bad pixel calibrations.
VF 008
Monitor: Contrast, bright-
ness, and geometry
Check ambient light; check display image quality. CAL 018
System: Exam in manual
mode + ACR score
Exercise application; use ACR phantom to verify Image
Quality
VF 009

Chapter 6 Page no. 248 PM001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card PM 001 - Senographe 2000 D Maintenance
TABLE 2 - ANNUAL CHECKS
TABLE 3 - SYSTEMATIC REPLACEMENTS: EVERY THREE YEARS
Test Description Document
reference
Cart: Mechanical checks Check cart mechanical operations (folding tables, slid-
ing doors, wheels, dust on fans, screen cleanliness,
etc.)
VF 002
UPS Battery Capacity
Recalibration
Recalibrate UPS battery capacity indication. VF 003
Replacement Item Description Document
reference
Compression system
belts
Install new compression system belts D/R 008
CPU Board batteries Install new batteries on the Gantry and Generator
CPU Boards
D/R 009, 021
X1 relay on the A.C. Dis-
tribution Board
Install a new X1 relay (91682082) on the 200 PL1 A.C.
Distribution Board
N/A

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
ERRORS.FM Page no. 249
CHAPTER 7 ERROR MESSAGE INFORMATION
This chapter lncludes job cards listing error messages generated by Senographe 2000 D sub-systems.

Page no. 250 ERRORS.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 001 - IDC Error Messages
ERR001.FM Page no. 251 Chapter 7
Job Card ERR 001 - IDC Error Messages Chapter 7
1. IDC DIAGNOSTICS
Each time that the IDC sub-system is booted (powered up or reset), it carries out a series of diagnostic
self-tests, first under the control of the BIOS, then under software control. As each test is performed, a
corresponding code is displayed on the four-digit green LCD panel at the rear of the IDC unit.
1-1. BIOS Diagnostic Display
While the BIOS tests are running, the LCD displays two blank digits followed a two-digit code corre-
sponding to the current test. If a test detects an error or fails to complete, the tests are halted and the
code of the failing test remains displayed.
1-2. Software Diagnostic Display
• Three sets of software diagnostics are carried out. While these test are running, the display shows
X000, where X may be 1, 2, or 3.
• If all tests are completed without failure, the display shows
FFFF.
• If a test fails, the display shows a code in the form XXFF for two seconds. XX is a code correspond-
ing to the failing test. The tests then continue.
• On completion of the tests after a failure occurs, the display shows
F0FF.
2. ERROR DETECTION
• If the IDC halts with a display in the format FFFF, all power-up tests have passed without failure.
• If the IDC halts with a display in the format
--XX, where XX can be any number, a BIOS test has
failed. If the problem occurs again after reset, the recommended action is to change the IDC unit.
• If the IDC halts with a display in the format
F0FF, a software test has failed. Reset the IDC and watch
the LCD display.
If the failure is repeated, a code
XXFF is displayed for two seconds. Note the error code XX.
Table 1 gives a list of IDC diagnostic error codes, with recommended sequences of actions.
3. ERROR LOG
All errors reported by the IDC diagnostic routines are recorded in the system error log.
To access the error log, right click on the AWS screen background; select
Service Tools/Service Desk-
top/Error Log
from the drop-down menu.

Chapter 7 Page no. 252 ERR001.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 001 - IDC Error Messages
TABLE 1 - IDC DIAGNOSTIC ERROR CODES
Error Code Displayed Actions (in recommended order)
FFFF No action; all tests passed.
- - XX A BIOS test (code
XX)has failed. If the problem persists, change the
IDC.
FOFF A software test has failed. Reset the IDC and watch the LCD display.
Note any error code displayed (XXFF for two seconds).
01, 02, 03 Reset IDC.
Restore IDC software.
11, 12, 13, 14, 15, 16, 23,
41, 53, 55
Reset IDC.
Change IDC.
21 Change IDC-Conditioner cable.
Change Conditioner.
Change Detector.
22 Check/top up condtioner coolant.
Change conditioner.
24 Change IDC-DMR serial link cable.
31 Check/change IDC-Detector Power Supply cable.
32 Change Detector Power Supply.
33 Change Detector Power Supply ac cable.
Change Detector Power Supply.
40 Check/change IDC-Detector fiber-optic cables.
Change Detector.
42 Check/change IDC-Detector fiber-optic cables.
51 Change IDC-DMR serial link cables.
Change DMR board 400PL2.
Change IDC.
54 Change IDC-DMR serial link cables.
56 Change IDC-DMR RT line cables.
60 Check Detector temp./humidity; if OK Change Detector.
64 Check/change IDC-Conditioner air and water pipes.
Change Conditioner.
Change Detector.
66 Repeat calibrations.
Change IDC.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 002 - AWS Error Messages
ERR002.FM Page no. 253 Chapter 7
Job Card ERR 002 - AWS Error Messages Chapter 7
1. AWS ERROR MESSAGES
Errors reported by AWS software applications are recorded in the system error log.
To access the error log, right click on the AWS screen background; select
Service Tools/Service Desk-
top/Error Log
TABLE 1 - AWS DIAGNOSTIC ERROR CODES
Error Code Error Text Action
1 Software failure None (automatic server restart).
2 Software integrity error Run
fsck.
Reinstall software.
3 Communication problem with
IDC subsystem
Reset IDC and/ or restart AWS browser.
Check cable connection between AWS and IDC).
4 Communication problem with
DMR subsystem
Check that DMR is on.
Check cable connection between AWS and Generator.
Change Generator Interface Board or change AWS.
6 Communication problem with
external network
Check network cable connection.
Check/change network hub.
7 Communication problem with
UPS
Check cable connection between AWS and UPS.
Try to restart the system.
Change UPS or AWS.
8 Battery low or discharged Charge UPS battery (press Cart ON switch on Generator
Cabinet, wait approx. 45 minutes).
Change UPS if necessary.
10 Failure accessing hardware
key
Check HW key validity and integrity.
Check HW key serial port on AWS.
Change HW key or AWS.
12 DICOM communication prob-
lem with worklist provider
The DICOM worklist provider is not DICOM compliant.
Check dicom compatibility with hospital network adminis-
trator.
Check dicom worklist provider declaration parameters.
18 Wrong value received from X
Generator
Check AWS and generator releases; if not compatible,
upgrade system.
19 Wrong value received from
IDC
Check AWS and IDC releases; if not compatible, upgrade
system.

Chapter 7 Page no. 254 ERR002.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 002 - AWS Error Messages This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 255 Chapter 7
Job Card ERR 003 - Gantry and Generator Error Codes Chapter 7
1. INTRODUCTION
This Job Card lists error codes generated by the Gantry and Generator sub-systems (DMR) and dis-
played on the DMR Console. Each message may contain two codes, the
User error code and the Ser-
vice error code
(refer to the explanation below). Error codes containing a Service Error Code are listed
here; codes containing a User Error code are listed in the Operator Manual. Many messages contain
both, and appear in both manuals.
Refer to section 2
Error Code Structure for an explanation of the error code display.
Section 3
Gantry and Generator Error Codes lists the error codes, with corresponding error descrip-
tions. Where possible (mainly for recently developed codes), the list also includes the text of a displayed
error message, and suggestions for corrective action.
2. ERROR CODE STRUCTURE
Error codes are displayed on the DMR Console in the format shown here:..
• The User Error Code is the code given in the Operator Manual; it is easy to use and communicate,
and gives sufficient information for use by the customer.
It includes a code for the type of error (
a) and a message number (bb). The example given, E55, indi-
cates that a tube angulation fault has occurred which caused exposure inhibition,
-
a: Type of error:
I: Warning; exposure not inhibited.
S: Warning; exposure inhibited.
E: Fault; exposure inhibited.
K: Fault; power shut down for safety.
-
bb: Message number indicating the origin of the fault.
• The Service Error Code is the code used in this document; it gives more information, and is recom-
mended for use in GE internal communication. It includes two reference codes:
-
ccc: Indicates the affected subassembly (hardware or software). The example given, 161, indi-
cates a tube housing fault.
-
ddd: Indicates the origin of the fault The example given, 013, indicates that no reference crossing
was detected during tube movement.
• The following abbreviations are used in the error code descriptions:
IST: Installation
MNT:Maintenance
VRTX: Real time operating system
• Software errors. Many of the codes are described as
software errors (e.g., 002/001; Gene. CPU soft-
ware error). These errors should never occur. If they do, and especially if they are repeated, please
contact Service Engineering with as much information as possible (including a copy of the error
codes file on diskette).
E 55 161 013Error text
a bb ccc ddd
User error code Service error code

Chapter 7 Page no. 256 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
3. GANTRY AND GENERATOR ERROR CODES

Code Message Text Description Recommended Actions
002/001 Gene CPU software error: VRTX error
during init
002/002 Gene CPU software error: reboot by
watch_dog_hard or reset button
002/003 Gene CPU software error: reboot by
watch_dog_soft
002/004 Gene CPU: reboot after power micro-fail-
ure
003/000 Gene CPU software error: bus error
003/001 Gene CPU software error: address error
003/002 Gene CPU software error: format error
003/003 Gene CPU software error: illegal instruc-
tion
003/004 Gene CPU software error: privilege viola-
tion
003/005 Gene CPU software error: trace error
003/006 Gene CPU software error: emulation line
1010 error
003/007 Gene CPU software error: emulation line
1111 error
003/008 Gene CPU software error: CHK error
003/009 Gene CPU software error: TRAPV error
004/001 Gene interface: reset error
004/002 Gene interface: read/write error on timer
PTM1
004/003 Gene interface: read/write error on timer
PTM1
004/004 Gene interface: read/write error on timer
PTM2
004/005 Gene interface: read/write error on timer
PTM2
004/006 Gene interface: interrupt error on DUART1
004/007 Gene interface: PTM1 timer 1 defective

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 257 Chapter 7
004/008 Gene interface: PTM1 timer 2 defective
004/009 Gene interface: PTM1 timer 3 defective
004/010 Gene interface: PTM2 timer 1 defective
004/011 Gene interface: PTM2 timer 2 defective
004/012 Gene interface: PTM2 timer 3 defective
004/013 Gene interface: interrupt error on PTM1
timer 1
004/014 Gene interface: interrupt error on PTM1
timer 2
004/015 Gene interface: interrupt error on PTM1
timer 3
004/016 Gene interface: interrupt error on PTM2
timer 1
004/017 Gene interface: interrupt error on PTM2
timer 2
004/018 Gene interface: interrupt error on PTM2
timer 3 D208
006/000 Generator: fault at exposure command
006/001 Generator: fault at exposure command
006/002 Generator: signal kV_Inf_5
(Less_than_5kV) not rising
006/003 Generator: Signal En_Pose
(Exposure_On) not rising
006/004 Generator: fatal security triggered during
exposure
006/005 Generator: fatal security triggered at end
of exposure or signal En_Pose
(Exposure_On) stays ’high’
006/011 Generator: signal Secu_Rattrap
(No_Fatal_Security) stays ’high’ after ’arc-
ing recovery’
007/xxx Gene init tests: one fatal security input <>
0; xxx identifies input.Access generator
error log display for full error decoding.
011/001 Focus bias software error: unknown com-
mand
011/002 Focus bias software error: unknown task
status
Code Message Text Description Recommended Actions

Chapter 7 Page no. 258 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
012/001 Focus bias: signal Pres_Polar
(Bias_Presence) ’high’ in standby status
012/002 Focus bias: signal Pres_Polar
(Bias_Presence) stays ’low’ when bias
generation starts
012/003 Focus bias: signal Pres_Polar
(Bias_Presence) goes ’low’ after stabiliza-
tion
012/004 Focus bias: signal Pres_Polar
(Bias_Presence) stays ’high’ after bias
generation stopped
016/001 Anode starter software error: unknown
command
016/002 Anode starter software error: unknown
task status
017/001 Anode starter: signal Pres_Rotation
(Rotation_Presence) ’high’ in standby sta-
tus
017/002 Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 9000 rpm
017/003 Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 6000 rpm
017/004 Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation at 9000 rpm
017/005 Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation at 6000 rpm
017/006 Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation (speed < 9000 rpm)
017/007 Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation (speed < 6000 rpm)
017/008 Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’high’ during
9000 rpm braking
017/009 Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’high’ during
6000 rpm braking
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 259 Chapter 7
017/010 Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’high’ during
rotor failure
017/011 Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’low’ during
rotation
021/001 Heater software error: unknown command
021/002 Heater software error: unknown task sta-
tus
021/003 Heater software error: Heater scaling
parameters (ICH) or heater values are not
valid
022/001 Heater track 1: signal Pres_Chauf
(Heater_Presence) ’high’ in standby sta-
tus
022/002 Heater track 1: signal Pres_Chauf
(Heater_Presence) stays ’low’ when
heater started
022/003 Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ during pre-
heater phase
022/004 Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ during 1st
trigger
022/005 Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ during
exposure
022/006 Heater track 1: mA value not significant
022/007 Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ after expo-
sure
022/008 Heater track 1: signal Pres_Chauf
(Heater_Presence) stays ’high’ when
heater stopped
023/001 Heater track 2: signal Pres_Chauf
(Heater_Presence) ’high’ in standby sta-
tus
023/002 Heater track 2: signal Pres_Chauf
(Heater_Presence) stays ’low’ when
heater started
Code Message Text Description Recommended Actions

Chapter 7 Page no. 260 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
023/003 Heater track 2: signal Pres_Chauf
(Heater_Presence) goes ’low’ during pre-
heater phase
023/004 Heater track 2: signal Pres_Chauf
(Heater_Presence) goes ’low’ during 1st
trigger
023/005 Heater track 2: signal Pres_Chauf
(Heater_Presence) goes ’low’ during
exposure
023/006 Heater track 2: mA value not significant
No mA measurement; change 701PL1
(kV/mA board).
023/007 Heater track 2: signal Pres_Chauf
(Heater_Presence) ’low’ after exposure
023/008 Heater track 2: signal Pres_Chauf
(Heater_Presence) stays ’high’ when
heater stopped
026/001 Grid sync software error: ACQ_NTR or
ACQ_OVR received or unknown reply
when starting
026/002 Grid sync software error: ACQ_NTR or
ACQ_OVR received or unknown reply
when stopping
026/003 Grid sync software error: unknown com-
mand
026/004 Grid sync software error: unknown task
status
027/001 Grid sync: signal Synchro_Bras
(Synchro_Arm) ’high’ in standby status
027/002 Grid sync: grid start signal not received or
grid fault detected
027/003 Grid sync: signal Synchro_Bras
(Synchro_Arm) ’low’ after grid start
027/004 Grid sync: grid stop signal not received or
grid fault detected
027/005 E04 Arm fatal fail-
ure
A Gantry error was signalled during the
exposure.
1. Reset DMR
2. Check communications
between Gantry and generator.
027/006 E04 Arm fatal fail-
ure
A Gantry error was signalled during the
exposure.
1. Reset DMR
2. Check communications
between Gantry and generator.
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 261 Chapter 7
031/001 Generator: delay task software error:
unknown command
031/002 Generator: delay task software error:
unknown task status
041/001 Gene thermal protection software: VRTX
error
041/002 Gene thermal protection software: VRTX
error
041/003 Gene thermal protection software: VRTX
error
041/004 Gene thermal protection software: VRTX
error
041/005 Gene thermal protection software:
unknown anode starter status
041/006 Gene thermal protection software: VRTX
error
046/001 Gantry thermal protection software: VRTX
error
051/001 Filter wheel software error: ACQ_NTR or
ACQ_OVR received or unknown reply
051/002 Filter wheel software error: unknown com-
mand
051/003 Filter wheel software error: unknown task
status
052/001 Filter wheel: filter change request not
transmitted or filter fault
056/001 Focus software error: ACQ_NTR or
ACQ_OVR received or unknown reply
056/002 Focus software error: unknown command
056/003 Focus software error: unknown task status
057/001 Focus: focus change request not transmit-
ted or focus fault
061/000 Gene to gantry communication driver: no
reply from gantry after 2 attempts
062/000 Gene to control panel communication
driver: no reply from control panel after 2
attempts
063/006 Gene init tests: no gantry status feedback
after 10 requests
Code Message Text Description Recommended Actions

Chapter 7 Page no. 262 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
066/000 Gene date/time: access timeout during
dater init
066/001 Gene date/time: access timeout during
read sequence
066/002 Gene date/time: access timeout during
write sequence
066/003 Gene date/time: access timeout during
time write sequence
067/000 Generator: Zero divide
068/xxx Generator: floating point calculation error.
xxx is VRTX task number
071/081 Generator: checksum error on battery
saved memory
071/082 Generator: timeout on signal
Alim_Puis_OK (Inv_Supply_OK) (going
’low’ or ’high’)
071/083 Generator: error reading ’kV law table’ in 0
pt technique (AOP)
072/001 Exposure software error: ACQ_NTR or
ACQ_OVR received at start of 1st trigger
072/003 Exposure software error: ACQ_NTR or
ACQ_OVR received at end of 1st trigger
072/004 Exposure software error: unknown origin
of aborted exposure
072/005 Exposure software error: aborted by PM
in 2 pt technique
072/006 Exposure software error: unknown expo-
sure technique
072/007 E06 Software Error
(gen. CPU)
One system task (at least) not stable after
a delay of 7s when a Prep is performed. A
DMR component (tube, rotor, etc.) or IDC
is not ready for the exposure (prepStatu-
sOk message not sent to DMR by IDC).
This error is either caused by
IDC or DMR. When a DMR sub-
system is not ready, there
should be a second error mes-
sage displayed relating to the
sub-system causing the error
072/008 Exposure software error: unknown task
status
072/009 Exposure software error: 0 pt or 1 pt expo-
sure and gantry absent
073/001 Gene software error (IST, standby):
parameter display problem
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 263 Chapter 7
073/002 Gene software error (IST, standby):
unknown tube type
073/003 Gene software error (IST, standby):
alpha=0.0 when calculating value
P_NDac_HTPM
073/004 Gene software error (IST, standby): incor-
rect thickness for correlation calculation
073/005 Gene software error (IST, standby): bias
calibration task problem
073/006 Gene software error (IST, standby): heater
scale calibration task problem
073/007 Gene software error (IST, standby): HV/
DAC calibration task problem
073/008 Gene software error (IST, standby):
FREQ/HV calibration task problem
073/009 Gene software error (IST, standby): PM
calibration task problem
073/010 Gene software error (IST, standby): PM
calibration task problem
073/011 Gene software error (IST, standby): PM
calibration task problem
073/012 Gene software error (IST, standby): PM
calibration task problem
073/013 Gene software error (IST, standby): gantry
connection task problem
073/014 Gene software error (IST, standby): gantry
connection fault
073/015 Gene software error (IST, standby): gantry
connection fault
073/016 Gene software error (IST, standby): gantry
connection fault
073/017 Gene software error (IST, standby): mA
measurement calibration problem
074/001 Gene software error (IST, exposure):
unknown exposure end
074/002 Gene software error (IST, exposure):
unknown gantry reply
074/003 Gene software error (IST, exposure):
unknown gantry reply
Code Message Text Description Recommended Actions

Chapter 7 Page no. 264 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
074/004 Gene software error (IST, exposure):
unknown exposure status
075/010 Gene software error (MNT): anode starter
speed impossible
075/011 Gene software error (MNT): anode
starter: abnormal test task status
075/020 Gene software error (MNT): heater:
impossible track
075/021 Gene software error (MNT): heater:
abnormal test task status
075/022 Gene software error (MNT): heater:
abnormal test task status
075/030 Gene software error (MNT): bias: abnor-
mal test task status
075/040 Gene software error (MNT): HV test:
abnormal test task status
075/041 Gene software error (MNT): gate test:
abnormal test task status
075/042 Gene software error (MNT): inverter test:
abnormal test task status
075/050 Gene software error (MNT): mA value:
abnormal test task status
075/051 Gene software error (MNT): mA measure-
ment: abnormal test task status
075/060 Gene software error (MNT): grid sync:
abnormal test task status
075/070 Gene software error (MNT): PM test:
abnormal test task status
075/071 Gene software error (MNT): PM test:
abnormal test task status
075/072 Gene software error (MNT): PM test:
measurement timeout
075/090 Gene software error (MNT): error codes:
display problem
075/091 Gene software error (MNT): error codes:
display problem
075/092 Gene software error (MNT): error codes:
display problem
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 265 Chapter 7
075/093 Gene software error (MNT): error codes:
display problem
075/100 Gene software error (MNT): transfer:
frame code error
075/101 Gene software error (MNT): transfer:
abnormal transfer task (save) status
075/102 Gene software error (MNT): transfer:
abnormal transfer task (save) status
075/103 Gene software error (MNT): transfer:
abnormal state transfer task (load) status
075/104 Gene software error (MNT): transfer:
abnormal state transfer task (load) status
075/105 Gene software error (MNT): transfer:
unknown frame
075/106 Gene software error (MNT): transfer: end
of loading error
075/107 Gene software error (MNT): transfer:
abnormal transfer task (protocol) status
078/002 Gene init tests software error: timeout on
system task
078/003 Gene init tests software error: defective
function will self-identify
078/004 Gene init tests software error: defective
function will self-identify
078/005 Gene init tests software error: defective
function will self-identify
078/128 Gene init tests software error: error follow-
ing VRTX request
079/001 Exposure aborted: timeout
079/002 Exposure aborted: heater fault
No mA measurement; change 701PL1
(kV/mA board).
079/003 Exposure aborted: anode starter fault
079/004 Exposure aborted: power fault
079/005 Exposure aborted: generator fault
079/006 Exposure: no gantry acknowledgement (at
start of 1st trigger) or gantry fault
Code Message Text Description Recommended Actions

