New OCT Slides for revascularisation and decision making

SadanandIndi 110 views 114 slides Jul 14, 2024
Slide 1
Slide 1 of 114
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79
Slide 80
80
Slide 81
81
Slide 82
82
Slide 83
83
Slide 84
84
Slide 85
85
Slide 86
86
Slide 87
87
Slide 88
88
Slide 89
89
Slide 90
90
Slide 91
91
Slide 92
92
Slide 93
93
Slide 94
94
Slide 95
95
Slide 96
96
Slide 97
97
Slide 98
98
Slide 99
99
Slide 100
100
Slide 101
101
Slide 102
102
Slide 103
103
Slide 104
104
Slide 105
105
Slide 106
106
Slide 107
107
Slide 108
108
Slide 109
109
Slide 110
110
Slide 111
111
Slide 112
112
Slide 113
113
Slide 114
114

About This Presentation

Optical coherence tomography
Vs intravascular ultrasound


Slide Content

OCT Optical Coherence Tomography Advanced Imaging to guide complex interventions

Basics of OCT OCT System overview OCT Catheter OCT Procedure OCT Image Interpretation Clinical Trials Case Based Discussion on Demo OCT Laptop What is in store for the day…

OCT BASICS

Time for High resolution imaging

What is OCT? Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images of tissue microstructure. Optical  of or relating to light (visual) Coherence  a measure of the correlation between phases of a wave or waves Tomography  method of producing images from a series of single planes or slices (Greek “TOMOS=Slice”) 5

What Is Intravascular OCT? OCT is an optical imaging modality that uses near-infrared light (1250-1350nm) for high-resolution imaging of vessel anatomy, tissue microstructure and stents. Key Features: Uses light, not sound Does not use X-ray Image acquisition is fast Images acquired are sharp, detailed and easy to interpret 6

Intravascular OCT Flexible fiber-optic catheter used for light delivery Fiber rotates to create image frames Fiber pulls back to map vessel segment The more frames per mm, or high frame density, the higher the resolution 7

Michelson Interferometer 1881 Albert Michelson (1852-1931) Nobel Prize Physics 1907

Time Domain OCT (TD-OCT): (older generation) Non-Commercially available for cardiovascular use 2001 to the present Moderate image quality slower imaging speeds (full pullback 30s) and made it necessary to totally block the blood flow in the vessel using an occlusion balloon Frequency Domain OCT (FD-OCT): Commercially available for cardiovascular use 2010 to the present Exceptional image quality Fast imaging (full pullback 3s) : 10 to 100x increase in speed Rapid contrast flush instead of balloon occlusion Time vs. Frequency Domain Intravascular OCT C7-XR system: FD-OCT 100 fps, 20 mm/s pullback M3 system: TD-OCT 20 fps, 1 mm/s pullback Gonzalo, N . Optical Coherence Tomography for the Assessment of Coronary Atherosclerosis and Vessel Response after Stent Implantation . Rotterdam, the Netherlands: Optima Grafische Communicatie; 2010. improved image quality

Light is too fast for direct echo measurement  interferometry Compares path length between known reference arm and sample arm Mechanical reference arm motion limits imaging speed Time Domain OCT (M2/M2x/M3) intensity axial distance Demod Amp Broadband Source D Tissue Mirror Reflections (moving)

Time Domain OCT: fixed laser signal, moving mirror

Measurement of interference pattern spectrum + Fourier transform Signal generated from all depths simultaneously Faster image acquisition without loss of quality greater penetration depth Frequency Domain OCT (C7-XR TM and ILUMIEN TM ) Swept Laser D λ intensity intensity distance FFT Amp

Frequency Domain OCT: sweeping laser signal, fixed mirror

OCT IMAGE ACQUISITION

OCT CATHETER MOVEMENT – REAL TIME

Pullback Generation One pullback  270 – 540 frames 16

Image Display 17 “L-Mode” longitudinal view “B-Mode” cross-sectional view

Intravascular OCT Rotate a sensor & Pullback distal proximal Catheter with fiber-optic core used for light delivery Fiber rotates to create image frames Fiber-optic core pulls back to map vessel segment During the pullback, blood clearing is required to get a good image quality. Sensor Lens Optic fiber OCT catheter Better image quality by: Faster rotation Slower pullback

