ENZYME-LINKED IMMUNOSORBENT ASSAY (ELISA) D r Dinakaran s 1 st year post graduate Biochemistry ESIC MC PGIMSR
Objectives Immunoassay definition and it’s types Principle/ definition Types of ELISA Applications Primary antibodies Detection strategies
IMMUNOASSAY Highly selective bioanalytical method measures the presence or concentration of analytes ranging from micro to macromolecules in a solution through the use of an antibody (usually) or an antigen (sometimes).
Types of Immunoassay Depending upon the label /binders Radioimmunoassay (RIA) Enzyme-Linked ImmunoSorbent Assay ( ELISA)
ELISA It is an immunological assay. One of the reaction components is non specifically adsorbed or covalently bound to the surface of a solid phase. ELISAs are typically performed in 96-well or 384-well polystyrene plates.
Principle: Detects interaction between specific antigen and antibody, by enzyme labelled antibodies. Substrate of enzyme is added which gets coupled with coloured reagent and change colourless reaction to coloured reaction. Intensity of colour is directly proportional to concentration of Antigen/Antibody in given sample.
Samples used in ELISA Serum Plasma Saliva Urine Tissue lysates
1. Coating/capture - Immobilization of antigen / antibody 2. Plate blocking - Addition of molecule to cover all unsaturated sites 3. Probing/detection- Incubation 4. Signal measurement- Detection of the signal generated. General steps in ELISA
ELISA plates The first step is to optimize the plate-coating conditions for the antigen or capture antibody. Clear polystyrene flat bottom plates - colorimetric signals Black or white opaque plates - fluorescent /chemiluminescent signals.
ELISA plates Thermo Scientific ELISA Plates - Size of our choice. Plate coating is by passive adsorption Hydrophobic interactions. Most common method for coating plates - Adding a 2–10 μg /ml solution of protein dissolved in an phosphate-buffered saline (pH 7.4) or carbonate-bicarbonate buffer (pH 9.4).
ELISA plates Incubation - several hours to overnight at 4–37° C. Coated plates can be used immediately or dried and stored at 4° C for later use. With the exception of competitive ELISAs, the plates are coated with less capture protein than can actually be bound during the assay.
ELISA plates Proteins are best coated on plates at a concentration lower than the maximum binding capacity - To prevent nonspecific binding in later steps by a phenomenon called "hooking". When hooking nonspecifically traps detection of primary and secondary antibodies high background signal results lowering the signal to noise ratio and sensitivity .
Different types of microplates for ELISA ELISA plate Coating Applications Modified polymer surfaces Increase hydrophobicity or hydrophilicity Enhance passive binding of biomolecules Antibody-binding plates Protein A, G, L, or A/G properly orient antigen binding capability Biotin-binding plates Streptavidin or neutravidin Binds small biotinylated peptides and other small molecules difficult to bind by passive adsorption Fusion-tag binding plates Glutathione (GST tag binding) or nickel or copper coated (His tag binding) Study of genetically engineered fusion proteins or protein-protein interactions
Types of ELISA Based on the basis of binding structure. 1. Direct ELISA (antigen coated plate, screening antigen) 2. Indirect ELISA (antigen coated plate, screening antibody ) 3. Sandwich ELISA (antibody coated plate, screening antigen) 4. Competitive ELISA (screening antigen )
Direct ELISA Sample antigen is adsorbed onto the walls of microtiter plate. Antibody linked enzyme is added. Antigen present in sample binds to antibody- enzyme complex. Substrate is added for colour change The colour intensity is proportional to the antigen concentration
Advantages: Fast and simple Eliminates possibility of non- specific binding to secondary antibody.
Disadvantages: Immunoreactivity of the primary antibody might be adversely affected Time-consuming Expensive – No flexibility Minimal signal amplification.
Application A direct ELISA test was developed to detect circulating antibodies specific to poliovirus types 1, 2 and 3.