Chapter 7 Page no. 266 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
079/008 Gantry fault detected before start of expo-
sure
079/009 Exposure: no sync signal from gantry to
start exposure
079/010 Exposure aborted by mA meter (0 pt or 1
pt technique)
079/011 Exposure: no gantry acknowledgement (at
end of 1st trigger) or gantry fault
079/014 E05 Failure during
exposure
Falling XRayAuthorizedRT signal not
received from IDC
Switch Off the system.Switch On
again; repeat the exposure.
Check IDC-DMR communication
079/015 E05 Failure during
exposure
Brightness value not received from IDC Switch Off the system.Switch On
again; repeat the exposure.
079/017 E16 AWS/SENO
Synchro fail
The acquisition was not correctly per-
formed on ADS (image not correctly trans-
ferred from IDC for example)
Reset system and try again. If
error persists, check AWS is run-
ning correctly; try rebooting
AWS and DMR.
080/001 Gene (IST, exposure): heater fault
080/002 Gene (IST, exposure): anode starter fault
080/003 Gene (IST, exposure): power failure
080/004 Gene (IST, exposure): interface fault
080/005 Gene (IST, exposure): no gantry reply -
timeout after 1 second
080/006 Gene (IST, exposure): gantry fault
080/007 Gene (IST, exposure): no gantry reply to
exposure command
080/008 Gene (IST, exposure): no grid sync feed-
back after timeout
082/000 Gene: ’Endurance’ error code
082/001 Gene: ’Endurance’ error code
082/010 Gene: ’Endurance’ error code
082/020 Gene: ’Endurance’ error code
082/030 Gene: ’Endurance’ error code
082/040 Gene: ’Endurance’ error code
083/002 I17 AWS./SENO
Comm. lost
DMR/AWS communication failure: the
DMR did not receive any message from
AWS during a period of 2 minutes.
Check DMR/AWS connections.
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 267 Chapter 7
084/008 E24 Failure before
exposure.
An error occurred before X-ray emission
but after the Radiography key has been
pressed, relating to a signal not correctly
received by the exposure automaton from
the IDC or Gantry.
This error may be caused by the
IDC or the Gantry. Reboot the
entire system and check IDC-
DMR communications.
084/009 E24 Failure before
exposure.
An error occurred before X-ray emission
but after the Radiography key has been
pressed, relating to a signal not correctly
received by the exposure automaton from
the IDC or Gantry.
This error may be caused by the
IDC or the Gantry. Reboot the
entire system and check IDC-
DMR communications.
084/013 E24 Failure before
exposure.
An error occurred before X-ray emission
but after the Radiography key has been
pressed, relating to a signal not correctly
received by the exposure automaton from
the IDC or Gantry.
This error may be caused by the
IDC or the Gantry. Reboot the
entire system and check IDC-
DMR communications.
090/000 Generator: PROM checksum error
092/000 I19 IDC/Seno
comm. lost.
Communication error between DMR and
IDC: the connection is lost (no answer
from IDC for 15 seconds). If the error con-
dition continues for 2 minutes, E19 is dis-
played.
Verify the link/cable between
DMR and IDC. Verify IDC is ON.
092/000 E19 IDC/Seno
down, DPS is off.
Communication error between DMR and
IDC: the connection is lost (no answer
from IDC for 2 minutes). When this error
occurs, the detector power supply is set
off.
Verify the link/cable between
DMR and IDC. Verify IDC is ON.
094/001 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
bad command received by the IDC task
Check DMR/AWS connections,
perform comm diagnostics
094/002 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
Prep command not received by the IDC
task
Check communication between
IDC and DMR, perform comm
diagnostics, perform IDC diag-
nostics
094/003 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
Stop command at end of exposure not
received by the IDC task.
Check communication between
IDC and DMR, perform comm
diagnostics, perform IDC diag-
nostics
094/004 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
XRayAuthorizedRT status not received by
the IDC task.
Check communication between
IDC and DMR, perform comm
diagnostics, perform IDC diag-
nostics
Code Message Text Description Recommended Actions

Chapter 7 Page no. 268 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
094/005 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
Stop command not received by the IDC
task at the end of the Radiography.
Check communication between
IDC and DMR, perform comm
diagnostics, perform IDC diag-
nostics
094/006 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
Prep command not received by the expo-
sure automaton (task T_APPLI)
Check communication between
IDC and DMR, perform comm
diagnostics, perform IDC diag-
nostics
094/007 E18 IDC/Seno syn-
chro failure.
An error occurs on IDC or on DMR which
generates an error on a signal/message
expected from the IDC task of the DMR:
XRayAuthorizedRT signal active in idle
status
Check communication between
IDC and DMR, perform comm
diagnostics, perform IDC diag-
nostics
096/019 I43 Prepa glnd Scf
out of range
Error on Scf glandular factor during the
Prep
none (warning)
096/020 I50 Prepa fat Scf
out of range
Error on Scf fat factor during the Prepa none (warning)
096/021 I51 Pre-exp. Scf out
of range
Error on Scf factor for the pre-exposure none (warning)
096/022 I52 Expo. AOP Scf
out of range
Error on Scf factor for the AOP exposure none (warning)
096/023 I53 Exposure
aborted by mAs
level
mAs value performed for the pre-exposure
out of range
Reset DMR.
096/024 I54 Exposure
aborted by bright.
level
Dose value returned by IDC out of range Reset DMR.
096/025 I55 Track/Filter
change
Track/filter configuration changed between
the pre-shot and the exposure
none (warning)
096/033 I56 Exposure
aborted, bright.
unknown.
IDC sends an invalid dose value to the
DMR.
Reset the entire system and
retry the exposure.
097/026 I48 U-turn before
end of pre-shot.
The exposure time for the pre-shot (AOP
mode) is greater than the grid motion
duration to the first turn-back point.
none (warning).
097/027 I49 U-turn before
end of exposure.
The exposure time (manual and AOP
mode) is greater than the grid motion
duration to the first turn-back point.
none (warning).
100/001 Gantry software error: VRTX error during
init
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 269 Chapter 7
101/001 Gantry interface: Reset error
101/002 Gantry interface: read/write error on timer
PTM1
101/003 Gantry interface: read/write error on timer
PTM1
101/004 Gantry interface: read/write error on timer
PTM2
101/005 Gantry interface: read/write error on timer
PTM2
101/006 Gantry interface: interrupt error on
DUART1
101/007 Gantry interface: PTM1 timer 1 defective
101/008 Gantry interface: PTM1 timer 2 defective
101/009 Gantry interface: PTM1 timer 3 defective
101/010 Gantry interface: PTM2 timer 1 defective
101/011 Gantry interface: PTM2 timer 2 defective
101/012 Gantry interface: PTM2 timer 3 defective
101/013 Gantry interface: interrupt error on PTM1
timer 1
101/014 Gantry interface: interrupt error on PTM1
timer 2
101/015 Gantry interface: interrupt error on PTM1
timer 3
101/016 Gantry interface: interrupt error on PTM2
timer 1
101/017 Gantry interface: interrupt error on PTM2
timer 2
101/018 Gantry interface: interrupt error on PTM2
timer 3
101/020 Gantry interface: PROM checksum error
101/023 Gantry interface: unexpected interrupt
received (level 2)
105/000 Gantry software error: unknown origin of
task call
105/001 Gantry software error: boolean expected
105/002 Gantry software error: unknown generator
grid request
Code Message Text Description Recommended Actions

Chapter 7 Page no. 270 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
105/003 Gantry software error: unknown generator
request for HTPM/IPM switch
105/004 Gantry software error: unknown 1st trigger
status
105/005 Gantry software error: unknown language
choice requested by generator
105/006 Gantry software error: unknown generator
command
105/009 Gantry software error: unknown VRTX
error code
105/016 Gantry software error: unknown acknowl-
edgement (light centering)
105/017 Gantry software error: unknown task com-
mand (light centering)
105/020 Gantry software error: unknown acknowl-
edgement (compression)
105/021 Gantry software error: unknown task com-
mand (compression)
105/023 Gantry software error: unknown compres-
sion pedal status
105/026 Gantry software error: unknown acknowl-
edgement (lateral collimator)
105/027 Gantry software error: unknown task com-
mand (lateral collimator)
105/030 Gantry software error: unknown acknowl-
edgement (rear collimator)
105/031 Gantry software error: unknown task com-
mand (rear collimator)
105/034 Gantry software error: unknown acknowl-
edgement (filter wheel)
105/035 Gantry software error: unknown task com-
mand (filter wheel)
105/038 Gantry software error: unknown acknowl-
edgement (tube angulation)
105/039 Gantry software error: unknown task com-
mand (tube angulation)
105/046 Gantry software error: unknown task com-
mand (format)
105/048 Gantry software error: unknown format
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 271 Chapter 7
105/049 Gantry software error: unknown acknowl-
edgement (column up/down)
105/050 Gantry software error: unknown task com-
mand (column up/down)
105/051 Gantry software error: unknown column
command button status code
105/052 Gantry software error: unknown acknowl-
edgement (arm rotation)
105/053 Gantry software error: unknown task com-
mand (compression arm rotation)
105/054 Gantry software error: unknown status
code of compression arm rotation control
button
105/055 Gantry software error: unknown acknowl-
edgement (tube arm rotation)
105/056 Gantry software error: unknown task com-
mand (tube arm rotation)
105/057 Gantry software error: unknown status
code of tube arm rotation control button
105/063 Gantry software error: unknown acknowl-
edgement (grid)
105/065 Gantry software error: unknown event
(grid)
105/066 Gantry software error: unknown acknowl-
edgement (power calculation)
105/067 Gantry software error: unknown event
(power calculation)
105/068 Gantry software error: unknown tube type
105/069 Gantry software error: unknown focus
change code
105/070 Gantry software error: unknown filter code
requested by generator
105/071 Gantry software error: unknown code for
’Exposure’/’No Exposure’ status manage-
ment
105/072 Gantry software error: unknown exposure
end code
105/073 Gantry software error: unknown ’Endur-
ance’ code
Code Message Text Description Recommended Actions

Chapter 7 Page no. 272 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
105/074 Gantry software error: unknown ’Endur-
ance’ event
105/080 Gantry/tube housing type configuration
incompatibility (angulation/angulation)
105/081 Gantry/tube housing type configuration
incompatibility (angulation/focus)
106/002 Gantry software error: column: unknown
task return
106/003 Gantry software error: column: unknown
task status
106/004 Gantry software error: column: unknown
task status
106/005 Gantry software error: column: unknown
task status
106/006 Gantry software error: column: unknown
task status
106/007 Gantry software error: column: unknown
order transmitted to task
107/001 Gantry software error: compression: too
long between task calls
107/002 Gantry software error: compression:
unknown task return
107/003 Gantry software error: compression:
unknown task status
107/004 Gantry software error: compression:
unknown order transmitted to task
108/001 Gantry software error: lateral collimator:
positioning timeout
108/002 Gantry software error: lateral collimator:
unknown task return
108/003 Gantry software error: lateral collimator:
unknown task status
108/004 Gantry software error: lateral collimator:
unknown task status
108/005 Gantry software error: lateral collimator:
unknown task status
108/006 Gantry software error: lateral collimator:
unknown order transmitted to task
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 273 Chapter 7
108/007 Gantry software error: lateral collimator:
unknown order transmitted to task
108/008 Gantry software error: lateral collimator:
position reference value is outside toler-
ance
109/001 Gantry software error: rear collimator:
positioning timeout
109/002 Gantry software error: rear collimator:
unknown task return
109/003 Gantry software error: rear collimator:
unknown task status
109/004 Gantry software error: rear collimator:
unknown task status
109/005 Gantry software error: rear collimator:
unknown task status
109/006 Gantry software error: rear collimator:
unknown order transmitted to task
109/007 Gantry software error: rear collimator:
unknown order transmitted to task
109/008 Gantry software error: rear collimator:
position reference value is outside toler-
ance
111/002 Gantry software error: tube angulation:
unknown task status
111/003 Gantry software error: tube angulation:
unknown task return
111/004 Gantry software error: tube angulation:
unknown task status
111/005 Gantry software error: tube angulation:
unknown task status
111/006 Gantry software error: tube angulation:
unknown order transmitted to task
111/007 Gantry software error: tube angulation:
unknown order transmitted to task
111/008 Gantry software error: tube angulation:
position reference value is outside toler-
ance
111/017 Gantry software error: tube angulation:
task triggered by timeout
Code Message Text Description Recommended Actions

Chapter 7 Page no. 274 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
112/000 Gantry software error: filter wheel:
unknown task return
112/001 Gantry software error: filter wheel:
unknown task status
112/002 Gantry software error: filter wheel:
unknown task status
112/003 Gantry software error: filter wheel:
unknown task status
112/004 Gantry software error: filter wheel:
112/005 Gantry software error: filter wheel:
unknown order transmitted to task
114/000 Gantry software error: light centering:
unknown order transmitted to task
117/000 Gantry software error: unknown set-up
message
117/002 Gantry set-up access: generator end of
exposure timeout
117/003 Gantry set-up access: communication
sync problem at set-up access
118/000 Gantry software error: grid: task triggered
by timeout (calibration time too long)
118/001 Gantry software error: grid: unknown task
return
118/002 Gantry software error: grid: unknown task
status
118/003 Gantry software error: grid: unknown task
status
118/004 Gantry software error: grid: unknown task
status
118/005 Gantry software error: grid: unknown
order transmitted to task
118/006 Gantry software error: grid: unknown
order transmitted to task
118/007 Grid movement: overcurrent detected
118/008 Grid movement: grid calibration time too
long
118/009 Grid movement: overcurrent detected
118/010 Grid movement: invalid offset
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 275 Chapter 7
118/011 Grid movement: invalid gain
118/012 Grid movement: limit of travel not found
120/000 Gantry software error: arm lock: unknown
task return
120/001 Gantry software error: arm lock: unknown
task status
120/002 Gantry software error: arm lock: unknown
task status
120/003 Gantry software error: arm lock: unknown
order transmitted to task
121/000 Gantry software error: rot. brake: unknown
order transmitted to task
121/001 Gantry software error: rot. brake: unknown
task status
121/002 Gantry software error: rot. brake: unknown
task status
121/003 Gantry software error: rot. brake: unknown
task return
121/004 Gantry software error: rot. brake: unknown
order transmitted to task
122/007 Gantry software error: Reset triggered by
Watch_dog_soft
122/013 Gantry software error: grid sensor
Presence_Trame (Grid_Presence) not
’high’ or ’low’
122/014 Gantry software error: tube housing ther-
mal contact not ’high’ or ’low’
151/000 Gantry: checksum error on battery saved
memory
151/021 Gantry interface: unexpected interrupt
received (level 0)
151/022 Gantry interface: unexpected interrupt
received (level 1)
153/075 Gantry ’Endurance’ error code: synchroni-
sation
153/076 Gantry ’Endurance’ error code: column
153/077 Gantry ’Endurance’ error code: compres-
sion
Code Message Text Description Recommended Actions

Chapter 7 Page no. 276 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
153/078 Gantry ’Endurance’ error code: compres-
sion arm rotation
153/079 Gantry ’Endurance’ error code: tube hous-
ing arm rotation
153/080 Gantry ’Endurance’ error code: lateral col-
limator
153/081 Gantry ’Endurance’ error code: rear colli-
mator
153/082 Gantry ’Endurance’ error code: light
centering device
153/083 Gantry ’Endurance’ error code: filter
153/084 Gantry ’Endurance’ error code: grid move-
ment
153/085 Gantry ’Endurance’ error code: tube hous-
ing angulation
153/086 Gantry ’Endurance’ error code: other func-
tion
155/082 Tube housing type/gantry configuration
incompatibility (focus/angulation)
155/083 Tube housing type/gantry configuration
incompatibility (focus/angulation)
156/008 Column up/down: Both end of travel sen-
sors are closed
156/009 Column up/down: is locked in both direc-
tions (from speed measurement)
156/010 Column up/down: signal Pres_Log_Asc
(Column_Soft_Pres) too long in ’low’ sta-
tus
156/011 Column up/down: signal Pres_Log_Asc
(Column_Soft_Pres) too long in ’low’ sta-
tus
156/012 Column up/down: signal Pres_Log_Asc
(Column_Soft_Pres) too long in ’high’ sta-
tus
156/013 Column up/down: signals Asc_Inactif &
R_On_Asc (Column_Power_Inactive &
On_Column_Return) both ’high’
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 277 Chapter 7
156/014 Column up/down:
Pres_Log_Asc*Pres_Oper_Asc*On_Asc
(Column_Soft_Pres*Column_Operator_Pr
es*On_Column) = 1 & R_On_Asc
(On_Column_Return) = 0
156/015 Column up/down:
Pres_Log_Asc*Pres_Oper_Asc*On_Asc
(Column_Soft_Pres*Column_Operator_Pr
es*On_Column) = 0 & R_On_Asc
(On_Column_Return) = 0 & Asc_Inactif
(Column_Power_Inactive) = 1
156/016 Power shutdown: uncontrolled movement
risk (column);Asc_Inactif & R_On_Asc
(Column_Power_Inactive &
On_Column_Return) both ’low’
156/017 Power shutdown: uncontrolled movement
risk (column);Asc_Inactif
(Column_Power_Inactive) ’low’ & column
defective
156/018 Power shutdown: uncontrolled movement
risk (column); signal Asc_Inactif
(Column_Power_Inactive) ’high’ during init
156/019 Column up/down: signal R_On_Asc
(On_Column_Return) ’high’ during init
156/020 Column up/down: direction of movement
is opposite to the desired direction
156/021 Column up/down: ’up’ end of travel sensor
is defective
156/022 Column up/down: ’down’ end of travel sen-
sor is defective
156/023 Power shutdown: uncontrolled movement
risk (column); R_On_Asc
(On_Column_Return) ’high’ without legal
command
156/024 Column up/down: column is driven up but
measured speed = 0 cm/s
156/025 Column up/down: column is driven down
but measured speed = 0 cm/s
156/027 Column up/down: emergency stop pro-
duced if movement continues more than
500 mS after command release
Code Message Text Description Recommended Actions

Chapter 7 Page no. 278 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
157/005 Compression: motor slipping during refer-
ence search
Check for play between com-
pression motor axis and pulley
(screw not tightened).
157/006 Compression: abnormal immobility of pad-
dle
Check for bad feedback from the
compression encoder.
157/007 Compression: loss of steps detected at
reference crossing
Check for defective cable on the
compression steel spring.
157/008 Compression: abnormal step generation
detected (during init)
157/009 Compression: signal Valid_Comp
(Comp_Valid) cannot inhibit paddle move-
ment (during init)
157/010 Compression: counter used by force sen-
sor no longer incremented
Loss of compression feedback; a square
wave should be always present between
TP14 and TP1 (0 V) of 800PL2.
157/011 Compression: position sensor closed too
long during compression
157/012 Compression: no reference crossing dur-
ing decompression
Check cable between 800PL8 and
800PL5.
157/013 Compression: signal Pres_Oper
(Operator_Presence) cannot inhibit pad-
dle movement (during init)
157/014 Compression: signal Pres_Ajust
(Comp_Adjust) too long in ’low’ status
157/015 Compression: signal Pres_Ajust
(Comp_Adjust) too long in ’high’ status
157/016 Compression: signal Pres_Log_Comp
(Comp_Soft_Pres) cannot inhibit paddle
movement (during init)
157/017 Compression: signal Pres_Log_Comp
(Comp_Soft_Pres) too long in ’low’ status
157/018 Compression: signal Pres_Log_Comp
(Comp_Soft_Pres) too long in ’high’ status
157/020 Compression: power supply problem (sig-
nal Alim_OK (Power_Supply_Active))
157/021 Compression: phase1 & phase2 cannot
be ’low’ at the same time (during init)
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 279 Chapter 7
157/022 Compression: signal Pres_Log_Comp
(Comp_Soft_Pres) ’high’ (during init)
157/023 Compression: slipping during decompres-
sion movement
Check for play between com-
pression motor axis and pulley
(screw not tightened).
158/009 Lateral collimator motor: loss of steps
detected at reference crossing
158/010 Lateral collimator motor: loss of steps
detected at reference crossing
158/011 Lateral collimator motor: loss of steps
detected at reference crossing
158/012 Lateral collimator motor: loss of steps
detected at reference crossing
158/013 Lateral collimator: no reference crossing
during movement
158/014 Lateral collimator: no reference crossing
during movement
158/015 Lateral collimator: no reference crossing
during movement
158/016 Lateral collimator: no reference crossing
during movement
158/017 Lateral collimator: optical sensor detection
problem
158/018 Lateral collimator: optical sensor detection
problem
158/019 Lateral collimator: slipping during collima-
tor movement.
Check for play between collima-
tor lateral motor axis and pulley
(screw not tightened). Possible
problem with 800PL1 (Gantry
power board)
159/009 Rear collimator motor: loss of steps
detected at reference crossing
159/010 Rear collimator motor: loss of steps
detected at reference crossing
159/011 Rear collimator motor: loss of steps
detected at reference crossing
159/012 Rear collimator motor: loss of steps
detected at reference crossing
159/013 Rear collimator: no reference crossing
during movement
Code Message Text Description Recommended Actions

Chapter 7 Page no. 280 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
159/014 Rear collimator: no reference crossing
during movement
159/015 Rear collimator: no reference crossing
during movement
159/016 Rear collimator: no reference crossing
during movement
159/017 Rear collimator: optical sensor detection
problem
159/018 Rear collimator: optical sensor detection
problem
159/019 Rear collimator: slipping during collimator
movement
Check for play between collimator rear
motor axis and pulley (screw not tight-
ened).
160/009 Front collimator motor: loss of steps
detected at reference crossing
160/010 Front collimator motor: loss of steps
detected at reference crossing
160/011 Front collimator motor: loss of steps
detected at reference crossing
160/012 Front collimator motor: loss of steps
detected at reference crossing
160/013 Front collimator: no reference crossing
during movement
160/014 Front collimator: no reference crossing
during movement
160/015 Front collimator: no reference crossing
during movement
160/016 Front collimator: no reference crossing
during movement
160/017 Front collimator: optical sensor detection
problem
160/018 Front collimator: optical sensor detection
problem
160/019 Front collimator: slipping during collimator
movement
Check for play between collima-
tor front motor axis and blade
(screw not tight).
Check for poor position of the
flag in optocoupler fork.
Check front blade calibration.
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 281 Chapter 7
161/000 Tube housing angulation: no reference
crossing during movement
161/001 Tube housing angulation: no reference
crossing during movement
161/009 Tube housing angulation: loss of steps
detected at reference crossing
161/010 Tube housing angulation: loss of steps
detected at reference crossing
161/011 Tube housing angulation: loss of steps
detected at reference crossing
161/012 Tube housing angulation: loss of steps
detected at reference crossing
161/013 Tube housing angulation: no reference
crossing during movement
161/014 Tube housing angulation: no reference
crossing during movement
161/015 Tube housing angulation: no reference
crossing during movement
161/016 Tube housing angulation: no reference
crossing during movement
162/006 Filter wheel: step number error between
two references
162/007 Filter wheel: step number error between
two references
162/008 Filter wheel: step number error between
two references
162/009 Filter wheel: step number error between
two references
162/010 Filter wheel: step number error between
two references
162/011 Filter wheel: no reference crossing during
rotation
162/012 Filter wheel: in standby, wheel moved from
stable position
164/001 Light centering device: signal
Centreur_OK (Light_Centering_OK) too
long in ’low’ status
168/020 Grid movement: power supply problem
(signal Alim_OK (Power_Supply_Active))
Code Message Text Description Recommended Actions

