2. Recording Instruments
Theseinstrumentsgiveacontinuousrecordofthegiven
electricalquantity
Theexamplesarevarioustypesofrecorders.Insuch
recordinginstruments,thereadingsarerecordedby
drawingthegraph.Thepointerofsuchinstrumentsis
providedwithamarker
3. Integrating Instruments
Integrating Instruments are those instruments which
totalize the events over a specified period of time. The
output of such instruments is the product of time and an
electrical quantity.
For example, a house energy meter , Unit of energy is
kwhr.
BASIC PRINCIPLE OF INDICATING
INSTRUMENTS
Three types of operating forces
i) Deflecting force
ii) Controlling force and
iii) Damping force
i) Deflecting Torque/Force
Thedeflectingtorque’svalueisdependentuponthe
electricalsignaltobemeasured;thistorque/forcehelpsin
rotatingtheinstrumentmovementfromitszeroposition.
Thesystemproducingthedeflectingtorqueiscalledthe
deflectingsystem.
Controlling System
Itisthesystemthatprovidesaforceequaland
oppositetothedeflectingforce.Controllingforcesare
appliedintwoways.
i)SpringControl(usedinmoderninstruments)
ii)GravityControl(notproperlyused)
i) Spring Control
A spring attached to the moving system produces a
controlling torque. The requirements for spring are
1. They should be non-magnetic.
2. They should be free from mechanical fatigue.
3. They should have a small resistance, where springs are
used to lead the current into moving system.
Generally, underdamped system is preferred for any
instrument.
Various methods used for producing damping torque
are,
i) Air friction damping
ii) Fluid friction damping and
iii) Eddy current damping
iii) Eddy Current Damping
The eddy current damping which is the most effective
way to provide damping. It is based on Faraday's law
and Lenz's Law.
Whenaconductormovesinamagneticfield,itcutsthe
magneticfieldandhenceemfisinduced.Thisinduced
emfopposesthecausesproducingit,thusopposing
themotionofthemovingsystem.
MOVING IRON INSTRUMENTS
There are two types
Moving iron attraction type instruments
Moving iron repulsion type instruments
Moving iron attraction type instruments
ItconsistoffixedcoilCandmovingironpieceD.
Thecoilisflatandhasanarrowslotlikeopening.
Themovingironisaflatdiscorasectoreccentrically
mountedonthespindle.
Thespindleissupportedbetweenthejewelbearings.
Thespindlecarriesapointerwhichmovesovera
graduatedscale.
Following consequences happens
Current in coil→ produce magnetic field → attract disc →
pointer moves → so measure current.
Tc provide by spring
Td provide as air friction damping
Moving iron repulsion type
instruments
Two vanes inside the coil, one is fixed and other is
movable
When the current flows in the coil, both the vanes are
magnetised with like polarities induced on the same
side.
Both vanes gets magnetized and get repulsive each
other. So pointer moves
It has two types
Radial vane type
Coaxial vane type
Torque Equation of Moving Iron
Instruments
The energy stored in the coil in the form of magnetic
field = (1/2)LI
2
.
As soon as the current changes to (I+dI), deflection in
the pointer becomes dƟ resulting into change in
inductance of coil from L to (L+dL).
Let this deflection in pointer is due to deflection torque
T
d.
Mechanical work done = T
d. dƟ ………………..(1)
Energy stored in Coil = (1/2)(L+dL)(I+dI)
2
Change in stored energy of coil = Final Stored Energy –
Initial Stored Energy
= (1/2)(L+dL)(I+dI)
2
–(1/2)LI
2
= (1/2)[ (L+dL)(I+dI)
2
–I
2
L]
= (1/2)[ (L+dL)(I
2
+2IdI+(dI)
2
–I
2
L]
= (1/2)[ LI
2
+2LIdI+L(dI)
2
+ dL.I
2
+2IdI.dL+dL.(dI)
2
–I
2
L]
Neglecting second order and higher terms of differential
quantities
i.e. L(dI)
2
, 2IdI.dL and dL . (dI)
2
= (1/2)[ 2LIdI+dL.I
2
]
= LIdI +(1/2)dL.I
2
……………………(2)
We can write as, e = d(LI) / dt
= IdL/dt + LdI/dt
But electrical energy supplied by the source = eIdt
= (IdL + LdI)
. I
= I
2
dL + LIdI
Accordingtolawofconservationofenergy,this
electricalenergysuppliedbythesourceisconverted
intostoredenergyinthecoilandmechanicalworkdone
fordeflectionofneedleofMovingIronInstruments.
