Impedance audiometry part 1

adityagroy 8,387 views 39 slides Feb 13, 2017
Slide 1
Slide 1 of 39
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39

About This Presentation

audiometry


Slide Content

IMPEDANCE AUDIOMETRY Aditya G hosh Roy PGT2 MS ENT

IMPEDANCE AUDIOMETRY USE TEST TYMPANOMETRY EUSTACHIAN TUBE FUNCTION STAPEDIAL REFLEX

SOUND WAVE SOUND WAVE SOLID RELECTED SOUND WAVE AIR AIR ABSORBED SOUND

SOUND AIR SOLID REFLECTED SOUND

IMPEDANCE SOUND MIDDLE EAR IMPEDANCE MATCHING DEVICE COCHLEA EAC TM

TYMPANOMETRY (DEF) CHANGE IN PRESSURE --- CHANGE IN TM STIFFNESS ---SOUND ENERGY REFLECTED BACK DEPENDS ON STIFFNESS OF TM OR PRESSURE CHANGE PROBE TONE AMPLIFIER MANOMETER S OU N D TYMPANIC MEMBRANE REFLECTED SOUND PRESSURE

THEREFORE TYMPANOMETRY MAY BE DEFINED AS MEASUREMENT OF CHANGE IN IMPEDANCE OF MIDDLE EAR CAVITY AT THE TM PLANE IN RELATION TO CHANGE IN PRESSURE IN EAC. TYMPANOGRAM Y or ORDINATE IMPEDANCE OR STIFFNESS X OR ABSCISSA AIR PRESSURE

IMPEDANCE NONE OF THE PARAMETERS MEASURED INDIVIDUALLY SO AS A RESULT ACOUSTIC ADMITTANCE IS MEASURED ADMITTANCE IS RECIPROCAL OF IMPEDANCE STIFFNESS MASS FRICTION IMPEDANCE OPPOSITION TO FLOW OF ENERGY ACOUSTIC OHM ADMITTANCE EASE TO FLOW OF ENERGY ACOUSTIC MHO

AUDIOMETERS STILL AUDIOMETERS DO NOT MEASURE ADMITTANCE DIRECTLY COMPLIANCE LOW FREQUENCY PROBE TONE USED IMPEDANCE EASE OF MOVEMENT OR SPRINGINESS OF MIDDLE EAR SYSTEM IS ALSO RECIPROCAL OF STIFFNESS

HENCE IF COMPLIANCE IS MEASURED USING LOW FREQUENCY TONE IMPEDANCE OF MIDDLE EAR SYSTEM MOST OF THE IMPEDANCE AUDIOMETERS MEASURE COMPLIANCE

IMPORTANT POINTS C1 – COMPLIANCE VALUE AT +200mm water C2 – COMPLIANCE VALUE AT (MAXIMUM) CX – STATIC COMPLIANCE (C2- C1=CX) STATIC COMPLIANCE IS DEFINED AS COMPLIANCE OF AUDITORY CONDUCTIVE PATHWAY AT TM.

MIDDLE EAR PRESSURE PRESSURE AT WHICH COMPLIANCE VALUE IS MAXIMUM IS CALLED AS MIDDLE EAR PRESSURE IE PRESSURE OF AIR IN MIDDLE EAR CAVITY.

PHYSICAL VOLUME TEST VALUE THE COMPLIANCE VALUE AT +200mM water IS THE VOLUME OF EAC .

TYMPANOMETRY MEASUREMENT OF STATIC COMPLIANCE MEASUREMENT OF MIDDLE EAR PRESSURE

MEASUREMENT OF STATIC COMPLIANCE cc/ ml if COMPLIANCE Millimho if SUSPECTANCE IF A 220 Hz PROBE– 1MILLIMHO OF SUSPECTANCE = 1 CC/ML OF COMPLIANCE TYMPANOGRAM

NORMAL RANGE COMPLIANCE IMP POINT– CLINICAL CONDITION OF TM TO BE ALWAYS NOTED STUDY BY A.S. FELDMAN– HEALED TM INVALIDATES COMPLIANCE AS ONE OF THE TESTS OF TYMPANOMETRY .35 TO 1.40 ML

