An ideal I.S.E. consists of a thin membrane across which only the
intended ion can be transported.
The transport of ions from a high conc. to a low one through a selective binding with some sites within the membrane creates a
Size: 663.27 KB
Language: en
Added: Sep 19, 2020
Slides: 23 pages
Slide Content
Ion selective electrodes ( ISE) Prepared by:- MANIKANT MANI
Ion selective electrodes(ISE) Also known as indicator electrodes . Respond directly to the analyte . Used for direct potentiometric measurements . Selectively binds and measures the activity of one ion (no redox chemistry) Examples pH electrode Calcium (Ca 2+ ) electrode Chloride (Cl - ) electrode
An ion-selective electrode ( ISE ), also known as a specific ion electrode ( SIE ), is a transducer (or sensor) that converts the activity of a specific ion dissolved in a solution into an electrical potential, which can be measured by a voltmeter or pH meter. An ideal I.S.E. consists of a thin membrane across which only the intended ion can be transported. The transport of ions from a high conc. to a low one through a selective binding with some sites within the membrane creates a potential difference.
when a charged species diffuses from a region of activity , A1, to a region of activity ,A2, there is a free energy change . This decrease in free energy should equal the electrical work done to the surroundings. Hence, Where Z is the charge of the species, including the sign.
Ion Selective Electrodes (including the most common pH electrode) work on the basic principal of the galvanic cell .By measuring the electric potential generated across a membrane by "selected" ions, and comparing it to a reference electrode, a net charge is determined. The strength of this charge is directly proportional to the concentration of the selected ion. The basic formula is given for the galvanic cell: E cell = E ISE - E Ref
TYPES OF ION SELECTIVE ELECTRODE (ISE) Glass Membrane Electrode (i.e. H+ electrode) (3) Liquid Membrane Electrode (e.g. Ca2+ electrode uses a liquid chelator ) (4) Gas Sensing Electrode (e.g. CO2 gas sensing electrode) (2) Solid State Electrode (e.g. F- electrode uses a Eu2+-doped LaF3 crystal)
The glass electrode Most of the metal cations (e.g. Na+ ) in the hydrated gel layer diffuse out of the glass membrane and into the solution, see Fig. Concomitantly, H+ from solution diffuse into the membrane. Response of glass electrode (at 25 °C). GLASS MEMBRANE ELECTRODE
During two-point calibration, the y-intercept can be corrected by the “calibrate” knob. The slope is corrected by the “slope” or “temperature” knob during pH meter calibration.
SOLID STATE ELECTRODE This is based on an inorganic crystal doped with a small amount of ions of a different valency to create vacant sites. For instance, in a F- ion electrode, a LaF ₃ crystal is doped with EuF ₂ so that there are anion vacancies for the migration of F through the LaF3 crystal, Response of F- ion electrode . E = constant - (0.05916) log (AF-, outside) = constant - (0.05916) log ᵞ F- - (0.05916) log [F- ] , At high pH, there is interference by OH- (Note kf -, OH- = 0.1) At low pH, F- is converted to HF ( pKa = 3.17) to which the electrode is insensitive.
The analyte solution is usually diluted in a high m buffer to keep the pH (at 5.5) and m constant. E = constant´ - (0.05916) log [F- ] Therefore, a plot of E versus log [F- ] should yield a straight line with a negative slope,
LIQUID MEMB R AN E ELE C TRODE Liquid membrane is a type of ISE based on water-immiscible liquid substances produced in a polymeric membrane used for direct potentiometric measurement. Used for direct measurement of several polyvalent cations (Ca ion) as well as a certain anions. The polymeric membrane made of PVC to separate the test solution from its inner compartment. Contains standard solution of the target ion. The filling solution contains a chloride salt for establishing the potential of the internal Ag/AgCl wire electrode.