I
2
dL + LIdI = Change in stored energy + Work done
⇒I
2
dL + LIdI = LIdI +(1/2)dL . I
2
+ T
d. dƟ
⇒T
d. dƟ = (1/2)dL.I
2
⇒T
d= (1/2)I
2
(dL/dƟ)
Thus deflecting torque in Moving iron Instruments is
given as
T
d= (1/2)I
2
(dL/dƟ)
In moving iron instruments, the controlling torque is
provided by spring. Controlling torque due to spring is
given as
T
c= KƟ in N-m
In equilibrium state,
Deflecting Torque = Controlling Torque
⇒T
d= T
c
⇒(1/2)I
2
(dL/dƟ) = KƟ
⇒Ɵ = (1/2)(I
2
/K)(dL/dƟ)
Ɵ α I
2
Thedeflectiontorqueisunidirectionalwhatevermay
bethepolarityofthecurrent.
Hence,theMIinstrumentscanbeusedforbothAC
andDC.
Errors in Moving Iron
Instruments
1. Errors with both A.C and D.C work:
(a) Hysteresis error.
(b) Stray magnetic field error.
(c) Temperature error.
(d) Friction error.
2. Errors with A.C work only:
(e) Frequency error.
(f) Error due to reactance of the instrument coil.
(g) Error due to eddy current.
(h) Error due to waveform.
Advantages of Moving iron
Instruments
UsedforthemeasurementofACandDCquantities.
Thesetypesofinstrumentshavehighvalueoftorqueto
weightratio.Duetothiserrorbecauseoffrictionisquite
low.
Itisverycheapduetosimpleconstruction.
Thereisnomovingpartintheinstrumentwhichcarries
current.
Theseinstrumentscanbedesignedtoprovideprecision
andindustrialgradeaccuracy.Awelldesignedmoving
ironinstrumentshaveaerroroflessthan2%orlessfor
DC.ForAC,theaccuracyoftheinstrumentmaybeof
theorderof0.2to0.3%at50Hz.
Notdamagedevenunderseveroverloadconditions.
Disadvantages of Moving Iron
Instruments
These instruments suffer from error due to hysteresis,
frequency change and stray losses.
Thescale of moving iron instrumentis not uniform.
Accurate readings are not possible at lower range.
If it is used at 50 Hz, calibration must also be done at
the same frequency i.e. 50 Hz.
Moving Iron Instruments are suitable for low frequency
application. Moving iron instruments are not suitable for
frequency above 125 Hz.
The reading of the instrument is affected by
temperature variation.
PERMANENT MAGNET MOVING
COIL (PMMC) INSTRUMENTS
The permanent magnet moving coil instrument is the
most accurate type for d.c. measurements.
Basic Principle
Theactionoftheseinstrumentsisbasedonthemotoring
principle.
Whenacurrentcarryingcoilisplacedinthemagneticfield
producedbypermanentmagnet,thecoilexperiencesa
forcedandmoves.
Asthecoilismovingandthemagnetispermanent,the
instrumentiscalledpermanentmagnetmovingcoil
instrument.
ThebasicprincipleiscalledD'Arsonvalprinciple.
Construction of PMMC
Instruments
Themovingcoiliseitherrectangularorcircularin
shape.
Thecontrollingtorqueisprovidedbythemethodof
springcontrolwiththehelpoftwophosphorbronzehair
springs.
