PACEMAKER BASIC AND TIMING CYCLE .pptx

awakush 1,254 views 140 slides Apr 19, 2023
Slide 1
Slide 1 of 140
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79
Slide 80
80
Slide 81
81
Slide 82
82
Slide 83
83
Slide 84
84
Slide 85
85
Slide 86
86
Slide 87
87
Slide 88
88
Slide 89
89
Slide 90
90
Slide 91
91
Slide 92
92
Slide 93
93
Slide 94
94
Slide 95
95
Slide 96
96
Slide 97
97
Slide 98
98
Slide 99
99
Slide 100
100
Slide 101
101
Slide 102
102
Slide 103
103
Slide 104
104
Slide 105
105
Slide 106
106
Slide 107
107
Slide 108
108
Slide 109
109
Slide 110
110
Slide 111
111
Slide 112
112
Slide 113
113
Slide 114
114
Slide 115
115
Slide 116
116
Slide 117
117
Slide 118
118
Slide 119
119
Slide 120
120
Slide 121
121
Slide 122
122
Slide 123
123
Slide 124
124
Slide 125
125
Slide 126
126
Slide 127
127
Slide 128
128
Slide 129
129
Slide 130
130
Slide 131
131
Slide 132
132
Slide 133
133
Slide 134
134
Slide 135
135
Slide 136
136
Slide 137
137
Slide 138
138
Slide 139
139
Slide 140
140

About This Presentation

PACEMAKER BASIC AND TIMING CYCLE .pptx


Slide Content

PACEMAKER BASICS AND TIMING CYCLE DR Awadhesh sharma Professor Cardiology LPS Institute of cardiology,Kanpur

PACEMAKER CIRCUIT Implantable pulse generator (IPG): Battery Circuitry Connector(s) Leads or wires Cathode (negative electrode) Anode (positive electrode) Body tissue IPG Lead Anode Cathode

Stuart Allen 06 Contains a battery that provides the energy for sending electrical impulses to the heart Houses the circuitry that controls pacemaker operations Circuitry Battery The Pulse Generator

The Pulse Generator Casing (can) Titanium (biocompatible, lightweight, stronger than steel) Connector (header) Leads plug into ports in the clear epoxy header Components Diodes, resistors, oscillator, microchips Battery The largest single component inside the pulse generator Lithium iodide

THE PULSE GENERATOR…… Common battery compositions include: Lithium-Iodine Lithium silver vanadium oxide with carbon monoflouride Starting battery voltage will vary depending on composition Longevity Dependent on impedance and output Commonly ranges from 6-12 years

LEADS

Leads are Insulated Wires Deliver electrical impulses from the pulse generator to the heart Sense cardiac depolarization Lead

Conductor Tip Electrode Insulation Connector Pin Pacing Lead Components Conductor Connector Pin Insulation Electrode

Lead Characterization Position within the heart Endocardial or transvenous leads Epicardial leads Fixation mechanism Active/Screw-in Passive/Tined Shape Straight J-shaped used in the atrium Polarity Unipolar Bipolar Insulator Silicone Polyurethane

Passive fixation The tines become lodged in the trabeculae

ACTIVE FIXATION The helix (or screw) extends into the endocardial tissue Allows for lead positioning anywhere in the heart’s chamber

Transvenous Leads - Fixation Mechanisms

CONDUCTORS - MP-35N – Alloy of nickel , cobalt, chromium, molybdenum advantage – high strength and resistance to corrosion disadvantage – high electrical resistance This is overcome by incorporating low resistance metals – silver ,titanium , platinum . These alloy can be constructed in two ways -

DFT vs DBS

COAXIAL LEAD VS CORADIAL LEADS SYSTEM

LEAD INSULATION - POLYURETHANE – High tensile strength resistance to mechanical abrasion has excellent lubricity with low frictional coefficient Thin layer of insulation can be used so reducing overall lead dimension Main disadvantage – in vivo polymer degradation

Two mechanism – ESC (environmental stress cracking ) at point of mechanical stress , at anchoring sleevs , venous entry point Metal ion oxidation (MIO)- at conductor insulator interface

SILICON INSULATION Completely inert and biostable More flexible – less risk of cardiac perforation Main disadvantage – low tensile strength susceptibility to abrasion and tear So thicker layer of insulation needed which lead to bulky design Cold flow phenomenon

Lead Insulation May Be Silicone or Polyurethane Advantages of Silicone-Insulated Leads Inert Biocompatible Biostable Repairable with medical adhesive Historically very reliable Advantages of Polyurethane-Insulated Leads Biocompatible High tear strength Low friction coefficient Smaller lead diameter