Indirect ELISA Antibody can be detected or quantitatively determined. Eg : Detection of HIV antibody A person ’ s serum - Applied to a multiwell (microtiter) plate. Antibodies in serum Bind to antigen . Wash Secondary antibody conjugated with HRP is added Wash Plate will contain enzyme in proportion to the amount of secondary antibody bound to the plate
Enzyme reaction A substrate containing H 2 O 2 and diaminobenzidine for the enzyme HRP is applied and catalysis by the enzyme leads to change in the colour . H 2 O 2 HRP H 2 O + (O) Diaminobenzidine (DAB) + (O) Oxidised DAB ( colourless ) (Brown colour ) The colour is directly proportional to the antibody concentrations
Advantages: Affordable. Versatile. Maximum immunoreactivity of the primary antibody is retained. Sensitivity is increased Signal will be amplified. Different detection methods can be used with the same primary antibody (colorimetric, chemiluminescent, etc.).
Disadvantages: Extra step. Time consuming. Cross- reactivity.
Sandwich ELISA Detect the presence/measure the concentration of the target antigen in samples. Detect the antigens between the two layers of antibodies 1) Capture antibody (polyclonal) 2) Detection antibody (monoclonal ) Antigens to be measured must also have at least two non-overlapping epitopes capable of binding to two antibodies.
Sandwich ELISA Advantages : High specificity because antigen/analyte is specially captured and detected Flexible and sensitive. Suitable for complex samples as antigen does not require any purification.
Disadvantages: Possibility of error due to many steps Necessary use of “matched pair” (divalent /multivalent antigen) and secondary antibody Antibody optimization is difficult if standard kit is not used. Time taking Expensive
Competitive ELISA Used to measure the concentration of an antigen in a sample. First incubate the antibody with a sample containing antigen. Antigen- antibody mixture is added to microtitre well which is coated with antigen. The more the antigen is present in the sample, more the primary antibody will bind to sample antigen. Therefore , there will be smaller amount of primary antibody available to bind to antigen coated in well and wash have to give. Enzyme conjugated secondary antibody specific for isotype of the primary antibody is added and Substrate is added The higher the concentration of antigen in the sample the lower is the absorbance.
Competitive ELISA
Advantages: Less sample purification Commonly used for small molecules. Highly sensitive even the sample is in small amounts.
Disadvantages: Less specific Cannot be used in diluent samples
Applications of ELISA Detect and Measure the Presence of Antibodies in the Blood Detect and Estimate the Levels of Tumor Markers Detect and Estimate Hormone Levels Tracking Disease Outbreaks Detecting Past Exposures Screening Donated Blood for Possible Viral Contaminants Detecting Drug Abuse
Primary antibodies for ELISA Either monoclonal or polyclonal antibodies can be used as the capture and detection antibodies. In addition to the use of traditional monoclonal antibodies, recombinant monoclonal antibodies may also be utilized for ELISA. Compared to traditional monoclonal antibodies derived from hybridomas, recombinant antibodies are not susceptible to cell-line drift or lot-to-lot variation, thus allowing for peak antigen specificity.
Primary antibodies for ELISA For sandwich ELISA, the capture and detection antibodies must recognize two different non-overlapping epitomes. When the antigen binds to the capture antibody, the epitome recognized by the detection antibody must not be obscured or altered. Capture and detection antibodies that do not interfere with one another and can bind simultaneously are called "matched pairs" and are suitable for developing a sandwich ELISA.
Ready-to-use ELISA kits available detect hundreds of specific cytokines, growth factors, neurobiology analytes phosphorylated proteins.