Thedampingtorqueisprovidedbythemovementofthe
aluminiumformerinthemagneticfieldproducedbythe
permanentmagnet.
The scale markings of the basic d.c. PMMC instruments
are usually linearly spaced
Torque Equation for PMMC
The deflecting torque is given by,
T
d= NBAI
T
d= GI
Where, G = NBA = constant
The controlling torque is provided by the springs
T
c= KØ
For the final steady state position,
T
d= T
c
ThereforeGI = KØ
Ø = (G/K)I or I = (K/G) Ø
Ø α I
Errors in PMMC Instrument
Errors due to permanent magnets
Error may appear in PMMC Instrument due to the
aging of the spring.
Change in the resistance of the moving coil with
the temperature
Advantages of Permanent Magnet
Moving Coil Instruments
The scale is uniformly divided
Power consumption is also very low
A high torque to weight ratio. So operating current
is small.
The sensitivity is high
It has high accuracy
Instrument is free from hysteresis error
Extension of instrument range is possible
Not affected by external magnetic field called
stray magnetic fields.
Disadvantages of Permanent
Magnet Moving Coil Instruments
These instruments cannot measure AC quantities.
The cost of these instruments is high
Ageing of permanent magnet and the control springs
introduces the errors.
The friction due to jewel-pivot suspension.
ELECTRODYNAMOMETER
WATTMETER
Fixed coil
Current Coil (C.C), which is connected in series with the
load and it carries the current through the load.
Moving coil
Across the load and it carries the current proportional to
the voltage across the load.
Pressure Coil (or) P.C.
Fixed Coil
Carry the load current of the circuit.
Generally they are divided into two halves but connected
in series.
The fixed coils are wound with heavy wire with less
number of turns
The maximum current range of wattmeter is 20A
Moving Coil
Themovingcoilisgenerallyattachedtothespindlewhich
isconnectedtothepointer.
Itismadeofthinwirebuthasmorenumberofturns
Aseriesresistorisusedinthevoltagecircuitinorderto
limitthecurrenttoasmallvalueintheorderof100mA.
Thevoltageratingofthewattmeterislimitedto600V.
ControlTorque-Controltorqueisprovidedbysprings
Damping-Air friction damping is used
Errors in electrodynamometer type
wattmeter
Error due to pressure coil inductance
Error due to pressure coil capacitance.
Error due to the effect of manual inductance.
Error due to wrong connection of current coil and
pressure coil.
Eddy current error.
Stray magnetic field error.
Error caused by vibration of moving system.
Temperature error.
INDUCTION TYPE ENERGY
METER
Energymetersisanintegratinginstrumentwhich
measuresquantityofelectricity.
Thesemetersrecordtheenergyinkilo-watt-hours
(kWh).
Energymeterisaninstrumentusedtomeasureenergy
whichisthetotalpowerconsumedoveraspecific
intervaloftime.
UnitofenergyiskWhorJoules.
Energy = Power x Time
Basic Principle
The operation of the induction type energy meter is based
on the passage of alternating current through two coils
Magnetic field which interacts with a aluminium disc
supported near the coils and make the disc rotates.
The current coil carries the line current and develops
magnetic field. This magnetic field is in phase with the
line current.
The pressure coil is highly inductive, hence the current
through it lags behind the supply voltage by 90̊.
Due to this, a rotating field develops which interacts with
the disc to rotate.
Construction Details
i) Driving System ii) Moving System
iii) Braking System iv) Registering System
iii) Braking System
The braking system consists of a permanent magnet
positioned near the edge of the aluminium disc.
The aluminium disc moves in the field of this magnet and
this provides a braking torque.
iv) Registering System / Counting System
Thefunctionofaregisteringorcountingmechanismisto
recordcontinuouslyanumberwhichisproportionaltothe
revolutionsmadebythemovingsystem.
Operation
The C.C carries the load current. It produces the magnetic
fields in phase with the line current.
