Unit 4 biomedical

AnuAntony19 9,027 views 109 slides Aug 22, 2016
Slide 1
Slide 1 of 109
Slide 1
1
Slide 2
2
Slide 3
3
Slide 4
4
Slide 5
5
Slide 6
6
Slide 7
7
Slide 8
8
Slide 9
9
Slide 10
10
Slide 11
11
Slide 12
12
Slide 13
13
Slide 14
14
Slide 15
15
Slide 16
16
Slide 17
17
Slide 18
18
Slide 19
19
Slide 20
20
Slide 21
21
Slide 22
22
Slide 23
23
Slide 24
24
Slide 25
25
Slide 26
26
Slide 27
27
Slide 28
28
Slide 29
29
Slide 30
30
Slide 31
31
Slide 32
32
Slide 33
33
Slide 34
34
Slide 35
35
Slide 36
36
Slide 37
37
Slide 38
38
Slide 39
39
Slide 40
40
Slide 41
41
Slide 42
42
Slide 43
43
Slide 44
44
Slide 45
45
Slide 46
46
Slide 47
47
Slide 48
48
Slide 49
49
Slide 50
50
Slide 51
51
Slide 52
52
Slide 53
53
Slide 54
54
Slide 55
55
Slide 56
56
Slide 57
57
Slide 58
58
Slide 59
59
Slide 60
60
Slide 61
61
Slide 62
62
Slide 63
63
Slide 64
64
Slide 65
65
Slide 66
66
Slide 67
67
Slide 68
68
Slide 69
69
Slide 70
70
Slide 71
71
Slide 72
72
Slide 73
73
Slide 74
74
Slide 75
75
Slide 76
76
Slide 77
77
Slide 78
78
Slide 79
79
Slide 80
80
Slide 81
81
Slide 82
82
Slide 83
83
Slide 84
84
Slide 85
85
Slide 86
86
Slide 87
87
Slide 88
88
Slide 89
89
Slide 90
90
Slide 91
91
Slide 92
92
Slide 93
93
Slide 94
94
Slide 95
95
Slide 96
96
Slide 97
97
Slide 98
98
Slide 99
99
Slide 100
100
Slide 101
101
Slide 102
102
Slide 103
103
Slide 104
104
Slide 105
105
Slide 106
106
Slide 107
107
Slide 108
108
Slide 109
109

About This Presentation

5th semester subject - medical electronics


Slide Content

DEEPAK.P
UNIT 5
Instrumentation for clinical
laboratory
1

DEEPAK.P2
Respiratory Transducers and
Instruments

Respiration
It is responsible for bringing oxygen in to the body and
discharging waste particles from body.
3 DEEPAK.P

Respiratory System Measurements
1.Vital capacity
2.Functional residual capacity
3.Inspiratory capacity
4.Total Lung capacity
4 DEEPAK.P

Respiratory System Measurements
Tidal volume = volume of air inhaled during normal quiet
breathing.
Residual volume (RV) = volume of air in the lungs after
maximal expiration.
Functional residual capacity (FRC) = volume of air in the
lungs after exhalation during quiet breathing.
Inspiratory capacity = volume of air from FRC to maximal
inhalation.
Vital lung capacity from RV to maximal inhalation.
5 DEEPAK.P

Respiratory System Measurements
1.Pneumograph is used to detect respiration.
2.Spiro meters are used for the following measurements.
1.Tidal Volume (TV)
2.Inspiratory Reserve Volume (IRV)
3.Expiratory reserve volume (ERV)
4.Residual Volume (RV)
5.Minute Volume
3.Pulmonary Capacities are also calculated by respiratory
machine.
6 DEEPAK.P

DEEPAK.P7
Pneumograph

Pneumograph
A Pneumograph, also known as a pneumatograph or
spirograph, is a device for recording velocity and force of
chest movements during respiration.
There are various kinds of pneumographic devices, which have
different principles of operation.
1.In one mechanism, a flexible rubber vessel is attached to the
chest and the vessel is equipped with sensors.
2.Others are impedance based.
•In first case a very thin elastic tube filled with mercury is
stretched across patient chest.
8 DEEPAK.P

Pneumograph
Impedance pneumography is a commonly-used technique
to monitor a person ’s respiration rate, or breathing rate.
The idea behind the Impedance pneumograph is that, the AC
impedance across the chest of a subject changes as
respiration occurs.
It is mainly used in neonatal respiration monitors.
9 DEEPAK.P

Pneumograph
In second case it is implemented by either using two
electrodes or Four electrodes
The objective of this technique is to measure changes in the
electrical Impedance of the person ’s thorax caused by
respiration or breathing.
10 DEEPAK.P

