Plates and screws 11

47,957 views 96 slides May 11, 2015
Slide 1
Slide 1 of 96
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79
Slide 80
80
Slide 81
81
Slide 82
82
Slide 83
83
Slide 84
84
Slide 85
85
Slide 86
86
Slide 87
87
Slide 88
88
Slide 89
89
Slide 90
90
Slide 91
91
Slide 92
92
Slide 93
93
Slide 94
94
Slide 95
95
Slide 96
96

About This Presentation

easy for orthopedic resident to understand


Slide Content

Plates and Screws Supervised by Dr. Marwan Abuhashem By Dr. Ismael Al- jabiri Al- Bashir Orthopedic Department Ministry Of Health Jordan

1 . Anatomical Reduction. 2. Stable internal fixation. 3. Preservation of Blood supply 4. Early mobilization. Principles of Fixation :

3 “ Surfaces of the fracture do not displace under functional load” Can only be achieved by interfragmentary compression Absolute stability A plate by itself rarely provides absolute stability The key tool of absolute stability is the lag screw Compression must sufficiently neutralize all forces[bending, tension, shear and rotation]

Relative Stability:  A fixation device that allows small amounts of motion in proportion to the load applied. The deformation or displacement is inversely proportional to the stiffness of the implant. Examples: Intramedullary rod, bridge plating, external fixation.

PLATES Introduction : Bone plates are like internal splints holding together the fractured ends of a bone. A bone plate has two mechanical functions . It transmits forces from one end of a bone to the other, bypassing and thus protecting the area of fractures. It also holds the fracture ends together while maintaining the proper alignment of the fragments throughout the healing process.

History Hansman’s Bone Plate (1886) Hansmann’s plates were: Bent at the end to protude through the skin Attched to bone by screw with long shanks that projected outside the soft tissues.

Bone Suture Stabilization Inserts (Koenig, 1905)

History Since 1958, AO has devised a family of plates for long bone fractures, starting with a round holed plate. In 1969 the Dynamic Compression Plate was developed. In 1994 LC DCP was created. In 2011 LCP with combination holes has come into use.

Names of plates. Shape (Semitubular, 1/3 rd tubular) Width of plate (Small, Narrow, Broad) Shape of screw holes. (Round, Oval) Surface contact characteristics. (LC, PC) Intended site of application (Condylar Plate) According to the function

Standard Plates Narrow DCP-4.5 mm Broad DCP – 4.5 mm 3.5 mm DCP

LC-DCP 3.5 & 4.5mm Reconstruction plate 3.5 & 4.5mm

1/3 tubular plate 2.7, 3.5 & 4.5 mm

Special Plates T Plates T&L Buttress plates

Lateral proximal Tibial buttress plates Condylar buttress plate Narrow lengthening plates

Broad Lengthening plate Spoon plate Clover leaf plate

DCP - 3.5 and 4.5 First introduced in 1969 by Danis Revolutionary concept of compression plating Featured a new hole designed for axial compression Broad 4.5 for Femur & Narrow 4.5 for Humerus & Tibia DCP 3.5 for Forearm, Fibula, Pelvis & Clavicle

Advantage of DCP : 1. Inclined insertion 25°longitudinal and 7° sideways 2. Placement of a screw in neutral position without the danger of distraction of fragments 3. Insertion of a lag screw for the compression 4. Usage of two lag screws in the main fragments for axial compression 5. Compression of several fragments individually in comminuted fractures 6. Application as a buttress plate in articular area

Problems with DCP Unstable fixation leads to fatigue & failure Strict adherence to principles of compression Compromised blood supply due to intimate contact with underlying cortex “Refractures” after plate removal

LC-DCP Represents a design change Overcome problems with DCP Plate footprint reduced Minimized kinking at screw holes, more countourable , reduced plate fatigue at hole Allows more inclination of screw in longitudinal plane and transverse plane.

In the DCP (A), the area at the plate holes is less stiff than the area between them so while bending, the plate tends to bend only in the areas of the hole. The LC-DCP(B) has an even stiffness without the risk of buckling at the screw holes.

The LC-DCP offers additional advantage Improve blood circulation by minimizing plate-bone contact More evenly distribution of stiffness through the plate Allows small bone bridge beneath the plate

Tubular plates 3.5 system - 1/3 rd Tubular 4.5 system - Semitubular Limited stability -Oval holes – Axial compression can be achieved. -Low rigidity (1mm thick). Lateral malleolus Distal ulna / Olecranon Distal humerus

limited stability. The thin design allows for easy shaping and is primarily used on the lateral malleolus and distal ulna. The oval holes allow for limited fracture compression with eccentric screw placement .