Principle of catheter based intravascular imaging 1 st line A-Line distal proximal Sensor

A-Line acquisition 2 nd line 3rd 4th 5th 1st 2nd 3rd 1 2 3 4 5 6 7 ………………………………………………………………………………………………..…n depth

R-Theta Conversion (Frame) 1 2 3 4 5 6 7 ……………………………………………………………………………………………..…n 1,2,3 4,5,6 One frame  560 axial lines Computer make lines to circle

A pulse of light is sent out that bounces off the different layers of tissue and returns back to be analyzed by the system. The catheter can be seen in the middle of the image, with the optical fiber at its core. Measure echo time delay of reflected light waves Ilumien each frames = 504 A lines Optis and optis integrated each frame = 560 A limes Image (Frame)

Difference of resolution . Direction of beam High resolution Low resolution Unable to distinguish. Distinguishable each component

Axial direction Lateral direction 2 types of Resolution(Axial/Lateral) Axial: parallel to the beam = along the vessel Lateral: perpendicular to both the beam and the catheter = cross-section of vessel ※JACC vol.37, No.5, 2001 1478-92(IVUS consensus report) The ability of an imaging system to resolve detail in the object that is being imaged

IVUS/OCT: smaller beam size & larger number of A-Lines To get better resolution (Lateral Resolution) Lateral resolution is around With OCT 20 to 40 µ m

Resolution 400 pix 600 pix 200 pix 130 pix

Resolution 400 pix 600 pix 200 pix 130 pix

OCT SYSTEM OVERVIEW

OCT Evolution 29 1999-2001 PTCA balloon + ImageWire TM R&D prototypes Not commercially available Inside PTCA balloon ‘Snapshot’ flush imaging 2007 M3 System CE mark 20 fps / 240 lines Occlusion + flush 2004 Soft occlusion balloon + ImageWire TM M2 System CE mark 15 fps / 200 lines Occlusion + flush 2009 C7 XR ™ System CE mark, FDA cleared 100 fps / 500 lines NO occlusion Occlusion-free Dragonfly TM

30 2011-12 2013-15 2016 ILUMIEN ™ System ILUMIEN ™ OPTIS™ System OPTIS™ Integrated System OPTIS™ Mobile System Commercially available 2011 100 fps / 54 mm pullback Combined FFR and OCT Wireless FFR 180 fps/75 and 54 mm pullback Advanced software tools Real time 3D reconstruction Pullback initialization from DOC Integrated in the Cath lab Angio Co-Registration With tableside FFR/ OCT controller Integrates with Multiple Cath labs Angio Co-Registration With tableside FFR/ OCT controller FFR and OCT System 2 nd Gen FFR and OCT System Cath lab integrated FFR and OCT system Latest system with 3D OCT, FFR, ACR & MSO 2015 Current Generation PCI Optimization Systems

31 2016/17 OCT product launches OPTIS™ Mobile System OPTIS™ Mobile Upgrade Kit (Ilumien™ OPTIS™ Systems only ) OPTIS™ Metallic Stent Optimization Software-MSO (Software Version E.4) OPTIS™ Mobile Workstation (Accessory for OPTIS™ Integrated System) Mobile System integrated into Cath lab during procedure to offer angio co-registration Angio Co-registration Wireless Tableside Controller Mobile cart connect into Cath Lab during procedure Enables the use in multiple labs Will include E.4 after LMR Enables angio co-registration function on Ilumien™ OPTIS System Angio Co-registration (ACR) Wireless Tableside Controller Mobile cart connect into Cath Lab during procedure Enables the use of OCT-ACR and FFR in multiple labs Leverage options to include E.4 New clinically relevant software enhancements for pre and post PCI Stent roadmap on angio co-registration New default view with automated measurements from lumen profile Acute Stent detection and apposition assessment Expanded 3D imaging Standard available for bug fixes only Offered as another point of control to operate the OPTIS Integrated system Optional control point for non-sterile personnel to perform all input/viewing functions available on OPTIS Integrated System Note: All systems having MSO must be updated to E.4.1 version by Service team.