Impedance Pneumograph
11 DEEPAK.P

Impedance Pneumograph Eqlt. Ckt.
12 DEEPAK.P

Pneumograph
•There are two type of pneumograph that uses piezoresistive
strain gauge transducers.
1.Mercury strainguage
2.Same wire foil or semiconductor piezoresistive devices.
•Thermistors are used as flow detectors in some
pneumographs.
•In one type a bed thermistor is placed just inside the
patients nostril.
•In another type thermistors are mounted on a patient who
is fitted with an endotracheal tube or is on a respirator or
ventillator.
13 DEEPAK.P

Thermistor airway Pneumograph
14 DEEPAK.P

Pneumograph
•In some transducer , the thermistor is placed with a thin
platinum wire stretched taut across a short section of
tubing.
15 DEEPAK.P

DEEPAK.P16
Spirometer

Spirometer
17 DEEPAK.P

Spirometer
•A bell or jar is suspended from top in a tank of water.
•An air hose leads from a mouthpiece to the space inside of
the bell above the water level.
•The weight maintains the bell at atmospheric pressure.
•When the patient exhales, the pressure inside the bell
increases above atmospheric pressure causing the bell to
rise.
•Similarly, When the patient inhales, the pressure inside the
bell decreases causing the bell to comes down.
18 DEEPAK.P

DEEPAK.P19
Bio Electric Amplifiers

Bio Electric Amplifiers
20 DEEPAK.P

Amplitudes and spectral ranges of some important bio-signals
21 DEEPAK.P

DEEPAK.P22
Instrumentation Amplifiers

Instrumentation Amplifier
An instrumentation (or instrumentational) amplifier is a type
of differential amplifier that has been outfitted with input
buffer amplifiers.
The input buffer eliminate the need for input impedance
matching and thus make the amplifier particularly suitable for
use in measurement and test equipment.
The gain of the circuit is given by
23 DEEPAK.P

Instrumentation Amplifier
24 DEEPAK.P

Instrumentation Amplifier
The two amplifiers on the left are the buffers.
The rightmost amplifier, along with the resistors is standard
differential amplifier circuit.
25 DEEPAK.P

DEEPAK.P26
Isolation Amplifiers

Isolation Amplifier
Isolation amplifiers are a form of differential amplifier that
allow measurement of small signals in the presence of a high
common mode voltage by providing electrical isolation and
an electrical safety barrier.
They protect data acquisition components from common
mode voltages, which are potential differences between
instrument ground and signal ground.
Isolation amplifiers are used in medical instruments to
ensure isolation of a patient from power supply leakage
current.
Amplifiers with internal transformers eliminate external
isolated power supply.
27 DEEPAK.P

Isolation Amplifier
28 DEEPAK.P

Symbol of Isolation Amplifier
29 DEEPAK.P

Isolation Amplifier
30 DEEPAK.P

Isolation Amplifier
31 DEEPAK.P

Isolation Amplifier
32 DEEPAK.P

Isolation Amplifier
33 DEEPAK.P

Isolation Amplifier
34 DEEPAK.P

Isolation Amplifier
The 3650 and 3652 are optically coupled integrated circuit
isolation amplifiers.
Compared to these earlier isolation amplifiers, the 3650 and
3652 have the advantage of smaller size, lower cost, wider
bandwidth and integrated circuit reliability.
35 DEEPAK.P

Isolation Amplifier
36 DEEPAK.P

DEEPAK.P37
Chopper Stabilized Amplifiers

Chopper Stabilized amplifiers
Chopper stabilization constantly corrects input offset voltage
errors, including both errors in the initial input offset voltage
and errors in input offset voltage due to time, temperature,
and common-mode input voltage.
A chopper-stabilized amplifier is actually two amplification
paths in parallel.
A high-accuracy, low-frequency path (A2) incorporates high
gain and chopping, while high-frequency signals are
amplified by the parallel wideband amplifier A1.
The outputs of both stages are subtracted in a summer
amplifier, whose output is fed back to the inputs of both
amplifiers through a feedback resistor.
38 DEEPAK.P

Chopper Stabilized Amplifiers
39 DEEPAK.P

Chopper Stabilized amplifiers
40 DEEPAK.P

Chopper Stabilized amplifiers
41 DEEPAK.P

Chopper Stabilized amplifiers
 Two problems arise when we tries to record low level bio-
potentials.
Noise
DC Drift
These are worse , if we are using high gain amplifiers to
amplify the weak bio-potentials.
Noises produced in amplifier circuit and human body makes
the problem worse.
Drift is the change in gain or dc offset caused by thermal
effects on amplifier components.
42 DEEPAK.P

Chopper Stabilized amplifiers
 Drift can be minimized with the use of negative feedback.
This can be avoid by convert a DC (or near dc, low frequency
analog) signal to an AC signal that will pass through the
amplifier.
The solution is to chop or sample the analog signal at a
frequency that will pass through the AC coupled amplifier.
43 DEEPAK.P

DEEPAK.P44
Input Guarding

Input Guarding
 Physiological signals are low amplitude signals.
In most cases physiological signals accompanied by large CM
signals.
 If Op Amps are used, both differential (E)and common mode
(Ecm) signals are present.
The op amp cannot distinguish artifact from real signal and
C.M signal.
To avoid these problem, input guarding is used.
Here , we are placing a shield at the CM signals.
Input cable is shielded to avoid CM signals.