Reconstruction plates Deep notches between holes Accurate contouring in any plane Pelvis Acetabulum Distal humerus Clavicle Olecranon

Reconstruction plates are thicker than third tubular plates but not quite as thick as dynamic compression plates. Designed with deep notches between the holes, they can be contoured in 3 planes to fit complex surfaces, as around the pelvis and acetabulum . Reconstruction plates are provided in straight and slightly thicker and stiffer precurved lengths. As with tubular plates, they have oval screw holes, allowing potential for limited compression.

LCP – Locking Compression Plate

LCP Latest in the evolution “ Internal fixator ” Combination of locking screw with conventional screw Extraperiosteal location of plate

LCP: internal external fixator

LCP Combines advantages of DCP principle and locking head principle. Flexibility of choice within a single implant. Screw hole have been specially designed to accept either: cortical screw and locking screw

The locking screws, by achieving angular stability within the plate holes are able to produce a similar hoop with just two unicortical screws. LCP

LOCKING COMPRESSION PLATE ( LCP) Principle : Angular-stability whereas stability of conventional plates is friction between the plate and bone Screw locking principle Provides the relative stability  Healing by callus formation (Secondary Healing)

Stability under load By locking the screws to the plate, the axial force is transmitted over the length of the plate secondary loss of the intraoperative reduction is reduced Blood supply to the bone No additional compression after locking Periosteal blood supply will be preserved

LCP used as internal fixator to bridge multifragmentary diaphyseal fracture zone. Locking compression plate is used. Standard cortical and cancellous screws are used as a traditional plate. LCP

Principle of internal fixation using LCP : 1. 1 st reduced the # as anatomical as possible 2. Cortical screw should be used 1 st in a fracture fragment 3. If the locking screw have been put, use of the cortical screw in the same fragment without loosening and retightening of the locking screw is not recommended 4. If locking screw is used first avoid spinning of plate 5. Unicortical screws causes no loss of stability

6. In Osteoporotic bone bicortical screws should be used. 7. In comminuted # screw holes close to the fracture should be used to reduce strain. 8. In the fracture with small or no gap the immediate screw holes should be left unfilled to reduced the strain. Principle of internal fixation using LCP :

Indications : Osteoporotic # Periprosthetic # Multifragmentry # Delayed change from external fixation to internal fixation.

Advantages : 1. Angular stability 2. Axial stability 3. Plate contouring not required 4. Less damage to the blood supply of bone 5. Decrease infection because of submuscular technique 6. Less soft tissue damage

Sizes of DCP Name of plate Small Narrow Broad Width 11 mm 13.5 mm 17.5mm Profile 4 mm 5.4 mm 5.4 mm Screw 2.7 , 3.5 cortex screw and 4 mm cancellous screw 4.5 mm cortex screw & 6.5mm canellous screw 4.5 mm cortex screw & 6.5mm canellous screw Sizes of LCDCP Name of plate Small Narrow Broad Width 11 mm 13.5 mm 17.5mm Profile 4 mm 5.4 mm 5.4 mm Screw 2.7 , 3.5 and 4 mm cancellous screw 4.5 mm & 6.5mm canellous screw 4.5 mm & 6.5mm canellous screw Name of plate Small Narrow Broad Width 11 mm 13.5 mm 17.5mm Profile 4 mm 5.0 mm 5.0 mm Screw 4 mm locking screw 5 mm locking screw 5 mm locking screw Sizes of LCP

LISS System Preshaped plates with self drilling self tapping screws with threaded heads. Through a small incision (using this jig ) plate is slid along the bone surface. position of plate and wire are checked radiologically before insertion of metaphyseal screw .

LISS-Less Invasive Stabilization System

LISS

Type of plate – Functional Regardless of their length, thickness, geometry, configuration and types of hole, all plates may be classified in to 4 groups according to their function. Neutralization plate. Compression plates. Buttress plate. Tension band plates.

NEUTRALIZATION PLATE Acts as a ""bridge”” protection No compression at the fracture site neutralization plate is to protect the screw fixation of a short oblique fracture a butterfly fragment a mildly comminuted fracture of a long bone fixation of a segmental bone defect in combination with bone grafting.

The Neutralization Plate Lag screws: compression and initial stability Plate: protects the screws from bending and torsional loads

NEUTRALIZATION PLATE

COMPRESSION PLATE produces a locking force across a fracture site plate is attached to a bone fragment then pulled across the fracture site , producing tension in the plate Direction of the compression force is parallel to the plate

Compression Plate Principle : - a self compression plate due to the special geometry of screw holes which allow the axial compression.