Chapter 7 Page no. 282 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
170/005 Inter-arm lock: signal Pres_Log_Ver
(Lock_Soft_Pres) too long in ’low’ status
170/006 Inter-arm lock: signal Pres_Log_Ver
(Lock_Soft_Pres) too long in ’high’ status
170/007 Power shutdown: uncontrolled movement
risk (arm lock); Eta_Verrou_Off
(Lock_Power_Inactive) ’low’ during init
170/008 Power shutdown: uncontrolled movement
risk (arm lock); Eta_Verrou_Off &
R_On_Verrou (Lock_Power_Inactive &
On_Lock_Return) both ’low’
170/009 Inter-arm lock: signals Eta_Verrou_Off &
R_On_Verrou (Lock_Power_Inactive &
On_Lock_Return) both in ’high’ status
170/010 Inter-arm lock: R_On_Verrou
(On_Lock_Return) = 0 &
Pres_Log_Ver*Pres_Oper_Ver*On_Verro
u
(Lock_Soft_Pres*Lock_Oper_Pres*On_Lo
ck) = 1
170/011 Inter-arm lock: signal R_On_Verrou
(On_Lock_Return) ’high’ without legal
command
170/012 Inter-arm lock: signal R_On_Verrou
(On_Lock_Return) ’high’ during init
170/013 Power shutdown: uncontrolled movement
risk (arm lock); R_On_Verrou
(On_Lock_Return) ’high’ without legal
command
171/005 Arm rot. brake: signal
Pres_Log_Frein_Rot
(Rot_Brake_Soft_Pres) from compres-
sion arm too long in ’low’ status
171/006 Arm rot. brake: signal
Pres_Log_Frein_Rot
(Rot_Brake_Soft_Pres) from compres-
sion arm too long in ’high’ status
171/007 Arm rot. brake: signal R_On_Frein
(On_Brake_Return) ’high’, but signal
Eta_Rot_Off (Rotat_Power_Inactive) ’low’
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 283 Chapter 7
171/008 Arm rot. brake:
Pres_Log_Frein_Rot*Pres_Oper*On_Frei
n_Rot
(Rot_Brake_Soft_Pres*Operator_Present
*On_Rotation_Brake) = 1 &
R_On_Frein_Rot
(On_Rotation_Brake_Return) = 0
171/009 Arm rot. brake: signal R_On_Frein_Rot
(On_Rotation_Brake_Return) ’high’ with-
out legal command
171/010 Power shutdown: uncontrolled movement
risk (arm brake); Eta_Rot_Off
(Rotat_Power_Inactive) &
R_On_Frein_Rot
(On_Rotation_Brake_Return) are ’low’
171/011 Power shutdown: uncontrolled movement
risk (arm brake); Eta_Rot_Off
(Rotat_Power_Inactive) ’low’ & comp. arm
defective
171/012 Power shutdown: uncontrolled movement
risk (arm brake); Eta_Rot_Off
(Rotat_Power_Inactive) ’low’ during init
171/013 Arm rot. brake: signal R_On_Frein_Rot
(On_Rotation_Brake_Return) ’high’ during
init
171/014 Power shutdown: uncontrolled movement
risk (arm brake); R_On_Frein_Rot
(On_Rotation_Brake_Return) ’high’ with-
out legal command
171/015 E76 Angle sensor
Failure
Rotation angle sensor frequency too high
or too low
Verify the rotation angle sensor
calibration, then the rotation
angle sensor itself (change if
necessary).
171/016 E76 Angle sensor
Failure
Rotation angle sensor in "dead" area for
more than 30 min
Verify the rotation angle sensor
calibration, then the rotation
angle sensor itself (change if
necessary).
171/017 E76 Angle sensor
Failure
Rotation angle sensor in "dead" area for a
short time
Verify the rotation angle sensor
calibration, then the rotation
angle sensor itself (change if
necessary).
171/018 E76 Angle sensor
Failure.
Angle sensor output not did not stabilize
during a minute following arm movement.
1. Reset DMR.
2. Verify sensor stabilization.
3. Check/change sensor.
Code Message Text Description Recommended Actions

Chapter 7 Page no. 284 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
171/019 E76 Angle sensor
Failure.
Angle sensor output change exceeded 5°
for more than a minute, with no arm move-
ment.
1. Reset DMR.
2. Check for friction point in sen-
sor; change if necessary.
3. If problem persists, note soft-
ware version and call your OLC.
171/020 E76 Angle sensor
Failure.
Angle sensor output change exceeded 5°
for more than a minute, with no arm move-
ment.
1. Reset DMR.
2. Check for friction point in sen-
sor; change if necessary.
3. If problem persists, note soft-
ware version and call your OLC.
171/021 E76 Angle sensor
Failure.
Angle sensor output changed by more
than 20°, with no arm movement.
1. Reset DMR.
2. Check sensor, connections,
and board 800PL5; change if
necessary.
171/022 E76 Angle sensor
Failure.
Angle sensor output changed by more
than 20°, with no arm movement.
1. Reset DMR.
2. Check sensor, connections,
and board 800PL5; change if
necessary.
171/023 E76 Angle sensor
Failure.
Angle sensor output changed by more
than 20°, with no arm movement.
1. Reset DMR.
2. Check sensor, connections,
and board 800PL5; change if
necessary.
171/024 E76 Angle sensor
Failure.
Angle sensor output changed by more
than 20°, with no arm movement.
1. Reset DMR.
2. Check sensor, connections,
and board 800PL5; change if
necessary.
172/003 Magnification sensor: unknown magnifica-
tion code
172/004 Gantry interface: arm lock temperature
calculation problem
172/005 Gantry interface: Bouclage_OK
(Loops_OK) signal ’low’ during init
172/006 Power failure due to CPU reset
If this appears at power on, the Gantry
interface board is probably defective.
172/007 Gantry interface: date write error
172/008 Gantry interface: time write error
172/009 Button interlock: signal
Presence_Operateur (Operator_Pres) too
long in ’high’ status
172/010 Button interlock: signal
Presence_Operateur (Operator_Pres) too
long in ’low’ status
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
ERR003.FM Page no. 285 Chapter 7
172/011 SID sensor: unknown SID code
172/012 Arm angulation sensor: unknown angula-
tion code
172/014 Gantry interface: date/time function read
error
172/015 Signal On_Asc (On_Column) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/016 Signal On_Centreur
(On_Light_Centering) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/017 Signal On_Comp ’high’ but Bouclage_OK
(Loops_OK) still ’low’
172/018 Signal On_Frein (On_Rotation_Brake)
’high’ but Bouclage_OK (Loops_OK) still
’low’
172/019 Signal On_Verrou (On_Lock) (Arm lock)
’high’ but Bouclage_OK (Loops_OK) still
’low’
172/020 Signal On_Trame (On_Grid) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/021 Signal Mot_Diaph_L_1
(Lat_Diaph_Motor_1) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/022 Signal Mot_Diaph_L_2
(Lat_Diaph_Motor_2) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/023 Signal Mot_Diaph_A_1
(Rear_Diaph_Motor_1) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/024 Signal Mot_Diaph_A_2
(Rear_Diaph_Motor_2) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/025 Signal Mot_Gaine_1 (Housing_Motor_1)
’high’ but Bouclage_OK (Loops_OK) still
’low’
172/026 Signal Mot_Gaine_2 (Housing_Motor_2)
’high’ but Bouclage_OK (Loops_OK) still
’low’
172/027 Signal Mot_Filtre_1 (Filter_Motor_1) ’high’
but Bouclage_OK (Loops_OK) still ’low’
Code Message Text Description Recommended Actions

Chapter 7 Page no. 286 ERR003.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card ERR 003 - Gantry and Generator Error Codes
172/028 Signal Mot_Filtre_2 (Filter_Motor_2) ’high’
but Bouclage_OK (Loops_OK) still ’low’
172/029 Signal R_Standby_Gaine
(Housing_Standby_Return) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/030 Signal R_Standby_Filtre
(Filter_Standby_Return) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/031 Signal R_Standby_Diap_L
(Lat_Diaph_Standby_Return) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/032 Signal R_Standby_Diap_P
(Rear_Diaph_Standby_Return) ’high’ but
Bouclage_OK (Loops_OK) still ’low’
172/055 Compression: one pedal pressed and sig-
nal Pres_Pedale (Pedal_Presence) ’low’
172/056 Compression: no pedal pressed and sig-
nal Pres_Pedale (Pedal_Presence) ’high’
172/064 Fuse fault: signal Fuse_OK ’low’
172/065 Power supply: ’ISO’ supply defective on
800-PL4
172/067 Power supply: ’+15V’ supply defective on
800-PL2
172/068 Power supply: one (or more) supplies
defective on 800-PL1 (except +27V)
172/088 Power micro-failure detected
Code Message Text Description Recommended Actions

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
DIAGS.FM Page no. 287
CHAPTER 8 DIAGNOSTIC INFORMATION
This chapter lncludes job cards describing Senographe 2000 D diagnostic operations.
Note:
While entering the Service Desktop or when doing diagnostics, a small Netscape WARNING win-
dow may appear in the upper left hand corner of the screen while the system is performing some
background functions. Ignore the message, it has no effect on the operation of the system.

Page no. 288 DIAGS.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 001 - AWS Sun-VTS Diagnostics
DIAG1.FM Page no. 289 Chapter 8
Job Card DIAG 001 - AWS Sun-VTS Diagnostics Chapter 8
1. INTRODUCTION
This Job Card describes Sun-VTS diagnostic procedures for the AWS (Acquisition WorkStation)
The AWS consists of a Solaris workstation, some components connected externally (e.g., modem) and a
set of subsystems. AWS diagnostics include diagnostics for the workstation hardware, software pro-
cesses in the workstation, and connection to the subsystems.
The objective of the diagnostics is to test AWS hardware components, locate any non-functional sub-
system, and then locate a non-functional FRU (Field Replaceable Unit).
1-1. Access to the tests
There are three ways to launch an AWS diagnostic:
• From a command window as
sdc or root user (local use only)
• From the prompt
ok (local use only)
2. COMMON PROCEDURES
2-1. Open a command window
Rightclick on the screen background; select Service tools and then select a Command window from root
menu.
2-2. Go to ok prompt
1. Open a command window
2. Log in as user
root
3. Type the command: halt
4. At prompt ok, type the command: setenv auto-boot? false
5. At prompt ok, type the command: reset
2-3. Relaunch AWS application
1. At prompt ok, type the command: setenv auto-boot? true
2. At prompt ok, type the command: reset
3. Log in as sdc user
2-4. Change user
Type the command: su - username and enter your password.

Chapter 8 Page no. 290 DIAG1.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 001 - AWS Sun-VTS Diagnostics
3. PROCESSOR TESTS
3-1. Short test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/fputest –v –n
3-2. Long test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/fputest –v –f
4. HARD DRIVE TESTS
4-1. Short test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the commands:
/opt/SUNWvts/bin/disktest –v -n -o dev=c0t0d0
/opt/SUNWvts/bin/disktest –v -n -o dev=c0t1d0
4-2. Long test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the commands:
/opt/SUNWvts/bin/disktest –v -f -o dev=c0t0d0
/opt/SUNWvts/bin/disktest –v -f -o dev=c0t1d0
5. PHYSICAL MEMORY TESTS
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/pmem –v -f -o size=100
6. VIRTUAL MEMORY TESTS
6-1. Short test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/vmet –v -f -o amount=100
6-2. Long test
• Type of test: SUN VTS
1. Open a command window as sdc user

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 001 - AWS Sun-VTS Diagnostics
DIAG1.FM Page no. 291 Chapter 8
2. Type the command: /opt/SUNWvts/bin/vmem –v –f -o reserve=100
7. VIDEO BOARD TESTS
• Type of test: SUN VTS
1. Open a command window as sdc
2. Type the command:
/opt/SUNWvts/bin/m64test
8. FLOPPY DRIVE TESTS
• Type of test: SUN VTS
1. Open a command window
2. Change user to root
3. Insert a formatted floppy disk
4. Type the command:
/opt/SUNWvts/bin/disktest –v -f -o dev=diskette
9. INTERNAL CD-ROM TESTS
9-1. Short test
• Type of test: SUN VTS
1. Open a command window as sdc
2. Type the command:
/opt/SUNWvts/bin/cdtest –v -n -o dev=c0t2d0
9-2. Long test
• Type of test: SUN VTS
1. Open a command window as sdc
2. Insert a formatted cd-rom disk (not blank)
3. Type the command:
/opt/SUNWvts/bin/cdtest –v –f -o dev=c0t2d0
10. EXTERNAL CD WRITER TESTS
10-1. Short test
• Type of test: SUN VTS
1. Open a command window as sdc
2. Type the command:
/opt/SUNWvts/bin/cdtest –v -n -o dev=c1t2d0
11. ADS-UPS LINE TESTS
11-1. Internal test
• Type of test: SUN VTS
1. Boot the station
2. Log as root
3. Type the commands:
ups stop
/opt/SUNWvts/bin/sptest –v -f -o dev=se0,L=I,T=b

Chapter 8 Page no. 292 DIAG1.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 001 - AWS Sun-VTS Diagnostics ups start
11-2. Full test without cable
• Type of test: SUN VTS
1. Boot the station
2. Log as root
3. Remove the UPS cable from the serial port b of AWS
4. Plug the loopback connector in serial port b of AWS
5. Type the commands:
ups stop
/opt/SUNWvts/bin/sptest –v -f -o dev=se0,L=P,T=b
ups start

6. Reconnect the cable when the test is complete
11-3. Full test with cable
• Type of test: SUN VTS
1. Boot the station
2. Log as root
3. Remove the cable from UPS serial port
4. Plug the DB9 loopback connector in the end of the cable
5. Re-connect the cable when the test is complete
6. Type the commands:
ups stop
/opt/SUNWvts/bin/sptest –v -f -o dev=se0,L=P,T=b
ups start
12. ADS-DMR LINE TESTS
12-1. Full test without cable
• Type of test: Unix command
1. Boot the station.
2. Log in as user roo.t
3. Remove the DMR cable from the port 1 of Digi Board (additional serial board.)
4. Plug the loopback connector in port 1 of Digi Board .
5. Type the command:
tip /dev/dty/a001s.
6. The system answers Connected.
7. Type some alphanumeric characters on the keyboard. If the board is good, they should be echoed
immediately on the screen
8. To exit the test, type the command:
<Return>~. . It may be necessary to repeat the sequence several
times.
9. Reconnect the cable when the test is complete
12-2. Full test with cable
• Type of test: Unix command

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 001 - AWS Sun-VTS Diagnostics
DIAG1.FM Page no. 293 Chapter 8
1. Boot the station
2. Log as user root
3. Remove the DMR cable from the DMR serial port
4. Plug the loopback connector in the end of the cable
5. Type the command:
tip /dev/dty/a001s
6. The system answers Connected.
7. Type some alphanumeric characters on the keyboard. If the cable is good, they should be echoed
immediately on the screen
8. To exit the test, type the command:
<Return>~. . It may be necessary to repeat the sequence several
times.
9. Reconnect the cable when the test is complete.
13. AWS-IDC LINE TEST (EXTERNAL)
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/nettest –v -f -o dev=hme0,target=IDC
14. MOTHER BOARD SERIAL PORTS TESTS
14-1. Internal test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/sptest –v –n -o L=I,T=a_b

Chapter 8 Page no. 294 DIAG1.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 001 - AWS Sun-VTS Diagnostics This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 002 - AWS Ethernet Port Diagnostics
DIAG2.FM Page no. 295 Chapter 8
Job Card DIAG 002 - AWS Ethernet Port Diagnostics Chapter 8
1. INTRODUCTION
This Job Card describes diagnostic procedures for the AWS (Acquisition WorkStation) Ethernet ports.
The AWS consists of a Solaris workstation, some components connected externally (e.g., modem) and a
set of subsystems. AWS diagnostics include diagnostics for the workstation hardware, software pro-
cesses in the workstation, and connection to the subsystems.
The objective of the diagnostics is to test AWS hardware components, locate any non-functional sub-
system, and then locate a non-functional FRU (Field Replaceable Unit).
1-1. Access to the tests
There are three ways to launch an AWS diagnostic:
• From a command window as
sdc or root user (local use only)
• From the prompt
ok (local use only)
2. COMMON PROCEDURES
2-1. Open a command window
Right-click on the screen background; select Service tools and then select Command window from the
root menu.
2-2. Go to ok prompt
1. Open a command window
2. Log in as user
root
3. Type the command: halt
4. At prompt ok, type the command: setenv auto-boot? false
5. At prompt ok, type the command: reset
2-3. Relaunch AWS application
1. At prompt ok, type the command: setenv auto-boot? true
2. At prompt ok, type the command: reset
3. Log in as sdc user
2-4. Change user
Type the command: su - username and enter your password.

Chapter 8 Page no. 296 DIAG2.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 002 - AWS Ethernet Port Diagnostics
3. AWS-IDC LINE TESTS
3-1. Internal test
• Type of test: Open boot command
1. Go to prompt ok
2. Type the command:
test net
3. Relaunch AWS application
• Result interpretation:
After the command has been entered, the following information is displayed:
Internal loopback test –
Result 1
Transceiver check – Result 2
where Result 1 and Result 2 are the results of each sub-test.
- If Result 1 is not
Succeeded, the AWS mother board is faulty.
- If Result 1 & Result 2 are
OK, the AWS Mother Board and the cable are good.
- If Result 1 is
OK & Result 2 is NOT OK, either the AWS transceiver or the cable may be faulty.
Repeat the test with a new cable to locate the bad device.
3-2. External test
• Type of test: SUN VTS
1. Open a command window as sdc user
2. Type the command:
/opt/SUNWvts/bin/nettest –v -f -o dev=hme0,target=IDC

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 002 - AWS Ethernet Port Diagnostics
DIAG2.FM Page no. 297 Chapter 8
4. AWS-NETWORK TESTS (INTERNAL)
4-1. Internal test
• Type: Open boot command
1. Go to prompt ok
2. Type the command:
watch-net-all
3. Relaunch AWS application
• Result interpretation:
After the command has been entered, the following information is displayed:
/pci@1f,0/pci@1/pci@1/SUNW,hme@0,1
Internal loopback test –
Result 1
Transceiver check – Result 2
Looking for Ethernet Packets
‘.’is a Good Packet. ‘X’ is a Bad Packet
Type any key to stop
...........
/pci@1f,0/pci@1,1/network@1,1
Internal loopback test –
Result 3
Transceiver check – Result 4
Looking for Ethernet Packets
‘.’is a Good Packet. ‘X’ is a Bad Packet
Type any key to stop
...........
The first group is for the additional ethernet board which manages communication with external network.
Result 1 and Result 2 are the results of each sub-test.
The second group is for the ethernet board of the mother board which manages communication with the
IDC. Do not consider it for this test.
- If Result 1 is different from
Succeeded, the AWS mother board is faulty.
- If Result 1 & Result 2 are
OK, the AWS Mother Board and the cable are good.
- If Result 1 is
OK, and Result 2 is NOT OK, either the AWS transceiver or the cable may be faulty.
Repeat the test with a new cable to locate the faulty device.

Chapter 8 Page no. 298 DIAG2.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 002 - AWS Ethernet Port Diagnostics This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 003 - AWS File System and SCSI Board Diagnostics
DIAG3.FM Page no. 299 Chapter 8
Job Card DIAG 003 - AWS File System and SCSI Board DiagnosticsChapter 8
1. INTRODUCTION
This Job Card describes diagnostic procedures for the AWS (Acquisition WorkStation) File system and
SCSI board.
The AWS consists of a Solaris workstation, some components connected externally (e.g., modem) and a
set of subsystems. AWS diagnostics include diagnostics for the workstation hardware, software pro-
cesses in the workstation, and connection to the subsystems.
The objective of the diagnostics is to test AWS hardware components, locate any non-functional sub-
system, and then locate a non-functional FRU (Field Replaceable Unit).
1-1. Access to the tests
There are three ways to launch an AWS diagnostic:
• From a command window as
sdc or root user (local use only)
• From the prompt
ok (local use only)
2. COMMON PROCEDURES
2-1. Open a command window
Rightclick on the screen background; select Service tools and then select a Command window from root
menu.
2-2. Go to ok prompt
1. Open a command window
2. Log in as user
root
3. Type the command: halt
4. At prompt ok, type the command: setenv auto-boot? false
5. At prompt ok, type the command: reset
2-3. Relaunch AWS application
1. At prompt ok, type the command: setenv auto-boot? true
2. At prompt ok, type the command: reset
3. Log in as sdc user
2-4. Change user
Type the command: su - username and enter your password.

Chapter 8 Page no. 300 DIAG3.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card DIAG 003 - AWS File System and SCSI Board Diagnostics
3. FILE SYSTEM FSCK TEST
• Type: Unix command
• Description:
This test checks the file system. It must be launched when file system is suspected to be corrupt (station
has been powered off manually, files can not be found, boot does not succeed, or some basic system
functions are not available). These problems are detected by the operating system during the boot and
the user is warned to run fsck.
• Procedure:
1. Shutdown the station
2. Log in as root
3. Type the command:
fsck -y
4. Type the command: echo $status
5. If the result is not equal to 0 or 40:
a. Backup the AWS configuration on floppy (see D/R Job Card
Backup and Restore) if not done
since the last configuration modification.
b. Re-install the AWS system from scratch (see D/R Job Card
AWS Software Install)
4. SCSI BOARD TESTS
• Type: Open boot command
1. Go to prompt ok
2. Type the command:
probe-scsi-all
3. Relaunch AWS application
• Result interpretation:
If no SCSI identifier is found, the SCSI board is faulty.
On a system with a cd-writer, two identifiers should be detected, the SCSI controller itself and the CD-
writer.
An example of a result showing two SCSI identifiers is given below:
/pci@1f,0/pci@1/pci@1/SUNW,isptwo@4

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renew.fm Page no. 301
CHAPTER 9 RENEWAL PARTS
1. INTRODUCTION
This chapter provides information for the identification of Renewal Parts for Senographe 2000 D equip-
ment. A list of contents is given overleaf.

Page no. 302 Renew.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Illustration 1 - Senographe 2000D System 2298472
Illustration 2 - Gantry Main Harness (white, from generator to gantry, 2225059)
Illustration 3 - Conditioner-Bulkhead Harness (black, from gantry to generator, 2270037)
Illustration 4 - Bulkhead/IDC-AWS-UPS Harness (black, from cart to gantry, 2231556)
Illustration 5 - Senographe Onyx (2302045)
Illustration 6 - Senographe Onyx (2302045) (continued)
Illustration 7 - Senographe Onyx Gantry (2302047)
Illustration 8 - Column Assy (2223734)
Illustration 9 - Column Assy (2223734) (continued)
Illustration 10 - Column Assy (2223734) (continued)
Illustration 11 - Column Assy (2223734) (continued)
Illustration 12 - Column Assy (2223734) (continued)
Illustration 13 - Onyx Gantry Board Assemblies (2302050)
Illustration 14 - Examination Arm (2228940)
Illustration 15 - Examination Arm (2228940) (continued)
Illustration 16 - Examination Arm (2228940) (continued)
Illustration 17 - Examination Arm (2228940) (continued)
Illustration 18 - Compression Arm (2215680)
Illustration 19 - Compression Arm (2215680) (continued)
Illustration 20 - Compression Arm (2215680) (continued)
Illustration 21 - Compression Arm (2215680) (continued)
Illustration 22 - Generator + Conditioner Cabinet (2302051)
Illustration 23 - Generator + Conditioner Cabinet (2302051) (continued)
Illustration 24 - Generator + Conditioner Cabinet (2302051) (continued)
Illustration 25 - Generator With HV Unit (2300422)
Illustration 26 - Generator With HV Unit (2300422) (continued)
Illustration 27 - Generator With HV Unit (2300422) (continued)
Illustration 28 - Generator With HV Unit (2300422) (continued)
Illustration 29 - Accessories Kit for ONYX (2302267)
Illustration 30 - Accessories Kit for ONYX (2302267) (continued)
Illustration 31 - Miscellaneous items
Illustration 32 - Console AWS Onyx (2302056)
Illustration 33 - Console AWS Onyx (2302056) (Continued)
Illustration 34 - Console AWS Onyx (2302056) (Continued)
Illustration 35 - Acquisition Workstation Onyx (2293685

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 303 Chapter 9
Renewal Parts List Chapter 9
2. FRU INFORMATION
• Information shown for each FRU (Field Replaceable Unit) is explained below.
• Wherever possible, the FRU or its position is illustrated for easy identification.
Left table with light box
2190112
FRU: 2
Job Card D/R031
• Light box Tube
2190113
FRU: 1
Job Card D/R032
• Lightbox AC Power Cable
2190114
FRU: 2
FRU name
Part number for ordering
FRU Codes:
FRU: 1 Critical
FRU: 2 Not Critical
FRU: N Not Available

The quantity is 1 if not specified
Notes:
These may include:
Reference to an illustration with more information.
Reference to a relevant Disassembly/Reassembly
Job Card or items included with the FRU, etc.
Items shown in this way are included in
the FRU shown immediately above, but
may also be ordered separately.