The P.C carries current proportional to the supply voltage.
The magnetic field due to pressure coil lags approximately
90̊ behind the supply voltage
Themagneticfieldduetocurrentcoildevelopseddy
currentinthealuminiumdiscwhichreactwithmagnetic
fieldduetothepressurecoil.
Thusatorqueisdevelopedinthediscthenitrotates.
Thebrakingmagnetproducesmechanismsothatthe
electricalenergyconsumedinthecircuitisdirectlygivenin
KWh(KiloWatthour)
Advantages of induction type energy meters
The construction is simple and strong.
It is cheap in cost.
It has high torque to weight ratio, so frictional errors are
less and we can get accurate reading.
It has more accuracy.
It requires less maintenance.
Disadvantages of induction type energy meters
The main disadvantage is that it can be used only for a.c.
circuits.
The creeping can cause error.
Lack of symmetry in magnetic circuit may cause errors.
CLASSIFICATION OF
TRANSDUCERS
Classification based on transduction principle
used
Classifiedasresistive,inductive,capacitivedepending
uponhowtheyconvertinputquantityresistance,
inductance,andcapacitancerespectively.
Primary and Secondary transducers
Primary
Transducers senses the input physical quantity directly and
convert directly into electrical quantity output.
Secondary
Input signal is sensed by other some detector or sensor and then
its output is given to transducer in other form then the transducer
converts the secondary signalinto electrical.
Active and Passive transducers
Active
Converts physical quantity into electrical quantity directly
So it is called self generating type transducers.
Passive
Inthistransducertheelectricalparametersresistance,inductance,
andcapacitancechangeswithchangeininputsignalarecalled
passivetransducers.
AnalogandDigitaltransducers
Analog transducers:
Output of analog transducer is continuous function of time.
Digital transducers:
Output of this type transducer is pulses or discrete form
Transducers and Inverse transducers
Transducers:
Transducer is a device, which converts input
physical quantity into output electrical quantity.
Inverse transducers:
Inverse transducer is a device, which converts
Input electrical quantity and output physical
quantity.
Capacitive transducer
The principle based on capacitance of a parallel
plate capacitor
C=εA/d= εoεrA/d
The change capacitance caused by
Change in overlapping area
Change in distance “d” between the plates
Change in dielectric constant
These are changes due to changing the force,
displacement and pressure
The change in capacitance causes change in
dielectric constant. Also measure the liquid level
Capacitance transducer –By
variation of overlapping area of
plates
C αA, capacitance changes linearly with change
in area of plates
The area changes linearly with the displacement
and also the capacitance
Char are linear. But initially non linearity due to
edge effects
Parallel plate capacitor, the capacitance is
C= εA/d= (εXW/d)* F
X= length of overlapping portion of plates in m
W= width of overlapping portion of plates in m
Sensitivity as
Cylindrical capacitor whose over lapping area is
varied by varying length of over lapping portion of
cylinder
Cylindrical transducer as shown
Capacitance as
S=const. Then relationship between capacitance
and displacement is linear
Fig shows two plate capacitor
Angular displacement to measured is applied
movable plate
The angular displacement changes the effective
area between area of plate and thus changes the
capacitance
Capacitive transducers-By variation
of distance between the plates
C α(1/d), used to measure linear displacement
Here one plate fixed and other plate moving
Moving plates moving away from or towards the
fixed plate as per displacement under
measurement, so capacitance decreases or
increases
Capacitance measured by AC bridges circuit, so
displacement of moving plate is determined
Curve is non linear. Sensitivity is high for initial
portion of curve
Capacitive transducer-
Differential arrangement
To achieve linear char, differential arrangement as
shown
It have three plates,
P1, P2→ Fixed plate
M → Movable plate
So two capacitor with differential o/p
M-midway between P1 & P2
AC voltage E applied between P1 & P2
C1=C2, E1=E2=E/2 (i.e) exactly midway between
2 plates
Advantages of capacitive
transducers
Have very high i/p impedance, so min loading
effect
Have good freq response. This response as high
as 50KHz and very useful for dynamic studies
Not affect by stray mag fields
High sensitivity, higher resolution
Force requirement of capacitive transducer is