OPTICAL COHERENCE TOMOGRAPHY - SYSTEM OVERVIEW

DOC (Drive Motor & Optical Controller) CPU (Central Processing Unit) ENGINE (Infrared Light Source) Physician and Operator Monitors Monitors

DOC ( Drive Motor and Optical Controller ) 1 Emergency shutoff – stops rotation, pullback and scanning 2 Enable scanning; second push will start pullback in manual trigger mode 3 Toggle between Live and Standby 4 Unload catheter 1 2 3 4

Dragonfly TM Duo / Dragonfly TM Optis OCT Catheter OCT ILUMIEN / ILUMIEN Optis ™ / Optis ™ range – both oct & ffr PressureWire TM Aeris TM PressureWire TM X ILUMIEN™ Console Ilumien Optis Console Wireless FFR

Comparison to ILUMIEN Engine ILUMIEN ILUMIEN OPTIS/ OPTISi & Mobile Parameter Nominal Settings Long Pullback High-resolution Pullback Engine speed 100 frames/sec 180 frames/sec 180 frames/sec Pullback speed 20 mm/sec 36 mm/sec 18 mm/sec Frame density 5 frames/mm 5 frames/mm 10 frames/mm Pullback length 54 mm 75 mm 54 mm Pullback time 2.7 sec 2.1 sec 3.0 sec Contrast (Max) 14 ml at 4 ml/sec 10 ml at 4 ml/sec 14 ml at 4 ml/sec File size 271 frames 270 MB 375 frames 375 MB 540 frames 540 MB Key Features of Abbott - OCT More Information Same Time Same Contrast

Ilumien Optis & MSO Software Reviewing an acquired Image Entering note Bookmark Calibration Performing the Measurements Lumen Profile MLA /AS% / DS% Mean Diameter Dynamic stent planning Automated Measurements Stent Roadmap Apposition Indicator 3D Bifurcation Mode Side-branch detection Carina view Expanded 3D Navigation 3D flythrough view Object segmentation 37 OPTIS™ Stent Optimization Software broadens the clinical utility for OCT to guide daily decision making for complex PCI both pre and post stenting though:

Clinical applications of oct 38

OCT CATHETER OCT Dragonfly Range Of Catheters

40

Usable length: 135 cm Outer diameter: 2.7 F (distal) Wire lumen: 0.014” Hydrophilic Coating Dragonfly™ DUO Imaging Catheter Specifications Lens

INTRODUCING LATEST OCT CATHETER Dragonfly TM OPTIS TM Simplifies the OCT procedure TIP MARKER DUAL LUMEN for GUIDEWIRE EXIT Easier loading ; reduce potential for lens damage and kinking Lens Marker Facilitates Angio Co-Registration ©2018 Abbott. All Rights Reserved. IN CP0007-EN 02/18

Dragonfly Duo Dragonfly OPTIS Dragonfly™ OPTIS™ changes vs. Dragonfly™ Duo Catheters Dual lumen Smoother transition zone Shorter rail Dedicated purge exit Tip Design Dragonfly™ OPTIS catheter

Duo vs optis Lens DUO OPTIS

Catheter design changes: Rapid exchange tip Improves catheter performance Reduces risk of damage and kinking Support for continuous calibration* Reduces steps necessary during calibration Shaft guide marker at 90 cm Aids in understanding distal tip position during radial procedures Gray hub Differentiates Dragonfly OPTIS catheter from previous Dragonfly catheters Dragonfly™ OPTIS™ Imaging Catheter *Refer to D.2 or E.2 Software Note: The Dragonfly OPTIS imaging catheter is designed for use with the ILUMIEN OPTIS PCI Optimization and OPTIS integrated systems equipment. It should not be used with older model OCT systems. ILUMIEN™, ILUMIEN™ OPTIS™ PCI Optimization and OPTIS™ integrated systems software Dragonfly™ OPTIS™ imaging catheter

OCT PROCEDURE Performing an OCT Procedure

OCT Procedure CONFIDENTIAL – FOR INTERNAL USE ONLY P reparation Console / Catheter / Contrast P osition Catheter P urge P uff P ullback Troubleshooting The 4 P’s