Chapter 9 Page no. 304 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
3. RENEWAL PARTS
ILLUSTRATION 1 - SENOGRAPHE 2000D SYSTEM 2298472
FRU detail: Illustration xx
Console AWS Onyx
2302056
FRU: N
SENOGRAPHE ONYX
2302045
FRU: N
Gantry main harness (packaged)
NOT ILLUSTRATED:
Conditioner-bulkhead harness
FRU: 2
2270037
(black, from gantry to generator)
Bulkhead/IDC-AWS-UPS harness
FRU: N
2231556
(black, from cart to gantry)
FRU: 2
2235109
FRU detail: Illustration 5, 6
FRU detail: Illustration 32
FRU detail: Illustration 2
FRU detail: Illustration 3 FRU detail: Illustration 4
Accessories kit for Onyx
FRU: N
2302267
FRU detail: Illustration 29, 30,
Gantry main harness
(white, gantry to generator)
FRU: N
2225059
Miscellaneous items
FRU detail: Illustration 31

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 305 Chapter 9
ILLUSTRATION 2 - GANTRY MAIN HARNESS (WHITE, FROM GENERATOR TO GANTRY, 2225059)
4-Pin HV Cable assembly
45475034
GANTRY MAIN HARNESS (NOT ILLUSTRATED):
FRU: 1
Gasket for C185129
C133902
FRU: 1
Set of cables, main harness
2231108
FRU: 1
• Optical fiber cable (Gene - Transmit)
45553497
FRU: 1
• O
ptical fiber cable (Gene - Receive)
45553498
FRU: 1
• O
ptical fiber cable (Arm Sync.)
45553499
FRU: 1
• O
ptical fiber cable (send PM)
45553500
FRU: 1

Chapter 9 Page no. 306 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 3 - CONDITIONER-BULKHEAD HARNESS (BLACK, FROM GANTRY TO GENERATOR,
2270037)
AWS - Generator link
2226623
CONDITIONER-BULKHEAD HARNESS (NOT ILLUSTRATED):
FRU: 2
IDC - Generator link
2226613
FRU: 2
cable detector ac su
pply (W2)
2225467
FRU: 2
Cable UPS ac supply (W5)
2225470
FRU: 2
Conditioner control
2226611
FRU: 2
Water hose assy. (blue)
2266811
FRU: 1
Air hose assies.
2231550 and 2231551
FRU: 2
Water hose assy.
2266812
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 307 Chapter 9
ILLUSTRATION 4 - BULKHEAD/IDC-AWS-UPS HARNESS (BLACK, FROM CART TO GANTRY, 2231556)
AC UPS Power
2231555
BULKHEAD/IDC-AWS-UPS HARNESS (NOT ILLUSTRATED):
FRU: 2
Detector PS-IDC Power
2226621
FRU: 2
AWS-DMR Link
2226625
FRU: 2
UPS-DMR Link
2226615
FRU: 2
IDC-DMR Link
2240227
FRU: 2
Conditioner control
2226612
FRU: 2
Fiber-o
ptic cable
2226619
FRU: 2
Conduit
2191670
FRU: 2

Chapter 9 Page no. 308 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 5 - SENOGRAPHE ONYX (2302045)
• Console top cover assy.
Senographe Onyx Gantry
RX Maxiray 70 TH-D Tube Housing
Image Receptor S2000D CDRH
Bucky S2000D CDRH
2228039
FRU: 1
2302047
FRU: N
2218380
FRU: 1
Job Card D/R 011
2225445
FRU: 1
2224137
FRU: 1
Job Card D/R 103
Image Receptor additional mounting hardware kit
(not illustrated)
FRU: 2
2231485
Generator + Conditioner Cabinet
2302051
FRU: N
FRU detail: Illustration 7
FRU detail: Illustration 22, 23, 24
Console CDRH EMC S2000D
2223736
FRU: 1

• Pushbutton without light
2227911
• Pushbutton with li
ght
2227912
FRU:1
• Button
45474409
FRU:1
Detector Water/Air Harness
2235107
FRU: 1
(packaged)
• Label (cleaning)
2153236
FRU:1
FRU:1
• FFDM Grid
2211408
FRU: 2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 309 Chapter 9
ILLUSTRATION 6 - SENOGRAPHE ONYX (2302045) (CONTINUED)
Leaded cover, left (packaged)
2235126
FRU: 1
Leaded cover, right (packaged)
2235125
FRU: 1
Front panel, screen printed
2225775
FRU: 2
Central cover (packaged)
2235124
FRU: 1
Cover, right, labelled (packaged)
2235120
FRU: 1
Compression cover, right
2290259
FRU: 1
Compression cover, left
2290258
FRU: 1
Column cover, right
2235117-2
FRU: 2
Column cover, left
2235118-2
FRU: 2
Ccover, left, labelled (packaged)
(not visible)
FRU: 1
2235121
Inspection door assembly (packaged)
(not visible)
FRU: 1
2235119
Cover mounting hardware kit (not illustrated)
FRU: 1
White cap (to hide cover screws)
(not illustrated)
FRU: 1
45559730
Set of four silk-screened
covers
45561897
FRU:1
45560702

Chapter 9 Page no. 310 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 7 - SENOGRAPHE ONYX GANTRY (2302047)
Examination Arm
Column Assembly (packaged)
Quadruple Footswitch (not illustrated)
2234897
FRU: 1
Double Footswitch
S2000D Display assembly (packaged)
2235105
FRU: 1
Onyx Gantry board assemblies
2302050
FRU: N
2228940
FRU: N
2235104
FRU: 2
2235874
FRU: 1
FRU detail: Illustration 14, 15, 16, 17
FRU detail: Illustration 8, 9, 10, 11, 12
FRU detail: Illustration 13
Column Assembly
2223734
FRU: N

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 311 Chapter 9
ILLUSTRATION 8 - COLUMN ASSY (2223734)
Elevator Motor wired
2151017
FRU: 1
Emergency stop button D30 RED 1 RT
45067225
FRU: 2
Detector Power Suppy
2210695
FRU: 1
Job Card D/R 102
Gas Spring
91726753
FRU: 1
Job Card D/R 016

Chapter 9 Page no. 312 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 9 - COLUMN ASSY (2223734) (CONTINUED)
Belt 73-tooth 12 T5/365
91726709
FRU: 1
Capacitor AL 33000 UF 63V
99091203
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 313 Chapter 9
ILLUSTRATION 10 - COLUMN ASSY (2223734) (CONTINUED)
A
A
Bearing, Ball, single
FRU: 2
99142807
RO 15 mm 32 mm
Bearing, Ball
FRU: 2
91726743
15 x 32 x 13 - 2 RO
Roller assembly
FRU: 1
2260917
single RO 6 mm 19 mm

Chapter 9 Page no. 314 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 11 - COLUMN ASSY (2223734) (CONTINUED)
A
B
A
B
Encoder Support,
FRU: 1
45554615
glued
Elevator Encoder cable
45553524
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 315 Chapter 9
ILLUSTRATION 12 - COLUMN ASSY (2223734) (CONTINUED)
B
A
CC
V1
V2
B
C
A
Optointerrupter (elevator at low)
45553520
FRU: 1
Optointerrupter (elevator at high)
45553521
FRU: 1

Chapter 9 Page no. 316 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 13 - ONYX GANTRY BOARD ASSEMBLIES (2302050)
800 PL2 Gantry interface board (packaged)
2247116
FRU: 1
800 PL1 Gantry power board
2102082-8
FRU: 1
Cable PL1/PL2-1
45553502
FRU: 1
Cable PL1/PL2-2
45553503
FRU: 1
Jumper, red (for CPU board)
99171545
FRU: 2
Battery, lithium 850 mAh
99171019
FRU: 1
800 PL3 CPU Gantry Onyx M3
2298456
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 317 Chapter 9
ILLUSTRATION 14 - EXAMINATION ARM (2228940)
Electro-magnet assy. with bracket
45560090
FRU: 1
Compression Arm
2215680
FRU: N
FRU detail: Illustration18, 19, 20, 21

Chapter 9 Page no. 318 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 15 - EXAMINATION ARM (2228940) (CONTINUED)
A
A B
B
C
C
Anti-rotation bushing
45554501
FRU: 2
Wired socket
(centering light)
2251680
FRU: 1
Optical box
45562593
FRU: 1
C ollim ator
2229927-2
FRU: 1
Job Card D/R 007
Mirror
FRU: 1
Belt 43-Tooth 10 T5/215
91726679
FRU: 1
Rocker switch 3 pos.
2186192
FRU: 1800 PL6 Tube housing
arm board (packaged)
2245986
FRU: 1
3-filter wheel assy.
45560092
FRU: 1
Job Card D/R 015
Screw (Tube rotation)
2232982
FRU:1
2128546
Job Card D/R 013
Halogen Lamp 12 V 1000 W
2188568
FRU: 1
A nti-scatter box
2223340
FRU: 2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 319 Chapter 9
ILLUSTRATION 16 - EXAMINATION ARM (2228940) (CONTINUED)
2139272 ADW REV. A
Optointerrupter (filter position)
45553543
FRU: 1
Tube Angulation Motor assy
45553539
FRU: 1
Optointerrupter (tube housing position)
45553546
FRU: 1
Tube Housing Fan assy.
36006829
FRU: 1
Tube Housing Arm Buttons
2254365
FRU:1

Chapter 9 Page no. 320 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 17 - EXAMINATION ARM (2228940) (CONTINUED)
B
D
A
A
B
C
C
D
Brake assy., wired
2178230
FRU: 1
Kit for compression cable
2248165
FRU:1
Examination Arm
Angle indicator
45554401
FRU:1
Ball bearing
91632919 FRU:2
Belt 49-tooth
91726680
FRU:1
Compression belt assembly
36006659
FRU:1
2210109
FRU:2
Pawl
Brake adjustment tool
56052522
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 321 Chapter 9
ILLUSTRATION 18 - COMPRESSION ARM (2215680)
A
A
D
C
B
800PL5 Compression distribution
FRU: 1
2198843-3
board (
packaged)
Angle sensor assy.
2229137
FRU: 1
Belt
36005885
FRU:1
Encoder support glued
45554615
FRU:1
800-PL11 Encoder board
2148752
FRU:1
Magnification stand holder left
36005971
FRU:2
Magnification stand holder
right
45554436
FRU:2
Release handle kit (packaged)
2235110
FRU:1

Chapter 9 Page no. 322 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 19 - COMPRESSION ARM (2215680) (CONTINUED)
B
C
D
800PL15 Tilt sensor board
2210221
FRU: 1
800PL14
Magnification
Sensor board
45553870
FRU:1
Ball bearing 6X19X6
99146577
FRU:1
Ball bearing, linear
91633373
FRU:2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 323 Chapter 9
ILLUSTRATION 20 - COMPRESSION ARM (2215680) (CONTINUED)
Compression motor assembly
45553529
FRU: 1
Cable (SID) W37
45553527
FRU:1
Tilting
45553534 FRU:1
Ma
gnification
45553535
FRU:1
Paddle position
45555537
FRU:1

Chapter 9 Page no. 324 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 21 - COMPRESSION ARM (2215680) (CONTINUED)
Paddle support assembly
2225076
FRU: 1
• Cable (
paddle fine movement)
45553536
FRU: 1
800-PL9 Encoder board
45554409
FRU: 1
800-PL8 Paddle board
45554410
FRU: 1
Set of handles
2227069
FRU: 1
• Set of 4 silk-screened covers
45561897
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 325 Chapter 9
ILLUSTRATION 22 - GENERATOR + CONDITIONER CABINET (2302051)
Right Cover Generator (packed)
2235116
FRU: 2
Front Cover Generator (packed)
2235114
FRU: 2
Rear Cover Generator (packed)
2235113
FRU: 2
Left Cover S2000D Generator (packed)
2235115
FRU: 2
Top Cover Generator (packed)
2235112
FRU: 2
Generator with HV unit
FRU: N
2300422
NOT ILLUSTRATED:
FRU detail: Illustration 25, 26, 27, 28

Chapter 9 Page no. 326 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 23 - GENERATOR + CONDITIONER CABINET (2302051) (CONTINUED)
200PL4 Relay board
2106963-2
FRU: 1
200PL2 Supply command board
2121398
FRU: 1
200PL1 Mains distribution board
2108592-2
FRU: 1
701HV unit
2107399
FRU: 1
Conditioner
2210405-5
FRU: 1
• Rela
y K1, K2
2100312
FRU:1
• Rela
y X1
91682082
FRU:1
• Coiled Resistor 280W
99182099
FRU:1
• Coiled Resistor 160W
99120760
FRU:1
• Fast Fuse 27 X 27 125A 150VAC
91684676
FRU:2
• Set of Fuses (6 x 32)
58079166
FRU:1
Coaxial cable
HV unit IC
45203215 or 45203148
FRU:1
CPU Power supply cable
45440115
FRU:1
Data control cable
45440116
FRU:1
Coolant kit for install or replace
FRU: 1
2260360
Line Filter Box
2102774
FRU: 2
250 V distribution cable (WG4)
2110070
FRU:2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 327 Chapter 9
ILLUSTRATION 24 - GENERATOR + CONDITIONER CABINET (2302051) (CONTINUED)
Circuit breaker 30 A
2228461
FRU: 1
Mains distribution
2221053-2
FRU: 1
Operator Interface
2221381
FRU: 1
Power supply 220V.23AMP
2228059
FRU:1
Line filter 250VAC
2229249
FRU:2
Service Interface 2221383 FRU:1

Chapter 9 Page no. 328 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 25 - GENERATOR WITH HV UNIT (2300422)
Transportation wheel D2X50mm DELRIN
91727258
FRU: 1
Fixed jack D=40 M10X25 HR:43
91719024
FRU: 2
A
Capacitor unit
2110227
FRU: 1
AFiltering inductor 2X125 UH 5%
45066952
FRU: 2
Diode module DD31N 1.4 KV 60A UL
58079525
FRU: 1
Thyristor (TH1) (TH2)
58079220
FRU: 1
Thermostat 55/45
o
C 1R 6A F6
58079252
FRU: 1
Thermostat 40/30
o
C 1T 6A F6
58079253
FRU: 1
300-PL2 Anode Starter Power Board
45553694
FRU: 1
300PL1 Inverter Board
2121509-2
FRU: 1

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 329 Chapter 9
ILLUSTRATION 26 - GENERATOR WITH HV UNIT (2300422) (CONTINUED)
Transformer 9kW 52A 415V
2262737
FRU: 1
Cable REL/CTRS (W10)
2110066
FRU: 2
Cable PSRS/CTRS (W12)
2110067
FRU: 2
Line Supply Contactor cable (W14)
2111145
FRU: 2
Capacitor AL 2200µF 350V
58079856
FRU:1
Diode DD31N400K (D24, D25)
91701990
FRU: 2
Diode DD61N400K
(D26, D27)
91707991
FRU: 2
Main Emergency switch
91717300
FRU:1
Contactor 3-P 660V/18A
91728240
FRU:1
Fuse Holder 2P 60x27 600V
91685123
FRU:2
Fast Fuse 27x60 35A 600V C 100KA
99184154
FRU:1

Chapter 9 Page no. 330 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 27 - GENERATOR WITH HV UNIT (2300422) (CONTINUED)
A
A
Power supply 50W 5V5A 12V2A - 12V0.5A
2264280
FRU: 1
400-PL1 Generator Command Board
2107747
FRU: 1
400-PL2 Generator Interface Board
2198369
FRU: 1
FRU: 1
400-PL3 CPU Gene. Onyx M3
2298455

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 331 Chapter 9
ILLUSTRATION 28 - GENERATOR WITH HV UNIT (2300422) (CONTINUED)
Heater & bias cable
45474577
FRU: 1
Cable 400PL1 - 200PL2
2110259
FRU: 2
Cable 400PL1 - 300PL1
2110262
FRU: 2
Cable 400PL1 - 300PL2
2110264
FRU: 2
Cable 400PL1 - 200PL2
2110271
FRU: 2
Cable 40-cond. 0.50 m
C801418
FRU: 1

Chapter 9 Page no. 332 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 29 - ACCESSORIES KIT FOR ONYX (2302267)
19 x 23 cm Compression paddle
FRU: 2
2218965-2
Axillary paddle for S2000D
FRU: 1
2228169-2
Calibration plate
FRU: 2
2233248
Magnification stand CDRH
FRU: 1
36006563
Magnet D4X20
FRU:2
91671876
Shoulder guard (packaged)
FRU: 1
2227387
• Plastic
plate for 19 x 23 paddle
2251745
FRU: 1
• Plastic plate for axillary paddle
2255606
FRU: 1
Installation kit Senographe 2000 D
2228457
FRU: 2
Square spot compression paddle
FRU: 2
Round spot compression paddle
FRU: 2
2218967-32218966-3
• Plastic
plate for square paddle
2251747
FRU: 1
• Plastic
plate for round paddle
2251746

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 333 Chapter 9
ILLUSTRATION 30 - ACCESSORIES KIT FOR ONYX (2302267) (CONTINUED)
Service terminal cable EMC
FRU: 1
2116257
Line supply cable
FRU: 2
2114797
Chassis
FRU: 2
2238771Loopback kit
FRU: N
2247879
Gate
FRU: 2
2238777
Purge tool
FRU: 1
2249359
• Loo
pback connector UPS
• Loopback connector DMR panel
2248430
FRU:2
• Loopback connector FO SC
2248316
FRU:2
• Wall connector (2 x O
ptical Fiber)
2231528
F R U :2
• Wall connector (2 x O
ptical Fiber,
2255935
F R U :2
w ith
plug)
• Adaptor sub D 9F/25M
2198175-2
FRU:2
2248665
FRU: 2
Console-GeneratorCable L=10m
FRU: 1
2150064
(packaged)
Emergency kit
FRU: 1
2223818

Chapter 9 Page no. 334 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 31 - MISCELLANEOUS ITEMS
Bucky FFDM without grid CDRH
FRU: 1
2222414
NOT ILLUSTRATED:
Seno
graphe 2000 D p.i.m.
FRU: 2
2228460-100
DMR+ & 2000D Schematics
FRU: 2
2269412-100
Seno 2000D DICOM Conformance statement
FRU: 2
2246811-100
Detector Temperature Troubleshooting kit
FRU: 1
2303052
Senographe 2000 D o.m. (English)
FRU: 2
2285097-100
Senographe 2000 D s.m.
FRU: 2
2302228-100
Documentation On
yx (2302059):
Light meter
FRU: 2
2109716
Breast Holder Cover FFDM
FRU: 1
2270410
• S
yringe kit
2270096
F R U : 1
Connector re
place kit for
compression motor cable
2247102
FRU:2
IQST Phantom
2222686
FRU: 2
Bucky support (packaged)
FRU: 2
2235128
FFDM Biopsy Paddle assy.
FRU: 1
2248758-2
FFDM Biopsy Localizing Pointers
FRU: 1
36005574

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 335 Chapter 9
ILLUSTRATION 32 - CONSOLE AWS ONYX (2302056)
Mouse support
FRU: 2
2190155
Accessory storage
FRU: 2
2232766
Set of wheels
FRU: 2
2190105
Right table and keyboard support
FRU: 2
2190110
Side covers, left and right
FRU: 2
2190106
Rear cover
FRU: 2
2232764
Left table with light box
FRU: 2
2190112
Light box tube
FRU: 1
2190113
Light box with AC power cable
FRU: 2
2190114
Monitor AWS Hi-Brite 21"
2305021
FRU: 1
UPS with support
FRU: 2
2295490 Battery for UPS
FRU: 2
2295553
Onyx Console
2302058
FRU: N
Cart adaptor for Sun keyboard
2265802
FRU: 2

Chapter 9 Page no. 336 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 33 - CONSOLE AWS ONYX (2302056) (CONTINUED)
Acquisition Workstation Onyx
FRU: N
2293685
IDC processor assembly
FRU: 1 2286100
FRU detail: Illustration 35
NOT ILLUSTRATED:
IDC Jumpstart diskette
2304217
FRU:N
IDC Application software M3
2303997
FRU:N
Cable, extension, M/F IEC 1.2M
FRU: 1
2189417
Cable, 0.5M, W/Multiplug 4 UTE
FRU: 1
2173762
Gn/Ylw cable C5-C5 AWG12 LG400
FRU: 1
2181484
Gn/Ylw cable Faston-Fork AWG LG150
FRU: 1
2191674
Gn/Ylw cable C5-C5 AWG12 LG150
FRU: 1
2181485
Gn/Ylw cable C5-Faston AWG12 LG150
FRU: 1
2181486
Gn/Ylw cable C5-C5 AWG12 LG800
FRU: 1
2181487
Gn/Ylw cable C5-Faston AWG12 LG250
FRU: 1
2181488
Gn/Ylw cable Faston-Faston AWG12 LG150
FRU: 1
2195199
Cable UPS-DS
2190154
FRU:1
Global Modem kit
FRU: 1
2245794
• Global Modem ada
pter assy. for Europe
FRU: 2
2249450
• Cable RJ11 25 ft.
FRU: 2
2216697
FFDM Bar Code Reader
FRU: 1
2258242
FFDM-IDC S/W Media set
FRU: 1
2303995

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
Spares-Onyx.fm Page no. 337 Chapter 9
ILLUSTRATION 34 - CONSOLE AWS ONYX (2302056) (CONTINUED)
AWS-Modem cable
FRU: 1
2226627
IDC-AWS cable
FRU: 1
2226628
Patch panel - local printer cable
FRU: 1
2226633
Patch panel - AWS (external ethernet) cable
FRU: 1
2226632
Patch panel - modem cable
FRU: 1
2226630
Proprietary key cable
FRU: 1
2244433
AW S H i- B r ight monitor cable
FRU: 2
2306204

Chapter 9 Page no. 338 Spares-Onyx.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Renewal Parts List
ILLUSTRATION 35 - ACQUISITION WORKSTATION ONYX (2293685)
• AW S jumpstart floppy
FRU: 1
2232049-4
Ethernet/SCSI combo adaptor, Sun
FRU: 1
2258672
2 x serial port adaptor
FRU: 1
2232092
AWS Fan U10 REF 370-3169 SUN
FRU: 1
2232050
Floppy drive REF 370-3159 SUN U10
FRU: 1
2232053
Power supply 250W REF 370-3169 SUN
FRU: 1
2232051
CD-ROM drive, SUN
FRU: 1
2299141
Hard disk, internal, 20 GB, 7200 RPM
FRU: 1
2299142
ADS Workstation Onyx
2299357
FRU: 1
Job Card D/R 107
CD-Writer with protective label
FRU: 1
2303925
Keyboard, Type-6 Unix, SUN
FRU: 1
2240569
FFDM Software backup kit
FRU: 1
2232045-4
Mouse, SUN
FRU: 1
2265191-2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
D-R.FM Page no. 339
CHAPTER 10 DISASSEMBLY/REASSEMBLY
This chapter contains information for the removal and replacement of Senographe 2000 D components.

