Instrumentation of DTA Maheen Zahra (1300143) BS IV (8 th )
Contents Differential thermal analysis Apparatus P rocedure Graphical representation Advantage of DTA
Differential thermal analysis In Differential Thermal Analysis, the temperature difference that develops between a sample and an in reference material is measured, when both are subjected to identical heat treatments.
INSTRUMENTATION
APPARATUS 1.Sample holder comprising thermocouples, sample containers and a ceramic or metallic block. 2.Furnace. 3.Temperature programmer 4.Recording system
Procedure Heart of the analysis – heating block Identical pair of cavities for the sample, ref. material Whole unit is set in an oven- control pressure Thermocouple is place directly in contact with the sample and another in contact with the reference
Temperature of the block is raised, the temperature of the sample & reference follow Zero temp. difference – no physical or chemical change If any reaction – difference in ∆T
Graphical Representation
NOTE.. Sharp Endothermic–changes in crystallinity or fusion Broad endotherms-dehydration reaction Physical changes usually result in endothermic curves Chemical reactions are exothermic
Advantage of DTA Instruments can be used at very high temperatures • instruments are highly sensitive •characteristic transition or reaction temperatures can be accurately determined.