Performing an OCT exam - REQUIREMENTS Required Material and Equipment • ILUMIEN™ or Optis System • Dragonfly™ or Dragonfly Duo Imaging Catheter • Sterile DOC Cover • 3 ml purge syringe • Contrast media indicated for coronary use, for purging and flush (allow 15 mL for each run planned) • 0.014 inch guidewire (with torque device if desired) • Guide catheter (6 French, 0.068 inch ID or larger, with no side holes) • Sheath introducer (to match guide catheter) • Hemostatic Y-Adapter/Connector • Heparinized , physiologic saline solution, for hydrophilic catheter preparation • Power injector pump for coronary angiography (capable of injecting 4.0 ml/sec for a total of 14 ml in 3.5 second) CONFIDENTIAL – FOR INTERNAL USE ONLY

Large Lumens increase device compatibility and allow more dye flow for better visualization FFR is strongly recommended in 6F OCT dragonfly is compatible with 6F large Lumen ID > 0.068” Guide catheter Inner lumen Mach 1 Boston Runway BSC Vista Brite Cordis Sherpa Active / Balanced MDT Launcher MDT Heartrail Climber Terumo 5F .058” .058” .058” .059” 6F .070” .070" .067” .070” .070” .071” .071” 7F .081” .078” .081” .081” .081” 8F .091” .088” .090” .090”

Allows for antegrade flow through the holes and out through the distal tip. Drawbacks: Decrease arterial visualization. Increase use of contrast Media Systemic drug delivery NB: GUIDE CATHETER WITH SIDE HOLES SHOULD NOT BE USED WITH FFR and OCT Side Holes in Guides

OCT procedure – Required materials 51 DOC – Bedside control to initiate pullback, able to run at fast speed Dragonfly Duo Imaging Catheter DOC Cover Ilumien / Optis Console Non Sterile Devices Sterile Devices 3ml Syringe

P reparation : STARTING Up THE OCT SYSTEM CONFIDENTIAL – FOR INTERNAL USE ONLY

Preparation : Add a new patient CONFIDENTIAL – FOR INTERNAL USE ONLY

Preparation : Patient data CONFIDENTIAL – FOR INTERNAL USE ONLY

Preparation : Step by step on the screen CONFIDENTIAL – FOR INTERNAL USE ONLY

Preparation DOC Cover Give the DOC Cover to the physician Pass the DOC through the DOC cover Pull the DOC cover “Telescoping style” fold DOC Cover Tab labeled “Insert DOC Here” DOC cover length increased from 60 inches to 96 inches Sterile blue cover wrapped around DOC cover Plastic ring around the DOC cover opening CONFIDENTIAL – FOR INTERNAL USE ONLY

P reparation: Catheter preparation Remove catheter from hoop Wipe shaft to activate hydrophilic coating Gently purge catheter with 3ml syringe filled with 100% contrast PRIOR to connection Inject until 3 drops exit tip Leave syringe attached for repeat purging in vivo CONFIDENTIAL – FOR INTERNAL USE ONLY

Don’t touch inner fiber optics! Preparation: Connect the catheter to the DOC CONFIDENTIAL – FOR INTERNAL USE ONLY

Preparation: Test Image To Test the Image hold the catheter CONFIDENTIAL – FOR INTERNAL USE ONLY

Preparation Summary Add a new patient Data Pass the doc through the DOC cover Activate catheter hydrophilic coating of Catheter Gently purge catheter with 3ml syringe filled with 100% contrast Connect the catheter to the DOC Test The image Auto-calibrate Prepare and Connects Automatic injector pump to the manifold CONFIDENTIAL – FOR INTERNAL USE ONLY

Procedure-Ready Automatic injector pump connects to manifold 3 cc contrast syringe remains connected DOC 6-7 F guide catheter (no side holes) 0.014" standard guidewire Dragonfly catheter loaded with guidewire Y should not be tighten too much CONFIDENTIAL – FOR INTERNAL USE ONLY