Page no. 340 D-R.FM
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 000 - Calibration after FRU Replacement
DR_000.fm Page no. 341 Chapter 10
Job Card D/R 000 - Calibration after FRU Replacement Chapter 10
1. CALIBRATION AFTER FRU REPLACEMANT
• The table below lists Calibrations and Checks which must be carried out after replacing the listed
FRUs (Field replaceable Unit).
• The Calibrations and Checks listed should normally be carried out in the order given.
D/R FRU Required Calibrations and Checks
002 Paddle board 800PL8 CAL 006 Comp. Force detector
007 Collimator CAL 011 Beam Alignment
CAL 012 Collimator Format
011 X-ray Tube Housing CAL 002 X-ray Bias Scale factor
CAL 001 Heater Scale Factor
CAL 003 Tube Heater Current
CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 017 AOP
CAL 016 Half-Value Layer
CAL 020 mAs Non-linearity
VF 008 Flat Field Test
012 Centering Lamp CAL 019 Light Beam
013 Mirror CAL 019 Light Beam
015 Fillter Wheel CAL 016 Half-Value Layer
025 HV Tank CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 003 Tube Heater Current
CAL 017 AOP
026 Mag Holder CAL 014 Detector Gain
027 Arm Angle Sensor CAL 009 Compression Arm Angle
103 Detector CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 012 Collimator Format
CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 015 Conversion Factor
CAL 017 AOP
CAL 020 mAs Non-linearity
VF 008 Flat Field Test
104 IDC CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 015 Conversion Factor

Chapter 10 Page no. 342 DR_000.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 000 - Calibration after FRU Replacement
Generator CPU Board 400PL3 CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale factor
CAL 003 Tube Heater Current
CAL 004 Tube mA Measurement
CAL 005 kV Scale Factor
CAL 017 AOP
CAL 016 Half-Value Layer
Generator I/F Board 400PL2 CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale factor
CAL 003 Tube Heater Current
Gene. Command board 400PL1 CAL 001 Heater Scale Factor
CAL 002 X-ray Bias Scale factor
CAL 003 Tube Heater Current
Gantry CPU Board 800PL3 Note: The followingf calibrations are required if
data has not been saved:
CAL 010 Tube Tilt
CAL 012 Collimator Format
CAL 019 Light Beam
CAL 009 Compression Arm Angle
VF 008 Flat Field Test
Gantry Power Board 800PL1 CAL 006 Comp. Force detector
VF 008 Flat Field Test
Compression Distri. Board 800PL5 CAL 009 Compression Arm Angle
CAL 007 Breast Thickness
Tube CAL 010 Tube Tilt
CAL 011 Beam Alignment
CAL 012 Collimator Format
CAL 013 Bad Pixel
CAL 014 Detector Gain
CAL 015 Conversion Factor
CAL 017 AOP
CAL 020 mAs Non-linearity
VF 008 Flat Field Test
Bucky CAL 014 Detector Gain
19 x 23 Paddle CAL 017 AOP
Breast Holder CAL 017 AOP
RCN Filter CAL 016 Half-Value Layer
D/R FRU Required Calibrations and Checks

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 002 - Paddle board 800PL8
DR_002.fm Page no. 343 Chapter 10
Job Card D/R 002 - Paddle board 800PL8 Chapter 10
1. SUPPLIES
• Paddle board 800PL8.
2. TOOLS
• 1 mm and 2.5 mm Allen wrenches.
• Circlip pliers (5 mm circlip).
• Flat screwdriver, 1/4".
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Removal of paddle board 800PL8
Refer to Illustration 1.
1. Remove buttons B1 and B2 (1 mm Allen wrench).
2. Remove screws S1 and S2 from the two plastic half-covers.
3. Remove the two plastic half-covers.
4. Set the arm in the horizontal position.
5. Switch OFF the DMR electrical power.
6. Slacken locking screw on encoder wheel C (21).
7. Remove circlip (18) from one end of the shaft only, and remove washers 34,17 and 16.
8. Remove the two mounting screws (23/24) from the support plate.
9. Gently extract the encoder wheel from the optointerrupter by pushing it downwards, and then push
the shaft downwards until the circlip groove is at the encoder wheel level.
10. Remove the paddle board support plate.
11. Cut the tyraps to release connector XJ1 and the ground cable.
12. Remove the three mounting screws on paddle board 800PL8 (2.5 mm Allen wrench).
13. Remove paddle board 800PL8.

Chapter 10 Page no. 344 DR_002.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 002 - Paddle board 800PL8
ILLUSTRATION 1 - PADDLE BOARD 800PL8 AND DETAIL OF BUTTON/SHAFT MOUNTING
25
23
24
21
34
18
17
16
S1
FRONT
3.5 mm pin protusion
S2
B1
B2
B1
Ball latch
adjustment screw

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 002 - Paddle board 800PL8
DR_002.fm Page no. 345 Chapter 10
5-2. Installation of paddle board 800PL8
Refer to Illustration 1.
1. Install paddle board 800PL8 on its support plate.
2. Reconnect connector XJ1.
3. Insert the equipped support plate into its housing.
4. Reset the shaft and encoder wheel to their initial positions.
CAUTION
The encoder wheel is flexible but FRAGILE.
5. Reinstall the washers (see Illustration 1, bottom) on the shaft, and reinstall the circlip.
6. Install the paddle board support plate (two mounting screws, 23/24).
CAUTION
The magnet assembly (25) must not be in contact with the Hall effect transducer (clearance
approx. 1 mm).
7. Mount the encoder wheel on its shaft, ensuring that it is central in relation to the optointerrupter.
8. Tighten the encoder wheel locking screw.
9. Reinstall the two tyraps.
10. Ball latch protusion (see illustration 1): check that the ball latch pin protudes 3 to 3.5 mm (about
0.14inch). If adjustment is necessary unclamp the nut located inside the compression support and
rotate the screw to obtain the right protusion value. Then reclamp the screw with the nut.
11. Install the two half-covers (screws S1 and S2).
12. Install buttons B1 and B2 (1 mm Allen wrench).
13. Check that compression fine adjustment operates correctly - the button shaft must not rotate too
freely. If the slight rotation resistance is not as required, check that the washers are correctly
installed (para.5.2.5 and Illustration 1, items 34, 17 and 16).
14. When pressure is applied with a paddle, check that a compression force is indicated.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
Note:
It is mandatory to check that the maximum compression force cannot be more than 30 daN (max-
imum allowed by regulations).
• Check that the compression system operates normally in application mode.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH A NEW PADDLE BOARD 800PL8.

Chapter 10 Page no. 346 DR_002.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 002 - Paddle board 800PL8 This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 003 - Magnification sensor board 800PL14
DR_003.fm Page no. 347 Chapter 10
Job Card D/R 003 - Magnification sensor board 800PL14 Chapter 10
1. SUPPLIES
Magnification sensor board 800PL14.
2. TOOLS
2.5 mm and 5 mm Allen wrenches.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Removal of magnification sensor board 800PL14
Refer to Illustration 1.
1. Remove the compression arm covers and sensor plate (see D/R Job Card "Gantry Arm and Tube
housing Spacer Covers", paragraph 5-2).
2. Switch OFF the DMR electrical power.
3. Remove the two mounting screws on the magnification unit support (left support only).
4. Remove the support.
5. On magnification sensor board 800PL14, disconnect connector XJ1.
6. Remove board 800PL14 and its support plate.
7. Separate board 800PL14 from its support plate by removing the two mounting screws (2.5 mm Allen
wrench).
5-2. Installation of magnification sensor board 800-PL14
Refer to Illustration 1.
1. To install magnification sensor board 800PL14, carry out the removal operations (above) in reverse
order.

Chapter 10 Page no. 348 DR_003.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 003 - Magnification sensor board 800PL14
ILLUSTRATION 1 - LOCATION OF MAGNIFICATION SENSOR BOARD AND SUPPORT PLATE
5-3. Functional check
1. Switch ON the DMR electrical power.
2. Set the magnification unit in each of its three positions, and check that the indication on the column
display corresponds to the selected position.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH NEW MAGNIFICATION SENSOR BOARD
800PL14.
LEFT SUPPORT PLATE
800PL14 MAGNIFICATIO N SENSOR BOARD

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 004 - Inter-arm tilt sensor board 800PL15
DR_004.fm Page no. 349 Chapter 10
Job Card D/R 004 - Inter-arm tilt sensor board 800PL15 Chapter 10
1. SUPPLIES
• Inter-arm tilt sensor board 800PL15.
2. TOOLS
• 2.5 mm Allen screwdriver (or wrench).
• 3 mm Allen wrench.
• 7 mm flat wrench.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Removal of inter-arm tilt sensor board 800PL15
Refer to Illustration 1.
1. Remove the compression arm covers and sensor plate (see D/R Job Card "Gantry Arm and Tube
housing Spacer Covers", paragraph 5-2).
2. Raise the column, to set board 800PL15 at a convenient working height.
3. Switch OFF the DMR electrical power.
4. Remove the lower screw S1 from the cable guide support plate, and remove the corresponding
spacer (7 mm flat wrench).
5. Remove the two mounting screws on board 800PL15.
CAUTION
There are two nylon spacers under the board
6. Disconnect the connector on board 800PL15 and remove the board.

Chapter 10 Page no. 350 DR_004.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 004 - Inter-arm tilt sensor board 800PL15
ILLUSTRATION 1 - CENTER PART OF COMPRESSION ARM
5-2. Installation of inter-arm tilt sensor board 800PL15
Refer to Illustration 1.
1. To install inter-arm tilt sensor board 800PL15, carry out the above operations in reverse order.
CAUTION
Do not forget to reinstall the two nylon spacers under the board.
CABLE GUIDE
BOARD 800PL15
S1
SUPPORT
PLATE

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 004 - Inter-arm tilt sensor board 800PL15
DR_004.fm Page no. 351 Chapter 10
5-3. Functional check
1. Switch ON the DMR electrical power.
2. Unlock the tube support arm and check that the three positions are correctly detected.
- Message S50: Incorrect X-ray incidence: at ±27 degrees
- No message at 0 degrees
3. Check that at 0 degrees, X-Ray exposure is available,
4. Check that at ±27 degrees, X-Ray exposure is forbidden.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH NEW INTER-ARM TILT SENSOR BOARD
800PL15.

Chapter 10 Page no. 352 DR_004.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 004 - Inter-arm tilt sensor board 800PL15 This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 007 - Collimator
DR_007.fm Page no. 353 Chapter 10
Job Card D/R 007 - Collimator Chapter 10
1. SUPPLIES
• Collimator.
• Tyraps.
2. TOOLS
• Flat screwdriver, 1/4".
• 4 mm Allen wrench.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Removal of collimator
1. Set tube housing arm to a convenient position.
2. Switch OFF the DMR electrical power.
3. Remove tube housing arm covers (see D/R Job Card
Removal of Covers, paragraph 5-3).
4. Remove collimator light shield box (one mounting screw).
5. Remove rod retaining front of collimator (2 locking pins).
6. Remove tyraps along cable harness between collimator and tube housing arm distribution board
800PL6.
7. On tube housing arm distribution board 800PL6, disconnect connectors XJ7, XJ8, XJ9, XJ10, XJ15
and XJ16 and separate these cables from the rest of the harness.
8. Remove rod retaining rear of collimator (2 locking pins) - push or tap the rod out to the right. This
releases the collimator.
9. Remove ground cable on rear of collimator (screwdriver).
10. Remove collimator.
5-2. Installation of collimator
1. Install new collimator (install rear rod and locking pins, followed by front rod and locking pins).
2. On tube housing arm distribution board 800PL6, reconnect connectors XJ7, XJ8, XJ9, XJ10, XJ15
and XJ16, reconstitute the cable harness, installing new tyraps.
3. Reinstall collimator light shield box (one mounting screw).
4. Reinstall tube housing arm covers (see D/R Job Card
Gantry Arm and Tube housing Spacer Covers,
paragraph 5-3).

Chapter 10 Page no. 354 DR_007.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 007 - Collimator
5-3. Functional check
1. Switch ON the DMR electrical power and check that the self-test is correctly executed.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW COLLIMATOR.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 008 - Compression Motor and Belts
DR_008.fm Page no. 355 Chapter 10
Job Card D/R 008 - Compression Motor and Belts Chapter 10
1. SUPPLIES
• Compression motor.
• Compression belts.
2. TOOLS
• Flat screwdriver, 1/4".
• 3 mm Allen wrench.
• 4 mm Allen wrench.
• 5 mm Allen wrench.
• 7 mm flat wrench.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
Carry out the following preliminary actions:
• Remove the compression arm covers and sensor plate (see D/R Job Card
Gantry Arm and Tube
housing Spacer Covers
).
• Set the compression paddle mounting in contact with the image receptor.
• Switch OFF the DMR electrical power.
5-1. Removal of compression motor and small belt
Refer toIllustrations 1 and 2.
1. Disconnect the connector from the compression motor.
2. Remove the cable support plate P1 (supporting the cable between 800PL5 and XJ4) and release the
harness by cutting the tyrap located under the motor.

Chapter 10 Page no. 356 DR_008.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 008 - Compression Motor and Belts
ILLUSTRATION 1 - COMPRESSION ARM WITH COMPRESSION MOTOR MOUNTING DETAIL
COMPRESSION MOTOR
SLIDE BAR
MAGNIFICATION RIGHT SUPPORT PLATE
P1
S2S1
S3 to S6
SLIDE BAR PROTRUSION

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 008 - Compression Motor and Belts
DR_008.fm Page no. 357 Chapter 10
ILLUSTRATION 2 - PULLEY AND MOTOR MOUNTING, WITH COMPONENT ASSEMBLY SEQUENCE
3. Remove the crank pulley bearing support plate (two 4 mm Allen screws, S1 and S2).
4. Remove the four screws, S3 to S6 (4 mm Allen wrench), from the motor support plate.
CAUTION
Check that the slide bar does not protrude from the front of the pulley, this prevents the
mounting of the small belt. If necessary, push the slide bar slightly downwards.
5. Carefully extract the motor and the small toothed belt
CAUTION
Take care not to damage the encoding wheel.
6. Remove the two left-hand motor mounting screws (S8 and S9 - see above), on the motor support
plate.
7. Remove the encoder board, thus releasing the toothed belt.
8. Remove the other two screws (S7 and S10, on the right), followed by the encoding wheel and the pul-
ley, if the motor is being changed.
RIGHT
LEFT
S7
S8 S9
S10
LARGE BELT PULLEY
SMALL BELT PULLEY
COMPRESSION
MOTOR
ENCODER WHEEL

Chapter 10 Page no. 358 DR_008.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 008 - Compression Motor and Belts
5-2. Encoder Board
• Before reinstalling the compression motor, check
for the presence of washers (A) between the fingers
of the black supporting bracket and the spacers on
which the encoder board is mounted.
• If these washers are present, remove them ( to
move the board further away from the motor axis).
• To assist in adjustment after reinstallation, it may be
useful to place a mark on the side of the LED
assembly to indicate the position of the center of
the LED.
5-3. Installation of compression motor and small belt
Refer to Illustrations 1 and 2.
• To install the compression motor and small belt, carry out the above operations in reverse order (8
thru 1).
• Do not tighten the four mounting screws (Illustration 2, items S7 to S10) until the adjustment
described below has been carried out. They are used to tension the small toothed belt by moving the
motor on its mounting plate, and to position the encoder board with respect to the encoder wheel.
5-4. Adjustment
1. Tension the belt by pushing the motor (to the left in the
illustration). Take care to avoid excessive belt tension.
2. Tighten the two right hand screws.
3. Position the encoder board so that the encoder LED is
positioned at the center of the encoder wheel slots(if you
have marked the side of the LED assembly use the mark
as a guide), by pusing the board assembly away from the
motor axis (down in the illustration).
• It is important that the rotating wheel does not touch
any part of the LED assembly . If it does, it will generate
dust which will obscure the LED and cause failures.
4. When you are satisfied that the position is correct, tighten the two left hand screws.
5-5. Removal of large compression belt
Refer to Illustrations 1 and 2.
WARNING
The compression motor and small belt must be removed before the procedure described
below can be executed (see section 5-1).
1. Release the lock, behind the arm, to offset the compression arm in relation to the tube housing arm.
2. Switch ON the DMR electrical power.
3. Set compression arm horizontal (the most convenient position).
4. Switch OFF the DMR electrical power.
A

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 008 - Compression Motor and Belts
DR_008.fm Page no. 359 Chapter 10
5. Remove the magnification right support plate (accessible behind arm - two 5 mm Allen screws).
6. Set paddle support opposite hole left free by removal of magnification right support plate.
7. Slacken the two screws on the paddle support, to release the belt joining piece.
Note:
In the case of routine maintenance; if the belt is being replaced; then the old belt can simply be cut.
8. On the end of the belt, unscrew the small tensioner screw (7 mm flat wrench - see Illustration 3), to
release the spring.
Note:
It is not necessary to remove the spring. The tensioner screw need only be unscrewed a few turns.
9. Remove the end piece from the adjustment end of the belt (spring end; two small slot-head
screws).Remove the spring housing/belt joining piece (two 3 mm Allen screws).
CAUTION
Check that the slide bar does not protrude out of the front of the pulley as this prevents the
mounting of the small belt. If necessary, push the slide bar slightly downwards.
10. After the belt has been removed or cut, remove the two pulleys and spindles.
ILLUSTRATION 3 - BELT JOINING PIECE, SHOWING BELT ENGAGEMENT AND TENSIONER SCREW
5-6. Installation of large compression belt
Refer to Illustrations 1, 2, and 3.
1. Install new belt.
Note:
Belt is already cut to exact length (1215 mm).
2. While leaving some slack on the belt, install the belt joining piece.
3. Engage the first four teeth, on each end of the belt, in the corresponding four notches in the belt join-
ing piece (see Illustration 3).
CAUTION
Take care not to twist the belt.
4. Engage belt on small pulley.
BELT TENSIONER SCREW
COIL SPRING
FIRST FOUR TEETH, ON EACH END, ENGAGED IN
FIRST FOUR NOTCHES IN BELT JOINING PIECE

Chapter 10 Page no. 360 DR_008.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 008 - Compression Motor and Belts Note:
This operation is facilitated by the insertion of a packing piece on the end of the spring, to compress
the spring and slacken the belt - Remove packing piece after belt is engaged on pulley.
5. Tension (compress) the spring (7 mm flat wrench), using a little locking compound (Loctite or similar)
on the tensioner threads.
6. Reinstall the bearings and pulleys.
Note:
To do this it will be necessary to pull on the belt.
7. Recentralize the paddle and mount the belt central piece on the paddle support.
8. Reinstall the motor support plate and motor (4 screws).
9. Reinstall small belt and cable support plate.
10. Reconnect small coding wheel connector.
11. Tension small toothed belt by moving motor.
12. Tighten all four screws (motor mount).
13. Reconnect motor connector.
14. Reinstall the magnification plate support.
Note:
See Functional check, below.
15. Install compression paddle.
16. Reinstall covers and small cable protector on motor flange (see D/R Job Card "Gantry Arm and
Tube housing Spacer Covers", paragraph 5-2 ).
CAUTION
Cables must pass outside protector plate, and plate must not be in contact with pulley.
5-7. Functional check
The functional check should be carried out both before and after the reinstallation of the compression
paddle and arm covers.
1. Switch ON the DMR electrical power.
2. Check that the belt runs correctly over its full travel range.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW COMPRESSION MOTOR AND
BELTS.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 009 - Gantry CPU Battery
DR_009.fm Page no. 361 Chapter 10
Job Card D/R 009 - Gantry CPU Battery Chapter 10
1. SUPPLIES
Lithium battery.
2. TOOLS
• 2.5 mm Allen wrench.
• Soldering iron.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
WARNING
Before removing gantry CPU board 800-PL3, save the gantry parameters, to ensure that
these parameters (saved using battery power) are not accidentally lost while the battery is
being changed - see D/R Job Card Back-up/Restore.
5-1. Removal of gantry CPU battery and installation of new battery
1. Remove the column cover (CPU side only) (see D/R Job Card Removal of covers).
2. On gantry CPU board 800PL3, remove the two mounting screws (2.5 mm Allen wrench), and remove
the board.
3. Solder the new lithium battery in the free position (BT1 or BT2), respecting the polarities.
4. Unsolder the old battery.
5. Reinstall gantry CPU board 800PL3 in its position in the gantry.
6. Reinstall the column cover (CPU side).
5-2. Functional check
1. Switch ON the Senographe DMR electrical power.
2. The following message should NOT be displayed after initialization:
SAVED MEMORY ERROR 151/000
If this message is displayed, reload the gantry parameters and repeat the initialization.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW GANTRY CPU BATTERY.

Chapter 10 Page no. 362 DR_009.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 009 - Gantry CPU Battery This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 010 - Optical Fiber (Gantry to Generator)
DR_010.fm Page no. 363 Chapter 10
Job Card D/R 010 - Optical Fiber (Gantry to Generator) Chapter 10
1. SUPPLIES
• Optical fibers.
• Tyraps.
2. TOOLS
• None.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
CAUTION
When disconnecting optical fibers, always pull on the connector AND NOT on the cable. If
in doubt, check that the core of the fiber is flush with the end of the connector.
5-1. Removal and reinstallation of optical fibers
1. Switch OFF the DMR electrical power.
2. Remove the generator covers and the gantry column covers (see D/R Job Card
Removal of covers).
3. Disconnect the optical fibers at both ends.
4. Disengage the harness connecting the generator and the gantry, particularly if the harness is coiled
under the generator.
5. Open the harness protection housing.
6. Disengage the black ribbed outer cover protecting the optical fibers.
Note:
It will be necessary to cut some of the tyraps in the generator and the gantry.
7. Remove the harness clamp at the bottom of the gantry column.
8. Gently pull the black ribbed outer cover out of the generator and the gantry.
CAUTION
Take care to avoid rubbing or dirt on the ends of the "good" fibers.
9. Open the ribbed outer cover, over its entire length, and remove the defective fiber.
10. Replace this fiber, respecting connector color coding.
11. Temporarily reconnect the four fibers to the generator and gantry, and check that the system operates
correctly when the electrical power is switched ON.
12. Disconnect the four fibers and reinstall the ribbed outer cover in the harness.
13. Carefully install new tyraps, particularly inside the gantry and between the base plate and gantry
interface board 800PL2.

Chapter 10 Page no. 364 DR_010.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 010 - Optical Fiber (Gantry to Generator)5-2. Functional check
1. Switch ON the DMR electrical power.
2. Raise and lower the column and visually check that the movement is not fouling any of the cables.
3. Reinstall the generator covers and the gantry column covers (see D/R Job Card
Removal of Covers).
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW OPTICAL FIBERS.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 011 - X-ray Tube Housing
DR_011.fm Page no. 365 Chapter 10
Job Card D/R 011 - X-ray Tube Housing Chapter 10
1. SUPPLIES
• X-ray tube housing.
• HT grease.
2. TOOLS
• Flat screwdriver, 1/4".
• 10 mm flat wrench.
3. SAFETY PRECAUTIONS
Be careful when touching the tube housing, as its temperature can be up to 80°C (about 180°F).
4. PREREQUISITES
None.
5. PROCEDURE
Notice:!
Take great care to avoid damage to the rhodium filter; it is extremely expensive!
5-1. Removal of X-ray tube housing
1. Select Aluminium filter (to avoid possible damage to rhodium or moly filters).
2. Remove tube housing spacer (see D/R Job Card
Gantry Arm and Tube housing Spacer Covers).
3. Switch OFF the DMR electrical power.
4. Disconnect HT connector on housing.
5. Disconnect stator cable connector (see illustration 1).
6. Remove HT positive cable and one ground cable from the grounding points.

