Venflon Pro Safety P2P

AdamMurr 1,173 views 2 slides Oct 11, 2015
Slide 1
Slide 1 of 2
Slide 1
1
Slide 2
2

About This Presentation

No description available for this slideshow.


Slide Content

BD Venflon
TM
Pro Safety
Points to Practise
Product Information
Code No. Gauge Colour Catheter Specification Units per

Length mm ODmm IDmm
Flow Rate Box/Case
H2O ml/min
393222 22G Blue 25 0.9 0.6 42 50/500
393224 20G Pink 32 1.1 0.8 67 50/500
393227 18G Green 45 1.3 1.0 103 50/500
393229 16G Grey 45 1.8 1.4 236 50/500
393230 14G Orange 45 2.0 1.6 270 50/500
BD Medical
The Danby Building
Edmund Halley Road
Oxford Science Park
Oxford OX4 4DQ
Tel: 01865 748844
Fax: 01865 717313
www.bd.com
Product Features and Benefits
1 Sharp polished needle to ensure easy skin and vein penetration
2 BD Vialon
TM
cannula with clinically proven benefits
3 Cannula tip has a tapered design to reduce pain during insertion
4 Injection port with snap cap
5 Grip plate allows a more ergonomic grip to aid insertion
6 Needle grip specially designed to accommodate
a variety of insertion techniques
7 End cap placed conveniently for easy handling
1
3
2
4
5
6
7
8
9
8 Passive safety mechanism automatically deploys on needle removal
9 A needle protection shield which:
- fully encases the needle tip
- is plastic and has no sharp edges
- is designed to minimise blood exposure
Rubik's Cube
®
used by permission of Seven Towns Ltd. www.rubiks.com

BD Venflon Pro
TM
Safety
Points to practise
Preparation
Insertion
Needle removal
• Comply to aseptic technique
throughout the procedure
• Disinfect the injection site according
to local policy
• Remove the cannula from the
packaging and lower the wings
• Adopt your preferred grip and remove
the needle cover
• Insert the cannula at a low angle
• Upon primary flashback lower the
angle almost parallel to the skin
• Advance the cannula slightly, two to
three millimetres, to ensure the cannula
tip is in the vein
• Stabilise the cannula
• Ease the needle back two to three
millimetres
• Secondary flashback between the needle
and cannula will confirm correct
placement in the vein
• Advance the cannula completely into
the vein
• Remove the tourniquet
• Stabilise the cannula
• Occlude the vein just above cannula
tip and withdraw the needle holding the
needle grip or grip plate
• Do NOT hold onto the needle
protection shield
• Dispose of the needle in a sharps
container
• Close the IV access with the end
cap, other add-on devices or add
administration set
• Apply sterile dressing
• Flush the cannula to verify correct
placement in the vein
BD Medical
The Danby Building
Edmund Halley Road
Oxford Science Park
Oxford OX4 4DQ
Tel: 01865 748844
Fax: 01865 717313
www.bd.com
Safety, in safe hands
2 3
4 5 6
7 9
10
8
BD, BD logo and all other trademarks are property of Becton, Dickinson and Company © 2011 BD.