Troubleshooting on mechanical ventilator
Alarm settings n mechanical ventilator
High peak pressure alarm on ventilator
mechanical ventilator basic
Size: 870.49 KB
Language: en
Added: Dec 30, 2019
Slides: 32 pages
Slide Content
Troubleshooting
On Mechanical Ventilator
Precautions that would reduce
troubles
I. Power:
Plug into a grounded AC power with
correct voltage receptacle.
Secure the power cord properly.
Battery Back up:
Check the battery level before connecting.
Charging should be carried out regularly.
Remember it is for short term use.
II. Gas Source
Preferable to have centralised supply.
If cylinders used, should be full
Spare cylinders should be available
III. Personnel
Properly trained personnel should only use.
Familiarizing staff with operator’s manual before using on
a patient.
(One manufacturer’s manual may not exactly match with
other brands).
Appropriate monitoring the functioning state of the
ventilator while in use. Familiarizing staff with alarm
system.
Do not place ventilators in a combustible or explosive
environment.
IV. Servicing and Testing
Qualified personnel should undertake servicing.
Follow specifications mentioned in service
manual
General servicing at regular intervals
Run prescribed tests and calibrations before using
ventilator on a patient.
Ensure that ventilators pass all the tests before
putting them in to clinical use.
ALARMS
All ventilators are equipped with visual
and audible alarms which notify user
problems.
Points to remember
Never ignore an alarm.
Find out for yourself what alarm is on.
Check the patient.
Silence the alarm.
Never mute the alarm on regular
basis.
Do not be like this !
But hear the alarm and respond
See the problem and
Ask if you do not know what to do
Troubleshooting
•Is it working ?
–Look at the patient !!
–Listen to the patient !!
–Pulse Ox, ABG, EtCO
2
–Chest X ray
–Look at the vent (PIP; expired TV; alarms)
CASE
•18 yr old man intubated for
organophosphorus poisoning and
intermediate syndrome was stable for 3 days
•He suddenly devlops distress.You notice that
his resp rate is 35/min, heart rate is 120/min,
BP is 90/70mmHg.
•What would you do?
Troubleshooting
•When in doubt, DISCONNECT THE PATIENT
FROM THE VENT, and begin bag ventilation.
•Ensure you are bagging with 100% O2.
•This eliminates the vent circuit as the source
of the problem.
•Bagging by hand can also help you gauge
patient’s compliance
•Ptis on VCV mode of ventilation, post op after
lap choleycystectomy. Suddenly you heard a high
peak pressure alarm. What will you do?
Ask sister to silence the alarm
Go yourself and silence the alarm
Change peak pressure alarm setting
Keep let it be ring
Sedate the pt
Give vecuronium
•Ptis on VCV mode of ventilation, post op after
lap choleycystectomy. Suddenly you heard a high
peak pressure alarm. What will you do?
Ask sister to silence the alarm
Go yourself and silence the alarm
Change peak pressure alarm setting
Keep let it be ring
Sedate the pt
Give vecuronium
Low pressure or Low min. Vent
Solution
Evaluate cuff pressure at regular intervals
Reinflateif leak
Ruptured is noticed change ET tube
Check circuit, junctionstighten or replace
Check water traps
Check ET tube placement Position it properly
Reconnect ventilator
Patient may require higher flow.
High Pressure Alarm
The measured peak inspiratory pressure is
great than set level because of Increased
airway resistance or decreased compliance
•If your Pplat is high, you are faced with a
COMPLIANCE problem
•If your Pplat is N, you are faced with a
RESISTIVE problem
•DD?
Begin Expiration
P
aw
(cm H
2
O)
Time (sec)
Begin Inspiration
PIP
P
plateau
(Palveolar)
Transairway Pressure (P
TA)}
Exhalation Valve Opens
Expiration
Inspiratory Pause
Water in the tube
Filter block
Secretions in airway
Partial / complete block –(ETt)
Kinking of tube
Biting the tube
Rt. Sided intubation
Fighting with ventilator
Cough
Bronchospasm
High Peak Pressure Alarm
(Resistance)
Approach to high pressure
•Ppeak is up
–Look at your Pplat
Solution
Suctioning, Irrigation
Release tubing
Bite block insertion
Empty the tubingsand water traps
Reposition ET tube
Reposition patient
Re assurance
Sedation & medication (pain)
•Pain
•Anxiety
•Metabolic acidosis
•Hypoxia
•Neurogenic hyperventilation
•Excessive trigger
•Water in tubing
High MV alarm
Solution
Check the patientArouse if needed
Activate back up facility if it was not
done already.
Consider switching over to any
mandatory mode
Set trigger level appropriately
Apnoea
No breath was delivered for the operator set
apnoeatime in spont, SIMV, AC, CMV & NIV
modes
patient effort is too minimal
Trigger level set improperly.