Backloading the Dragonfly DUO onto the Guidewire Entry portal is blue. Exit portal for guidewire is distal to proximal marker. Tortuous Anantomy or Long stent : Disconnect the DOC and the catheter / Or use 2 guide wires Proximal marker Lens located at end of torque wire CONFIDENTIAL – FOR INTERNAL USE ONLY When manipulating the catheter the system must be in Standby mode

Insertion and Positioning Dragonfly Optis CONFIDENTIAL – FOR INTERNAL USE ONLY Lens Guidewire is prevented from going down image core and wall improves kink resistance Pinhole to reduce blood ingress Simplifies the OCT procedure

P ositioning Dragonfly Duo/ Optis Distance from the Distal marker to Lens Marker = 26mm Distance from Lens Marker to the proximal Marker= 50mm Frame the specific region of interest in the High-resolution Mode Distance from lens to Lens marker is = 2 mm Proximal and Lens markers are on the torque wire and then the markers move during the pullback Markers 26 mm apart Pullback length: 54mm Or 75mm Lens 2 mm Distal Marker Lens Marker Proximal Marker Markers 50 mm apart POI 54MM POI 75MM Insertion and positioning of the Dragonfly catheter should be done in the “Stand by” Mode CONFIDENTIAL – FOR INTERNAL USE ONLY

Pullback Prep: P urging Blood Purged If blood creeps into catheter lumen, purge with attached contrast syringe. CONFIDENTIAL – FOR INTERNAL USE ONLY To purge the catheter the system must be in Standby mode

Pullback Prep: P uff Marginal, Blood Swirls Optimal Clearance During live scan, use a puff of contrast to evaluate clarity. CONFIDENTIAL – FOR INTERNAL USE ONLY

Blood Clearance CONFIDENTIAL – FOR INTERNAL USE ONLY

RBC mixed into flush fluid Flush contrast diluted with saline, less viscous, does not flush all RBC Data on file at LLI Blood Speckle

Turbulent flow between flush and blood Flush not filling lumen, perhaps going into other arteries OR End of flush bolus Blood swirls CONFIDENTIAL – FOR INTERNAL USE ONLY

P ullback : Automatic trigger When the Trigger Type is set to Automatic Press Live View, Press the “Enable Pullback” button to allow the system to detect initiation of the imaging flush. Inject Flush Now.” The imaging pullback will start automatically when clear image frames are detected. Note that the Enabled state lasts for 15 seconds. CONFIDENTIAL – FOR INTERNAL USE ONLY

P ullback : Manual trigger When the Trigger Type is set to Manual Press Live View, => The Enable Pullback button changes to Start Pullback Inject Flush Media Press start Pullback or F5 or Button on the DOC when blood is cleared from the vessel lumen Note that the Enabled state lasts for 15 seconds. When the Trigger Type is set to Pressure , the DOC motor does not speed up until a signal is received from the pressure transducer. CONFIDENTIAL – FOR INTERNAL USE ONLY

For a New & PROPER acquisition FOLLOW The 5 P’s P REPARATION of the system & catheter P OSITION the catheter relative to target lesion/stent P URGE blood from catheter lumen, if present P UFF of contrast to evaluate clearance P ULLBACK! Image acquisition CONFIDENTIAL – FOR INTERNAL USE ONLY

OCT IMAGE INTERPRETATION

TIPS for Image interpretation BACTE-RIA B - Backscatter A - Attenuation C - Composition T - Texture E - Edge RIA July 8, 2019 Enter presentation title via "insert>header and footer>footer" | 74

OCT Image Interpretation Terminology Backscatter The reflection of light waves off the tissue and back to the Dragonfly catheter High backscatter means a brighter pixel Also described as a “signal rich” region Low backscatter means a darker pixel Also described as a “signal poor” region Attenuation The reduction in intensity of the light waves as they pass through tissue due to absorption or scattering High attenuation means the light cannot penetrate very deep Low attenuation means the light can pass through to allow visualization of deeper tissue

OCT Image Interpretation Terminology Composition Homogeneous Uniform in structure ( Uniform or similar pixels) Heterogeneous Structure consists of dissimilar elements ( Non-uniform or dissimilar pixels) Texture Coarse Fine 76