very small and require small power to operate
them
Disadvantages of capacitive
transducers
Very high o/p impedance. So complicated
measuring circuit
Stray capacitance including that cables etc in
parallel with o/p impedance of transducer also
causes error and introduces non linearity
The cable connecting the transducer to the
measuring point is also a source of error. The
cable may br source of loading resulting in loss of
sensitivity. Also loading makes the low freq
response error
The instrumentation circuitry used with these
transducer is very complex
Application of capacitive
transducers
Use to measure both linear and angular
displacement
Use to measure force and pressure. Here first
convert displacement causes change of
capacitance
Able to measure pressure directly in all those
cases in which permittivity of a medium changes
with pressure, such as in case of benzene
permittivity vary by 0.5%, in pressure range of 1 to
1000 times the atm pressure
Use to measure humidity. Since the permittivity of
gases varies with variation in humidity. Though the
variation in capacitance due to variation in
humidity is quite small but is detectable
Commonly used in conjunction with mechanical
Capacitor microphone
Most commonly used as studio
Thin electrically conductive diaphragm is
suspended over back plate forming a flexible
capacitor
One plate is diaphragm, it mounted not touching.
Other plate is back plate
Battery connected to both plates, which produces
electrical potential or change between them
Sound wave exit the diaphragm. Distant between
plate change the capacitance, due to change the
voltage
This is excellent choice for mixing vocals, acoustic
guitar, piane, sound effect.
Inductive transducer
Either self generating or passive type
Self generating type utilize basic generator
principle
An inductive transducer is a device that convert
physical motion into a change in inductance
The principle used as
No of turns
Geometric configuration
Permeability of magnetic material or magnetic
circuit
Transducer based on principle of change
in self inductance with no of turns
o/p changes w.r.to no of turns
Measure displacement of linear and angular
movement as shown
Here no of turns changes, inductance changes,
then o/p changes
Transducer working on principle of change in
self inductance with change in permeability
Inductive transducer on principle of variation of
permeability causing change in self inductance as
shown
Iron core surrounded by winding. Here
permeability changes, then L-changes
Iron move out of winding, permeability↓, then L↓
in coil. Source to measure displacement
Variable reluctance inductance
transducer
Use to measure linear displacement
Hence length of magnetic path varies with the
displacement and reluctance of magnetic circuit
changes causing in self inductance of the coil
Linear Variable Differential
Transformer (LVDT)
Construction
It is widely used inductive transducer to translate
linear motion into electrical signal
LVDT is differential transducer consist of one
primary (P) and two secondary (S1 & S2). Both
wound on non magnetic material
S1 & S2 have equal no of turns and identical placed
on either side of pri winding
Displacement to be measure is applied to arm
attached to the soft iron core
In order to overcome the problem of eddy current
losses in the core, nickel-iron alloy is used as core
material and is slotted longitudinally
Construction as shown
Working
Primary winding voltage range as 5-25V and freq
as 50Hz-20kHz
Primary winding exited AC current source, so
produces AC mag field which induces AC voltages
Es1-o/p voltage of S1
Es2-o/p voltage of S2
Both S1 & S2 are in series opposing.
Differential o/p voltage = Eo=Es1-Es2
Case (i): When the core is at its normal (NULL)
position
Core is normal null position
Both sec. have equal flux linkages (i.e) Es1=Es2
So Eo=Es1-Es2=0
Case (ii): The core is moved to the left of the NULL
position
Core moved left of NULL position (i.e) at A
Flux linkages more in S1 and less in S2
So Es1>Es2, Eo= Es1-Es2
Eo=+ve which is in phase wih o/p
Case (iii): The core is moved to the right of the null
position
Core move right of NULL position (i.e) at B
Flux linkages less in S1 and more in S2
So Es2>Es1, Eo=Es2-Es1
Eo α(movement of core) (i.e) linear motion
Eo↓ or Eo↑ depends on direction of motion
o/p of one secondary increases and other
secondary decreases. So which use to measure
displacement
Variation of Eo w.r.to displacement of core as
shown. For small displacement only linear char.