Dynamic compression principle: a The holes of the plate are shaped like an inclined and transverse cylinder. b–c Like a ball, the screw head slides down the inclined cylinder. d–e Due to the shape of the plate hole, the plate is being moved horizontally when the screw is driven home. f The horizontal movement of the head, as it impacts against the angled side of the hole, results in movement of the plate and the fracture fragment already attached to the plate by the first screw (1). This leads to compression of the fracture .

51 Compression plate: eccentric DC (dynamic compression) hole Removable device: compression device Interfragmentary compression by plate

52 External compression device

METHODS OF ACHIEVING COMPRESSION With tension devise By overbending With dynamic compression principle (DCP/LC-DCP) By contouring plate Additional lag screw thro plate

BUTTRESS PLATE is to strengthen (buttress) a weakened area of cortex The plate prevents the bone from collapsing during the healing process. A buttress plate applied a force to the bone which is perpendicular (normal) to the flat surface of the plate.

The fixation to the bone should begin in the middle of the plate, closest to the fracture site on the shaft. The screws should then be applied in an orderly fashion, one after the other, towards both ends of the plate . example : the T-plate used for the fixation of fractures of the distal radius and the tibial plateau. BUTTRESS PLATE

Bridge Plating Bridge Plating for comminuted fracture instead of individually fixing each fragment minimal disruption to blood supply reduction is performed indirectly - compression is only sometimes possible

TENSION BAND PRINCIPLE Tension-band principle.

Tension Band Principle :- Its describes how the tensile forces are converted into compressive forces by applying a devise eccentrically or to the convex side of a curved tube or bone. Indications :- Fracture Patella, olecranon , medial malleolus , greater trochanter of the femur. Static Dynamic

Dynamic and static tension band In dynamic tension band the tensile forces are converted to compression on the convex side of an eccentrically loaded bone Examples : Patella Olecranon Greater tuberosity Tension band principle to the medial malleolus example of static tension band

HOW MANY SCREWS ? Bones No. of Cortices No. of Holes Type of Plate Forearm 5 to 6 Cortex 6 holes Small 3.5 Humerus 7 to 8 Cortex 8 holes Narrow 4.5 Tibia 7 to 8 Cortex 7 holes Narrow 4.5 Femur 7 to 8 Cortex 8 holes Narrow 4.5 Clavicle 5 to 6 Cortex 6 holes` Small 3.5

Timing of Plate Removal Malleolar fractures. 8-12mo The tibial pilon. 12-18mo The tibial shaft. 12-18mo The tibial plateau. 12-18mo The femoral condyles. 12-24mo The femoral shaft. 24-36mo Upper extremity. 12-18mo Shaft of radius / ulna. 24-28mo Distal radius. 8-12mo Metacarpals. 4-6mo

SCREW: INTRODUCTION An elementary machine to change the small applied rotational force into a large compression force Function Holds the plate or other prosthesis to the bone Fixes the # fragments ( Position screw) Achieves compression between the # fragments (Lag screw)

SCREWS 4 functional parts Head Shaft\ Shank \Core Thread Tip

Head: Recess Types 1. Slotted 2. Cruciate 3. Philips 4. Hex/ Allen 5. Torx ( eg Stardrive of Synthes )

Screw: Shaft/ Shank/ Core Smooth link Almost not present in standard cortex screw Present in cortical SHAFT SCREW or cancellous screw

Screw: Run out Transition between shaft and thread Site of most stress riser Screw break Incorrectly centered hole Hole not perpendicular to the plate

Screw: Thread Inclined plane encircling the root Single thread May have two or more sets of threads V-thread profile: more stress at sharp corner Buttress thread profile: less stress at the rounded corner

Core Solid section from which the threads project out wards. The size of core determines the strength of screw and its fatigue resistance. The size of drill bit used is equal to the core diameter.