45 DEEPAK.P

Input Guarding
 Technique for increase both the input impedance of the
amplifier of bio-potentials and the CMRR.
Instrumentation amplifier providing input guarding

46 DEEPAK.P
Driven-right-leg circuit
reducing common-mode
interference.

Input Guarding

47 DEEPAK.P

DEEPAK.P48
Flow volume Transducers

Flow volume Transducer
•It measures the flow volume in litres per minute.
•The transducer assembly consists of differential pressure
transducer and an airway containing a wire mesh
obstruction.
•The wire mesh produces a pressure drop when it placed in
an airway.
•This pressure drop is measured as a differential pressure
across the mesh.
49 DEEPAK.P

Flow volume Transducer
50 DEEPAK.P

DEEPAK.P51
Blood Cell counting

Blood Cell Counting
52 DEEPAK.P

Blood Cell Counting
53 DEEPAK.P

Blood Cell Counting
54 DEEPAK.P

Blood Cell Counting
55 DEEPAK.P

Blood Cell Counting
56 DEEPAK.P

Coulter Method: Blood Cell Counting
57 DEEPAK.P

Coulter Method: Blood Cell Counting
58 DEEPAK.P

Blood Cell Counting
Using this technology, cells are sized and counted by detecting
and measuring changes in electrical resistance when a particle
passes through a small aperture.
This is called the electrical impedance principle of counting
cells.
A blood sample is diluted in saline, a good conductor of
electrical current, and the cells are pulled through an aperture
by creating a vacuum.
Electrical resistance or impedance occurs as the cells pass
through the aperture causing a change in voltage.
59 DEEPAK.P

Blood Cell Counting
This change in voltage generates a pulse .
The number of pulses is proportional to the number of cells
counted.
The size of the voltage pulse is also directly proportional to the
volume or size of the cell.
This was the principal parameter used in earlier analyzers for
characterizing all cell types, but it is now used primarily for
counting and sizing red blood cells and platelets.
60 DEEPAK.P

Coulter Method: Blood Cell Counting
61 DEEPAK.P

Pico Cell Blood Cell Counting
62 DEEPAK.P

Pico Cell Blood Cell Counting
63 DEEPAK.P

Optical Blood Cell Counting
64 DEEPAK.P

Optical Blood Cell Counting
65 DEEPAK.P

Optical Blood Cell Counting
66 DEEPAK.P

DEEPAK.P67
Blood Flow Meter

Blood Flow Meter
•Blood flow is the continuous circulation of blood in the
cardiovascular system.
•The science dedicated to describe the physics of blood
flow is called hemodynamics.
•Usually the blood flow measurements are more invasive than
blood pressure measurements / ECG
The abnormal changes in the blood flow or blood velocity
gives rise to malformation of vessels.
68 DEEPAK.P

Blood Flow Meter
•Blood flow is nothing but the volume of blood per time
[ml/min].
Typical values for blood flow [cm/s]:
1.Aorta 100 – 250
2.Abdominal 100
3.Vena Cava 5 – 40
69 DEEPAK.P

DEEPAK.P70
EM Blood Flow Meter

EM Blood Flow Meter
•The commonly used instrument for blood flow is
Electromagnetic type.
71 DEEPAK.P

EM Blood Flow Meter
•The operation principle behind the electromagnetic blood
flow meters is Faraday’s law of electromagnetic induction
which states that if electrical current carrying conductor
moves at right angle through a magnetic field, an
electromotive force is induced in the conductor.
Electromagnetic blood flow meters can be classified in to
1.Sine wave electromagnetic blood flow meter
2.Square wave electromagnetic blood flow meters
72 DEEPAK.P

EM Blood Flow Meter
•The sine wave electromagnetic blood flow meter uses
wave alternating current to generate the required
magnetic field.
•The flow voltage (induced voltage) is also sinusoidal.
•Even if for lower frequency application, the circuit becomes
complex and high frequency application results a problem of
stray capacitance effect, it can’t be used for a wide range of
frequency values.
•In the square wave electromagnetic blood flow meter, the
excitation is square wave alternating current, and the
induced voltage is square wave too.
73 DEEPAK.P