OCT Image Interpretation Terminology Edge/Border The creation of a border is due to the interface between different tissue types One of the parameters used to differentiate plaque types Calcium Lipid

Image Orientation Normal coronary artery Uniform silhouette 3 layers visible in vessel wall Data on file at LLI Imaging catheter Guidewire shadow Adventitia Media Intima 78

Fibrous Plaque 79 Homogeneous High backscatter signal rich brighter pixel Finely textured Low attenuation deeper tissue can be visualized

Fibrous Rich Plaque, Homogeneous EEM 80 Not to be distributed or reproduced

Calcified Plaque Sharp edges Heterogeneous Low backscatter signal poor Low attenuation deeper tissue can be visualized 81 Data on file at LLI

Calcium Fibrous 82 Not to be distributed or reproduced Calcium and Fibrous Plaque

Lipid Plaque High attenuation low tissue penetration Diffuse shadowy edges High backscatter on surface Low backscatter deeper 83 Data on file at LLI

Systematic Approach for OCT image interpretation / Major Plaque Characteristics (Signal Rich /poor) (Backscatter high or low ) Bright / Dark Edge Sharp v s diffuse Attenuation (Absorption) low vs high Homogeneous / Heterogeneous Yes or no

Systematic approach for OCT Image interpretation Lesion Signal /Backscatter Brighter Signal Backscatter Darker Attenuation (Absorption) High (signal poor) Low Sharp edge Calcium Diffuse edge Heterogeneous Homogeneous Lipid Attenuation (Absorption) Low Homogeneous Fibrous Gonzalo N. 2010 ”Optical Coherence Tomographty for the Assessment of Coronary Atherosclerosis and Vessel Response after Stent implantation”. (Thesis)

Plaque Types - Recognition Fibrous Bright pixels Finely textured Deep penetration Homogeneous Lipid Dark pixels Diffuse edge Low penetration Homogeneous Calcium Dark pixels Sharp edge Deep penetration Heterogeneous Fibrous Lipid Calcium 86 Gonzalo N. 2010 ”Optical Coherence Tomographty for the Assessment of Coronary Atherosclerosis and Vessel Response after Stent implantation”. (Thesis)

PCI Follow up Bioabsorbable vascular scaffold Strut appearance is translucent Metal stents Struts are opaque Bright reflection on the surface of the stent strut Struts cast a shadow into the vessel wall Data on file at LLI Images: European Heart Journal (2011):32;294–304

PCI Follow up

PCI Follow up

PCI follow up Vessel Wall Damage Tissue prolapse Convex shaped protrusion of tissue between adjacent stent struts towards the lumen without disruption of the continuity of the luminal vessel surface Data on file at LLI Image: Image: Gonzalo N. Optical Coherence Tomography for the Assessment of Coronary Atherosclerosis and Vessel Response After Stent Implantation (Thesis) 2010

Stent Strut Apposition Strut malapposition Undersized Stent

PCI Follow up Stent Strut Coverage 2 months 7 months Data on file at LLI Images: Dr. Suzuki, Toyohashi Heart Center Data on file at LLI

PCI Follow up Neointimal Hyperplasia In-stent restenosis Thick layer between stent struts and lumen Data on file at LLI Stent struts

In-Stent Restenosis 94 Not to be distributed or reproduced

PCI follow up Thrombosis Data on file at LLI Thrombus Stent struts Data on file at LLI

PCI Follow up Thrombus – Red (Acute) Thrombus – red Absorbs near-infrared light High backscatter on surface due to signal attenuation Appears as a bright mass Shadow (cannot see behind it) Data on file at LLI Red thrombus Data on file at LLI

White thrombus PCI Follow up Thrombus – White (Chronic) Thrombus – white High backscatter Low attenuation Can see behind it Data on file at LLI

PCI Follow up Vessel Wall Damage Edge dissection A disruption of the vessel luminal surface in the edge region Easy to interpret using cross-sectional and longitudinal views Data on file at LLI Image: Columbia Presbyterian Hospital

PCI follow up Vessel Wall Damage Intimal tear Clear visualization of even small irregularities Data on file at LLI

WHAT’s NEW in OCT

Automated Measurements Stent Roadmap Apposition Indicator 3D Bifurcation Mode Side-branch detection Carina view Expanded 3D Navigation 3D flythrough view Object segmentation OPTIS™ Stent Optimization Software broadens the clinical utility for OCT to guide daily decision making for complex PCI both pre and post stenting though: What’s New?