MEASUREMENT OF MIDDLE EAR PRESSURE EXPRESSED IN mm WATER / DEKA PASCAL ( daPa ) 1mm WATER = 1.20 DEKA PASCAL RANGE = +50 TO -50 mm WATER MEP NEGATIVE WHEN P < -100 mm WATER PAED AGE GROUP-- +25 TO -100mm WATER

IMP POINT--POSITIVE PRESSURE PEAK SWALLOWING CHILD CRYING PHYSIOLOGICAL VALSALVA HOLD BREATH IN EXPIRATION IN EARLY AOM PUS FORMED IN MEC AND ET DYSFUNCTION DECREASE VOLUME + INCREASE PRESSURE OF MEC POSITIVE PRESSURE PEAK

TYPES AND SHAPES OF TYMPANOGRAMS

TYPE A SHARP MAXIMUM AT 0 mm WATER NORMAL EAR AND SOME CASES OF OTOSCLEROSIS

TYPE AD NORMAL MIDDLE EAR PRESSURE WITH HIGH COMPLIANCE SEEN IN OSSICULAR CHAIN DISCONTUINITY OR SCARRING OF TM NOTCHED PEAK OR DOUBLE MAX AT 0 mm USUALLY SEEN WITH A HIGH FREQUENCY PROBE IN OSSICULAR DISCONTUINITY

TYPE AS NORMAL MIDDLE EAR PRESSURE WITH LOW COMPLIANCE OTOSCLEROSIS AND THICKENED TM

TYPE B FLAT TYMPANOGRAM LITTLE OR NO COMPLIANCE AND NO SHARP PEAK SEEN IN ADHESIVE OTITIS MEDIA, OTITIS MEDIA WITH EFFUSION, PERFORATION OF TM

TYPE C NEGATIVE MIDDLE EAR PRESSURE WITH NORMAL COMPLIANCE ET DYSFUNCTION WITHOUT EFFUSION

TYPE CS NEGATIVE MIDDLE EAR PRESSURE WITH LOW COMPLIANCE ET DYSFUNCTION WITH OTITIS MEDIA WITH EFFUSION

CLASSIFICATION BY A.S. FELDMAN PEAK PRESSURE TYPE OF CONFIGURATION OF TYMPANOGRAM COMPLIANCE OR AMPLITUDE

1.NORMAL TYMPANO GRAM GOOD COMPLIANCE AND PEAK PRESSURE AT 0 mm WATER CORRESPONDS TO JAGER TYPE A

2 .NORMAL MIDDLE EAR PRESSURE WITH HIGH COMPLIANCE IF 226HZ PROBE USED THEN EITHER OSSICULAR CHAIN DISCONTUINITY OR SCARRING OF TM BUT IF A 660 OR 800 HZ PROBE USED THEN A NOTCHED PEAK OR DOUBLE MAX AT 0 mm USUALLY SEEN IN OSSICULAR DISCONTUINITY

3.NORMAL MIDDLE EAR PRESSURE WITH LOW COMPLIANCE OTOSCLEROSIS OR OSSICULAR FIXATION AND THICKENED TM

4. FLAT TYMPANOGRAM LITTLE OR NO COMPLIANCE AND NO SHARP PEAK SEEN IN ADHESIVE OTITIS MEDIA, OTITIS MEDIA WITH EFFUSION, PERFORATION OF TM . CORRESPONDS TO JAGER TYPE B.

5.NEGATIVE MIDDLE EAR PRESSURE WITH LOW COMPLIANCE ET DYSFUNCTION WITH OTITIS MEDIA WITH EFFUSION CORRESPONDS TO JAGER TYPE CS

6. POSITIVE MIDDLE EAR PRESSURE NORMAL COMPLIANCE RARELY SEEN IN EARLY AOM IN EARLY AOM PUS FORMED IN MEC AND ET DYSFUNCTION DECREASE VOLUME + INCREASE PRESSURE OF MEC POSITIVE PRESSURE PEAK

7.NEGATIVE MIDDLE EAR PRESSURE WITH NORMAL COMPLIANCE ET DYSFUNCTION WITHOUT EFFUSION

TO SUMMARISE

TO BE CONT..