Presentation1, basic principle of ultrasound.

57,420 views 50 slides May 29, 2019
Slide 1
Slide 1 of 50
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50

About This Presentation

health&Medicine


Slide Content

DR/ABD ALLAH NAZEER. MD. Basic Principles of Ultrasound.

Ultrasound Definition Physics. Transducer. Artifacts. Terminology.

Sound is a mechanical, longitudinal wave that travels in a straight line and measured by cycle per second by Hertz(Hz) unit. Audible Sound 20-20,000 Hz. Sound requires a medium through which to travel. Ultrasound is a mechanical, longitudinal pressure wave with a frequency exceeding the upper limit of human hearing, which is 20,000 Hz or 20 kHz. Medical Ultrasound is > 2MHz,(2MHz to 16MHz).

Basic Ultrasound Physics Amplitude oscillations/sec = frequency - expressed in Hertz (Hz). Wave length &frequency

ULTRASOUND – How is it produced? Produced by passing an electrical current through a piezoelectrical (material that expands and contracts with current) crystal.

Ultrasound Production. Transducer produces ultrasound pulses (transmit 1% of the time) These elements convert electrical energy into a mechanical ultrasound wave. Reflected echoes return to the scan head which converts the ultrasound wave into an electrical signal

Frequency vs. Resolution The frequency also affects the QUALITY of the ultrasound image The HIGHER the frequency, the BETTER the resolution The LOWER the frequency, the LESS the resolution A 12 MHz transducer has very good resolution , but cannot penetrate very deep into the body A 3 MHz transducer can penetrate deep into the body, but the resolution is not as good as the 12 MHz Low Frequency 3MHz. High frequency 12 MHz.

Image Formation. Electrical signal produces ‘dots’ on the screen. Brightness of the dots is proportional to the strength of the returning echoes Location of the dots is determined by travel time. The velocity in tissue is assumed constant at 1540m/sec Distance = Velocity Time

Interactions of Ultrasound with Tissue. Acoustic impedance (AI) is dependent on the density of the material in which sound is propagated - the greater the impedance the denser the material. Reflections comes from the interface of different AI’s greater  of the AI = more signal reflected works both ways (send and receive directions)

Acoustic impedance(Z). Is the resistance to the propagation of sound and this depend on density and velocity of sound and measured by MegaRayls (Z). Air = 0.0004 Z.(Low acoustic impedance). Bone = 7.8 Z.(High acoustic impedance). Adipose = 1.34 Z. Liver = 1.65 Z.

Interactions of Ultrasound with Tissue Reflection. Refraction. Transmission. Attenuation.

Reflection The ultrasound reflects off tissue and returns to the transducer, the amount of reflection depends on differences in acoustic impedance . The ultrasound image is formed from reflected echoes.

Refraction

Transmission Some of the ultrasound waves continue deeper into the body. These waves will reflect from deeper tissue structures.

Attenuation Defined - the deeper the wave travels in the body, the weaker it becomes -3 processes: reflection, absorption, refraction Air (lung)> bone > muscle > soft tissue >blood > water. Speed of sound in Air = 330 m/s, Bone =4030 m/s, Tissues =1040m/s.

Machines and Transducers.

Ultrasound frequencies in diagnostic radiology range from 2 MHz to approximately 15 MHz. It is important to remember that higher frequencies of ultrasound have shorter wavelengths and are absorbed/attenuated more easily. Therefore, higher frequencies are not as penetrating . This explains why high frequencies are used for the superficial body structures and low frequencies are used for those that are deeper. The following frequencies are a guide to frequencies typically used for ultrasound examination: 2.5 MHz: deep abdomen, obstetric and gynecological imaging 3.5 MHz: general abdomen, obstetric and gynecological imaging 5.0 MHz: vascular, breast, pelvic imaging 7.5 MHz : breast, thyroid 10.0 MHz: breast, thyroid, superficial veins, superficial masses, musculoskeletal imaging. 15.0 MHz: superficial structures, musculoskeletal imaging.

Artifacts.

Orientation Longitudinal Plane. Transverse Plane.

  Longitudinal Plane.

  Terminology. Echogenicity. Hypo-Hyper-Iso-An Echoic The circle is Hyperechoic to the surrounding tissue. The circle is Hypoechoic to the surrounding tissue.

The circle is Isoechoic to the surrounding tissue. The circle is An Echoic to the surrounding tissue.

Multiple Echogenicity.

Thank You.