Introduction, Fundamental modes of vibrations in poly atomic molecules, Sample handling, Factors affecting vibrations, Instrumentation - Sources of radiation, wavelength selectors, Detectors
Applications
Size: 1.06 MB
Language: en
Added: Sep 24, 2023
Slides: 25 pages
Slide Content
Introduction,
Fundamental modes of vibrations in poly atomic
molecules,
Sample handling,
Factors affecting vibrations
Instrumentation -Sources of radiation,
wavelength selectors, Detectors
Applications
CONTENT
It is a type of absorption spectroscopy.
Analytical technique used for qualitative and quantitative analysis of sample by
using IR region of EMR.
Basically used in identification of compounds by determining its structure and
functional group.
Functional group have unique and separate fingerprint structure.
IR region is divided into 3 regions:
1.Near: 14000 –4000 cm
-1
2.Mid: 4000 –400 cm
-1
3.Far: 400 –10 cm
-1
Each functional group has specific wave number (frequency of absorption)
INTRODUCTION
When frequency of radiations is equal to vibration frequency of molecule,
absorption of IR radiation takes place and peak is recorded.
IR Radiations Detector
2783 –2843 cm
-1
PRINCIPLE
TYPICAL IR ABSORPTION FREQUENCIES
Only applicable for compounds having matching frequencies (Range of wave number).
Change in dipole moment is also essential.
A molecule will absorb IR radiation if the change in Vibrationalstates is associated
with a change in the dipole moment of the molecule.
Vibrations which do not change the dipole moment are IR inactive (Homonuclear
diatomics)
Homonucleardiatomic molecules: O
2, N
2, H
2, Cl
2
Heteronucleardiatomic molecules: HF, HCl
SELECTION RULE
FUNDAMENTAL MODES OF VIBRATIONS
Movement of molecule when it absorbs IR radiations.
They are of 2 types of vibrations.
1.Stretching Vibrations
Distance between 2 atoms increases or decreases but in the same axis.
It is further subdivided into 2 types:
a)Symmetric stretching
Movement of atoms in the same directions.
b)Asymmetric stretching
Movement of atoms in the opposite directions.
2. Bending Vibrations
In this, position of atoms changes with respect to original bond axis.
Stretching absorption of bond appears at higher frequencies than bending
absorption.
It is further subdivided into 4 types:
a)Scissoring
2 atoms approaches each other.
b)Rocking
Movement of atoms in the same directions.
c) Wagging
2 atoms moves up & down the plane with respect to central atom.
d) Twisting
1 atoms moves up the plane & other moves down the plane with respect to
central atom.
FACTOR AFFECTING VIBRATIONS
1. Coupled interaction
Normally there is only one Vibrational frequency i.e. C-H & C=O.
But in case of –CH
2and CO
2
H-C-H O=C=O
It has two fundamental stretching vibration:
Symmetric (Lower wave no.) & Asymmetric stretching (Higher wave no.)
2. Hydrogen Bonding
Special type of dipole-dipole interaction between the molecules.
H
H
O
H
O
H
Intermolecular
Intramolecular
Bond between 2 different molecule
(Broad band)
Bond between same molecule
(Sharp & well defined band)
H-bonding decreases frequency
e.g. O-H: 3650cm
-1
, after H-bonding O-H-: 3200cm
-1N
H
O
H
O
3. Resonance effect
It will increase the bond length which decreases the frequency.
(Bond sharing: Increase bond length)
e.g.
1715 cm
-1
1690 cm
-1
4. Electronic effect (Inductive effect)
Frequency will change by changing the substituents of neighbors group.
Inductive Effect
+ I effect (ERG) -I effect (EWG)
FrequencyBond length FrequencyCH3 C
O
CH CH
2 CH3
H
C O H
H
C O
1750 cm
-1
1745 cm
-1CO
CH3
CH3 ClCH
2
C
ClCH
2
O 1725 cm
-1
1740 cm
-1
Solid sampling technique
Preparation of solid sample:
A)Direct sampling: Solid sample is put in sample holder.
B)Pellatizationtechnique:
Solid sample + KBr
Pass under high pressure in press
Formation of small thick (1-2 mm) pellets. (1 cm in diameter)
Pellets are transparent to IR radiations.
KBr, NaClare only used for IR suitable compounds.
C) Mulling technique
Solid sample mix with mineral oil (Nujol)
Formation of paste
Paste is placed between two salt plate
Nujolis transparent in most of IR region.
Hexachloridieneis also used as mulling agent.
Liquid sampling technique
In this, sample is sandwiched between 2 NaClplate and form a thin film (0.1 –0.3 mm).
If sample contains water then Calcium fluroideplates are used.
Gas sampling technique
In this, aqueous sample is placed into gas cell (10 cm long) made from NaClwall.
Solution sampling technique
In this, aqueous sample is dissolved in suitable solvents then analyzed in the form
of solution.
