Classification of BSCs
BSCareclassifiedintothreeclasses:-
•ClassIcabinetsprovidepersonnelandenvironmentalprotectionbutnot
sample/productprotection(cabinetairisnotfiltered)
•ClassIIcabinetsprovidepersonnel,product,andenvironment
protectionsincetheairinthecabinetisfiltered.
•ClassIIIcabinetsaregas-tight,andallmaterialsenterandleavethrougha
dunktankordouble-doorautoclave.
Types of Protection
Typesofprotection Personnel Product Environment
ChemicalFumeHoods(vapors&gases) X
Laminar Flow CleanBenches
(protectionfromparticulate)
X
ClassIBiologicalSafetyCabinets
(protectionfromparticulate)/TBsafety
hoods
X X
ClassII&IIIBiologicalSafetyCabinets
(protectionfromparticulate)
X X X
Biosafety Cabinet
Proper Use of BSCs
1.Place equipment to be used in
cabinet
2.Be sure not to over crowd (one
person per use)
3.Keep back and front grills clear
4.Keep items away from sides of
work deck
5.Ensure cabinet is running before
start of work
6.Disinfect surface with 70% ethanol-
before and after use
7.Allow cabinet to run for 10 min
before and after use
8.Verify windows are sealed or at the
proper level.
9.Avoid overcrowding of the work area
in a BSC.
10.Avoid constant in and out motion at
face of cabinet. This will draw air out
of the cabinet.
Decontamination of BSCs
BSCs should be decontaminated:-
•Prior to filter changes
•If access to the contaminated area is required
•If the cabinet is to be moved or to be decommissioned
•If there is a change in research or diagnostic profile
•Laboratory personnel must decontaminate the BSC before and after
use.
BSC Placement and Selection
PlacementofBSCshouldbeawayfrom:-
•Doors
•Airsupplyvents
•Heavytrafficareas
SelectionofBSCshouldbebasedriskassessment.
Care and Maintenance of Autoclave (1/2)
•DevelopSOPsforproperuseandmaintenanceofthe
autoclave,followingthemanufacturer’sguideline.
•QualityControlforautoclavesisimportantandshouldbe
performedasperthefacility’sSOP.
oChemicalindicators
ProcessIndicators–autoclavetape
ProcessIntegrators-
oBiologicalIndicators
Care and Maintenance of Autoclave (2/2)
Note:
•TheBiologicalindicatorsarerecommendedasthegoldstandardforconfirmation
ofsterility.
•AutoclavetapedoesNOTconfirmthesterilityofmaterials.
Care and Maintenance of Centrifuge
•Allcentrifugesmustbeoperatedandservicedaccordingtomanufacturers’
instructionsandservicedbyappropriatelyqualifiedpersonnel.
•Centrifugesmustbecleanedanddisinfectedregularly,orimmediately
decontaminatedafteraspill,withanappropriatedisinfectant.
4. Pipettes and Pipette Aid
•Apipetteisalaboratorytoolcommonlyusedinthelaboratorytotransfera
measuredvolumeofliquid.
•Itisimportantforthesafetyoftheuserandenvironmentthatproper
techniqueisused.
•Topreventthegenerationofaerosols,pipettesmustnotbeusedtoblowair
orforciblyexpelliquids/solutionsthatcontainbiologicalagents.
•Allpipettetipsshouldhavecottonplugstoreducecontaminationofpipetting
devices.
•Mouthpipettingisprohibited,andtheuseofapipetteaidisencouraged.
Care and Maintenance of Pipette
•ThefacilityshouldhaveanSOPforthe
useandmaintenanceofpipettesand
pipetteaid.
•Followthemanufacturers’guide.
•Maintenanceandcalibrationofpipette
aidmustbedonebyaqualified
personnel.
Note:Calibrationofpipettesandpipetteaidscanbedoneatthe
NationalCalibrationCentreatNationalPublicHealthlaboratoriesin
Nairobi.