Small changes
At ‘O’ position of core, Eo not equal to zero due to
have some residual magnetism (i.e) 1% of Emax
Residual voltage due to mag unbalance or
electrical unbalance
Due to harmonics & saturation of iron core
contribute residual voltage. Also due to stray mag
field.
Advantages of LVDT
Upto 5mm, it have linear displacement
High sensitivity, range as 10mV/mm –40mV/mm
Give high o/p. No need of amplification
Use freq upto 20kHz, more reliable
Have low hysterisis, hence repeatability is
excellent under all condition
Rugged construction, vibration without any
adverse effect
Power consume < 1W, small weight
Stable and easy maintanance
Disadvantages of LVDT
Require large displacement of o/p
Sensitive with stray mag field
Performance affected by vibration
Receiving instrument select to operate AC signal
or a demodulator network must be used if a DC
o/p is required
The dynamic response is limited mechanically by
mass of the core and electrically by freq of applied
voltage. The freq of the carrier should be at least
10 times the highest freq component to be
measured
Performance is affected with temperature
Application of LVDT
LVDT use to measure
Displacement
Force
Weight
Pressure
Position
STRAIN GAUGES
Piezo resistive Effect
Ifametalconductorisstretchedorcompressed,its
resistancechangesonaccountofthefactthatbothlength
anddiameterofconductorchange.
Alsothereisachangeinthevalueofresistivityofthe
conductorwhenitisstrainedandthispropertyiscalled
piezoresistiveeffect.
Uses of strain gauges
Used for measurement of strain and associated stress in
experimental stress analysis.
Many detectors and transducers notably the load cells,
torque meters, pressure gauges, temperature sensors,
accelerometers and flow meters, employ strain gauges as
secondary transducers.
Classification of Strain gauges
Wire strain gauge
Foil strain gauge
Thin film strain gauge
Semiconductor strain gauge
1. Wire strain gauges
It is small size, min leakage, employ high temp
It has two types
Unbounded resistance wire strain gauge
Bonded resistance wire strain gauge
Unbounded resistance wire strain gauge
It consist of wire stretched between 2-point of insulating
medium (i.e) air
Dia=25µm
Wire have high tension. So that no sag & no vibration
Load applied, resistance changes, unbalances the bridges.
So V0 changes, V0 αstrain, displacement ≈ 50µm
Bonded resistance wire strain gauge
The schematic as shown
Dia of wire≈25µm
Loop as back and forth
The grid of fine wire is cemented on a carriers which
may be a thin sheet of paper, backelite or teflon
Wire converted on the top with thin material, so not
damaged mechanically
Spreading of wire permits uniform distribution of
stress
2. Foil strain gauge
It is extension of resistance wire strain gauge
Metal & alloys use for foil. Nichrome, constantant
use for wire
It have high dissipation capacity. So use high temp
gauge. It have better bonding due to larger area
Advantage as fabricate to larger scale, any shape.
Etched foil gauge construction consist of first
bonding layer of strain sensitive material to a thin
sheet of paper of paper or bakelite
Etched foil strain gauge made thinner than
comparable wire units. More flexible. So it placed
remote & restricted places and curved placed.
3. Thin film strain gauges
This can be produced by depositing a thin layer of
metal alloy an elastic metal specimen by means of
vacuum deposition
This technique, relatively new and extensively
used to produces a strain gauge that is
molecularly bondes to the specimen under test
and so the drawback of epoxy adhesive bond are
eliminated
Thin technique is most widely used for transducer
application such as in disphragm type pressure
gauges.