Chapter 10 Page no. 366 DR_011.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 011 - X-ray Tube Housing
ILLUSTRATION 1 - TUBE HOUSING CONNECTIONS
7. On board 800PL6, disconnect connector XJ4.
8. Slacken two screws at rear of housing and remove two at front.
9. Remove housing by gently pulling forwards.
CAUTION
Do not touch the molybdenum and rhodium filters with fingers.
5-2. Installation of X-ray tube housing
1. Prepare new housing. 2. Remove protection cover from tube housing window. 3. Remove cap from HT pot and insert grease. 4. Carefully slide the new housing into position. 5. Reinstall the two front screws on housing and tighten all four screws. 6. Reinstall the three cables on the housing grounding point. 7. Reconnect green connector XJ1 (anode motor power). 8. Reconnect connector XJ4 on board 800PL6 (other end of thermal cable). 9. Reconnect HT cable into rear of housing.
+HT
screw
HT connector (female)
Ground connector screw
Stator, pressostat
Thermal switch
Pressostat connector
& thermal switch
connectors
Stator connector

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 011 - X-ray Tube Housing
DR_011.fm Page no. 367 Chapter 10
5-3. Check and adjustment of cradle centering, check for lateral play
• Lateral play may exist between Items 1 and 2 (Illustration 2) due to loosening of the screw (Item 3)
which causes the setscrew to loosen.
• In Application Mode, and following tilt of the arm, there is a risk of the X-ray field shifting to the left or
to the right in relation to the image receptor. There is also a risk of motion blurring due to vibration
from tube rotation.
Note:
The two setscrews (Illustration 2, Item 2) should touch the two tube joint shafts at the left and right
of the tube, without applying pressure to them.
ILLUSTRATION 2 - TOP VIEW OF TUBE HOLDER CRADLE (THROUGH ROTATION SHAFT)
2
3
1
2
3
1

Chapter 10 Page no. 368 DR_011.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 011 - X-ray Tube Housing
5-4. Calibration and adjustment of X-ray tube
1. Disconnect connector XJ4 from board 800PL6 and check that the message: TUBE HOUSING OVER
TEMPERATURE is displayed on the control panel.
2. Reconnect connector XJ4 and check that the message disappears.
3. Carry out the following Calibration Job Cards:
•
X-Ray Tube Focal Bias Voltage Scale Factor.
•
Heater Current Scale Factor.
•
X-Ray Tube Heater Current.
4. Before proceeding, perform a Clear Calibrations operation. From the Service Desktop, click the
Cali-
brations
button, then select Detector, and Clear Calibration.
5. Carry out the following Calibration Job Cards:
•
Tube Tilt Calibration.
•
X-ray Beam Alignment Calibration.
•
Collimator Format Calibration.
•
Bad Pixel Calibration.
•
Detector Gain Calibration.
•
Conversion Factor Measurement.
6. Carry out the VF Job Card
X-Ray Format Geometry check.
7. Carry out the Calibration Job Card
AOP Calibration.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW X-RAY TUBE HOUSING.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 012 - Centering Device Lamp
DR_012.fm Page no. 369 Chapter 10
Job Card D/R 012 - Centering Device Lamp Chapter 10
1. SUPPLIES
• New centering device lamp (only genuine part number 2188568 must be used, to ensure correct light
intensity, easy filament centering and reliability)
2. TOOLS
• Standard screwdrivers, open–end wrenches, etc.
• Ruler calibrated in mm.
3. SAFETY PRECAUTIONS
Not applicable.
4. PREREQUISITES
None.
5. PROCEDURE
Note:
The objective is to replace the centering device lamp and set it to the correct vertical height.
5-1. Remove old lamp.
1. Turn the Senographe DMR off.
2. Remove the two half-covers from the gantry head.
3. Remove the anti-scatter box (2 parts); first remove the side part then loosen slightly the attachment
screw (at center underside) of the box and remove the anti-scatter box by pulling it gently forward.
4. Remove collimator assembly retaining pins by first removing the two cotter pins from one end of each
retaining pin. Pivot the collimator assembly downwards.
5. Loosen the two light box securing screws slightly; remove the light box by pulling it gently forward.
CAUTION
Wait for the lamp to cool before removing it.
6. Remove the old lamp by pulling firmly but gently downwards.

Chapter 10 Page no. 370 DR_012.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 012 - Centering Device Lamp
5-2. Install and Position New Lamp
CAUTION
When handling the new lamp, be sure not to touch the glass surface with your fingers.
Always use a clean dry cloth or paper handkerchief.
1. Insert the leads of the new lamp into its socket SLOWLY, and PUSH THEM IN ALL THE WAY.
Note:
The distance of 30 mm between filament center and pin end is guaranteed by the lamp manufac-
turer, see Illustration 1
.
2. Reinstall the light box but do not tighten the securing screews.CAUTION
The light box is for protection in case of any explosion of the new lamp being installed.
3. Turn on the Senographe DMR and verify that
the new lamp lights when activated, then turn
the Senographe DMR off again.
4. Tighten the two light box securing screws.
5. Check that the CENTER of the filament is
aligned with the CENTER of the small slot in
the light box when your eye is at the level of
the small slot (see illustration 1). If not, start
5.2 again.
6. Pivot the collimator assembly back up and
reinstall its retaining pins and cotter pins.
CAUTION
CORRECT LAMP FILAMENT HEIGHT: 30 ± 0.5 mm
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 013 - Mirror
DR_013.fm Page no. 371 Chapter 10
Job Card D/R 013 - Mirror Chapter 10
1. SUPPLIES
Mirror.
2. TOOLS
• Flat screwdriver, 1/4".
• 3 mm Allen wrench.
• 10 mm flat wrench.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
WARNING
GREAT CARE SHOULD BE TAKEN TO AVOID TOUCHING THE MIRROR ( NO FINGER
PRINTS
) AND TO AVOID DAMAGING THE FILTER WHEEL, WHICH IS EXTREMELY FRAGILE
AND LOCATED JUST ABOVE THE MIRROR
5-1. Mirror Removal
Refer to Illustration 1.
1. Affix a piece of adhesive tape on the Bucky to mark the rear edge of the light field.
2. Set the tube housing arm to convenient position.
3. Switch OFF the DMR electrical power.
4. Remove tube housing arm covers (See D/R Job Card
Gantry Arm and Tube housing Spacer Covers).
5. Remove collimator light shield box (one mounting screw).
6. Remove rod retaining front of collimator (2 locking pins).
7. Open the collimator.
8. Remove the mirror adjustment screw and its spring.
9. Remove the locknuts on the two mirror mounting screws.
10. Remove the two mirror mounting screws (3 mm Allen key) and the mirror.
5-2. Mirror Installation
• To install the new mirror, carry out the above operations in reverse order (10 thru 1).

Chapter 10 Page no. 372 DR_013.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 013 - Mirror
ILLUSTRATION 1 - TUBE HOUSING ARM WITH MIRROR INSTALLATION DETAIL
5-3. Adjustment of mirror
• Adjust the mirror angle so that the rear edge of the light field coincides with the adhesive tape refer-
ence position made in 5-1-1 above.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW MIRROR.
TUBE HOUSING ARM
MIRROR FRAME
MIRROR MOUNTING SC REWS
AND LOCKNUTS
MIRROR ANGLE ADJUS TMENT SCREW
AND SPRING

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 015 - Filter wheel
DR_015.fm Page no. 373 Chapter 10
Job Card D/R 015 - Filter wheel Chapter 10
1. SUPPLIES
Filter wheel.
2. TOOLS
• Flat screwdriver, 1/4".
• 5 mm Allen wrench.
• 8 mm flat wrench.
• Circlip pliers.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Removal of filter wheel
1. Set tube housing arm to convenient position.
2. Switch OFF the DMR electrical power.
3. Remove tube housing arm covers (see D/R Job Card
Gantry Arm and Tube housing Spacer Covers).
4. Remove collimator light shield box (one mounting screw).
5. Remove the front rod retaining the collimator (two spring clips).
6. On tube housing arm distribution board 800PL6, disconnect connectors XJ6 and XJ11.
CAUTION
In the following operations concerning the filter wheel and motor, take great care TO
AVOID TOUCHING THE MIRROR.
7. Remove the filter motor mounting screws, one each side of motor (5 mm Allen wrench).
8. Remove the motor, complete with filter wheel.
9. Remove the nut and washer, retaining the filter wheel spindle to the support plate (8 mm flat wrench).
10. Separate the spindle from the wheel by removing the circlip.
5-2. Installation of filter wheel
To install the new filter wheel, carry out the above operations in reverse order (section 5-1 10 thru 1).

Chapter 10 Page no. 374 DR_015.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 015 - Filter wheel
5-3. Functional check
• Switch on the DMR electrical power supply and check that the self-test is correctly executed.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW FILTER WHEEL.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job card D/R 016 - Gas Spring
DR_016.fm Page no. 375 Chapter 10
Job card D/R 016 - Gas Spring Chapter 10
1. SUPPLIES
Gas spring.
2. TOOLS
• Service Terminal.
• Flat screwdriver, 1/4".
• 8 mm Allen wrench.
• Set of steps (for access to top of column).
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. GAS SPRING REPLACEMENT
Refer to Illustration 1.
1. Remove the column covers (see D/R Job Card
Removal of covers).
2. Switch OFF the DMR electrical power.
3. Remove three plastic caps on column top cover.
4. Remove the three cover mounting screws and the cover.
5. Disconnect cable from XJ1 on board 800PL12 (inside, on cover).
6. Raise column approx. 30 cm, to raise the safety plate above the fixed part of the gas spring.
7. Remove jack (gas spring) safety plate inside column (two 8 mm Allen screws).
CAUTION
For safety, the two 8 mm screws are installed using thread locking compound (loctite or
similar). An extension will probably be required on the Allen wrench when slackening the
screws.
8. Remove clamp on front edge, at bottom of column (two bolts; 7 mm flat wrench).
9. Lower column and insert new gas spring (inner plunger downwards), next to the existing gas spring,
in the corner where the motor is located.
10. Raise column to maximum height (gas spring is then not under load).
11. Try to move the gas spring. If no movement is possible on the upper and lower mounting rods, use
the column maintenance menu on the Service Terminal to raise the column,
very gently, to the upper
mechanical stop.
12. Remove the upper and lower gas spring mountings to release the spring (two small rods, secured by
locking pins - no tool required).

Chapter 10 Page no. 376 DR_016.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job card D/R 016 - Gas Spring
ILLUSTRATION 1 - GAS SPRING INSTALLATION IN COLUMN
13. Install the new gas spring in place of the old.
14. Reinstall upper and lower gas spring mounting rods, complete with locking pins (align pins vertically).
15. Place old gas spring in one corner of column.
16. Check cable harness is correctly positioned. Adjust as required.
17. Lower column to approx. 30 cm above lower end of travel stop (check movement of cable harness).
18. Reinstall clamp on front edge, at bottom of column (two bolts, 7 mm flat wrench).
19. Reinstall the gas spring safety plate inside column (two 8 mm Allen screws).
CAUTION
Apply thread locking compound (loctite or similar) to the two 8 mm screws before installa-
tion .
GAS SPRING
COLUMN MOTOR
FRONT VIEW SIDE VIEW
CLAMP

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job card D/R 016 - Gas Spring
DR_016.fm Page no. 377 Chapter 10
20. Lower the column to the fully down position.
21. Remove the old gas spring.
22. Reconnect cable to XJ1 on board 800PL12 (inside, on cover).
23. Reinstall the cover, and the three cover mounting screws.
24. Reinstall the three plastic caps on column top cover.
25. Reinstall the column covers (see D/R Job Card
Removal of covers).
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH NEW GAS SPRING IN COLUMN.

Chapter 10 Page no. 378 DR_016.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job card D/R 016 - Gas Spring This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 019 - Back-up/Restore procedure
DR_019.fm Page no. 379 Chapter 10
Job Card D/R 019 - Back-up/Restore procedure Chapter 10
1. SUPPLIES
• Blank DOS formatted HD 3.5 floppy disk.
2. TOOLS
• None.
3. SAFETY PRECAUTIONS
• No specific safety precautions are applicable.
4. PREREQUISITES
• None.
5. BACKUP/RESTORE PROCEDURE
Backup and Restore procedures are described on the following pages.
5-1. Introduction
When a Backup procedure is performed, the following operations occur:
•If a
Backup 1 file already exists on the AWS disk, it is renamed to Backup 2.
• A new backup file named
Backup 1 is created on the AWS disk. This contains:
- AWS configuration data.
- IDC configuration data.
- DMR data (Generator and Gantry parameters).
• The AWS configuration data may also be copied to a floppy disk. No IDC or DMR backup data are
written to the floppy.
It is possible to carry out a Restore procedure in several ways:
• AWS configuration data can be restored from
Backup 1 or Backup 2, from the backup floppy diskette.
• IDC data can be restored from the AWS disk file
Backup 1 or Backup 2.
• DMR data can be restored from the AWS disk file
Backup 1 or Backup 2.
Notice:!
A system backup is recommended after each change of configuration of any part of the system,
and after each set of calibrations.
A system backup is
not recommended after a single calibration or configuration change when the
change is to be followed by more changes. Each time that a system backup is made, the current
AWS file backup
Backup 2 is lost (overwritten by Backup 1), and this must be taken into account
when considering a new backup.

Chapter 10 Page no. 380 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 019 - Back-up/Restore procedure
5-2. Backup procedure
1. Select Root menu /Service Tools /Service Desktop; click Submit, then click the Utilities button.
2. Insert a blank 3.5" HD floppy disk into the floppy drive on the AWS unit (not the IDC). To access the
floppy drive, remove the Cart rear cover and pull back the Ultra 10 unit. Click the
Floppy Format but-
ton, and wait for the format to complete.
Note:
The presence of a floppy disk is not essential for the backup. If no floppy is inserted, AWS param-
eters are still copied to the hard disk.
3. Click the
Backup button. The Back up screen is displayed (see illustration below).
4. Enter a suitable name into the text field
System Backup Description Text; click the START button to
continue. If no floppy disk is present, the message
Floppy disk error or not formatted is displayed;
click
Continue to continue the backup. The process takes about 10 minutes.
5. An AWS reset is carried out automatically; click
OK when it completes.
6. The message
System Backup Completed is displayed at the end of the backup operation.
7. To check that the backup copy is available, click the
RESTORE button: The name of the system
backup,
Backup 1, appears on the screen.
8. If the backup is not completed successfully, refer to section 6-3
Backup Troubleshooting.
9. Remove the floppy disk; label it with the name and date of the backup and keep it safely.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 019 - Back-up/Restore procedure
DR_019.fm Page no. 381 Chapter 10
5-3. Restore procedure
1. Select Root menu /Service Tools /Service Desktop; click Submit, then click the Utilities button..
2. Click the
Restore button: the Restore screen is displayed (see illustration below).
3. Select the appropriate restore procedure: you may restore separate sub-system data (AWS, IDC, or
DMR) as required.
4. Select the data source (Backup1 or Backup 2 on the AWS disk, or 3.5" floppy).
5. If restoring from floppy, insert the diskette containing the backup data into the drive on the AWS unit.
To access the floppy drive, remove the Cart rear cover and draw back the Ultra 10 unit.
6. Click the
START button to proceed, and wait for restore completion.
7. A message indicating successful completion is displayed at the end of the restore procedure.
8. If IDC data has been restored, the IDC must be rebooted. See section 5-4.
9. If the restore is not completed successfully, check the following:
• That the backup source is functioning correctly (AWS or floppy disk).
• That communication is correct between the AWS and the sub-system being restored.

Chapter 10 Page no. 382 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 019 - Back-up/Restore procedure
5-4. IDC Restore
• After IDC data has been restored, the IDC must be rebooted so that it can recognize the changed
data. This can done either by power OFF/ON, or using a command window.
• To reboot the IDC using a command window:
- Open a command window
- Enter the command: telnet idc<Return>.
- Press <Ctrl>+X (Control key and X simultaneously).
- The IDC shuts down and reboots; the message
Connection closed by foreign host is displayed. If
you wish to view IDC startup messages you must reconnect by entering the command telnet
idc<Return> again.
- The DMR console first displays the message
S86 AWS/IDC Communication lost, then displays
the message
S83 Detector environment not OK until the reboot is complete.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 019 - Back-up/Restore procedure
DR_019.fm Page no. 383 Chapter 10
6. TROUBLE-SHOOTING NOTES
Use the information in the following sections as a guide in trouble-shooting backup/restore problems.
6-1. AWS Backup from Service Desktop
The diagram below illustrates the backup process.
6-2. Backup Process
When the Backup button in service desktop is clicked:
1. Files from a previous backup stored in
~/current (backup 1) are copied into ~/previous (backup 2).
2. Files in
~/previous (backup 2) are then lost.
3. The IDC directory is copied to backup 1.
4. Numerous AWS config files spread in many locations are gathered into four files and copied into
backup 1.
5. DMR calib files are copied through AWS/DMR serial link to backup 1.
6. If a formatted floppy was inserted (optional), three AWS config. files are copied to the floppy.
Backup 1
/export/home/iqst/subsystems/ads/backup/current/
/export/home/iqst/subsystems/dmr/backup/current/
/export/home/iqst/subsystems/idc/backup/current/
Next Backup
Backup 2
/export/home/iqst/subsystems/ads/backup/previous/
/export/home/iqst/subsystems/dmr/backup/previous/
/export/home/iqst/subsystems/idc/backup/previous/
AWS Hard Disk
AWS Config files
browser preferences, hosts declared, printer config.,
config from Service Desktop, etc.
ads.tar (1 M)
ads_hd.tar (1 M)
iip.data.tar (1 M)
iop.tar (100 k)
Backup
DMR CPUs
brgen.dat (17 k)
brgan.dat (0.6 k)
IDC
Complete directory
/idc (160 M)
about 200 files
AWS config floppy
ads.tar (1 M)
iip.data.tar (10 k)
iop.tar (100 k)
Optional

Chapter 10 Page no. 384 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 019 - Back-up/Restore procedure
6-3. Backup Troubleshooting
6-3-1. If IDC Backup Fails
• Check communication between the AWS and the IDC.
• Check that there is enough space on the AWS disk to receive the backup data.
6-3-2. If AWS Backup to Hard Disk Fails
• Check that there is enough space on the AWS disk to receive the backup data.
6-3-3. If AWS Backup to Floppy Fails
• Check that the floppy is correctly formatted.
• Check that the floppy is not damaged.
• Check that there is enough space on the floppy to receive the backup data.
Note:
If backup to floppy is not possible, record the AWS parameters manually, as detailed on the follow-
ing pages.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
DR_019.fm Page no. 385 Chapter 10
AWS Parameters manual backup Chapter 10
7. MANUAL PARAMETER BACKUP
If backup to floppy is not possible, it may be necessary to record AWS parameters manually so that they
can be restored later (e.g., after an upgrade procedure).
The following pages provide information and space to note each of the following sets of parameters:
1. Site Information and General Configuration
2. Filter Management
3. Browser Preferences
Layout
Filter
Sort
4. Medical Application Preferences
Auto Push
Auto Print
Annotations
Auto Delete
Names
5. Worklist Management
6. Network Management
7. Printer Management

Chapter 10 Page no. 386 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
7-1. Site Information and General Configuration
• Access: Tools menu/Service Desktop. Click on Submit to enter the Service Desktop then click the
Configuration button. Click Site to view and note Site information, then General to view and note
the General Configuration.
Site Information Copy the current settings into the table below. (note that your hardware key must be
installed to display all settings):
General configuration. Copy the current settings into the table below.
Parameter Current setting
Site Name:
Site Address:
Phone:
Fax:
Contact Person:
Phone:
Station Name:
System ID:
Installation date:
Contract Number:
Contract Expiry Date:
Auto Delete: Language:
QAP timeout Upper threshold: English
. . . . . . . days . . . . . . . . % Français
Lower threshold: Español
. . . . . . . . % Deutsch
Screen Blank Timeout Maximum Image Storage Time: Italiano
. . . . . . . days . . . . . . . . hours Portugès
Edit Patient: <> On
<> Off

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
DR_019.fm Page no. 387 Chapter 10
7-2. Filter Management
• Access: Tools menu/Filter management. For each filter name, click Update to view the settings,
and copy them into one of the tables below. Copy this sheet if you need additional space.
1.
Filter name:
Circle the current value Note current value of the parameter
Patient level:
Patient Name Begins with / Contains / Is
Patient ID Begins with / Contains / Is
Exam level:
Exam description Begins with / Contains / Is
Performing Physician Begins with / Contains / Is
Referring physician Begins with / Contains / Is
Exam modality MG / DX N/A
Exam date Between / Before / After / On
Exam time
Series level:
Series description Begins with / Contains / Is
Series Type SCPT / RAW /PROCESSED N/A
Application level:
Apply this filter up to
the selected level
Patient / Exam / Series N/A

Chapter 10 Page no. 388 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup 2.
Filter name:
Circle the current value Note current value of the parameter
Patient level:
Patient Name Begins with / Contains / Is
Patient ID Begins with / Contains / Is
Exam level:
Exam description Begins with / Contains / Is
Performing Physician Begins with / Contains / Is
Referring physician Begins with / Contains / Is
Exam modality MG / DX N/A
Exam date Between / Before / After / On
Exam time
Series level:
Series description Begins with / Contains / Is
Series Type SCPT / RAW /PROCESSED N/A
Application level:
Apply this filter up to
the selected level
Patient / Exam / Series N/A

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
DR_019.fm Page no. 389 Chapter 10
3.
Filter name:
Circle the current value Note current value of the parameter
Patient level:
Patient Name Begins with / Contains / Is
Patient ID Begins with / Contains / Is
Exam level:
Exam description Begins with / Contains / Is
Performing Physician Begins with / Contains / Is
Referring physician Begins with / Contains / Is
Exam modality MG / DX N/A
Exam date Between / Before / After / On
Exam time
Series level:
Series description Begins with / Contains / Is
Series Type SCPT / RAW /PROCESSED N/A
Application level:
Apply this filter up to
the selected level
Patient / Exam / Series N/A

Chapter 10 Page no. 390 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
7-3. Browser Preferences
• Access: Tools menu/Browser preferences.
Select each set of preferences in turn: Layout; Filter; Sort.
Layout:
Record the current settings on the illustration.
Mark each selected button clearly with a cross.
Filter:
Record the current filter names for local and remote lists. If none, write "None".
Local lists. Current settin
g: . . . . . . . . . . . . . . . . . . . .
Remote lists. Current setting: . . . . . . . . . . . . . . . . . . . .
Sort:
For each parameter, circle the currently selected option:
Sort patients by: Name / Identifier / Last study date
Sort exams by: Date / Identifier / Description / Performing Physician / Referring Physician
Modality: MG / DX
Sort series by: Type / Number
Sort images by: Number / Old-new date / New-old date
Split series: Yes / No

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
DR_019.fm Page no. 391 Chapter 10
7-4. Medical Application Preferences
• Access: Tools menu/Medical application preferences.
Select each set of preferences in turn:
Auto Push; Auto Print; Annotations; Auto Windowing; Quality Check; Auto Delete; Names.
Auto push
Circle the current selection: Auto Push ON / OFF .
If Auto Push is ON, use the table below to enter the names in the Remote Host window, and circle the
selected push criterion for each host:
Auto print
Circle the current selection: Auto Print ON / OFF .
If Auto Print is ON, use the table below to enter the Dicom printer information For each named printer,
note the name, the selected print mode and the selected number of copies, and click the Printer proper-
ties button to obtain the selected print parameters: The parameters shown in square brackets in the table
(e.g., [Fit to film]) are default values which normally should not change. Copy this sheet if you need addi-
tional space.
Remote Host Name Criterion
Raw / Proc / Both / None
Raw / Proc / Both / None
Raw / Proc / Both / None
Raw / Proc / Both / None
Raw / Proc / Both / None
Dicom printer Name Criterion Criterion
Print mode [Fit to film]
Number of Copies
Printer properties
Format [standard]
Media type Paper / Clear film / Blue film
Destination Magazine / Processor
Film orientation Portrait / Landscape
Magnification type [Cubic]
Film size [8 in by 10 in]