New Default with Lumen Profile Automated measurements Close out X added for both Lumen Profile and L-mode TSC also easy to toggle Menu/Settings to change this default

Stent roadmap with optional bookmarks Displays lumen profile markers: distal, proximal landing and MLA on the co-registration angiogram image Optional bookmarks Icon to toggle this view on and off 2.8 mm tolerance ( ±1.4mm on each end) Note: OCT Frame Indicator changes Thinner for less angio -obstruction Length reduced from 3mm to 2mm (±1.0mm on each end)

Stent display for Apposition indicator Displays sent apposition information in four locations Apposition indicator bar L-mode Stent roadmap Cross-sectional view (dots)

Apposition Thresholds: Optional adjustment Measurement is taken from the leading edge of the stent or what is closest to the lenses of the OCT catheter Software identifies struts that are >25% malapposed across a 1mm length Default Settings : >300 microns red >200 microns yellow <200 microns silver Could also take the red down to 200 if user would rather see only two colors Note: this functionality is not available for BVS

3d Bifurcation mODE Side branches are displayed as pink dots on the lumen profile and dotted lines trailing into the L-mode 3D Bifurcation mode will open with the view that matches the closes side branch view of where the OCT frame marker is currently positioned Bifurcation mode is hidden if no side branches are detected that are approximately 1.5mm diameter or greater The software default zoom level for bifurcation mode is optimized for the OCT system not the demo laptop .  You may want to zoom out when you first open the bifurcation mode. If you don't have a TSC, you can hold down the right mouse button and use the zoom bar at the far left of the 3D screen to adjust the zoom

3d nAVIGATION with NEW 3D OPTIONs Menu

3D Object Display View Options: Lumen Mode Tissue + Lumen Mode Stent Only Mode Additional Options: Individually turn up to two guidewire (s) on/off Turn side-branch (es) on or off Note: Guidewire rendering may vary due to shadow, artifact or out of the cut-plane Flythrough *any of the above modes can be viewed in the flythrough view

AptiVue™ Software for OPTIS™ (version E.5.1) Features: Automated Stent Expansion Calculations Under expansion Indicator for BMS/DES 2 Methods – Dual and Tapered Resting Index, Resting Full Cycle Ratio (RFR) 1 Identifies ischemia-causing lesions like FFR without hyperemia Allows pullbacks to assess serial lesions and diffuse disease Personally Identifiable Information (PII) Security Full Encryption of patient data User authentication (login) Current Algorithm Improvements 2 nd Guide wire detection Stent rendering in 3D 1. Svanerud et al. VALI.DATE RFR. EuroIntervention 2018.

Summary: OCT for PCI Optimization Pre- and post-stent assessment Stent size selection and landing zone planning Position, expansion, apposition Intimal tear, dissection, false lumen Tissue prolapse Calcification PCI follow-up Strut coverage Neo-intimal growth Restenosis/thrombosis Data on file at LLI

Optimize your pci with OCT OCT has clear advantages for stent optimization – PCI OPTIMIZATION It’s all about the lumen Fast, easy to use, easy to read 10 times Superior Image Resolution than IVUS SEE BETTER… TREAT BETTER….

112 THANK YOU FOR ATTENDING

OCT IN CATH LAB Pre - PCI OCT Pull Back Check Clearance and then only do Pull Back – PUFF Calibration If ACR –CO Register Morphology – Calcified, Lipid & Fibrotic lesions zones, healthy 3 layered zones (180D) Landing Zone – Distal & Proximal Luminal & EEL Dimensions (Mean Diameter), Length of the lesion Dissection Thrombus presence 113

OCT IN CATH LAB Post - PCI OCT Pull Back Calibration Lumen Profile, Rendered Stent, If ACR – Co Register Lumen Profile - Place the Distal & Proximal Markers in the edge of the stent Check for Edge dissections Check for MLA – Ilumien 3 Crieteria Apposition 114