4. Semiconductor strain gauge
It have high sensitivity have gauge factor
It required high value of gauge factor. It is 50 time
higher then wire strain
Resistance change w.r.to applied strain
Semiconductor used as germanium & silicon
The schematic as shown
Consist of strain material and leads placed in
protective box. Thickness of wafer 0.05mm used
Bonded on suitable insulating subsrate, such as
teflon
For making contact use gold leads
For soldering leads use cadmium material
It have both +ve and –ve gauge factor for p and n-
type silicon respectevely
Advantages of semiconductor
strain gauges
Measure very small strain as well as 0.01 micron.
Also high gauge factor between -100 and +150
Manufacturing very small size range of 0.7-7mm
use to measure high localized strain
Chemically inert and low sensitivity
Have excellent hysteresis char.
Disadvantages
Sensitive to change w.r.to temp, more expensive
Poor linearity char
Hall effect
Ic flows downwards in semiconductor pellet which
placed in magnetic field perpendicular to pellet
surface, an VHcreated in pellet in direction
perpendicular in both Ic and magnetic field. This
process called as hall effect.
Electromagnetic force act on charged particle
according to F.L.H.R, the charged particle are
biasing to left side of semiconductor pellet.
The magnitude of emf VH, which is called the hall
voltage
VH=1/d(BIcRH)
RH= hall constant
B=flux density
D=thickness of semiconductor
Semiconductor device which are made use in
detecting magnetic field called “hall element or hall
Conceptual Diagram of Hall
Effect Transducer
(15)(15)
Advantages
Can operate high speed than mechanical points
Operating frequency as 100KHz
Measure wide range of magnetic fields
Stable, reliable, long lasting
High resolution and small size
Disadvantages
Very low o/p drive capability
Difficult to operate in strong external magnetic field
Less accurate
Application
BLDC motor, Proximity detector, Speed sensor
(motor control)
Vending machine, Shaft position sensor, valve
position detector
Piezoelectric transducer
Piezoelectric material is one which an electric
potential appears across certain surfaces of a
crystal surfaces of a crystal if the dimension of the
crystal are changed by the application of
mechanical force. This potential produced by
displacement of changes.
The effect is reversible also, varying potential
applied to proper axis of crystal, it will changes the
dimension of crystal thereby deform it. This
phenomenon is known as piezoelectric effect
Piezo is greek word meaning force or pressure.
Element exhibiting piezoelectric quantity are called
electro resistive elements
Material for piezoelectric transducer
Common used material as rochelle salt,
ammonium, dihydrogen phosphate, quartz and
ceramics made with barium titanate, dipotassium ,
lithum sulphate
The piezoelectric effect can be made to respond to
mechanical deformation of material in many
different modes. These modes are
Thickness expansion
Transverse expansion
Thickness shear
Face shear
Mechanical deformation generates a charge and
this charge appears as voltage across electrodes
A tensile force produce a voltage of one polarity while a
compressive force produces a voltage of opposite
polarity
A crystal between a solid base and the force summing
member. An extremely applied force, entering the
transducer through its pressure, applies pressure to top
of crystal. This produces a voltage across the crystal
proportional to the magnitude of applied pressure.
Magnitude and polarity of induced surface charges
proportional to mag and direction of force
Q=F*d
d= crystal charge sensitivity in coulombs per newton and is
constant for a given crystal cut
F= force in newton
The Voltage sensitivity g = E
0/tP or g=ε/P ...........
here ε = E
0/t
Modes of operation of
piezoelectric crystal
The different modes as
Thickness shear
Face shear
Thickness expansion
Transverse expansion
Advantages of piezoelectric transducer
Small size, light weight, rugged construction
It has self generating type and no need of external
power
o/p is quite large
Very good high freq response. Range as 1Hz to
20KHz. Natural frequency as 50KHz
Disadvantages of piezoelectric transducer
Eo affect with temp variation of crystal
Use for dynamic measurement only
Application of piezoelectric transducer
Use to measure of force, pressure, temp
Employ high freq accelerometer