Commonly used solvents: Chloroform, Carbon tetrachloride, Carbon disulfide,
etc.
INSTRUMENTATION
Typically it consist of following parts:
1.Radiation source
2.Monochromator
3.Sample cell
4.Detector
5.Read out system
1.RADIATION SOURCE
IR spectrometer requires a source of radiant energy for narrow frequency bands.
Incandescent solid is chosen as Radiation source.
i) Nernst glower
Consist of rod or hollow tube (2 cm long, 1 mm in diameter) made by sintering a
mixture of cerium, zicronium, thorium, etc.
Heated between 1000-1800
0
C (emit radiations)
Max. radiations: 7100 cm
-1
ii) Globar
It is silicon carbide rod (5 cm long and 0.5 cm in diameter)
Heated between 1300-1700
0
C
Max. radiations: 5200 cm
-1
iii) Nichromewire
A coil of thin wire heated by passing current.
It is a tightly wound coil of nichrome(Ni-Cr) wire, electrically heated
to 1100 K. It produces a lower intensity of radiation than the Nernst or
Globarsources, but has a longer working life.
iv) Rhodium wire
Wire is sealed in a cylinder.
It is used when the required wavelength range and intensity
are not too great.
v) Tungsten filament lamp
Mainly used for near IR region.
The lamp consists of a tungsten filament in a evacuated glass or
quartz envelope that contains a small amount of iodine vapor to
increase the lifetime of the filament.
2. MONOCHROMATOR
Radiation source emits radiations of various frequencies.
Used for selection of certain frequencies and for rejection of other frequencies.
Types:
1.Prism: Dispersive element made from NaCl
2.Grating: Causes linear and higher dispersion.
3. SAMPLE CELLS
Different sampling techniques are used for insertion of sample into sample cell.
4. DETECTOR
Thermal detectors are used for IR (depends on heat energy)
1.GOLAYCELL
Principle:WhenIRradiationisabsorbedbythegaspresentinthechamber,thegas
expandsanddeformstheflexiblemembranewhichresultsinthedeflectionoflight
fromthemembranewhichisthendetectedbythephotodiodearraydetector.
Ithassmallercylinderwithoneendofblackenedmetalplateandotherflexible
diaphragmmirror.
ItisfilledwithXenongas.
WhenIRradiationsfallsonblackenedmetalplatethengasheatupandexpandsthat
increasethepressure&thendeformthemetalizeddiaphragmmirror.
2. BOLOMETER
It works on the principle, increase in temp. results in increase in electrical
resistance of metal.
i.e. 1
0
C increase –0.4 % increase in Resistance
It consist of a thin metal conductor whose temp. changes when IR radiations falls
on it which further changes the resistance.
On the basis of resistance, IR intensity is calculated.
3. THERMOCOUPLE
It works on the principle that 2 dissimilar metal wires are connected together at both the
ends, so a temp. differential exist and electric current flows between two ends.
Thermocouple detector consists of two similar metal junctions such as Bismuth (Bi) and one
dissimilar metal placed in between these metals such as Antimony (Sb).
One Bismuth junction is termed a detector junction or hot junction which is exposed to the IR
radiation and the other Bismuth junction is termed a reference junction or a cold junction.
Reference junction has a higher heat capacity than the detector junction so that the detector
junction can respond to the minute changes in the IR radiation.
When the IR radiation falls on the detector junction, the junction heats up. This causes a
difference in the temperature between the detector junction and the reference junction.
This generates a potential difference between these two and is detected by the voltmeter.
Electricity flow is ∝Energy difference between 2 connections.
4. THERMISTOR
It is made up of fused mixture of metal oxides.
It is same as that of Bolometer, but in this electrical resistance of mixture
decreases with increase in temperature.
T ∝1 / R, 1
0
C increase –5 % Resistance decrease
One advantage of a thermistor over a Bolometer is that the response time is rapid
which gives improved resolution and faster screening rates. But the sensitivity of
the thermistor is poorer as compared to the bolometer.
5. PYROELECTRIC
It contain a pyroelectricmaterial which produce electricity on changing thermal
energy.
In this dielectric material is placed between electrodes.
When IR radiations exposed to black coating, it generates thermal energy and
temperature changes.
Due to this, charge developed in dielectric and form pyroelectricmaterial and this
leads to flowing electricity.
Effect of this detector depends upon the rate of temp. changes.
It is mainly used in FTIR spectrometer.
5. READ OUT SYSTEM
APPLICATIONS
IR Spectroscopy is widely used in industries and research work.
Identification of compound by using Fingerprint region.
Structural determination
Detection of impurities (Standard and test)
Distinguish between two types of H –bond, if IR frequency changes after dilution
then it is intermolecular & vice versa.
Study of progress of chemical reactions (Bond formation impact on functional
group)