Screw: Core Diameter Narrowest diameter across the base of threads Also the weakest part Smaller root  shear off Torsional strength varies with the cube of its root diameter

Screw: Thread Diameter Diameter across the maximum thread width Affects the pull out strength Cancellous have larger thread diameter

Screw: Tip Designs 1. Self-tapping tip: Flute Cuts threads in the bone over which screw advances Cutting flutes chisel into the bone and direct the cut chips away from the root

Screw: 2.Non self tapping Lacks flutes Rounded tip Must be pre-cut in the pilot hole by tap Pre-tapped threads help to achieve greater effective torque and thus higher inter-fragmental compression Better purchase

Screw: 3.Corkscrew tip Thread forming tips In Cancellous screws which form own threads by compressing the thin walled trabecular bone Inadequate for cortical bone

Screw: 4.Trochar Tip Like self tapping Displaces the bone as it advances Malleolar screw Schanz screws Locking bolts for IMIL

Screw: 5.Self drilling self tapping Like a drill bit In locked internal fixator plate hole Pre-drilling not required Flute Good purchase in osteoporotic and metaphyseal area

Locking Screws vs Cortical Screws Creates Fixed Angle Generates Friction/Compression 4.4mm Core Dia. 3.5mm Core Dia. 5.0 mm Locking Screw 4.5 mm Cortical Screw

Bending stiffness proportional to the core diameter Pull out strength is proportional to the size of the thread Cannulated screws have less bending stiffness

Biomechanics Locking Screw Design: Shaft Why smaller threads? Not used to generate compression between plate and bone Why larger core diameter? Decreased thread diameter allows increased core diameter Greater bending and shear strength Dissipates load over larger area in bone Cortex Screw Locking Screw

AO/ASIF Screws: Types Cortical Fully threaded Shaft screw 1.5:phalanx *drill bit 1.1 mm 2.7: mc and phalanx *bit:2.0 3.5: Radius/ Ulna/ Fibula/ Clavicle*bit:2.5 4.5: Humerus/ Tibia/ femur *bit:3.2

AO/ASIF Screws Cortical screws: a machine type Smaller threads Lower pitch Large core diameter Smaller pitch higher holding power greater surface area of exposed thread for any given length better hold in cortical bone

AO/ASIF Screws Cancellous screws: a wood type core diameter is less the large threads Higher pitch Greater surface are for purchase Untaped pilot hole Pilot hole equals the core diameter lag effect option with partially threaded screws theoretically allows better fixation in soft cancellous bone.

Cancellous Fully threaded Cannulated or Non- cannulated Partially threaded 16mm or 32 mm Cannulated or Non- cannulated 4.0,  drill bit 2.5mm humeral condyle 6.5  drill bit 3.2mm tibial and femoral condyle

MALLEOLAR SCREW: - smooth shaft - partially threaded - trephine tip : no tapping needed - was designed as lag screw for malleoli fixation NOW small cancellous screws preffered - distal humerus and lesser trochanter - size : 25mm – 75 mm

Cannulated screws 3.0 4.0 4.5 6.5 7.0 7.3

Special Screws Locking bolt Herbert Screw Dynamic Hip Screw Malleolar Screw Interference screw Acutrak screw

Pedicle screw Suture anchor

Headless Screws Herbert screw bridging a scaphoid fracture Acutrak screw

Bioabsorbable Screws The most common materials used are polylactic acid (PLA), poly-L-lactic acid (PLLA), and polyglycolic acid.

Advantages of bioabsorbable screws • Does not interfere with MRI. • Does not interfere with future revision surgery if needed. • Decreased incidence of graft laceration. • Does not need implant removal

Disadvantages of bioabsorbable screws • Major disadvantage is screw failure during insertion. Special screw drivers that span the entire length of screw reduce incidence of screw breakage. • Foreign body reaction may be seen in some.

Function or mechanism. Neutralization screws – neutralizes forces on the plate in plate fixation. Lag screws – For inter-fragmentary compression . Reduction screw – To reduce displaced fracture by pushing or pulling . Position screw – Holds two fragments in position without compression. Eq. Syndesmotic screw Anchor screw – Acts as an anchor for wire or suture. In tension band wiring Locking head screw – In locking plates Locking screw – In interlocking nails Poller screw – To guide the nail path in interlocking nailing of fractures close to the bone ends.

TAP To cut threads in bone of same size as the screw to facilitate insertion Flutes : to clear the bone debris Two turns forward and half turn backward recommended to clear debris Used with sleeve Done manually Power tapping NOT recommended For cancellous bone : short and wide thread , slightly smaller dia than screw

For cortical screws : - as fixation screw : both cortices - as lag screw : only far cortex For cancellous screw: - only near cortex - sometimes in young patients tapping entire screw length needed

LAG SCREW TECHNIQUE : - to achieve interfragmentary compression - this technique is used if a screw is to be inserted across a # , even through a plate. - screw has no purchase in near fragment, thread grips the far fragment only -achieved either with screw with shaft or fully threaded screw

Positioning of screws: -max. interfragmentary compression : placed in middle of fragment, right angle to fracture plane - max. axial stability: right angle to long axis of bone

Thank You