Epicardial Leads Leads applied directly to the surface of the heart Utilized in pediatric patients and patients contraindicated for transvenous leads Fixation mechanisms include: Epicardial stab-in Myocardial screw-in Suture-on Applied via sternotomy, thoroscopy, or limited thoracotomy

Flows through the tip electrode (cathode) Stimulates the heart Returns through body fluid and tissue to the PG (anode) A UNIPOLAR PACING SYSTEM CONTAINS A LEAD WITH AN ELECTRODE IN THE HEART Cathode Anode - +

Anode Flows through the tip electrode located at the end of the lead wire Stimulates the heart Returns to the ring electrode above the lead tip A BIPOLAR PACING SYSTEM CONTAINS A LEAD WITH 2 ELECTRODES IN THE HEART Cathode

Unipolar leads Unipolar leads have a smaller diameter than bipolar leads Unipolar leads exhibit larger pacing artifacts on the surface ECG One electrode on the tip & one conductor coil Conductor coil may consist of multiple strands - ( multifilar leads)

Bipolar leads Bipolar leads are less susceptible to oversensing noncardiac signals (myopotentials and EMI) Coaxial Lead Design Circuit is tip electrode to ring electrode Two conductor coils (one inside the other) Inner layer of insulation Bipolar leads are typically thicker than unipolar leads

Unipolar Bipolar Advantages Smaller diameter Easier to implant Large spike No pocket stimulation Less susceptible to EMI Programming flexibility Disadvantages Pocket stimulation Far-field oversensing No programming flexibility Larger diameter Stiffer lead body Smal l spike Higher impedance Voltage threshold is 30% higher

CHARACTERISTICS OF AN PACEMAKER CIRCUIT: Voltage Current Impedance 33

VOLTAGE Voltage is the force, or “push,” that causes electrons to move through a circuit In a pacing system, voltage is: Measured in volts (V) Represented by the letter “V” Provided by the pacemaker battery Often referred to as amplitude or pulse amplitude

CURRENT The flow of electrons in a completed circuit In a pacing system, current is: Measured in milliamps (mA) Represented by the letter “I” Determined by the amount of electrons that move through a circuit

IMPEDANCE The opposition to current flow In a pacing system, impedance is: Measured in ohms ( W) Represented by the letter “R” The measurement of the sum of all resistance to the flow of current Impedance is a function of the characteristics of the conductor (wire), the electrode (tip), and the myocardium

Impedance is the sum of all resistance to the flow of current. The resistive factors to a pacing system include: Lead conductor resistance The resistance to current flow from the electrode to the myocardium POLARIZATION IMPEDANCE, which is the accumulation of charges of opposite polarity in the myocardium at the electrode-tissue interface.

Impedance Pacing lead impedance typically stated in broad ranges, i.e. 300 to 1500 Ω Factors that can influence impedance Resistance of the conductor coils Tissue between anode and cathode The electrode/myocardial interface Size of the electrode’s surface area Size and shape of the tip electrode

Voltage, Current, and Impedance are Interdependent The interrelationship of the three components is analogous to the flow of water through a hose Voltage represents the force with which . . . Current (water) is delivered through . . . A hose, where each component represents the total impedance: The nozzle, representing the electrode The tubing, representing the lead wire

Voltage and Current Flow Electrical Analogies Spigot (voltage) turned up, lots of water flows (high current drain) Spigot (voltage) turned low, little flow (low current drain) Water pressure in system is analogous to voltage – providing the force to move the current

Resistance and Current Flow Electrical Analogies Normal resistance – friction caused by the hose and nozzle More water discharges, but is all of it going to the nozzle? High resistance – a knot results in low total current flow Low resistance – leaks in the hose reduce the resistance

POLARIZATION After an output pulse, positively charged particles gather near the electrode. The amount of positive charge is Directly proportional to pulse duration Inversely proportional to the functional electrode size (i.e. smaller electrodes offer higher polarization) Polarization effect can represent 30–40% of the total pacing impedance As high as 70% for smooth surface, small surface area electrodes

Within the electrode, current flow is due to movement of electrons (e−). At the electrode–tissue interface, the current flow becomes ionic & (-) vely charged ions ( Cl −, OH−) flow into the tissues toward the anode leaving behind oppositely charged particles attracted by the emerging electrons. It is this capacitance effect at the electrode tissue interface, that is the basis of polarization

Lead wire fracture Increased resistance High Impedance Conditions A Fractured Conductor A fractured wire can cause Impedance values to rise Current flow from the battery may be too low to be effective Impedance values may exceed 3,000 W Other reason for high impedance: Lead not seated properly in pacemaker header (usually an acute problem).