Chapter 10 Page no. 392 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup Annotations
Record the annotation settings: place a cross in the box beside each option currently selected:
Auto Windowing
Circle the current selection: Standard / High / Low / User .
Quality check
Circle the current selection: OK / NOT OK .
Auto Delete
Circle the current selection: Auto Delete ON / OFF .
If Auto Delete is ON, use the table below to record the auto delete profile: place a cross in the box beside
each of the options currently selected:
Names
Use the table below to record the currently entered names:
Annotation level Selected annotations Printing annotations
Patient information
Acquisition date and time
Full Hospital information Model 1
Partial X-ray parameters Model 2
None Anatomical information Model 2
Custom Processing Information
Display parameters
Measurements
Raw images sent Raw images committed Both images committed
Processed images
sent
Processed images com-
mitted
Processed images
printed
Both images sent
Operators Performing Physicians Referring Physicians

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
DR_019.fm Page no. 393 Chapter 10
7-5. Worklist Management
• Access: Tools menu/Worklist management.
For each Host name listed, click Update to view tht parameter settings, and copy them into the table
below. Copy this sheet if you need additional space:
7-6. Network Management
• Access: Tools menu/Network management.
For each Host name listed, click Update to view the remote host parameter settings, and copy them
into the table below. Where multiple choices are shown, circle the selected parameter. Copy this
sheet if you need additional space
Parameter Entry Entry
Host label
Host name
Application Entity Title
Network address
Port number
Comments
Parameter Host 1 Host 2 Host 3
Host label
Host name
Application Entity Title
Network address
Network protocol
Port number
Query retrieve Patient / Study / No provider Patient / Study / No provider Patient / Study / No provider
Storage commitment Yes / No Yes / No Yes / No
Application Entity Title
Network address
Port number
Manage send Yes / No Yes / No Yes / No
Comments

Chapter 10 Page no. 394 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup 7-7. Printer Management
• Access: Tools menu/Worklist management.
For each Dicom printer listed, click Update to view the printer parameter settings. Copy the labels,
names, etc., into one of the illustrations below and mark all selected check boxes with a cross. Copy
this sheet if you need additional space:
- Printer 1:
DICOM printer parameters
DICOM printer label :
Hostname :
Application Entity Title (AET) :
Network address :
Port number :
Layouts :
Slide formats : 35mm 40mm
Pixel depth : 10 12
Film size :
8 in x 10 in W H
10 in x 12 in W H
10 in x 14 in W H
11 in x 14 in W H
14 in x 14 in W H
14 in x 17 in W H
24 cm x 24 cm W H
24 cm x 30 cm W H
Printer pixel size (micron) :
Configuration information :
Density : Min Max
Magnification type : Cubic
Smoothing factor :
Trim : No
Save Clear Done

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup
DR_019.fm Page no. 395 Chapter 10
- Printer 2:
DICOM printer parameters
DICOM printer label :
Hostname :
Application Entity Title (AET) :
Network address :
Port number :
Layouts :
Slide formats : 35mm 40mm
Pixel depth : 10 12
Film size :
8 in x 10 in W H
10 in x 12 in W H
10 in x 14 in W H
11 in x 14 in W H
14 in x 14 in W H
14 in x 17 in W H
24 cm x 24 cm W H
24 cm x 30 cm W H
Printer pixel size (micron) :
Configuration information :
Density : Min Max
Magnification type : Cubic
Smoothing factor :
Trim : No
Save Clear Done

Chapter 10 Page no. 396 DR_019.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
AWS Parameters manual backup This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 021 - Generator CPU Battery
DR_021.fm Page no. 397 Chapter 10
Job Card D/R 021 - Generator CPU Battery Chapter 10
1. SUPPLIES
Lithium battery.
2. TOOLS
• Soldering iron.
• DMR data diskette.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
None.
5. PROCEDURE
CAUTION
Before removing generator CPU board 400PL3, back-up the generator parameters, to
ensure that these parameters (normally saved using battery power) are not lost while the
battery is being changed - see D/R Job Card
Back-up/Restore Procedures.
5-1. Removal of generator CPU battery and installation of new battery
1. Remove the generator front panel (CPU side) D/R Job card Removal of covers.
2. Pull generator CPU board 400PL3 gently forward and remove it.
3. Solder the new lithium battery in the free position (BT1 or BT2), respecting the polarities.
4. Unsolder the old battery.
5. Reinstall generator CPU board 400PL3 in its position in the generator.
6. Reinstall the generator front panel (CPU side) D/R 101.
5-2. Functional check
1. Switch ON the Senographe DMR electrical power.
2. Check that the message:
SAVED MEMORY ERROR 071/08 is not displayed after initialization
3. If the above message is displayed, restore the generator parameters from the previously saved dis-
kette or from the AWS disk (see D/R Job Card
Back-up/Restore Procedures).
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW GENERATOR CPU BATTERY.

Chapter 10 Page no. 398 DR_021.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 021 - Generator CPU Battery This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 022 - Capacitor box
DR_022.fm Page no. 399 Chapter 10
Job Card D/R 022 - Capacitor box Chapter 10
1. SUPPLIES
Capacitor box.
2. TOOLS
• 8 mm flat wrench.
• 10 mm flat wrench.
• 4 mm Allen wrench.
3. SAFETY PRECAUTIONS
• Switch OFF mains power supply.
• Check residual voltage (inverter supply).
4. PREREQUISITES
None.
5. PROCEDURE
5-1. Removal of filtering capacitors
Refer to Illustrations1 and 2.
1. Remove the four generator side panels (see D/R Job Card
Removal of covers).
2. Switch OFF the electrical power supply, both in the main cabinet and on the rotary switch in the gen-
erator cabinet (position 0).
3. Using a voltmeter (rating 1000 V DC), check that there is no residual voltage between screws S1 and
S2 before proceeding.
4. Remove screws S1 and S2 (10 mm flat wrench), and remove the two resonant capacitors, RC1 and
RC2, together with their upper plate.
5. Remove screw S3, followed by screws S4, S5 and S6 (8 mm flat wrench).Remove screws S7 and S8,
followed by screws S9 and S10 mounting the capacitor box on the "Inverter" panel (4 mm Allen
wrench).
6. Remove the capacitor box.
5-2. Installation of filtering capacitors
• To install new filtering capacitors, carry out the above operations in reverse order .
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH NEW INVERTER FILTERING CAPACITORS.

Chapter 10 Page no. 400 DR_022.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 022 - Capacitor box
ILLUSTRATION 1 - CAPACITOR BOX - TOP VIEW (WITHOUT RESONANT CAPACITORS)
ILLUSTRATION 2 - CAPACITOR BOX - SIDE VIEW
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
C7 C5 C3 C1
C8 C6 C4 C2
C7+
C8-
C1+
C1-/C2+
C2-
C8+/C7-
S4
S5
S6
S1
S3
S2
RC1 RC2
S1 S2
S3
S7, S8
S9, S10
CAPACITOR BOX
RESONANT CAPACITORS

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 025 - HV and kV/mA tank
DR_025.fm Page no. 401 Chapter 10
Job Card D/R 025 - HV and kV/mA tank Chapter 10
1. SUPPLIES
• HT block, equipped kV/mA.
• HT grease.
2. TOOLS
• Ammeter.
• Mammo kV peak meter.
• 8 mm flat wrench.
• X-ray waveform meter RMI 233 and oscilloscope (see Section 3, below).
3. SAFETY PRECAUTIONS
• X-ray emission safety precautions are applicable.
•If only the kV/mA board is being changed (not recommended) then the HT probe frequency
response must be calibrated. This requires the use of tool RMI 233 and an oscilloscope. Not cali-
brating the HT probe frequency response may result in non-compliance with CDRH standards (see
below) and will affect the AOP behaviour (incorrect breast thickness measurement during pre-expo-
sure).
4. PREREQUISITES
None.
5. PROCEDURE
CAUTION
The HT block and kV/mA board assembly is factory calibrated. DO NOT ATTEMPT TO
ADJUST POTENTIOMETER P1 (HT probe frequency compensation).
5-1. Removal and installation of HT and kV/mA block
1. Switch OFF the DMR electrical power.
2. Remove the generator covers (see D/R Job Card
Removal of covers).
3. On the HT block, disconnect the following, in the sequence listed:
- HT connector,
- heating/bias connector GKB,
- ground connections (3 wires) (8 mm flat wrench),
- inverter GKA/1 and GKA/2 (8 mm flat wrench),
- HT probe cable: GKC.
4. On the kV/mA board, disconnect XJ2, XJ3 and XJ1.
5. Tighten the air-bleed screw.
6. Slacken the HT block mounting clamp and remove the block.
7. Install the new HT block and reconnect the cables (grease the HT end-piece if necessary).
8. Slacken the air-bleed screw.

Chapter 10 Page no. 402 DR_025.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 025 - HV and kV/mA tank 6. CHANGING THE KV/MA BOARD ONLY
• Changing the kV/mA board alone IS NOT RECOMMENDED (see Section 3, above). However, if the
change is absolutely necessary, the calibration procedure is given below:
1. Configure the Senographe DMR as follows:
- magnification device in position 1.9; Large focal spot; 2-point mode; 30 kV, 30 mA.
2. Place the RMI 233 probe on the magnification plate, in the center of the X-ray beam. Connect the
probe to an oscilloscope with memory facility; select scan 5 ms/division.
3. Execute an exposure and observe the beginning of the exposure waveform on the oscilloscope.
Adjust P1 to obtain the shortest rise time, with no overshoot, as shown in the waveform 1 below:
CAUTION
Incorrect adjustment results in wrong radiological thickness measurement during the pre-
exposure in AOP mode. When the adjustment is correct a normal exposure with 40 mm of
plexiglas in AOP mode, and an exposure aborted just after pre-exposure in the same con-
ditions, should display the same radiological thickness on the technical parameter display.
7. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW HT BLOCK AND kV/mA BOARD
V
t
V
t
V
t
1
2
3
CORRECT
RESPONSE
UNSATISFACTORY
FOR AEC/AOP
NON-COMPLIANCE
WITH CDRH !
UNSATISFACTORY
FOR AEC/AOP &
HV RISETIME

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 026 - Magnification device
DR_026.fm Page no. 403 Chapter 10
Job Card D/R 026 - Magnification device Chapter 10
1. SUPPLIES
• None.
2. TOOLS
• 2.5 mm Allen wrench.
3. SAFETY PRECAUTIONS
• None.
4. PREREQUISITES
• None.
5. PROCEDURE
Refer to illustration 1.
1. Insert the plate in the guideway with the teeth engaged (Item 3).
2. Using the 2.5 mm Allen wrench, loosen the M5 Allen screw (Item 2) securing the left side of the guide
shaft (Item 1)
.
3. Keeping the magnification plate in contact with the bearing plate, and with the tooth (Item 3) in its
housing, pull the latching handle assembly (Item 1) to bring the tooth into contact with the edge of its
housing, and so removing play.
4. Tighten the M5 Allen screw (Item 2) securing the left side of the guide shaft.
5. Check that there is no play in the two magnification positions.
ILLUSTRATION 1 - ADJUSTMENT OF MAGNIFICATION DEVICE

Chapter 10 Page no. 404 DR_026.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 026 - Magnification device
Co mpletion
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 027 - Arm Angle Sensor (potentiometer)
DR_027.fm Page no. 405 Chapter 10
Job Card D/R 027 - Arm Angle Sensor (potentiometer) Chapter 10
1. SUPPLIES
• Arm Angle Sensor.
2. TOOLS
• Small flat screwdriver.
3. SAFETY PRECAUTIONS
• No specific safety precautions are applicable.
4. PREREQUISITES
Note:
Automatic view selection: The Senographe 2000 D system normally provides automatic selec-
tion of the view name, which is displayed on the senographe console. View selection depends on
the value of the compression arm angle and on the laterality (L or R) selected by the operator. This
function is active by default; it depends upon correct operation of the Arm Angle Sensor. Continued
operation in degraded mode (as described in the o.m.) is possible after failure of the angle sensor,
by disabling automatic view selection, as follows:
- Set the bit 8 of switch B1 on Generator CPU board to access INSTALL menu of DMR Console;
- Enter menu: SETUP/ARM/VALID/INSTAL/CONFIG/ANGLE/NO
• Remove Compresson Arm front covers. See D/R Job Card
Gantry arm and Tube housing covers.
5. PROCEDURE
5-1. Disassembly of Arm Angle Sensor
1. Locate Arm Angle Sensor at the front bottom of Compression arm.
2. Disconnect connector (1) of Angle Sensor cable
(at top of compression arm).
1

Chapter 10 Page no. 406 DR_027.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 027 - Arm Angle Sensor (potentiometer)3. Remove plate (2) and ties which attach the angle
sensor cable along the compression arm.
4. Remove the two screws (4, 5) securing the Arm
Angle Sensor (3).
5. Remove and discard the defective sensor (3).
5-2. Reassembly
1. Reverse disassembly procedure.
CAUTION
Take care to remount correctly the angle
sensor. Potentiometer is seen from opera-
tor side.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW ARM ANGLE SENSOR.
2
4
5
3

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 101 - Removal of Covers
DR_101.fm Page no. 407 Chapter 10
Job Card D/R 101 - Removal of Covers Chapter 10
1. SUPPLIES
None
2. TOOLS
• Metric Allen wrenches (including 5.5 mm).
• Large flat screwdriver.
• Slim flat blade (knife or similar).
3. SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment and switch OFF the Senographe hospital
supply circuit breaker. Lock it in the OFF position or mark it clearly with a warning sign to
prevent others from applying power.
NEVER switch off at the UPS except in emergency (risk of data loss).
4. PREREQUISITES
None
5. REMOVAL AND REPLACEMENT OF COVERS
5-1. Remove/replace generator covers
Refer to Illustration 1 for the locations of securing screws.
1. Remove all power from the DMR.
2. Remove the four plastic plugs from screws of generator top panel.
3. Use an Allen wrench to remove the four generator top panel retaining screws (5).
4. Remove generator top panel by lifting upwards.
5. For EACH of the four generator side panels:
a. Slightly lift the panel upwards and towards yourself.
b. Disconnect the ground wire
at the Generator end to avoid risk of creating a short circuit.
6. Perform the procedure in reverse order for replacement.
Be sure to reconnect the ground wire on
each of the four side panels
Take care when replacing the Generator Cabinet covers, especially after connecting air and
water pipes to the Conditioner. Check that the covers do not block the pipes and cause a
blockage.
5-2. Remove/replace gantry covers
Refer to Illustration 1 for the locations of securing screws.
1. Remove all power from the DMR.
2. Use a hex wrench to remove screw (1).
3. Remove plastic plugs and screws (2) and (3) from each column half-cover.

Chapter 10 Page no. 408 DR_101.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 101 - Removal of Covers 4. Do not touch screws (4). Undo hex bolt (6) by 2 or 3 turns (accessible through a slot at rear of the
base); this method retains correct alignment of covers with safety stop switches.
5. Slide the complete half-cover towards you to remove it.
6. Perform the procedure in reverse order for replacement.
BE CAREFUL NOT TO PINCH OR DAMAGE CABLES.
Note:
It is possible to remove only one column half-cover.
ILLUSTRATION 1 - COVER SECURING SCREW LOCATION
4
1
2
3
5
CPU AND OTHER
BOARDS ACCESSIBLE
FROM THIS SIDE
6
2
3
5
1
4
6

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 101 - Removal of Covers
DR_101.fm Page no. 409 Chapter 10
5-3. Gantry arm and Tube housing covers
WARNING
Be careful when touching the tube housing, as its temperature can be up to 80 °C (about
180 °F)
5-3-1. Removal and reinstallation of tube housing arm lower covers
Refer to Illustration 2.
1. Using a flat blade, remove the plastic caps - four around the rotation axis and two at the bottom end
of the arm (three in each half-cover).
2. Set the arm in the horizontal position, tube on the right, and remove the half-cover mounting screws
(1, 2, and 3).
3. Lift the half-cover and remove it.
4. Rotate the arm through 180 degrees.
5. Repeat 2 and 3, above, for the other half-cover, but removing mounting screws 4, 5, and 6.
6. To reinstall the arm lower covers, carry out the above operations in reverse order.
ILLUSTRATION 2 - TUBE HOUSING ARM LOWER COVERS - MOUNTING SCREW LOCATIONS
4
5
1
2

Chapter 10 Page no. 410 DR_101.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 101 - Removal of Covers
5-3-2. Removal and reinstallation of compression arm covers
1. Set the compression arm in the horizontal position, tube
on the right, and unlock the tube housing arm to set it at
27 degrees, with the tube below the horizontal.
2. Switch OFF the DMR electrical power.
3. Refer to the illustration. Use the flat screwdriver and 4 mm
Allen wrench to remove the two left hand cover securing
screws (1 and 2).
4. Remove the left half-cover by slightly lifting it on the com-
pression motor side, and then gently pulling it to the left.
5. Switch the DMR electrical power back on.
6. Reset the tube housing arm to 0 degrees.
7. Rotate the compression arm through 180 degrees and
unlock the housing arm to set the tube 27 degrees down-
wards.
8. Switch OFF the DMR electrical power.
9. Use the flat screwdriver and 4 mm Allen wrench, remove
the two right hand cover securing screws.
10. Remove the right half-cover by slightly lifting it on the com-
pression motor side, and then gently pulling it to the right.
11. Recover spacer plate E (on the top of the sensor indicator plate P) and remove plate P by gently pull-
ing it towards the bottom of the compression arm.
12. To reinstall the compression arm covers, carry out the above operations, in the reverse sequence.
CAUTION
Reinstall spacer plate E as quickly as possible. It will stay in place when one of the half-
covers has been reinstalled. To prevent plate P from falling, do not set the compression
arm in the vertical position until at least one half-cover has been installed.
1
2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 101 - Removal of Covers
DR_101.fm Page no. 411 Chapter 10
5-3-3. Removal and reinstallation of tube housing spacer covers
Refer to Illustration 4.
1. Remove strip B (clipped in position).
2. Set the tube housing arm horizontal, with the head on the right.
3. Remove the plastic covers and screws 3, 5 and 7.
ILLUSTRATION 3 - TUBE HOUSING SPACER COVERS - MOUNTING SCREW LOCATIONS
4. Remove the left half-cover.
5. Rotate the arm through 180 degrees (head on the left).
6. Remove the plastic covers and screws 4, 6 and 8.
7. Remove the right half-cover.
8. To reinstall the tube housing spacer covers, carry out the above operations in reverse order.
CAUTION
Take care to install the short screws in positions 3 and 4.
4
8
6
7
5
B
P
3

Chapter 10 Page no. 412 DR_101.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 101 - Removal of Covers 5-4. Cart covers
5-4-1. Rear cover
To remove the rear cover :
a. Remove the two securing screws B at the rear of
the Cart (cross–head screwdriver).
b. Release the cover by tilting it towards you.
c. Disconnect the ground connections which are
made from the various units in the Cart to the rear
cover.
d. Remove the cover by lifting.
Reverse the procedure to replace the cover.
Remember to
reconnect all
ground connections!
5-4-2. Side covers
To remove a side cover:
a. First remove the rear cover (see above).
b. Remove the external securing screw C (cross–
head screwdriver).
c. Remove the internal securing screw D (cross–head
screwdriver).
d. Remove the cover by lifting.
Reverse the procedure to replace the cover.
B
C
D

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 102 - Image Receptor Power Supply
DR_102.fm Page no. 413 Chapter 10
Job Card D/R 102 - Image Receptor Power Supply Chapter 10
1. SUPPLIES
• Image Receptor Power Supply
2. TOOLS
• 5 mm wrench.
•7 mm wrench.
• small screwdriver.
3. SAFETY PRECAUTIONS
WARNING
DMR senographe must be powered off.
4. PREREQUISITE
• Remove DMR column covers. See D/R Job Card Removal of covers.
5. PROCEDURE.
5-1. Disassembly of Power Supply
Refer to Illustration 1.
1. Switch off the DMR senographe.
2. Locate the defective power supply (6)
mounted on the left of the DMR column.
3. Disconnect the AC mains connector (3)
and the two low voltage connectors (1)
(one to Image Receptor and one to IDC).
4. Release the two securing clamps (4, 5)
located at bottom of power supply chassis
(two hex 5 mm nuts).
5. Remove the two hex 5 mm securing
screws and washers (2, 7) holding the top
of the power supply chassis. Take care not
to lose the screws.
1
2
3
4
5
7
6

Chapter 10 Page no. 414 DR_102.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 102 - Image Receptor Power Supply 6. Remove the defective power supply by holding both the
top and bottom of its chassis with both hands.
Note:
Tilt the power supply assembly slightly forward to
ease removal.
5-2. Reassembly
• Reverse the disassembly procedure.
5-3. Functional check
1. Switch ON the Senographe DMR electrical power.
2. Check that the normal boot message is shown.
3. Make a test exposure.
.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE DMR IS NOW OPERATIONAL WITH THE NEW POWER SUPPLY.
8

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 103 - Image Receptor
DR_103.fm Page no. 415 Chapter 10
Job Card D/R 103 - Image Receptor Chapter 10
1. SUPPLIES
• Image Receptor and its floppy disk with parameters.
2. TOOLS
• Right-angled flat screwdriver.
• Set of metric Allen wrenches.
• 7 mm hex wrench.
• Thread locking compound (Loctite blue).
3. SAFETY PRECAUTIONS
WARNING
If the casing is punctured, remove the detector using protective gloves and dust masks.
Send the protective items for disposal along with the defective detector.
WARNING
The DMR senographe must be powered OFF.
4. PREREQUISITES
• Remove Bucky (or Breast holder).
• Remove Compression Arm covers for access (refer to the D/R Job Card
Removal of Covers).
5. PROCEDURE
CAUTION
When disconnecting and connecting coolant hoses on the image receptor connector
panel, the panel should be facing downwards (i.e., normal position), to minimize the risk of
coolant spilling or splashing onto the adjacent electrical connector.