Antibody pair kits Uncoated ELISA kits Coated ELISA kits* Instant ELISA kits Need to coat plate Yes, an overnight coating process is required Yes, an overnight coating process is required No No Incubation time** 24 h 24 h 2.5–4 h 3 h Hands-on time 1 hr 30 mins 1 hr 30 mins 1 hr 40 mins Readout HRP-TMB (colorimetric) HRP-TMB (colorimetric) HRP-TMB (colorimetric) HRP-TMB (colorimetric) Instrumentation needed Microplate reader, absorbance Microplate reader, absorbance Microplate reader, absorbance Microplate reader, absorbance Instrumentation read time 2 min 2 min 2 min 2 min ELISA Kits
Detection strategies for ELISA Chromogenic (colorimetric) Fluorescence Chemiluminescence Sensitivity Equipment required Standard absorbance plate reader Fluorometer Luminometer plate reader Enzyme HRP or AP Fluorescent tag or HRP (with chemifluorescent substrates) HRP or AP Advantages Direct visualization, high reproducibility between plates High reproducibility between plates, wide dynamic range Most sensitive detection strategy, wide dynamic range Considerations Requires black microplates Requires opaque or black microplates
Detection strategies for ELISA Chemiluminescence is a chemical reaction that generates energy released in the form of light. Most chemiluminescent substrates are Horseradish PerOxidase (HRP)- dependent, although some and Alkaline Phosphatase (AP) equivalents are available. The most common approach is to use luminol in the presence of HRP and a peroxide buffer. Light emission occurs only during the enzyme-substrate reaction, therefore when the substrate becomes exhausted, the signal ceases.
Detection strategies for ELISA Chemiluminescent detection more sensitive One drawback of using chemiluminescent substrates the signal intensity can vary with other substrates. For assays requiring many plates to be read, this can present a problem if the signal begins to decay before plates are read. Make sure the assay has been optimized with the substrate in order to avoid misinterpreting signal-fade in a sample as low antigen abundance. Chemiluminescent substrates for HRP include Thermo Scientific Super Signal ELISA Pico and ELISA Femto substrates.
Detection strategies for ELISA Fluorescent ELISA substrates are not as common require a fluorometer that produces the correct excitation beam signal emission to be generated from the fluorescent tag. Chemifluorescent detection is also enzyme-based, but the generated product is fluorescent rather than colorimetric. Examples of chemifluorescent substrates for HRP are Thermo Scientific QuantaRed and QuantaBlu substrates.
CEA CarcinoEmbryonic Antigen(CEA) is normally found in embryonic endodermal epithelium. Increased serum CEA levels have been detected in persons with primary colorectal cancer and other malignancies. Elevated serum CEA levels have also been detected in patients with nonmalignant disease, especially patients who are older or who are smokers. So, CEA levels are not useful in screening the general population for undetected cancers, But CEA is a useful tool for monitoring and managing cancer therapy.
Principles of the Procedure The Atellica IM CEA assay is a 2-site sandwich immunoassay using direct chemiluminometric technology. The first antibody, Lite Reagent rabbit polyclonal anti-CEA antibody labeled with acridinium ester. The second antibody, Solid Phase mouse monoclonal anti-CEA antibody covalently coupled to paramagnetic particles. A direct relationship exists between the amount of CEA present in sample and the amount of relative light units (RLUs) detected by the system.
Assay Procedure The system automatically performs the following steps: Dispenses 50 µL of sample into a cuvette. Dispenses 50 µL of Lite Reagent and 250 µL of Solid Phase, then incubates for 12 minutes at 37°C. Washes the cuvette with special reagent water. Dispenses 300 μL each of Atellica IM Acid and Atellica IM Base to initiate the chemiluminescent reaction. Reports results.
Conclusion Based on sensitivity Sandwich ELISA > Indirect ELISA > Direct ELISA ELISA technologies continue to grow play a major role in clinical research development of more diagnostic and screening tests.
References Teitz fundamentals of clinical chemistry 6 th edition Textbook of biochemistry 4 th edition – Dr Rafi MD Textbook of biochemistry for medical students 10 th edition –DM Vasudevan, Sreekumari S, Kannan Vaidyanathan . https://www.thermofisher.com/overview-elisa.html dated 27/02/24 . Murthy N, Nair KM, Bhaskaram P. A direct ELISA technique to detect antibodies against polioviruses. Indian J Biochem Biophys . 1995;32(5):249-253.