12.06.2007
1
Evaluation of automated RNA extraction
methods during an AI outbreak
Markus Durban
Friedrich Loeffler Institut
12.06.2007
2
Outline
• Introduction
• Results of the evaluation of different extraction methods
and kits
• Summary
12.06.2007
3
Basics
• Real time PCR became more and more important in
diagnostics of viruses
• Ready-to-use kits for real time PCR are available
• Sensitivity of the assay is influenced by the yield and the
purity of the nucleic acid
• Nucleic acid extraction is now the most critical and labor-
intensive step
12.06.2007
5
Closed systems
•Positive
– Validated methods for use in clinical diagnostics
– Easy to handle
– Short training period
• Negative
– Kits and plastics mostly offered by one supplier
– Mostly long durations for extractions
– Only one extraction method can be used
12.06.2007
6
Open platforms
•Positive
– Free in choosingthe extraction method and the kit
– Optimization of the extractionprotocols possible
– The use of different methods and kits for different
problems is possible
• Negativ
– Mostly no validated methods
– Longer training period
– Programming is necessary
12.06.2007
7
Configuration of our extraction robot
Pipetting arm
Barcode-Reader
Gripper
Troughs for buffers
Cooled rack
Vacuum station
Waste and washing station
Lysed samples
Magnetic separator
Heater
12.06.2007
8
Silica membranes
• Nucleic acids are bound at silica membranes
• Removal of liquids by vacuum or centrifugation
12.06.2007
12
Macherey & Nagel Nucleospin Multi 96 Virus
• RNA extraction using Trizol as lysis reagent
- Samples are mixed with Trizol
- Chloroform is added and the samples are centrifuged
- The upper phase is used for further extractions
12.06.2007
18
Magnetic beads
• Nucleic acidis bound on coated magnetic particles
• Particles are washed in tips or cavities
Separation in
the tip
Separation on a
magnet
Separation in
the cavity
12.06.2007
19
• Using the BioRobot M48
• AIV H12N5-dilution curve 10
-2
to 10
-6
QIAGEN MagAttractVirus Mini M48 Kit
+ Good results already in the
first run
+ Validated for clinical
diagnostics
- Almost 5 h for 48 Samples
12.06.2007
20
• Using the Tecan freedom evo
• AIV H12N5-dilution curve 10
-1
to 10
-6
QIAGEN MagAttractVirus Mini M48 Kit
Eluate incubated for 10 min at 70°C
12.06.2007
21
• Using the Chemagen MSM 1
• AIV H12N5-dilution curve 10
-1
to 10
-6
Chemagic Viral DNA/RNA Kit
FAM HEX
12.06.2007
22
• Using the Tecan freedom evo
• AIV H12N5-dilution curve 10
-1
to 10
-6
Chemagic Viral DNA/RNA Kit
12.06.2007
23
• Using the Tecan freedom evo
• AIV H12N5-dilution curve 10
-1
to 10
-6
Macherey & Nagel NucleoMag 96 Virus Kit
12.06.2007
24
• Using the Tecan freedom evo
• AIV H12N5-dilution curve 10
-1
to 10
-6
AGOWA mag Maxi DNA Isolation Kit
12.06.2007
25
Summary for kits based on magnetic beads
38.19 38.39 37.81 38.47 37.08 37.65 10
-6
37.36 36.05 33.86 37.18 35.93 37.56 10
-5
34.55 33.54 30.56 33.38 33.56 33.27 10
-4
31.73 29.72 27.18 30.05 30.56 29.39 10
-3
28.38 26.67 23.74 26.81 26.97 25.55 10
-2
24.83 23.07 19.71 23.08 - 22.07 10
-1
AGOWA mag
Maxi DNA
Isolation Kit
NucleoMag 96
Virus Kit
Chemagic
Viral
DNA/RNA Kit
(on MSM 1)
Chemagic
Viral
DNA/RNA Kit
MagAttract
Virus Mini
M48 Kit
(on M 48)
MagAttract
Virus Mini
M48 Kit
12.06.2007
26
• Extracted until now:
- around 30.500 samples of dilution curves, animal
experiments and diagnostic samples
~14.300 AIV/NDV samples
~9.300 BTV-EDTA-blood samples
~1.300 serum samples
~2.000 samples with trizol as lysis buffer
~2.100 tissue samples
~2.000 different samples for kit evaluation
SummaryI
12.06.2007
27
• The automated RNA extractions from liquid samplesis
actually done with NucleoSpin Multi 96 Virus Kit
• QIAamp Virus BioRobot MDx Kit, MagAttract Virus Mini M48
Kit and Chemagic Viral DNA/RNA Kit can be used as
alternative
• The extraction from liquids is validated for AIV, NDV,FMD,
SVD, CSF, and BTV
• A combination of the used kit with trizol als lysis buffer is
possible and established for serum, blood and tissue
samples
• The extraction of virus RNA in tissue is almost validated for
CSF, AIV and NDV using the NucleoSpin 96 RNA Tissue Kit
• In case of an outbreak more than 1000 samples can be
extracted and analyzed per day
SummaryII
12.06.2007
28
Thank You
Martin Beer
Bernd Haas
Bernd Hoffmann
Ulrike Polenz
Christian Korthase
Anja Schulz
Federal Ministry of Food, Agriculture and Consumer Protection
…and all others for bringing me samples