Chapter 10 Page no. 416 DR_103.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 103 - Image Receptor 5-1. Disassembly of Image Receptor
1. Refer to the upper illustration:
• Release (two screws) and remove the front con-
nector (1).
2. Refer to the lower illustration:
• Disconnect the two optical fiber cables (2) by
extracting the optical connectors by hand.
CAUTION
Do not apply any force to fiber cables and take
care not to bend them excessively.
• Disconnect the air and water supply pipes (3) by
puling the connectors (not the pipes!).
CAUTION
Be careful to avoid spillage of coolant, especially onto or near electrical connections. If
any coolant is spilled, it must be cleaned up immediately.
3. Remove the bucky (two screws) for access to
the bucky interconnection plate (5).
4. Remove the two screws (4) securing the
bucky interconnection plate (5) to the rear wall
plate of the Image Receptor support (6).
1
2
3
4
5
6

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 103 - Image Receptor
DR_103.fm Page no. 417 Chapter 10
5. Remove three of the four Image Receptor securing
screws (7) located at the bottom of the compression
arm (5 mm allen wrench).
Note:
It is necessary to tilt the exam arm first to the left
and then to the right for access to the securing
screws (two on each side).
6. Loosen the last screw, then hold the receptor to pre-
vent it from falling while removing the screw. It may
be advisable to lower the arm to a more convenient
height before performing this step.
7. Remove the Image Receptor (8) by holding it with
both hands and pulling it backward from its guides.
8. Release the hex screw securing the ground cable ter-
minal (7 mm hex wrench and 5 mm allen wrench).
9. Place the defective Image Receptor on a stable stand.
WARNING
Drain all coolant from the detector before it is stored or shipped. Shipping a detector
which contains water may cause irreparable damage. This is especially important in win-
ter, when there is a danger of freezing.
7
8

Chapter 10 Page no. 418 DR_103.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 103 - Image Receptor
5-2. Reassembly
• Reverse the disassembly procedure
• Check the Conditioner fluid level (refer to D/R Job Card
Conditioner Fill and Maintenance).
CAUTION
Be careful to avoid spillage of coolant, especially onto or near electrical connections. If
any coolant is spilled, it must be cleaned up immediately.
5-3. Clear Calibrations and Load Image Receptor Parameters
The new Image Receptor is supplied with a floppy disk containing parameters specific to the detector. To
clear the current detector calibrations and download the new parameters into the system, proceed as fol-
lows:
1. Power on the system.
2. Select
Service desktop/Calibration/Detector; then select Clear Calibrations and perform the clear
procedure as indicated on the screen.
3. Select
Service desktop/Utilities, then click on the Detector Replacement button and follow the instruc-
tions. Insert the floppy disk supplied with the new Image Receptor as requested.
5-4. First power-on check
1. Cycle OFF/ON the system,
2. Check that the message
All devices are ready appears on the screen,
3. If the above message is displayed, continue with step 5.5 below. If not, verify that the various connec-
tions are correctly made.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
6-1. Data Backup
• After performing the specified calibration and check procedures, a system backup must be made to
safeguard the new detector and calibration data.
6-2. CAD License
• If the CAD option is present, the new detector ID must be declared on the DMPU. Follow the instruc-
tions given in the Senographe 2000D CAD Option s.m. Job Card
License Management.
THE SENOGRAPHE 2000 D IS NOW OPERATIONAL WITH THE NEW IMAGE RECEPTOR.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 104 - IDC unit
DR_104.fm Page no. 419 Chapter 10
Job Card D/R 104 - IDC unit Chapter 10
1. SUPPLIES
• IDC unit.
2. TOOLS
•5 mm wrench.
• 4 mm Allen wrench
• Small flat and cruciform screwdrivers.
• Blank 3.5" HD floppy diskette.
3. SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment correctly and switch OFF the
Senographe 20000 D hospital supply circuit breaker. Lock it in the OFF position or mark it
clearly with a warning sign to prevent others from applying power.
4. PREREQUISITES
• Remove Cart rear cover. See D/R Job Card Removal of covers.
5. PROCEDURE.
5-1. Disassembly of IDC unit
1. Locate the IDC unit (1) inside FFDM Cart.
2. Disconnect all cables from the IDC unit rear panel.
Note:
Note locations of disconnected cables.
1

Chapter 10 Page no. 420 DR_104.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 104 - IDC unit 3. Remove the screw (2) securing the IDC drawer to the
Cart left upright.
4. Disconnect the IDC tray ground wire from the terminal at
the right of the tray (see the rear panel connection illustra-
tion, section 7 item 8).
5. Pull out the IDC unit tray (3).
6. Undo the Velcro strips (4) attaching the IDC unit to the
tray.
7. Remove the defective IDC unit (5).
5-2. Reassembly of IDC unit
• Reverse the disassembly procedure, but do not connect the fiber optic cables.
See section 7 for IDC rear panel connections.
Notice:!
The fiber optics must not be connected before loading software.
5-3. Software Restore
1. To restore all data and software to the IDC unit, perform the IDC Restore operation as described in
the D/R Job Card
Back-up/Restore Procedures.
Notice:!
Be sure to carry out only the Restore operation - if you make a new backup you will destroy the
existing backup information!
2. Boot/reset (power OFF/ON) the IDC unit, then reconnect the fiber optic cables.
5-4. Functional check of new IDC unit
1. Switch OFF, then back ON the Senographe 2000D electrical power.
2. Check that the message "system is operational" appears on the screen,

3. If the above message is displayed, the Senographe 2000D is now operational. If not, verify that the
various connections are correctly made.
4. Make an acquisition.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
2
3
4
5

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 104 - IDC unit
DR_104.fm Page no. 421 Chapter 10
7. IDC REAR PANEL CONNECTIONS
Reference Function Reference Function
1 DMR Link RT (1) 2240227 5 Ethernet (To ADS)
2 Conditioner 2226612 6 AC mains
3 Detector Power Supply 2226621 7 Optical Fiber cables (to detector)
4 DMR Link Serial 2240227 8 Tray ground wire
2
3
1
4
6
7
5
2
3
1
4
6
7
5
8
IDC - Dione II
IDC - Dione I

Chapter 10 Page no. 422 DR_104.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 104 - IDC unit This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 105 - Conditioner Unit
DR_105.fm Page no. 423 Chapter 10
Job Card D/R 105 - Conditioner Unit Chapter 10
1. SUPPLIES
• Conditioner unit.
Kit including:
- Conditioner pre-mixed coolant (part # 2178181).
- Funnel.
- Drain tube (short length of water pipe with connector).
• Empty bottle.
2. TOOLS
• Drain tube
• Small flat screwdriver.
3. SAFETY PRECAUTIONS
CAUTION
Read and follow the Safety Precautions given in the D/R Job Card Conditioner Fill and
Maintenance when draining the Conditioner and handling coolant.
• Follow the instructions given for shut down of the system before beginning disassembly
4. PREREQUISITES
• Remove Generator Cabinet covers as required for access. See D/R Job Card Removal of covers.
5. PROCEDURE
5-1. Preparation - Conditioner Drain
• Drain the Conditioner Unit as described in the Conditioner Drain section of the D/R Job Card Condi-
tioner Fill and Maintenance
.
5-2. Disassembly of Conditioner
• Before beginning work, shut down correctly the equipment and switch OFF the DMR hospital supply
circuit breaker. Lock it in the OFF position or mark it clearly with a warning sign to prevent others from
applying power.

Chapter 10 Page no. 424 DR_105.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 105 - Conditioner Unit 1. Locate the Conditioner unit (1) in the Mains
Distribution Rack.
2. Disconnect serial line connector (2) from the
Conditioner unit.
3. Disconnect the two air pipes and two water pipes (4)
from the Conditioner unit.
Note:
The water connectors are self-sealing.
4. Disconnect the AC Mains cable connector (3).
1
3
4
2

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 105 - Conditioner Unit
DR_105.fm Page no. 425 Chapter 10
5. Remove the two front screws (6) and the two rear
screws (5) securing the Conditioner unit to the
generator support.
6. Remove defective Conditioner unit from the Gen-
erator cabinet by pulling it through the left side of
the generator chassis.
5-3. Reassembly
Reverse the disassembly procedure.
5-4. Coolant Fill
• Fill the Conditioner Unit and cooling system as described in the Coolant Fill section of the D/R Job
Card
Conditioner Fill and Maintenance.
5-5. Functional check
• Switch ON the Senographe electrical power
• Check that the conditioner display shows the message IDC CONTROL for about 5 seconds shortly
after the system is switched on.
• Check that within about 15 minutes of operation manifold temperature shown on the Conditioner dis-
play is 20°C +
2°C.
• About 15 minutes after the system is operational, make an exposure; verify that the exposure pro-
ceeds correctly, with no warning message displayed after the acquisition.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.
THE SENOGRAPHE IS NOW OPERATIONAL WITH THE NEW CONDITIONER UNIT.
5
6

Chapter 10 Page no. 426 DR_105.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 105 - Conditioner Unit This page is blank.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 106 - Conditioner Fill and Maintenance
DR_106.fm Page no. 427 Chapter 10
Job Card D/R 106 - Conditioner Fill and Maintenance Chapter 10
1. SUPPLIES
Cooling liquid kit (currently p/n 2260360, to be replaced by p/n 2295518) including:
- Pre-mixed coolant for conditioner. This is provided in two containers; marked
Flushing and Top-
ping
. Their contents are identical, the labelling is only for convenience of use.
- Funnel.
- Safety goggles.
- Drain tube (short length of water pipe with connector).
- Empty container labelled
Empty.
2. TOOLS
• Small flat screwdriver.
3. SAFETY PRECAUTIONS
• When filling or topping up the unit, take great care to keep water away from the vents, the ON/OFF
switch and the power cord connection. Do not immerse the unit in any liquid.
• Dispose of old and surplus coolant with care as described in the section
Coolant Disposal; refer to
the
Caution below.
CAUTION
• Before adding coolant to the Conditioner, always switch OFF the Generator and MDR cir-
cuit breakers in the Generator Cabinet. Switch on the MDR breaker only as instructed.
• Always use a funnel when adding coolant to the Conditioner.
• Always wear goggles when working with the coolant.
• The coolant provided for filling the Conditioner circuit is water protected by a fungicide.
In its undiluted form this fungicide is toxic, but in the solution provided it is very dilute,
and is not dangerous to health if handled correctly. Waste solution in the concentration
used will not cause pollution if disposed of in a non-sensitive environment.
• However, when handling the solution you should treat it with care; avoid unnecessary
contact and wear goggles. See below for First Aid advice.
First Aid in case of contact with Conditioner coolant:
Immediately flush eyes or skin with plenty of water for at least 15 minutes.
If coolant contacts the eye, call a physician immediately.
If coolant is swallowed, drink one to two glasses of water or milk. Avoid alcohol. Call a physician
immediately.
Remove and wash contaminated clothing before reuse
4. PREREQUISITES
• Remove Generator Cabinet covers as required for access to the unit. Refer to D/R Job Card Removal
of covers
.

Chapter 10 Page no. 428 DR_106.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 106 - Conditioner Fill and Maintenance
5. PROCEDURE
• In addition to the following instructions, refer to document 2295518INS (included in the coolant kit) for
detailed information on coolant change in Senographe 2000 D systems.
CAUTION
When performing any of the following procedures, be careful to avoid spillage of coolant,
especially onto or near electrical connections. If any coolant is spilled, it must be cleaned
up immediately.
5-1. Normal maintenance
• Check the water level weekly or when the CHECK FLUID message appears.
• Be sure to use only GE pre-mixed coolant provided in the Cooling Liquid kit p/n 2295518 for filling the
reservoir. Use the container labelled
Topping.
• Clean the exterior of the unit with a soft cloth and warm, soapy water. Never use abrasive or solvent-
based cleaners to clean the unit.
5-2. Coolant Fill
When any part of the cooling circuit has been disconnected (e.g., during system installation, or after
replacement of the Conditioner unit or the Digital Detector), or if the liquid level is observed to be low, the
cooling liquid circuit must be correctly filled. Proceed as described below.
Notice:!
If the cooling circuit is being filled as part of the first system power-up, the sequence given in the
section
First Power On in Installation Job Card System Power On must be followed.
1. Check that all hoses are correctly connected.
2. Remove the reservoir cap, then slowly add liquid
from the container labelled
Topping to the reservoir
until the liquid reaches the bottom of the cap
threads. The total capacity of the cooling system is
approximately 500 ml (1 pint).
3. Switch on the unit (switch ON the MDR circuit
breaker and press the Conditioner ON button in the
generator cabinet).
4. As the water circulates and air bubbles are flushed
from the system, the water level falls. Add liquid as
necessary until the level is steady at the bottom of
the cap threads.
5. Replace the reservoir cap when you are satisfied
that the level is correct.
MDR Breaker
ON/OFF
Conditioner
Detector supply
Cart
UPS
Generator
Breaker
push buttons
:
Conditioner
and
MDR
Generator

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 106 - Conditioner Fill and Maintenance
DR_106.fm Page no. 429 Chapter 10
5-3. Conditioner Drain
Use this procedure to drain the cooling system before replacing any part of the circuit:
1. Shut down the entire system by clicking the
System Shutdown button on the Browser window.
2. Switch OFF the Generator and MDR circuit breakers in the Generator Cabinet.
3. Prepare the
Empty container provided in the Cooling Liquid kit to receive the used coolant. Place the
end of the drain tube into the container.
4. Disconnect the coolant input pipe from the Conditioner. The pipe connector seals automatically.
5. Connect the disconnected coolant input pipe to the drain tube. The connector is now unsealed, and
coolant can drain into the prepared container.
6. Switch ON the MDR circuit breaker and press the Conditioner ON button in the generator cabinet.
7. Coolant drains into the container, until the “Check Fluid” alarm stops the flow.
8. Switch OFF the MDR circuit breaker.
9. The Conditioner is now empty. Note that coolant may still be present in other parts of the circuit.
When disconnecting pipes (for example from the Digital Detector), be ready to empty any remaining
coolant carefully into the
Empty container.
10. Label the container containing the drained coolant to indicate the nature of its contents; cross out the
word
Empty on the label, and copy the information from one of the coolant containers (Flush or Top-
ping
) provided in the kit. See below for disposal.
11. Switch ON the MDR and Generator circuit breakers; replace the Generator Cabinet covers.
5-4. Conditioner Flush
The coolant must be changed every 6 months. Proceed as follows:
1. Follow the drain procedure given above (
Conditioner Drain) to the point where the “Check Fluid”
alarm stops the flow (step 7).
2. Switch OFF the MDR circuit breaker and refill the conditioner reservoir with clean pre-mixed coolant
from the container labelled
Flush.
3. Switch ON the MDR circuit breaker and press the Conditioner ON button in the generator cabinet.
4. Coolant drains into the container, until the “Check Fluid” alarm again stops the flow.
5. The circuit now contains fresh clean liquid.
6. Switch OFF the MDR circuit breaker; fill the conditioner reservoir to the bottom of the cap threads.
7. Disconnect the coolant input pipe from the drain tube; reconnect it to the Conditioner input .
8. Top up the coolant as required from the container labelled
Flush, using the procedure given above
(
Coolant Fill, steps 3, 4, and 5), until the level is steady and you can replace the reservoir cap.
9. Label the container containing the drained coolant to indicate the nature of its contents; cross out the
word
Empty on the label, and copy the information from one of the coolant containers (Flush or Top-
ping
) provided in the kit. See below for disposal.
10. Switch ON the MDR and Generator circuit breakers; replace the Generator Cabinet covers.

Chapter 10 Page no. 430 DR_106.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 106 - Conditioner Fill and Maintenance
5-5. Cleaning the Air Filter
To clean the air filter, proceed as follows
1. Follow the instructions of the D/R Job card
Conditioner Unit to power down, access and release the
Conditioner Unit.
2. Remove the air filter from the air filter housing. Pull the plastic tab towards the front of the unit then
pull the tab out to begin dislodging the air filter.
3. Wash the air filter by holding the clean side of the air filter under running water for either two minutes
or until the dirty side of the filter Is clean.
4. Allow the filter to dry. Remove any lint in the filter by tapping gently until the lint dislodges.
5. Replace the air filter in its housing
6. Secure the unit in its normal position.
5-6. Coolant Disposal
• In a normal non-sensitive environment, used and surplus coolant may be disposed of by pouring
down a drain.
• Triple-rinse containers before placing them in appropriate waste receptacles.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 107 - AWS unit
DR_107.fm Page no. 431 Chapter 10
Job Card D/R 107 - AWS unit Chapter 10
1. SUPPLIES
• AWS unit.
2. TOOLS
• 5 mm wrench.
• 4 mm Allen wrench
• small flat screwdriver.
• CD-ROM
Senographe 2000 D AWS Application Software (supplied with the system).
3. SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down correctly the equipment and switch OFF the
Senographe 2000 D hospital supply circuit breaker. Lock it in the OFF position or mark it
clearly with a warning sign to prevent others from applying power.
4. PREREQUISITES
• Remove Cart rear cover. See D/R Job Card Removal of covers.
5. PROCEDURE.
5-1. Disassembly of AWS unit
1. Locate AWS unit (1) inside the AWS Cart..
2. Remove all connected cables from the AWS unit rear
panel.
Note:
Note locations of all disconnected cables.
1

Chapter 10 Page no. 432 DR_107.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 107 - AWS unit 3. Remove screw (2) securing the AWS tray to the Cart
left upright.
4.Disconnect the ground wire from the terminal at
the right of the AWS tray (see illustration of AWS
Rear Panel Connectors, section 5-8).
5. Pull out the AWS unit tray (3).
6. Undo the Velcro strips (4) securing the AWS unit to
the tray.
7. Remove the defective AWS unit (5).
5-2. Reassembly of new AWS unit
• Reverse the disassembly procedure.
Note:
See section 5-3 for AWS rear panel connections.
2
3
4
5

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 107 - AWS unit
DR_107.fm Page no. 433 Chapter 10
5-3. AWS Rear Panel Connectors
Reference Function
1 Service-specific cable (connected at AWS
panel only); 2244433
2 External Ethernet; 2226632
3 CD-ROM Writer; 530-2115-02
4 Generator Link; 2226625
5 Not used
6 UPS; 2226615
7 Ethernet IDC; 2226628
8 Cart Monitor Display
9 Serial A (Modem); 2175122
10 Cart Keyboard
11 AC mains
12 Tray ground connection
2
3
5
6
128911 10
14
7

Chapter 10 Page no. 434 DR_107.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 107 - AWS unit
5-4. Software Installation
• Install the AWS Application from the AWS Application Software CD-ROM supplied with the system.
CAUTION
It is important to load the same software version as that installed in the old AWS.
• Follow the instructions given in the CD-ROM booklet to install the software.
5-5. Back-Up and Restore of data
Refer to the D/R Job card Back-Up and Restore Procedure.
1. Restore AWS parameters from the back-up diskette. If the back-up diskette is not available, the
patients in the Worklist will be lost, and the procedures given in the CFG Job Card
AWS Configura-
tion
must be carried out.
2. Because the IDC and DMR parameter back-up copies were held in the AWS, a new back-up must be
performed in order to make IDC and DMR back-up parameters available.
Perform a system back-up.
Notice:!
If this is not done, no back-up for IDC and DMR parameters will exist.
5-6. Re-installation of CD-R writer
After removal and re-installation of the AWS, the CD-ROM unit must be re-installed. Follow the proce-
dures in Job Card IST
Installation of the CD-ROM Writer.
5-7. Declaration of Network Devices
Open the Remote Hosts Selection window (refer to CFG Job Card Declare Network Hosts) and verify
that all network devices are still correctly declared. If not, follow the procedures given in the CFG Job
Cards
Declare Network Hosts and Declare Printers to declare hosts as required.
5-8. Functional check of new AWS unit
1. Power up the Senographe 2000 D and check that the message All devices are ready appears on the
AWS screen,
2. If the above message is displayed, the Senographe 2000D is now operational. If not, verify that the
various connections are correctly made.
3. Perform an image acquisition, then a network transfer to other devices.
6. COMPLETION
• Carry out the procedure(s) specified as Required Calibrations and Checks for this Job Card in Job
Card
Calibration Interactions.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 108 - Software reload
DR_108.fm Page no. 435 Chapter 10
Job Card D/R 108 - Software reload Chapter 10
1. SUPPLIES
• Software kits supplied with the system
2. TOOLS
• None.
3. SAFETY PRECAUTIONS
No specific safety precautions are applicable.
4. PREREQUISITES
• Wherever possible, a full system backup must be available. Refer to the D/R Job card Back-Up/
Restore Procedure
.
5. AWS SOFTWARE
5-1. AWS Full Software Reload
• A full AWS software reload (operating system and application software) may be necessary if, for
example, the AWS unit has been replaced, or if lost or corrupted files make it impossible to run the
application after rebooting, and reloading the application only is not successful.
Use the same AWS software CD-ROMs as used for the current installation (supplied with the system
or supplied for a software update).
CAUTION
If reloading software after a system crash or after replacement of the AWS, it is important
to load the same software version as that previously installed.
• Follow the
Install from scratch instructions given in the booklet supplied with the AWS Application
Software CD-ROM to install the software. Before starting the procedure, take note of the require-
ments for
Material needed.
• After completing the installation, refer to the D/R Job card
Back-Up/Restore Procedure:
- Restore AWS parameters from the backup diskette. If this diskette is not available the procedures
given in the CFG Job Card
AWS Configuration must be carried out.
- Perform a system backup.
Notice:!
If the system backup is not done, no backup for IDC and DMR parameters will exist.

Chapter 10 Page no. 436 DR_108.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Job Card D/R 108 - Software reload
5-2. AWS Application Reload or Update
• To reload the Application software only, or if a new version of application software must be installed
on your system, use the AWS Application Software CD-ROM used for the current installation, or that
provided for the update, and follow the
Update from CD-ROM instructions given in the CD-ROM
booklet.
6. IDC SOFTWARE
6-1. IDC Software Reload from Backup
• If an IDC software reload is necessary (for example, if the IDC unit has been replaced, or if lost or
corrupted files make it impossible to run the application after rebooting) restore the IDC directory
data from the current system backup. Refer to the D/R Job card
Back-Up and Restore Procedure.
6-2. IDC Software Reload or Update from CD-ROM
• Software reload from CD-ROM may be necessary if for some reason the backup is not available, or if
a new version of the IDC software must be installed on your system. Use the IDC Application Soft-
ware CD-ROM used for the current installation or supplied for a software update, and follow the
instructions given in the CD-ROM booklet to install the software.
7. DMR PARAMETERS
7-1. DMR Software Reload
• If DMR parameters have been lost or corrupted, reload them from the current system backup. Refer
to the D/R Job card
Back-Up/Restore Procedure.
• If the backup is not available, it wil be necessary to recalibrate the DMR.

GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
rev.fm Page no. 437
Revision History
0 August 2, 2001 Onyx M3 release, based on FFDM 2269413-100 rev. 02.
Changes include: IST005, modified to include installation of
Siemens Hi-Bright monitor; IST017, modified to reflect new
installation procedure to avoid coolant spillage; CFG 009, now
specifies printer parameters for use with ScrapBook; CAL000,
new document listing CAL interactions, all CAL Job Cards mod-
ified to refer to CAL 000 for pre- and post-requisites; CAL017,
correction (replace instruction to open worklist by instruction to
close exam list if open); CAL020, new job card added for auto-
matic mAs non-linearity calibration; VF012 sec. 5-2, corrected
value for item 5 (was 3%); Renewal Parts extensively revised in
accordance with latest FRU specifications; D/R000, new docu-
ment specifying required calibrations and checks after FRU
changes, all D/R Job Cards modified to refer to D/R000 for post-
requisites; D/R001 deleted (replaced by D/R000); D/R019,
minor corrections and added forms for manual backup;
D/R 103, added cautions about coolant spillage; D/R104,
added illustration of IDC Dione II; D/R106, modified to more
fully describe the coolant kit and coolant disposal, added cau-
tions about coolant spillage; D/R108, minor corrections; Appen-
dix 1 deleted ( Physicist Checks now treated in separate
documents).
438

Page no. 438 rev.fm
GE Medical Systems Senographe 2000 D Acquisition System
Revision 0 Service Manual 2302228-100
Number Size Revision
2302228-100 A4 0

Senographe 2000 D Acquisition System
Service Manual
2302228±100 Revision 0
Senographe 2000 D Acquisition System
Service Manual 2302228±100 Revision 0
Technical
Publications
2302228±100
Revision 0
Senographe 2000 D Acquisition System
sm
Service Manual
do not duplicate
Copyright

2001 by General Electric Co.
Tags