G A S SENSING E LECTRODE They are used to assay the gases dissolved in aqueous solutions. It is constructed by enclosing the glass pH membrane in a second, gas- permeable hydrophobic membrane. A thin layer of an electrolyte solution is held between the two membranes. They also have a small reference electrode enclosed within the gas- permeable membrane. The change is detected by a combination pH sensor within the housing. This type of electrode does not require an external reference electrode.
Advantages of Ion Selective Electrode ( ISE) Technique When compared to many other analytical techniques, Ion-Selective Electrodes are relatively inexpensive and simple to use and have an extremely wide range of applications and wide concentration range. Under the most favorable conditions, when measuring ions in relatively dilute aqueous solutions and where interfering ions are not a problem, they can be used very rapidly and easily. They are particularly useful in applications where only an order of magnitude concentration is required, or it is only necessary to know that a particular ion is below a certain concentration level.
They are invaluable for the continuous monitoring of changes in concentration for example in potentiometric titrations or monitoring the uptake of nutrients, or the consumption of reagents.They are particularly useful in biological/medical applications because they measure the activity of the ion directly, rather than the concentration. ISEs are one of the few techniques which can measure both positive and negative ions. They are unaffected by sample colour or turbidity. ISEs can be used in aqueous solutions over a wide temperature range. Crystal membranes can operate in the range 0 C to 80 C and plastic membranes from 0 C to 50 C.
LIMITATION 1. Precision is rarely better than 1%. 2. Electrodes can be fouled by proteins or other organic solutes. 3. Interference by other ions. 4. Electrodes are fragile and have limited shelf life. 5. Electrodes respond to the activity of uncomplexed ion. So ligands must be absent or masked. m must be kept constant.
Empirical Calibration Plot Potential (mV) Slope = 59/z i mV z i = charge of ion Called Nernstian slope Used to determine the unknown concentration of analytes . Departure from linearity is observed at low concentrations
APPLICATION Ion-selective electrodes are used in a wide variety of applications for determining the concentrations of various ions in aqueous solutions. The following is a list of some of the main areas in which ISEs have been used. Pollution Monitoring: CN, F, S, Cl, NO3 etc., in effluents, and natural waters.Agriculture : NO3, Cl, NH4, K, Ca, I, CN in soils, plant material, fertilisers and feedstuffs. Food Processing: NO3, NO2 in meat preservatives. Salt content of meat, fish, dairy products, fruit juices, brewing solutions. F in drinking water and other drinks.
K in fruit juices and wine making. Corrosive effect of NO3 in canned foods. Detergent Manufacture: Ca, Ba, F for studying effects on water quality. Paper Manufacture: S and Cl in pulping and recovery-cycle liquors. Explosives : F, Cl, NO3 in explosive materials and combustion products. Biomedical Laboratories: Ca, K, Cl in body fluids (blood, plasma, serum, sweat ). F in skeletal and dental studies. Education and Research: Wide range of applications. Ca in dairy products and beer.
Determination of Fluoride Ion In the lanthanum fluoride electrode, the sensing element is a crystal of lanthanum fluoride LaF 3 , doped with europium fluoride EuF 2 to create lattice vacancies . Such a crystal is an ionic conductor by virtue of the mobility of fluoride ions which jump between lattice vacancies. An electrochemical cell may be constructed using such a crystal as a membrane separating two fluoride solutions. This cell acts as a concentration cell with transference where the fluoride transport number is 1. As transference of charge through the crystal is almost exclusively due to fluoride, the electrode is highly specific to fluoride
The only ion which significantly interferes is hydroxide (OH - ). Generally such "alkaline error" can be avoided by buffering the sample to a pH below 7 The cell diagram of a typical experimental arrangement is: Cu ' | Ag,AgCl | KCl || solution | LaF 3 | KF,KCl | AgCl,Ag | Cu where: Cu ' | Ag, AgCl | KCl is an external reference electrode KF,KCl/AgCl,Ag/Cu is an internal reference inside the fluoride selective electrode