Low Impedance Conditions An Insulation Break Insulation breaks can cause impedance values to fall Current drain is high and can lead to more rapid battery depletion Current can drain through the insulation break into the body or other lead wire, not through myocardium Impedance values may be less than 300 W

Ohm’s Law Describes the relationship between voltage, current, and resistance V = I X R I = V / R R = V / I V I R V R I V R I R V I = = = X

Ohm’s law tells us: If the impedance remains constant, and the voltage decreases, the current decreases. If the voltage is constant, and the impedance decreases, the current increases.

Stimulation Threshold Pacing Voltage Threshold The minimum pacing voltage at any given pulse width required to consistently stimulate the heart outside the myocardial refractory period causing it to contract

E = Energy delivered by Pulse to the Pacing Circuit and Cardiac Tissue - V . I . t I 2 Rt V 2 t/R

Rheobase - ( the lowest point on the curve) by definition is the lowest voltage that results in myocardial depolarization at infinitely long pulse duration Chronaxie ( pulse duration time ) by definition, the chronaxie is the threshold pulse duration at twice the rheobase voltage

Strength interval curve

Effect of Lead Design on Capture Lead maturation Fibrotic “capsule” develops around the electrode following lead implantation May gradually raise threshold Usually no measurable effect on impedance

STEROID ELUTING LEADS Steroid eluting leads reduce the inflammatory process Exhibit little to no acute stimulation threshold peaking Leads maintain low chronic thresholds

Effect of steroid on stimulation thresholds

Evolution of Pacing Threshold Voltage Threshold (V) Observation Time (weeks) Acute Phase Chronic Phase Safety Margin 6 s 4 3 2 1 4 8 12 16

INJURY CURRENT……. Due to pressure exerted by electrode on myocardium Seen in both active and passive leads(less pronounced) Lack of injury current – denotes poor contact / scarred myocardium Disappears over minutes to hours this may leads to high threshold soon after lead placement which subside at end of procedure

Magnet mode Activates magnetic reed switch Model dependent behavior with magnet Asynchronous pacing – most common No apparent rate or rhythm change Brief asynchronous pacing and then return to programmed value Continuous or transient loss of pacing Magnets will cause most DDDs to convert to DOO at about 85 with a BOL (beginning of life) battery, and to VOO at a rate of about 65 at ERI (effective replacement interval) is reached to conserve power.

The Pacing Pulse t Pacing Pulse Pulse Duration (Width) Output Voltage V = Pulse Amplitude in Volts (V) (say 2.5 V) t = Pulse Duration or Width in milliseconds ( ms ) (say 0.5 ms ) R = Impedance of Pacing Circuit (ohms) (say 500 ohms) I = V/R = Current through pacing circuit (mA) = 2.5 V/ 500 ohms = 0.005 A = 5 mA V t

Why measure stimulation threshold? To enable programming stimulus voltage amplitude and pulse width such that Consistent capture & Patient Safety is ensured Battery drain minimized, Pacemaker longevity maximized Good thresholds Ventricle - <1V @ 0.5ms Atrium - <1.5V @ 0.5ms

SENSING OF INTRINSIC HEARTBEATS Sensing is the ability of the pacemaker to “see” when an intrinsic depolarization is occurring Pacemakers record the Intracardiac Electrogram (EGM) by constantly recording the potential difference between the cathode and anode Depolarization Wave Processed by Device

Sensitivity Setting Sensitivity settings less than 2.5 mv – High sensitivity – can lead to oversensing Sensitivity settings greater than 2.5 mV – Low sensitivity – can lead to undersensing Amplitude (mV) Amplitude (mV) Time Time 5.0 2.5 1.25 5.0 2.5 1.25

Undersensing . . . Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately Intrinsic beat not sensed Scheduled pace delivered VVI / 60

Oversensing An electrical signal other than the intended P or R wave is detected Pacing is inhibited Marker channel shows intrinsic activity... ...though no activity is present

FUSION AND PSEUDOFUSION COMPLEX

TIMING CYCLE

VVI/60(1000msec) ? Last paced event occur later than expected

HYSTERESIS…… Goal is to encourage own intrinsic activity . One of the cause of rate variation in paced ECG Hysteresis work best in patient with intrinsic rate near the programmed base rate.

V VP VP VP VOO TIMING VP VP

V VP VP VP VP VS VVI TIMING

PACEMAKER MEDIATED TACHYCARDIA Main trigger are – PVC loss of atrial capture

ECG- pacing at MTR no evidence of atrial capture retrograde P wave in inferior leads