Haemoglobinometry ⚫ BY MAHESH KUMAR ⚫ ROLL NO. :: 06 ⚫ BSC MLT 1 ST YEAR [DR. RPGMC TANDA KANGRA]
HAEMOGLOBINOMETRY ◼ method used to measure the concentration of hemoglobin in blood. It's essential for diagnosing conditions like anemia and monitoring certain medical treatments. VARIOUS METHODS: 1.Calorimetric Methods: In these methods, color comparison is made between the standard and the test sample by calorimetric methods. *Visual Method: Sahli's Acid-Hematin Method * Photoelectric Methods: Cyanmethemoglobin Oxyhemoglobin method: HB liq. NH3 Oxyhemoglobin color. Obs. Photocalorimeter 540nm *Alkaline Hematin method
2.Gasometric Method: oxygen carrying capacity of blood is measured in a van slyke apparatus. 1gm HB carries 1.34 ml of oxygen. This method measures only physiologically active hemoglobin ,which can carry oxygen. Result 2% less than other methods. 3.Chemical Method: Iron content of HB is estimated first. 100 grams of HB contain 374 mg of iron. 4. Specific Gravity Method: This method is useful in mass screening like selection of blood donors copper sulphate technique is used to determine specific gravity of blood and rough estimate of HB is obtained
SAHLI'S ACID HEMATIN METHOD: Principle: Blood is added to N/10 HCL which converts Hb into Acid-Hematin The brown color of Acid-Hematin is matched against the brown color of the standard of the comparator. Apparatus: Sahli's Hemoglobinometer: Comparator, Sahli's HB Tube Dropper Glass Rod as Stirrer Reagent: N/10 HCL and Distilled Water Specimen: EDTA-Anticoagulated venous blood or from a finger prick.
Method: Add 0.N HCl with a dropper into the square HB meter tube up to the mark of 2 gm% Fill HB pipette with 0.02 ml of blood and wipe off the excess blood at the nozzle of the pipette with moistened cotton. Blow off the blood in the pipette into the HB meter tube. Rinse the pipette by drawing in and discharging the blood acid solution at least twice. Allow acid to act on RBCs for 10 minutes to lyse the red cells and convert HB into acid hematin Match the color of the solution with that of the comparator in natural light. If the color is darker, add distilled water with the dropper and stir the solution with the glass rod. Continue the process until the color of the solution matches the comparator color. Take out the stirrer from the solution. Take reading from the upper meniscus in gm%
Sources of Error: If blood in EDTA tube is not mixed properly before filling the pipette, low/ high values may be obtained depending on whether the plasma or sedimented RBCs are taken in the pipette. Advantages: Easy to perform and quick . Disadvantages: Carboxyhemoglobin, methemoglobin, Sulfhemoglobin , and fetal hemoglobin are not converted to acid hematin by 0.1 N HCl. Color of brown glass standard fades with time. Personal visual error in matching colour is 10% Normal Range: Males: 13-18 gm% Females: 11-16 gm% Newborns: 14-19 gm%
Clinical significance of HB: Low Hb values: Deficiency or hemolytic anemia High Hb values: Polycythemia vera Secondary erythrocytosis Hypoxia Cigratte smoking Precautions: Rinse the Hb pipette using tap water in a beaker. This prevents blocking of the pipette.
CYANMETHAEMOGLOBIN METHOD : PRINCIPLE : When blood is mixed with Drabikin's reagent containing Potassium Cyanide and potassium ferricyanide. Hemoglobin reacts with K.ferricyanide to form methemoglobin and is then converted to stable cyanmethemoglobin (HICN) by potassium cyanide. The color of the solution is compared with known cyanmethemoglobin standard at 540 nm, and its intensity is directly proportional to the concentration of hemoglobin in blood. Apparatus: Hb pipette Spectrophotometer
Reagents : Drabkin's solution at ph ( 7.0-7.4) Contains; POTASSIUM CYANIDE =100mg POTASSIUM FERICYANIDE=400mg POTASSIUM DIHYDROGEN PHOSPHATE =280mg STEROX SE =1ml D/W TO =2ltr's
Sample: EDTA anticoagulated venous blood or blood from a finger prick or capillary blood. METHOD Take 5 ml of Drabkin's solution in a test tube. Add 0.02 ml of blood to it using an HB pipette. Wait for 10 minutes. Switch on the calorimeter with a green filter / 540 nm, and wait for 5 minutes. Adjust OD to 0 using Drabikin's solution as a blank. Pour the test tube solution into the cuvette and note down the OD of the test sample. OD of the standard cyanmethemoglobin solution is taken against the blank. Calculate hemoglobin concentration using the standard curve.
Preparation of standard curve 1. Make serial dilutions of the reference solution eg 1in2, 1in4, 1in8 and so on with Drabkin’s solution. As the Hb present in the solution is known, the Hb concentration of each dilution will also be known. 2. Take the OD of each dilution in the colorimeter against a blank of Drabkin’s solution. 3. Plot the OD against the Hb concentration on a linear graph paper. The absorbance is plotted on the vertical axis and the hemoglobin concentration on the horizontal axis.The points should lie in a straight line that passes through the origin. 4. From this graph a table of readings and the corresponding Hb value can be prepared. This is more convenient than the graph specially when a large number of readings have to be taken. After OD of the sample is taken, the corresponding Hb value can be directly read from the table
Advantages: All forms of hemoglobin except sulfhemoglobin are converted to HICN, so accurate HB values are obtained. Visual error is eliminated as no color matching is required. Disadvantages: Potassium cyanide (KCN) is a poisonous substance, and therefore Drabkin solution must never be pipetted by mouth. Abnormal plasma proteins may cause turbidity when blood is diluted with Drabikin solution, leading to erroneous results. Precautions: Never pipette Drabikin solution directly. Lipemic blood with high total leukocyte count (TLC) can cause turbidity, resulting in erroneous results.
MODERN HB ESTIMATION METHOD: Automated Hematology Analyzers: These are sophisticated instruments that use flow cytometry, impedance, and/or spectrophotometry techniques to measure hemoglobin levels along with other blood cell parameters. They offer high throughput and accuracy, requiring minimal manual intervention.