EM Blood Flow Meter
74 DEEPAK.P

EM Blood Flow Meter
75 DEEPAK.P

EM Blood Flow Meter
76 DEEPAK.P

Types of EM Blood Flow Meter
77 DEEPAK.P

EM Blood Flow Meter
78 DEEPAK.P

EM Blood Flow Meter
79 DEEPAK.P

DEEPAK.P80
Ultrasonic Blood Flow Meter

Ultrasonic Blood Flow Meter
81 DEEPAK.P

Ultrasonic Blood Flow Meter
•An ultrasonic flow meter is a type of flow meter that
measures the velocity of a fluid with ultrasound to
calculate volume flow.
•The blood cells in the fluid reflects the ultrasound signal
with a shift in the ultrasonic frequency due to its
movement.
•Using ultrasonic transducers, the flow meter can measure the
average velocity along the path of an emitted beam of
ultrasound, by averaging the difference in measured
transit time between the pulses of ultrasound propagating
into and against the direction of the flow or by measuring
the frequency shift from the Doppler effect.
82 DEEPAK.P

Ultrasonic Blood Flow Meter
83 DEEPAK.P

Ultrasonic Blood Flow Meter
84 DEEPAK.P

Ultrasonic Blood Flow Meter
85 DEEPAK.P

Ultrasonic Blood Flow Meter
86 DEEPAK.P

Ultrasonic Blood Flow Meter
87 DEEPAK.P

Ultrasonic Blood Flow Meter
88 DEEPAK.P

Ultrasonic Blood Flow Meter
89 DEEPAK.P

Ultrasonic Blood Flow Meter
90 DEEPAK.P

Ultrasonic Blood Flow Meter
91 DEEPAK.P

Doppler Blood Flow Meter
92 DEEPAK.P

DEEPAK.P93
Blood pH Measurement

Blood pH Measurement
•In chemistry, pH is the negative log of the activity of the
hydrogen ion in an aqueous solution.
•Solutions with a pH less than 7 are said to be acidic and
solutions with a pH greater than 7 are basic or alkaline.
• Pure water has a pH of 7.
•Mathematically, pH is the negative logarithm of the activity
of the (solvated) hydronium ion, more often expressed as the
measure of the hydronium ion concentration
94 DEEPAK.P

Blood pH Measurement
•Blood pH is an important factor to determine the acid base
balance in the human body.
•Blood pH level plays an important role for your overall
health, because if your blood pH level is acidic, your cells
cannot function properly.
•One of the major contributors to acidosis is carbon dioxide, a
byproduct of metabolism.
•Carbon dioxide combines with water to form carbonic acid.
•The normal pH of blood is between 7.35-7.45.
95 DEEPAK.P

Blood pH Measurement
•The increase in hydrogen ion concentration causes the pH of
the body fluids to decrease.
•If the pH of the body fluids falls below 7.35, symptoms of
respiratory acidosis become apparent.
•Blood gas analyzers are used to measure pH, pCO2 and pO2
etc
•pH of biological fluids is measured using a glass electrode.
96 DEEPAK.P

Blood pH Measurement
97 DEEPAK.P

Blood pH Measurement
98 DEEPAK.P

Blood pH Measurement
99 DEEPAK.P

Blood pH Measurement
•A blood sample for measuring pH parameters is collected in
a glass syringe.
•The blood pH level can be determined by
1.Saliva pH test
2.Urine pH test
3.Blood pH test
100 DEEPAK.P

Saliva Blood pH Test
•Saliva pH test: It is one of the fastest and simplest tests for
measuring blood pH level.
•However, it is not the most accurate.
•Before taking the test, you should wait at least two hours
after eating.
•Depending on the food you eat, it can increase or decrease
your blood pH level.
•To do the saliva pH test, go buy the pH paper test strip at
your local drug stores and health food stores and follow the
directions on the package.
101 DEEPAK.P

Urine Blood pH Test
•This test works similar to the saliva pH test, but it tests urine
instead of saliva.
•The pH test strip for testing urine is the same as the saliva
pH test strip.
•Blood pH test
•This test is not as convenient as the saliva and urine pH test,
but it is usually more accurate.
•To do the blood pH test, you will need to visit your doctor
and get your blood drawn.
102 DEEPAK.P

Blood pH Measurement
•One of the best ways to bring your blood pH back to its
normal state is to eat mostly organic alkaline food. Another
way is to drink distilled water.
•Distilled water is very pure water, so if you want to reduce
your risk of acidosis, you will need to add some minerals to it.
103 DEEPAK.P

Blood pH Measurement
104 DEEPAK.P

Glass Electrode
105 DEEPAK.P

PH SENSITIVE FET
106 DEEPAK.P

Conductimetric PH Sensor
107 DEEPAK.P

Fiber Optic PH Sensor
108 DEEPAK.P

Fiber Optic PH Sensor
109 DEEPAK.P